Title: Liquidborne Particle Counting using Light Obscuration and Light Scattering Methods
1Liquidborne Particle Counting using Light
Obscuration and Light Scattering Methods
2What has been . . .
- Focus has been on injectable liquids
- Possibility to block capillaries and arteries
- Red Blood cells are about 5 µm
- Capillary (5 to 10 µm)
- Large veins (10 to 50 µm)
- Threat of microbial infection
- Allergic reaction to foreign substances
3Contaminants Reported in IV Solutions
Extrinsic
4Definition of Particulate Contaminants
- Unwanted insoluble matter that exist as
randomly-sourced extraneous substances - Excludes homogeneous monotonic materials that
exist as a precipitate or suspension - i.e. colloids, drug degradation or otherwise
derived from a defined source and can be analyzed
by chemical means - Regarded as contamination and adulteration
under Food and Drug Act - the chemical composition of the particulate is
varied, and - would not be declared on the label
- Examples bits of paper fiber, fragments of
filler material, etc
5 Liquid Particle Counting Applications
- Final Product Testing USP lt788gt
- SVP or SVI (Small Volume Parenteral/Injectable)
- Ampoules, Vials
- LVP or LVI (Large Volume Parenteral/Injectable)
- IV (Intravenous) solutions
- Process contamination studies
- Decomposition studies (stability)
- DI or WFI Water
- Precision Cleaning Medical Devices
- Aqueous
- Other Chemicals
6Other Applications for Particle Counting
- Medical Devices
- Cleanliness of manufacturing environment
- Cleanliness of device before implantation
- pacemakers, stents, artificial arteries
- Cleanliness of reclaimed devices
-
- Design of particulate-based medicines
- Inhalation therapies
- Intentional occlusion of blood flow to cancers
- Time-based dosages
- Transdermal absorption
7Global Regulations Particles in Liquids
- USP lt788gt, EP 2.9.19, JP XV, KP, CP
- Primary method
- Optical Particle Counter OPC
- Light Obscuration Counter
- Secondary method
- Optical microscope
- Subjective
- Labor intensive
- Requires more time to process samples
8Proposed USP 787, USP 1787
- USP lt787gt
- Under discussion
- Focused on reducing necessary test volumes due to
concerns of biotechnology manufacturers of cost
for test - Expensive and often very small dose factory
- for example 500 uL pre-filled syringe
9Proposed USP 787, USP 1787
- USP lt787gt
- Primary method ?
- Optical Particle Counter OPC
- Light Obscuration Counter
- Secondary method ?
- Optical microscope
- Subjective
- Labor intensive
- Requires more time to process samples
10Proposed USP 787, USP 1787
- USP lt787gt
- Small sample volume
- - 1 mL ??
- Concerns with variability
- - within production lots
- - in analytical methods
11Optical Particle Counter
- Optical Instrument
- Must move fluid through sensor
- Can quantify particles from 100 nm to 5000 µm
- Counts particles individually (one at a time)
- Cannot tell you composition
- But results are immediate
12Many shapes and sizes
Talc
Alumino-silicate with K and Ti
13Sizing Particles by Microscope
14Challenges of Protein-based Products
- Handling can change material !!!
- Agitation
- Heat and Light
- Contaminates
- Container Vials versus syringes/cartridges
- Shear forces
- Key concern is Aggregation
- Reduction of native form (impacts efficacy)
- Introduction of homogeneous aggregates
- Introduction of heterogeneous aggregates
15Challenges of Protein-based Products
- Transparency of most proteineous entities
- Refractive index
- NIST working on calibration material
- Not contamination but instead a shift from
native form - Not a solution as with small-molecule
therapeutics - Formation of quaternary structures dimers, etc.
