Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry - PowerPoint PPT Presentation

About This Presentation
Title:

Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry

Description:

Advantages of a Syringe Driven Sample Introduction Accessory ... Moly. Zinc. Thallium. Glass Expansion. www.geicp.com. Niagara II Plus. Selectable sample loop ... – PowerPoint PPT presentation

Number of Views:27
Avg rating:3.0/5.0
Slides: 18
Provided by: jerryd4
Category:

less

Transcript and Presenter's Notes

Title: Advantages of a Syringe Driven Sample Introduction Accessory for ICP Spectrometry


1
Advantages of a Syringe Driven Sample
Introduction Accessory for ICP Spectrometry
  • Jerry Dulude and Scott Bridger
  • Glass Expansion

2
Assist Automated Productivity Accessory
3
Assist components
  • Niagara II Plus Rapid Rinse accessory
  • Sample syringe plus drive
  • Built in switching valve
  • Built in peristaltic pump
  • Optional internal standard syringe plus drive

4
Niagara II Plus
5
Niagara II Plus Flow Diagram
6
Niagara II Plus Flow Diagram-2
7
Screen display of Niagara II Plus software
8
ICP-MS Carryover Comparison
9
Niagara II Plus
  • Selectable sample loop
  • Optional internal standard lines
  • Positive displacement pump
  • Multi-stage valve
  • Sample does not contact pump tubing

10
Time Saving for Niagara II Plus
Standard operation Standard operation Niagara II Plus Niagara II Plus
Cycle time Rinse time Cycle time Rinse time
Probe to sample 5 0 5 5
Uptake delay 15 0 5 5
Stabilization 10 0 10 0
Read 15 0 15 0
Probe to rinse 5 0 0 0
Rinse 30 30 0 0
TOTAL 80 30 35 10
11
Additional Assist Features
  • Precision syringe and drive for sample
  • Optional precision syringe and drive for internal
    standard

12
Syringe vs. peri pump
13
Advantages of Syringe Sample Delivery
  • Computer control of delivery speed
  • Broad selection of sample/IS ratio
  • High precision for scanning systems

14
e.g. 20x dilution with peri pump _at_ 0.5ml/min
uptake
  • 0.38mm ID (orange/green)
  • 1.75mm ID (blue/green)
  • Pump speed2rpm
  • Very Noisy Signal

15
Assist Performance on a Elan9000 ICP-MS


16
Assist Productivity Enhancement
Without Assist Without Assist With Assist With Assist With Assist
Cycle time Rinse time Cycle time Spray chamber rinse Probe/uptake line rinse
Probe to sample 5 0 5 5 0
Uptake delay 15 0 5 5 0
Stabilization 10 0 4 0 4
Read 15 0 15 0 15
Probe to rinse 5 0 0 0 0
Rinse 30 30 0 0 0
TOTAL 80 30 29 10 19
17
Productivity comparison
  • Standard operation
  • Niagara II (w/o FIA)
  • Niagara II Plus
  • Assist
  • 19
  • 27
  • 43
  • 52

of cycle devoted to measuring signal
Write a Comment
User Comments (0)