Title: Spectrochemical Measurements
1CHAPTER 2
- Spectrochemical Measurements
2COMPLETE SPECTROCHEMICALMEASUREMENT
- Steps involved in determination of the
concentration of the analyte in a sample - acquisition of the initial sample,
- sample preparation or treatment to produce the
analytical sample, - presentation of the analytical sample to the
instrument, measurement of the optical signals, - establishment of the calibration function with
standards and calculations, - interpretation,
- feedback.
3Spectrochemical measuremennt process
- A sample introduction system presents the sample
to the encoding - system, which converts the concentrations c1,
c2, c3 into optical signals O1 - O2, O3.
- The information selection systems selects the
desired optical signal O1 for presentation to the
radiation transducer. - This device converts the optical signal into an
electrical signal (current i, voltage e,
frequency f, etc.) that is processed and read out
as a number.
4Optical Electrical signal
Spectrum
Mainly ? selector Spectrum
Manual or automatic
Human operator is being replaced by
microcomputers
Sample is treated before introduction
Convert the transducer output into a form
appropriate for readout as numerical values
5Expression of Optical Intensity
- Optical intensities are expressed in two systems
- Radiometric system
- Photometric system
6Radiometric System Basic Definitions
- radiometric system of units is based on the
actual radiant energy emitted by a source or
striking a receiver (e.g., optical transducer)
and is preferred in the International System of
Units (SI). - The basic quantity in this system is the radiant
energy Q in joules (J). - In the radiometric system there are general
quantities used to describe radiation sources,
and radiation receiver. - radiant intensity, emittance, emissivity, and
radiance - refer specifically to radiation from a source
- volumes, areas, and solid angles
- refer to properties related to source
- Irradiance and exposure
- describe the receiver and its area
7(No Transcript)
8- All quantities are in general functions of
spectral position (wavelength, wave number,
frequency, etc.) in that they are usually
employed to represent the magnitude of the
quantity over some spectral interval. - In general these values represent the cumulative
magnitude of the quantity over the wavelength
interval from 0 to ?. - If the term "total" is employed, as in total
radiance, it implies the radiance over the
wavelength interval from 0 to?. - Generally, radiometric quantities are considered
within small spectral intervals.
9- Spectral quantities
- radiometric quantities per unit spectral
interval and given a subscript ? (for
wavelength), ? (for the frequency)and
(for wave number) - spectral radiance B ?, is the radiance per unit
wavelength interval (per nm) - Partial radiance, B??
- radiance in the wavelength interval ?2 - ?1
-
Cumulative radiance
- Total radiance, B
- the radiance from a source related to spectral
radiance
10- Sources that emit narrow spectral lines (typical
halfwidths ltlt 1 Ao) are usually characterized by
reporting the radiance B of each line which is
the integrated spectral radiance over the total
width of the line. - A broadband source is normally characterized by
its spectral radiance B? because only part of its
emitted spectral range is selected or observed as
determined by a wavelength selector.
11Geometric Factors
- Often, radiometric quantities include the
geometric factors of solid angle and projected
area.
- (a) Plane angle and one radian of angle are
illustrated. - One radian is the angle at the center of a
circle that - intercepts an arc equal in length to the
radius.
12- (b) Solid angle is defined by the cone generated
by a line that passes through the vertex O and a
point moved along the periphery of the surface. - One steradian is the solid angle at the center of
a sphere of radius r that subtends an area of r2
units on the surface
13Examples of Use of radiometric terms
- In most spectroscopic situations one is
eventually interested in the radiant power that
is incident on a receptor - Consider, for example, a point source with
dimensions that are small compared to the
distance (d) from the source to the receptor of
projected area Ap. - The source could be characterized by the total
radiant power ? that it emits in all directions. - In this case, it is more useful to use the
radiant power per unit solid angle (the radiant
intensity), which is given by
14A Source of significant area
The radiant power, ?I incident on area A2 of the
receptor is the source radiant times the area
times the solid angle viewed times the area
viewed
15Photometric System(will not be used further)
- It is a relevant system based on the apparent
intensity of av source as viewed by the average
bright adapted human eye. - Quantities in this system have meaning only in
the visible region - The basic unit of this system is the lumen
- A source of 1 candela emits 1 limen per steradian
- Photometric and corresponding radiometric
quantities are given in the following table
16(No Transcript)
17Relationships between the radaint quantities and
the spectrochemical methods
18Emission measurements Emission and
chemiluminescence (bioluminescence) methods
The energy changes that occur during excitation
(dashed lines) or emission (solid lines)
Typical spectrum
e.g., sodium atoms are excited in a flame by
Collisional processes and emit characteristic
radiation.
