Title: Discrepancy Between Scheimpflug and Placido-Based Topographies and Correlation with Astigmatism After Scheimpflug Topography-Guided LASIK
1Discrepancy Between Scheimpflug and
Placido-Based Topographies and Correlation with
Astigmatism After Scheimpflug Topography-Guided
LASIK
- Mohamed Abou Shousha, MD,1,2 Osama S. Ibrahim,
MD, PhD 2 - 1.Bascom Palmer Eye Institute, University of
Miami, Miller School of Medicine, FL, USA - 2. Faculty of Medicine, University of Alexandria,
Alexandria , EGYPT
The authors have no financial interest in the
subject matter of this poster.
2Purpose
- Keratometry (K) readings
- K reading axes
- Topographic astigmatism
obtained by
Scheimpflug- based topographer
Placido-based topographer
Then correlating presence of discrepancies to
postoperative astigmatism after Scheimpflug
topography-guided LASIK
3Methods
- Retrospective non-randomized study
- 40 eyes of 20 patients who underwent
- Scheimpflug topography-guided LASIK (Oculink)
- for the correction of primary ametropia.
- Scheimpflug (ALLEGRO Oculyzer) and Placido-based
(ALLEGRO Topolyzer) - topographies were obtained.
- Adjusting the head position to avoid head tilt
was done in all measures.
- K readings, their axes and topographic
astigmatism obtained from the two topographers
were compared - Presence of discrepancy was correlated to
postoperative astigmatism after Scheimpflug
topography-guided LASIK (Oculink).
4Results K readings
- In 50 of eyes (n20), the two topographers
showed more than 0.5 Diopter (D) difference in K
readings, while in 5 of eyes (n2) the
difference was more than 1 D. - In 95 of those eyes (n19), the K readings
obtained from the Scheimpflug topographer were
steeper.
5Difference in Avg K is 1.2 D, Steeper by
Scheimpflug topographer Difference in K
astigmatism is 0.8 D, more astigmatism by
Scheimpflug topographer
Placido-based topographer
Scheimpflug- based topographer
Avg K 42.1 D Astigmatism -3 D Axis 23.8
Avg K 43.3 D Astigmatism -3.8 D Axis 16.8
6Results Topographic astigmatism
- In 15 of eyes (n6), there was more than 0.5 D
difference in topographic astigmatism between the
two topographers. - In 67 of those eyes (n4) it was more in the
Scheimpflug topographer.
7Difference in K astigmatism is 0.8 D more
astigmatism by Scheimpflug topographer
Placido-based topographer
Scheimpflug- based topographer
Avg K 44.7 D Astigmatism -0.4 D Axis 2.8
Avg K 44.9 D Astigmatism -1.2 D Axis 179.6
8Results K readings axes
- In 12.5 of eyes (n5), there was more than 10
degrees difference - In 7.5 of eyes (n3) there was more than 20
degrees difference in K readings axes.
9Difference in Axes is 10.5 Difference in Avg K
is 0.1 D Difference in K astigmatism is 0.2 D
Scheimpflug- based topographer
Placido-based topographer
Avg K 42.2 D Astigmatism -2.8D Axis 178
Avg K 42.3 D Astigmatism -3 D Axis 168.5
10Difference in Axes is 22.3 Difference in Avg K
is 0.1 D Difference in K astigmatism is 0.3 D
Scheimpflug- based topographer
Placido-based topographer
Avg K 42.1 D K astigmatism -0.3 D Axis 164.4
Avg K 42.9 D Astigmatism -0.6 D Axis 6.7
11Results Scheimpflug topography-guided LASIK
(Oculink)
- Postoperative astigmatism of more than 1 D was
present in 30 of eyes (n12). - 75 of those eyes (n9) showed discrepancies
between the two topographers with regard K
readings, their axes and/or topographic
astigmatism.
12Conclusion
- Discrepancies exist between Scheimpflug and
Placido based topographers with regard K
readings, their axes and topographic astigmatism.
- Cases that showed discrepancies had a higher
incidence of postoperative astigmatism after
Scheimpflug topography-guided LASIK.