Liquidborne Particle Counting using Light Obscuration and Light Scattering Methods PowerPoint PPT Presentation

presentation player overlay
1 / 67
About This Presentation
Transcript and Presenter's Notes

Title: Liquidborne Particle Counting using Light Obscuration and Light Scattering Methods


1
Liquidborne Particle Counting using Light
Obscuration and Light Scattering Methods
2
What has been . . .
  • Focus has been on injectable liquids
  • Possibility to block capillaries and arteries
  • Red Blood cells are about 5 µm
  • Capillary (5 to 10 µm)
  • Large veins (10 to 50 µm)
  • Threat of microbial infection
  • Allergic reaction to foreign substances

3
Contaminants Reported in IV Solutions
Extrinsic
4
Definition of Particulate Contaminants
  • Unwanted insoluble matter that exist as
    randomly-sourced extraneous substances
  • Excludes homogeneous monotonic materials that
    exist as a precipitate or suspension
  • i.e. colloids, drug degradation or otherwise
    derived from a defined source and can be analyzed
    by chemical means
  • Regarded as contamination and adulteration
    under Food and Drug Act
  • the chemical composition of the particulate is
    varied, and
  • would not be declared on the label
  • Examples bits of paper fiber, fragments of
    filler material, etc

5
Liquid Particle Counting Applications
  • Final Product Testing USP lt788gt
  • SVP or SVI (Small Volume Parenteral/Injectable)
  • Ampoules, Vials
  • LVP or LVI (Large Volume Parenteral/Injectable)
  • IV (Intravenous) solutions
  • Process contamination studies
  • Decomposition studies (stability)
  • DI or WFI Water
  • Precision Cleaning Medical Devices
  • Aqueous
  • Other Chemicals

6
Other Applications for Particle Counting
  • Medical Devices
  • Cleanliness of manufacturing environment
  • Cleanliness of device before implantation
  • pacemakers, stents, artificial arteries
  • Cleanliness of reclaimed devices
  • Design of particulate-based medicines
  • Inhalation therapies
  • Intentional occlusion of blood flow to cancers
  • Time-based dosages
  • Transdermal absorption

7
Global Regulations Particles in Liquids
  • USP lt788gt, EP 2.9.19, JP XV, KP, CP
  • Primary method
  • Optical Particle Counter OPC
  • Light Obscuration Counter
  • Secondary method
  • Optical microscope
  • Subjective
  • Labor intensive
  • Requires more time to process samples

8
Proposed USP 787, USP 1787
  • USP lt787gt
  • Under discussion
  • Focused on reducing necessary test volumes due to
    concerns of biotechnology manufacturers of cost
    for test
  • Expensive and often very small dose factory
  • for example 500 uL pre-filled syringe

9
Proposed USP 787, USP 1787
  • USP lt787gt
  • Primary method ?
  • Optical Particle Counter OPC
  • Light Obscuration Counter
  • Secondary method ?
  • Optical microscope
  • Subjective
  • Labor intensive
  • Requires more time to process samples

10
Proposed USP 787, USP 1787
  • USP lt787gt
  • Small sample volume
  • - 1 mL ??
  • Concerns with variability
  • - within production lots
  • - in analytical methods

11
Optical Particle Counter
  • Optical Instrument
  • Must move fluid through sensor
  • Can quantify particles from 100 nm to 5000 µm
  • Counts particles individually (one at a time)
  • Cannot tell you composition
  • But results are immediate

12
Many shapes and sizes
Talc
Alumino-silicate with K and Ti
13
Sizing Particles by Microscope
14
Challenges of Protein-based Products
  • Handling can change material !!!
  • Agitation
  • Heat and Light
  • Contaminates
  • Container Vials versus syringes/cartridges
  • Shear forces
  • Key concern is Aggregation
  • Reduction of native form (impacts efficacy)
  • Introduction of homogeneous aggregates
  • Introduction of heterogeneous aggregates

15
Challenges of Protein-based Products
  • Transparency of most proteineous entities
  • Refractive index
  • NIST working on calibration material
  • Not contamination but instead a shift from
    native form
  • Not a solution as with small-molecule
    therapeutics
  • Formation of quaternary structures dimers, etc.
  • Protein complexes
  • Reconstitution of lyophilized product

16
Refractive Index
  • Key is the ability to distinguish between the
    particle and the surrounding fluid
  • - needs to be great enough
  • Optical response is proportional to comparative
    index

17
Refractive Index
  • Key is the ability to distinguish between the
    particle and the surrounding fluid
  • - needs to be great enough
  • Optical response is proportional to comparative
    index

18
Refractive Index
  • NIST working on protein-like calibration material
  • Probably 2 years away
  • Exploring 2 methods of manufacture
  • Need thread-like material
  • Indices near water
  • Stable over reasonable period

19
II. Sample Handling
20
Settling/Agitation
  • Entrained gas
  • - sonication probably not ideal with protein
    structures
  • - light vacuum seems to work OK
  • Settling
  • Limits collection of particles
  • - especially of greater mass
  • - dependent on time and viscosity
  • - improved collection with slanted containers