- Protein complexes
- Reconstitution of lyophilized product
16Refractive Index
- Key is the ability to distinguish between the
particle and the surrounding fluid - - needs to be great enough
- Optical response is proportional to comparative
index
17Refractive Index
- Key is the ability to distinguish between the
particle and the surrounding fluid - - needs to be great enough
- Optical response is proportional to comparative
index
18Refractive Index
- NIST working on protein-like calibration material
- Probably 2 years away
- Exploring 2 methods of manufacture
- Need thread-like material
- Indices near water
- Stable over reasonable period
19II. Sample Handling
20Settling/Agitation
- Entrained gas
- - sonication probably not ideal with protein
structures - - light vacuum seems to work OK
- Settling
- Limits collection of particles
- - especially of greater mass
- - dependent on time and viscosity
- - improved collection with slanted containers
21Consistency of sample characteristics
- Temperature
- Settling
- Probe position
22Issues with Sampling Particles in Liquids
- Sampling Errors Account for most problems
- Accidental Contamination or Alteration by
Technician
- 1. System Preparation
- Initial Cleanliness
- Calibration
- 3. Sample Handling
- Aggregation
- Settling
- Cavitation
2. Sample Preparation Contamination -
Particles - Gases - Liquids
23Particle Counter vs. Microscope
- Sizing of Particles
- Microscope Subjective size measurement
- Electronic consistent for specific particle
type - Particle orientation and movement
- Electronic large, long particles can be
mis-read - Calibration
- How to establish accurate sizing and counting
- Consistency (repeatability)
24Sizing Particles by Microscope
25HIAC Liquid Particle Counters
- Example HIAC 9703
- The industry standard liquid particle counter
since 1997 - USP lt788gt was written specifically around HIAC
technology - Every major manufacturer of particle calibration
standards uses the HIAC 9703 -
26HIAC Liquid Particle Counters
- Example HIAC 9703
- Improved sample mounting method for small vials
or containers - Detection of usual conditions such as bubbles or
contamination - Proven syringe sampler
- SVI and LVI sampling
- Addresses non-compendial applications, e.g. RD
and other low frequency, small sample volume
applications
- Reproducibility
- Repeatability
27Detection Ranges
1µm
0.1µm
10µm
100µm
Light Obscuration
Light Scattering
nm
28Light Obscuration
- Light Obscuration Sensors and system
- also known as Light Extinction
- also known as Light Blocking
29Principles Light Obscuration
30Principles Light Obscuration
31Principles Light Obscuration
32Particle Detection
- Like an air particle counter, the larger the
particle, the larger the pulse that is created
33Principles Light Scattering
Detector Output
Detector
Light Trap
Laser Diode
Mirror
34Principles Light Scattering
Detector
Particle
Light Trap
Laser Diode
Mirror
35Advantages Light Scattering
- Good sensitivity from 0,1µm to 50µm
- Wide range of sample concentration
- Good rejection of false counts
- High sample flow rates
36Disadvantages Light Scattering
- More complicated construction higher cost
- Characteristics of particle surface (shiny,
color) affect response
37Effect of colors and surfaces on Light Scattering
38Talc
Alumino-silicate with K and Ti
39Sizing Particles by Microscope
40General Comments on Liquid Counting
- Particle Counters Report Size
- But measure an Optical Response
- Difference in reported size compared to
microscope - Calibration Relates the Optical Signal to Size
- Difference between calibration material
characteristics and real world particles
41General Comments on Liquid Counting
- Particle Counters Report Size
- But measure an Optical Response
- Differences in reported size compared to
microscope - Calibration Relates the Optical Signal to Size
- Difference between calibration material
characteristics and real world particles
42LO results versus LS results
- Light Obscuration LO
- Good immunity to variations of surface and
morphology - Very stable
- Limit of quantitation circa 1.2 1.3 microns
- Light Scattering LS
- Results affected by surface characteristics and
coloring - Good stability
- Limit of quantitation sub-micron
- Problem can occur in the attempt to correlate
results of these two methods above 1 micron
43Sampling Particles in Liquids
- Good, consistent results depend on
- Well-trained operators
- Careful technical methodology
- Constant effort needed
- Review ways to avoid or reduce errors due to
sampling methods - Usually cause false positives
44System Preparation
- 2-step Verification - optional
- Run 2 test solutions
- Blank (particle-free water)
- Aqueous solution containing known counts
- at 10 µm or 15 µm
- In the range of 1000 to 3000 counts per mL
- Frequency based on risk analysis
- Each morning
- Shift change
- Change of operator
- Other interval
45System Preparation
- Check for bubbles in sample lines and syringe
- Affects flow rate and calibration
- Verify correct calibration curve installed
- Different flow rates for same sensor
- Change of syringe size
- Some companies have multiple sensors
- Verify calibration is current
- Sensor resolution and response curve
- IST tests conducted USP, JP
46System Preparation
- Instrument Standardization Tests IST
- Five tests of system
- Volume accuracy
- Flow rate accuracy
- Calibration of sensor
- Resolution
- Count accuracy
- Required by USP and JP but not EP
- USP lt788gt 31
- at intervals of not more than six months.
- JP lt24gt
- at least once a year.