- Addition of thermal, electrical or chemical
- energy causes nonradiational excitation of the
- analyte and emission of radiation in all
- directions (isotropic emission)
19- The frequency of the emitted radiation
corresponds to the discrete energy differences
between levels, as shown in the figure - When thermal equilibrium is maintained, the
number of atoms per cm3 in level i, ni is related
to the total number of atoms per cm3, nt, by the
Boltzmann distribution
Excitation energy relative to the ground state
Statistical factor of state i
Partition function
- Na and other alkali metals have excited levels
close to the - ground state levels. Thus their resonance
lines occur in the - visible and near IR regions and are readily
observed in media - such as flames.
20- The radiant power of emission ?E from state j to
state i is given by the population density of
excited atoms nj times the probability Aji (s-1)
that an excited atom will undergo the transition,
times the energy per emitted photon h?ji, times
the volume element observed V (cm3). Or
- The equation shows that the radiant power of
emission - is proportional to the excited-state
population density - and thus to the analyte concentration through
the - previous equation.
212. Absorption measurement
- For absorption to occur, the frequency of the
incident radiation must correspond to the energy
difference between the two states involved in the
transition as shown in the figure. - For many conditions the absorption of radiation
follows Beer's law
22Absorptivity
a
Conc.
Absorption pathlength
Absorbance
Transmittance
Molar absorptivity
Also,
a
233. Luminescence measurement
- Luminescence is radiation emitted from relatively
cool bodies. - There are several classes of luminescence
spectrochemical methods - Chemiluminescence and bioluminescence
- excited analyte species are produced by
chemical reactions, and the resulting emission is
measured. - Electroluminescence
- It results from the movement of electrons in a
sample and may be caused by an electrical
discharge, by recombination of ions and electrons
at an electrode, and by interactions of materials
with accelerated electrons as in a cathode ray
tube. - Triboluminescence
- It results from the mechanical separation of
charges followed by a discharge (e.g., broken
crystals of sugar). - Thermoluminescence
- It is the enhancement of other types of
luminescence by the addition of heat. - Chemiluminescence and bioluminescence are
employed in - analytical procedures. The
excitation/emission transitions for these were
illustrated in a previous figure.
24Photoluminescence methods Molecular and atomic
fluorescence
- Methods that utilize an external radiation source
for excitation (as in absorption methods), but
the sought-for information is the radiation
emitted by the sample as shown in the figure
Loss of energy by emission Of photons
Radiationless processes
25Measurement of luminesced radiant power
- When a portion of the incident radiant power ?o
is absorbed so that the transmitted radiant power
? is less than the incident radiant power - Under many conditions the radiant power
luminesced (for all wavelengths) ?L is
proportional to the absorbed radiant power (?o -
?). Thus,
The transmitted radiant power is related to the analyte concentration by Beer's law
Thus,
Expansion of the above eq. in a Taylor series
gives,
When the term abc is lt 0.01, higher-order terms in the expansion contribute less than 1 to ?L, and under these conditions,
26Scattering measurement
- Radiation from an external source can also be
scattered by the sample - The intensity, frequency, and angular
distribution of scattered radiation can be used
in spectrochemical methods. - In molecular scattering methods, particles
smaller than the wavelength of the incident
radiation can scatter that radiation elastically
without a change in its energy. - Small-particle scattering is called Rayleigh
scattering - it typically occurs with atoms or molecules.
- Rayleigh scattered radiation occurs in all
directions from the scattering particle.
27- Debye Scattering
- It is the scattering that takes place from larger
- particles with dimensions on the order of the
wavelength of the incident radiation. - Here the scattered radiation is of the same
frequency as the incident radiation, but the
angular distribution of the scattered radiation,
unlike Rayleigh scattering, is not uniform. - Mie scattering
- Scattering from much larger particles
- Large-particle scattering (Debye or Mie) can be
used to determine particle sizes and is important
in turbidimetry and nephelometry where suspended
particles are the scatterers.
28- Brillouin and Raman scattering
- These are forms of inelastic scattering which
involve a change in the frequency of the incident
radiation. - Brillouin scattering results from the reflection
of radiant energy waves by thermal sound waves - Raman scattering involves the gain or loss of a
vibrational quantum of energy by - molecules.
- The scattering signal is proportional to the
incident radiant power.
29Selection of Optical Information
- In analytical procedures the selection step
allows us to separate the analyte optical signal
from a majority of the potential interfering
optical signals. - The vast majority of analytical techniques select
the desired information based only on its
wavelength - Thus, wavelength selection is essential!
