The BS W Portablecoulometric Karl Fischer
titrator
2
KF Titration Cell Differences
VOLUMETRIC
COULOMETRIC
Detector
Detector
Generator
Burette
3
Why Have Two Different KF Methods?
  • Volumetric KF usually used for
  • High water content gt 10.
  • Coulometric KF better suited for
  • Low water content 1ppm 10.

4
Advantages of Coulometric KF
  • Fast analysis most analysed in lt 1min
  • Low reagent usage multiple samples can be
    analysed with a single charge of reagent which
    means less waste and less handling of chemicals,
    good news for Health and Safety
  • Simple operation no need for a chemist
  • No reagent calibration

5
Analysis of Petroleum Products
  • ASTM D1533, D4928, D6304, IP 386 and API MPMS
    chapter 10.9, stipulate that Karl Fischer anode
    reagent should be mixed 6040 with xylene
  • Intended to improve sample miscibility and limit
    problems with the generator electrode due to waxy
    deposits and asphaltines dropping out of the
    sample

6
Problems with Petroleum Products
  • Modification of the reagent and addition of oil
    samples affects the resistance and capacitance of
    the titration cell

7
Why Is This a Problem?
  • Coulometers work on fundamental law
  • 10.71 coulombs will titrate 1mg of water
  • When the resistance in the electrolysis cell is
    increased the coulometer cannot operate correctly
    and gives incorrect results, usually high.

8
What Is This Problem?
  • Every coulometer has a maximum titration current
    and count rate
  • If the resistance in the cell is high it cannot
    generate required current
  • But it continues to count as if the required
    current was being generated

THIS PRODUCES AN INCORRECT RESULT
9
What Happens at Present?
  • The method states that the performance of a
    coulometer should be monitored after every 10
    samples by injecting water standard
  • What if result is incorrect?
  • Do you assume that all the previous results are
    correct?
  • Can you afford to make that assumption?

10
A.C.E. Control
  • Automatically
  • Compensated
  • Errors
  • Guarantees that the electrolysis current produced
    and the count rate displayed are always correctly
    synchronized, regardless of changes to the
    electrolysis cell resistance
  • At the heart of the BS W Portable titrator

11
Advantages of ACE control
  • Reduction in reagent consumption
  • More cost effective
  • Better conditions for personnel
  • Most importantly -
    The ACE
    control system

GUARANTEES THE CORRECT RESULT
12
BS W Portable
Offers more flexibility
13
Why Do I Need BS W Portable?
  • Increased flexibility
  • Can analyse a wider range of samples
  • Allows selection of different
  • Volumes
  • Densities
  • Weights
  • Result formats
  • Delay times

14
Target Applications
  • Crude oils
  • Petroleum products
  • Lube oils
  • Brake fluids
  • Hydraulic fluids

15
BS W -Truly Portable
  • Carry case
  • Built in printer
  • Built in battery
  • Car adapter
  • Moisture tight low drift cell
  • All standard

16
Low Drift Cell (LDC)
  • Specifically designed for outdoor use
  • Unique dual lock system
  • Ground glass joint
  • Tight seal screw caps
  • Anti stick joints (no grease required)

17
(No Transcript)
18
Cou-Lo Formula Reagents
  • Single shot bottles
  • Safer to use
  • Safer to store
  • Low cost
  • Faster precondition
  • Reduced downtime
  • Improved sample miscibility / solubility
  • For all coulometers