Title: FRM: Nitrogen System
1FRM Nitrogen System Validation
AAR-440 Fire Safety BranchWm. J. Hughes
Technical CenterFederal Aviation Administration
2Background
- FAA Fire Safety Research developed prototype
OBIGGS to demonstrate its potential for fuel
tank ullage flammability reduction to support FRM
rulemaking - Built and tested inerting system based on the two
flow concept - Validated system operation during ground and
flight testing - Developed instrumentation and methods to validate
both inerting system operation and FRM
effectiveness - Measured inerting system parameters during
aircraft operation and compared data with
existing static performance data - Developed instrumentation system to measure
oxygen concentration during flight testing
3Prototype Installation Ground Testing
- System installed in a Boeing 747SP with fully
functioning systems and ground service equipment - Decommissioned from airline service and purchased
by the FAA for Ground Testing Only - All major systems fully operational
- Has independent power for test equipment and
instrumentation
- System installed between two structural members
in empty pack bay as designed - Completely Instrumented
- Oxygen sampling, pressure taps, and thermocouples
on FRM to measure performance - Thermocouples in Pack Bay
- Some Weather Data Available
4Measured Results Ground Testing
- Ground tested the inerting system on the 747SP
test article - Operated inerting system with static conditions
(best as possible) for the purposes of validating
the system performance - Focused on the volume flow of 5 and 11 NEA that
could be generated with varying ASM pressure at
sea level
- Testing Illustrated
- Stable system performance is difficult to achieve
in a dynamic aircraft environment - Inerting system has potential of being used as a
flammability reduction method (FRM)
5Prototype Installation In-Flight Demonstration
- Installed FAA prototype inerting system on Airbus
- System was installed on a cargo pallet and
mounted in the aft cargo bay of an Airbus A320
used for the purpose of flight test - System and aircraft fully instrumented with
extensive DAS capabilities - System modified to use only 1 ASM to match size
to aircraft
- Operated out of Airbus, France Flight Test
Toulouse - Measured both system performance and ullage O2
- Testing validated concept of using a simplified
inerting system as an FRM
6Measured Results In-Flight Demonstration
- Compared dynamic ASM performance with static
measurements - Operated on A320 flight test and compared to
static lab data - Data comparison generally good with some big
discrepancies
Single ASM Test
- Illustrates the difficulty in obtaining stable
conditions during flight test - Obtaining 180 degree F air temperature inlet
problematic - This makes model validation difficult
7Measured Results In-Flight Demonstration
- Observed oxygen concentration progression and
distribution for typical short flight with a
rapid descent - Need to show ullage oxygen concentration stays
below 12 under flight cycle scenarios deemed
important by engineering analysis
- Data illustrated the effectiveness of NEA and
reducing ullage flammability - Inert gas distributed easily with no mixing
devices - Ullage behavior consistent with perfect mixing
(easy to model)
8Prototype Installation Total System Installation
- Installed system in highly modified Boeing
747-100 - Modified and operated by NASA for the purposes of
carrying a Space Shuttle Orbiter for operations
and maintenance - Fully operational, standard, fuel system with
unmodified pack bay - Standard 747 Multiple-bay / compartmentalized
center-wing tank - Operated by excellent test pilots/crew and
dedicated maintenance group - System Installed in pack bay as would be in fleet
aircraft
- Instrumentation
- Aircraft is Fully Instrumented
- Oxygen sampling, pressure taps, and thermocouples
on system for OBIGGS performance - Thermocouples in Pack Bay Area
- Pressure altitude measured
9Measured Results Total System Installation
- Dynamic performance during takeoff/ascent and
descent/landing portion of flight was analyzed - Each flight different, with a range of varying
system performance - These variations in performance had a small but
measurable effect on the resulting average oxygen
concentration observed in the ullage at the end
of the flight cycle (at touch down) - System undersized for this aircraft
- Average ullage oxygen concentration behavior very
typical of previous observations - Peak average ullage oxygen concentrations around
12 - Peak forward bay (vented bay) oxygen
concentrations were well above 12 on descent
10Comparison of System Performance During Takeoff
11Comparison of System Performance During Landing
12Comparison of System Performance During Takeoff
13Measured Results Multi-Bay NEA Distribution
- Inert gas distribution with single point deposit
more problematic during flight test than in scale
tests in lab - Observations illustrate inert gas should be
distributed differently on descent to keep peak
oxygen concentrations below 12 in forward tank
vent bays
- Distribution models duplicate acquired data
poorly - Inert gas flow did not split from bay to bay IAW
bay wall areas - Some bays do not illustrate perfect mixing (hard
to model)
14Background - Measuring Oxygen Concentration
- To demonstrate the potential for flammability
reduction, Fire Safety developed an in-flight
oxygen concentration measurement system - FAA (OBOAS) used during several flight tests to
validate that the OBIGGS did reduce the
flammability of the fuel tank ullage - FAA has studied the different methods of
measuring oxygen concentration and has garnered
many lessons - There are many ways to measure oxygen
concentration - Many different sensors available on the market
- Reactive/soluable (Galvanic, Zirconium Oxide
- Non-consumable (Paramagnetic, Light Absorption)
- Many different ways to apply an oxygen sensor
- Continuous gas sample (regulated or unregulated)
- Discrete gas sample
- In situ sensor placement
15Block Diagrams of Continuous Gas Sampling Methods
16Vacuum Bottle Gas Sample Assembly for Flight Test
17On-board Oxygen Analysis System Development
- Wanted to apply a traditional oxygen
sensor/analyzer used in lab for a flight test
application - Must provide constant condition
(pressure/temperature) gas sample - Developed a gas sample system that can do this
from 0-40,000 feet for a wide range of ullage
temperatures - Used powerful diaphragm pumps with active
pressure controllers to provide a constant
pressure and stable temperature gas sample - Integrated with remote flow through galvanic cell
sensors - System used flash arrestors, liquid traps,
desiccating filters, and a ventilated containment
shroud to ensure safety of flight operations - Results from test series were published in DOT
reports - Data lag significant, loud, heavy, sample train
not reliable - With some refinement, was easy to calibrate and
gave good repeatable, results and filled an
otherwise empty niche
18FAA Oxygen Concentration Measurement Method
Fuel Tank Vapor Fuel Tank Liquid Pressurized
Air Electrical Power Electronic Signals
19OBOAS Validation Data from Airbus Flight Test
20The Trouble with Accurate Measurements . . .
- Your measurement can be very accurate, but if
your gas sample location is not representative of
the volume you are measuring, all is for naught - Observed steep (1-3 O2) vertical gradients in
fuel tank bay for extended periods of time (½ - 1
hour) - OBIGGS flow tends to take care of the horizontal
mixing whether it be one large bay or multiple
partitions in the tank - Only observed steep gradients in a few isolated
ground inerting cases where the fuel tank was
otherwise quiescent - Multiple-bay tanks pose a challenge to determine
what ullage areas will act as one area and what
areas will not - Can always add more sample locations, but where?
- Ideally applicant will use a system of modeling
in advance with flight test validation to
illustrate inert gas distribution
21Good and Poor Mixing Data from Ground Inerting
Test
22New Inovations Light Absorption with TLD
- Oxygen concentration can be measured by examining
the amount of light absorbed in a photo cell
filled with the gas - This methodology has been around for years but
was made more practical with the advent of
tunable laser diodes (TLDs) - Recently developed instrument for flight test
- Unit acquires an unregulated gas sample and
passes it through a cell - System has elaborate calibration to compensate
for sample pressure and temperature - Technology has been applied in-situ