30Wavelength Selection
Instrumentation for spatial dispersion and
detection of optical signals
- Some of the radiation from the spectrochemical
encoder enters the - entrance slit and strikes the dispersion
element. - The dispersion element and image transfer
system cause each - wavelength to strike a different position in
the focal plane where - different photo detector configurations can be
used - According to the phtodetector configuration
various names were given - to these optical devices
31Specific names given to optical instruments
- spectrograph, a large aperture in the focal plane
allows a wide range of wavelengths to strike a
spatially sensitive detector such as a
photographic plate. - In recent years, solid-state video-type detectors
have become available and are often employed in
spectrographs in place of film. - These detectors are actually an array of a large
number of closely spaced miniature photoelectric
detectors. - They have the advantage that the spectrum can be
obtained immediately without the time required
for film development, for obtaining the density
of the lines recorded, and so on. - A spectroscope is a device that allows a visual
observation of the spectrum. It is a spectrograph
that uses a viewing screen for observing the
spectrum in the focal plane.
32- In a monochromator, an exit slit about the same
size as the entrance slit is used to isolate a
small band of wavelengths from all the
wavelengths that strike the focal plane. - One wavelength band at a time is isolated and
different wavelength bands can be selected
sequentially by rotating the dispersion element
to bring the new band into the proper orientation
so that it will pass through the exit slit. - If the focal plane contains multiple exit slits
so that several wavelength bands can be isolated
simultaneously, the wavelength selector is called
a polychromator.
33- A spectrometer is a spectrochemical instrument
which employs a monochromator or a polychromator
in conjunction with photoelectric detection of
the isolated wavelength band(s). - The photodetector is placed just outside the exit
slit. - If a polychromator is employed with a separate
photodetector for each exit slit, the instrument
is often called a direct-reading spectrometer. - Some spectrometers use optical components to
sweep the spectrum quite rapidly across a single
exit slit. - These rapid-scanning spectrometers can obtain a
spectrum in a few milliseconds.
34- A spectrophotometer is an instrument similar to a
spectrometer except that it allows the ratio of
the radiant power of two beams to be obtained, a
requirement for absorption spectroscopy. - A photometer is a spectrochemical instrument
which uses an optical filter for wavelength
selection in conjunction with photoelectric
detection.
35- Interferometers are nondispersive devices in
which the constructive and destructive
interference of light waves can be used to obtain
spectral information.
36Measurement of optical signals
- All spectrochemical techniques that operate in
the UVvisible and IR regions of the spectrum
employ similar instrumental components, as
mentioned before. - The major instrumental differences between
emission, photoluminescence, and absorption
techniques occur in the arrangement and type of
sample introduction system, encoding system, and
information selection system. - All techniques depend upon the measurement of
radiant power. - The specific transducers and signal processing
devices used in various regions of the spectrum
in specific spectrochemical techniques are
described later. - In this section we explore how the analytical
signal is extracted from the readout data in
spectrochemical methods.
37Radiant power monitor
. The radiant power monitor provides a numerical readout that is related to the radiant power (number of photons per second or watts) impingent on the transducer.
38Analytical Signal
- The analytical signal is rarely obtained directly
as a result of one spectrochemical measurement. - Because of the presence of background and other
extraneous signals, the analytical signal must be
extracted from the raw readout data. - The analytical signal for emission and
chemiluminescence techniques is defined as the
signal to be displayed by the readout device due
only to analyte emission. - It is given the symbol EE, and we presume that EE
is directly related to the radiant power of
emission?E. - Similarly, the analytical signal in
photoluminescence techniques,? L, is the measured
signal due only to radiationally produced
emission of the analyte. - In the case of absorption methods, the analytical
signal is the absorbance A due only to absorption
of radiation by the analyte species.
39- Because of the presence of extraneous signals,
such as signals from concomitants, the sample
cell, and room light, at least two measurements
are required to obtain the analytical signal. - The background or extraneous signal that
registers on the readout device is due to two
primary sources. - The first source is the dark signal Ed of the
radiant power monitor, which is the signal
present when no radiation is impingent on the
transducer. - The second source is the background signal, EB
due to background radiation that strikes the
transducer. - The background radiation is composed of radiation
from all sources other than the desired optical
phenomenon from the analyte. - The transducer can convert this optical signal to
an electrical current, voltage, or charge. - Normally, the output of the signal processing
system to be displayed on the readout device is
an electrical voltage - Generally, analyte and background signals will be
written as voltages E.
40Analytical signal in Emission and
Chemiluminescence Spectrometry
Instrumentation for emission spectrochemical
methods.
Spectrochemical encoder
- The excitation source is the spectrochemical
encoder - The emission that results from excitation of the
analyte species - by a flame, a plasma, or a chemical reaction
encodes the - concentration of the analyte as the radiant
power of emission ?E. - In some spectrochemical methods the excitation
source and sample - container are an integral unit, as in the
nebulizer-burner used in - flame emission and the reaction cell used in
chemiluminescence
41- The analytical signal of the sample is usually a
total or composite signal EtE - This total signal is the sum of analytical
signal EE the dark signal Ed and the background
emission signal EbE - To extract the analytical signal, a second
measurement is required to obtain the sum of the
dark signal and the background emission signal. - This second measurement is normally made by
replacing the analytical sample with a blank,
then
Blank signal Eb Ebk
- If desired, the dark signal can be obtained
separately by blocking all - radiation from reaching the radiant power
monitor. - The background emission signal could then be
obtained from Ebk - Ed. - In many instruments the blank solution is used
to adjust the readout - device to read zero by suppression of the
blank signal. - This establishment of the zero position is
still, however, a measurement - of the blank signal
42Analytical signal in Photoluminescence
Spectrometry
- An external source of EMR excites the analyte.