47Sample Preparation
- Removing residue of previous sample
- If previous sample contained particles, may be
good plan to run a particle-free blank before
the next sample - Use liquid that is compatible with sample fluid
- An aqueous blank could trigger false counts in an
oil-based sample by causing immiscible droplets - Potential residue from previous sample can cause
change of counts - Data from first run of series is often discarded
48Sample Preparation
- Contamination
- Particles
- Gases
- Liquids
49Sample Preparation
- True Particle Sources
- Residue on glassware and equipment
- Ambient air
- Paper dust
- Glass
- Diluent
- Residue from previous sample
- Colloidal suspensions
- False Particle Sources
- RF signals or other electronic interference
- Bubbles from entrained gases
50Sample Preparation
- Work in controlled environment
- Use particle-free gloves
- Let water flow for 200 mL or more after opening a
valve / tap - Opening / closing valve generates particles
- Wipe the outside of containers before sampling
- Particles on surface of vials or ampoules
- Open vials and ampoules away from beaker or flask
- Particles from activity can fall into open
container - Wash outside of containers to reduce potential
particle source
51Degassing sample
- Three common methods
- Allowing to stand in ambient air
- Risk of large particles settling
- Sonification ultrasound
- 80 to 120 watts USP lt788gt
- 30 seconds USP lt788gt
- Vacuum
- Bell jar or dessicator
- 0.6 0.8 atmospheres for 2 to 10 minutes
52Sample Preparation
- Possibility of decreasing true particle counts
- Settling
- Lack of agitation
- Position of probe in sample vessel
- Remaining material from previous sample run
- Sample with lower counts
- Blank
53Sample Handling
- Aggregation
- Settling
- Entrained gases
54Sample Handling
- Aggregation
- Samples held over time or at extremes of
temperature can develop aggregates of smaller
particles - Exposure to light can also trigger this reaction
- Sub-micron particles can thus add to particle
counts above 10 µm - Suggestion
- Profile counts under 10 µm e.g. at 2 or 5 µm in
addition to standard count sizes at 10 and 25 µm
55Sample Handling
- Settling
- Undercounting caused by
- Gravitational settling
- Failure to suspend particulate matter adhering to
walls or stopper of container - Standards have recommended procedure for
agitation - Multiple inversions of original container before
decanting - Constant motion of liquid during sampling
- Gently stir the contents of the container by
hand-swirling or by mechanical means USP lt788gt
56Optical Particle Counter
- Instrument sensor
- Must move fluid through sensor
- Can quantify particles from 100 nm to 5000 µm
- Counts particles individually (one at a time)
- Cannot tell you composition
- But results are immediate
57Questions?
58Patent Pending
59New Hardware 9703
- Key features
- Auto stop for sensor elevator arm
- Small vial holding clamp
- Sample probe with reduced dead volume
- Back-flush and forward flush from front panel
- Supports MC-05 sensor (0.5 micron sensitivity)
60Software PharmSpec 3
- Key new features
- Compendial test support continues
- USP, EP, JP, KP looks same as previous PharmSpec
versions - Uses same log on as for Windows
- Improved Report format
- Improved Error Detection and Display
61Still the HIAC 9703 you know and trust only
better!
- Syringes
- 1 ml, 10 ml, 25 ml
- Flow rate settings
- 10 to 100 ml
- Sensors
- MC-05 is added
- Sampling Probes
- added shorter small-bore probe
- Instrument size / shape
- 50 of instruments are placed in laminar flow
cabinets. - Smooth, curved surfaces create less turbulence
for the air flow
62Easier, Faster and Confident Sampling
- Use less sample, save valuable time - protect
your investment - Small vial clamp ensures that sample does not
spill during testing - Probe needle safety switch prevents probe damage
- New small needle probe with industrys smallest
tare volume
63Easier, Faster and Confident Sampling
- Reduce uncertainty from data anomalies
- Bubble alarm notification
- Contamination alarm notification
- Advanced notification when service or
calibration is due - Invalid configuration notification
64HIAC just got easier!
- Less time needed for clean-up
- Automated flushing and cleaning routines
- Push a button, walk away and return to a clean
sensor - Export your data with ease
- Select one, several or all of your historical
data records with our batch export utility - Select PDF, Word, Excel, or text files
- Save time with electronic signature
- Stricter interpretation of 21CFR Part 11
electronic signature process. WITHOUT more
manual inputs - Remembers user Login ID
65HIAC 9703 Flexibility
- Interchangeable sampling probes, syringes, and
sensors - Ensure you have one instrument to manage all
applications - Now supports MC05 sub-micron sensor
- Change configuration with no impact to instrument
validation - Customized reporting
- Customize the number of reviewers and approvers
for compendial test reports - Add company logo, user-defined descriptors
- Customized test recipes
- Procedure Builder enables the development of
unique test recipes for your application - Enables testing to marketing license-specific
applications
66New Sampling Probe
- 3 probes available
- Tare
- ¼ ID 1.2 ml tare volume
- 1/16 ID 0.172 ml tare volume
- New small / short probe
- 1/16 ID 0.09 ml tare volume
- Tests can be performed on 1 ml of product!
67Small Vial Clamp
- Small Vial Clamp Platform
- Can be retrofitted
- Can be removed
- Ease of Use
- Use one hand to compress lever arms
- Use other hand to place sample in central
location - Decompress hand
- Clamp auto-centers and holds sample container
68Docking Module
- Docking Module
- Enables removal/disabling of the stir bar
mechanism - Enables field installation of small vial clamp
outside of the laminar flow cabinet - Avoid potential of re-qualification that can
occur if instrument is moved - Future developments to expand 9703 applications
69Sampling Safety Switch
- Sampling safety switch
- Ensures the sampling probe does not crash (and
bend or break) into the docking module - Ensures the probe does crash into or tip the
sample container
70Liquidborne Particle Counting using Light
Obscuration and Light Scattering Methods