The analyte concentration is optically encoded as
the luminescent radiant power ?L, which is
measured with the radiant power monitor. - The emission wavelength selector that views the
luminescence of the sample is typically placed to
collect radiation at 90 with respect to the
excitation axis.
Instrumentation for photoluminescence
spectrometry
Specific wavelengths from an external radiation source are isolated by the excitation wavelength selector to excite the analyte in the sample cell. The emission wavelength selector selects the wavelength band where analyte luminescence is concentrated and passes it to the radiant power monitor
43- The total analytical signal EtL is expressed
Blank
Analytical luminescence signal
Analytical thermal emission signal
background
dark
Scattering
Background luminescence
- Analyte and background emission in the
UV-visible region are - usually significant only in atomic
spectroscopy.
44- The analyte luminescence signal EL can be
obtained with two measurements only if the
analyte emission signal EE is small compared to
EL, which is often the case. - If EE is significant, subtraction of the blank
signal gives a measured analyte luminescence
signal EL that differs from EL
- To obtain the true analyte luminescence signal
EL when EE is significant, - the excitation source must be turned off.
Then the two measurements EtE - and Ebk are made to obtain EE.
- Subtraction of EE from EL gives the true
analyte luminescence signal. - In some cases it is possible to eliminate the
measured contribution from - analyte emission optically or electronically.
- For example, if the excitation source is
modulated and alternating-current - (ac) amplification is used, the ac
luminescence signal can be distinguished - from the do emission signal. Often the blank
measurement is used to set - the zero position of the readout device.
45Analytical signal in Absorption Spectrometry
- Typical absorption spectrometer
- It is similar to the luminescence spectrometer
except that all - components are placed on the same optical axis
- The shutter allows the user to block the
radiation source in order to - obtain the dark signal. Usually, only one
wavelength selector is - required.
- Absorption measurements can be made as
transmittance T where - absorbance A is calculated manually or the
logarithmic conversion can - be done electronically or with computer
software and the absorbance A - displayed by the readout device.
461. Transmittance readout
- T values could be obtained by
- 1. measuring the signal ES that results from
the - source radiant power passing through the
- analytical sample
- 2. measuring the signal Er that results from
the - source radiant power passing through the
- ideal blank or reference solution
- 3. obtaining the transmittance as in
sample
reference
In practice, the presence of other signals (dark signal, background emission) necessitates a third measurement The measured transmittance T' is defined by the equation
47- ESt is the total sample signal obtained with the
source shutter - open and the analytical sample in the sample
container, - Eot is the zero percent transmittance (0 T)
signal obtained with the shutter closed and the
blank in the sample container, - Ert is the 100 T signal obtained with the
shutter open and the blank (reference) in the
sample container - The 0 T signal Eot is composed of any background
emission EbE and dark current Ed - When the blank is in the sample container and the
shutter open, the measured total reference signal
Ert called the 100 T signal, is composed of the
reference transmission signal Er, the 0 T
signal, and any background luminescence EbL
48- When the analytical sample is in the sample
container and the shutter is open, the measured
signal is Est, the total sample signal. This
signal is given by
Sample transmission signal
emission signal
luminescence signal
- From the above equations, the measured
transmittance is
492. Direct absorbance readout
- Many modern absorption spectrometers can display
absorbance directly. - The true absorbance A is given by
Voltage proportional to the analyte absorbance
Log conversion factor in volts per A unit
- The voltage EA and hence A are found from two
measurements - A reference logarithmic voltage or zero
absorbance voltage Elr is - obtained with the shutter open and the blank
in the sample container - 2. Then a sample logarithmic voltage Els is
obtained with the shutter open and the analytical
sample in the sample container
50- The voltage is then given by
The voltages Els and Elr are logarithmically related to ES and Er
Constant reference voltage
- Often Elr is set to zero on the readout device
so that Els is read out - directly as EA
- Note that in the two-step absorbance
measurement scheme, a - measurement is not made with the lightsource
shutter closed (0 T) - since A would be infinity.
- Thus (Ed EbE) must be negligible compared to
ES and Er or - electronically or optically set to zero by
other means. - Also, EE EbL EL must be negligible so
that ES Est and Er Ert - otherwise, the measured absorbance A' only
approximates the true - absorbance A.