21
Consistency of sample characteristics
  • Temperature
  • Settling
  • Probe position

22
Issues with Sampling Particles in Liquids
  • Sampling Errors Account for most problems
  • Accidental Contamination or Alteration by
    Technician
  • 1. System Preparation
  • Initial Cleanliness
  • Calibration
  • 3. Sample Handling
  • Aggregation
  • Settling
  • Cavitation

2. Sample Preparation Contamination -
Particles - Gases - Liquids
23
Particle Counter vs. Microscope
  • Sizing of Particles
  • Microscope Subjective size measurement
  • Electronic consistent for specific particle
    type
  • Particle orientation and movement
  • Electronic large, long particles can be
    mis-read
  • Calibration
  • How to establish accurate sizing and counting
  • Consistency (repeatability)

24
Sizing Particles by Microscope
25
HIAC Liquid Particle Counters
  • Example HIAC 9703
  • The industry standard liquid particle counter
    since 1997
  • USP lt788gt was written specifically around HIAC
    technology
  • Every major manufacturer of particle calibration
    standards uses the HIAC 9703

26
HIAC Liquid Particle Counters
  • Example HIAC 9703
  • Improved sample mounting method for small vials
    or containers
  • Detection of usual conditions such as bubbles or
    contamination
  • Proven syringe sampler
  • SVI and LVI sampling
  • Addresses non-compendial applications, e.g. RD
    and other low frequency, small sample volume
    applications
  • Reproducibility
  • Repeatability

27
Detection Ranges
1µm
0.1µm
10µm
100µm
Light Obscuration
Light Scattering
nm
28
Light Obscuration
  • Light Obscuration Sensors and system
  • also known as Light Extinction
  • also known as Light Blocking

29
Principles Light Obscuration
30
Principles Light Obscuration
31
Principles Light Obscuration
32
Particle Detection
  • Like an air particle counter, the larger the
    particle, the larger the pulse that is created

33
Principles Light Scattering
Detector Output
Detector
Light Trap
Laser Diode
Mirror
34
Principles Light Scattering
Detector
Particle
Light Trap
Laser Diode
Mirror
35
Advantages Light Scattering
  • Good sensitivity from 0,1µm to 50µm
  • Wide range of sample concentration
  • Good rejection of false counts
  • High sample flow rates

36
Disadvantages Light Scattering
  • More complicated construction higher cost
  • Characteristics of particle surface (shiny,
    color) affect response

37
Effect of colors and surfaces on Light Scattering
38
Talc
Alumino-silicate with K and Ti
39
Sizing Particles by Microscope
40
General Comments on Liquid Counting
  • Particle Counters Report Size
  • But measure an Optical Response
  • Difference in reported size compared to
    microscope
  • Calibration Relates the Optical Signal to Size
  • Difference between calibration material
    characteristics and real world particles

41
General Comments on Liquid Counting
  • Particle Counters Report Size
  • But measure an Optical Response
  • Differences in reported size compared to
    microscope
  • Calibration Relates the Optical Signal to Size
  • Difference between calibration material
    characteristics and real world particles

42
LO results versus LS results
  • Light Obscuration LO
  • Good immunity to variations of surface and
    morphology
  • Very stable
  • Limit of quantitation circa 1.2 1.3 microns
  • Light Scattering LS
  • Results affected by surface characteristics and
    coloring
  • Good stability
  • Limit of quantitation sub-micron
  • Problem can occur in the attempt to correlate
    results of these two methods above 1 micron

43
Sampling Particles in Liquids
  • Good, consistent results depend on
  • Well-trained operators
  • Careful technical methodology
  • Constant effort needed
  • Review ways to avoid or reduce errors due to
    sampling methods
  • Usually cause false positives

44
System Preparation
  • 2-step Verification - optional
  • Run 2 test solutions
  • Blank (particle-free water)
  • Aqueous solution containing known counts
  • at 10 µm or 15 µm
  • In the range of 1000 to 3000 counts per mL
  • Frequency based on risk analysis
  • Each morning
  • Shift change
  • Change of operator
  • Other interval

45
System Preparation
  • Check for bubbles in sample lines and syringe
  • Affects flow rate and calibration
  • Verify correct calibration curve installed
  • Different flow rates for same sensor
  • Change of syringe size
  • Some companies have multiple sensors
  • Verify calibration is current
  • Sensor resolution and response curve
  • IST tests conducted USP, JP

46
System Preparation
  • Instrument Standardization Tests IST
  • Five tests of system
  • Volume accuracy
  • Flow rate accuracy
  • Calibration of sensor
  • Resolution
  • Count accuracy
  • Required by USP and JP but not EP
  • USP lt788gt 31
  • at intervals of not more than six months.
  • JP lt24gt
  • at least once a year.

47
Sample Preparation
  • Removing residue of previous sample
  • If previous sample contained particles, may be
    good plan to run a particle-free blank before
    the next sample
  • Use liquid that is compatible with sample fluid
  • An aqueous blank could trigger false counts in an
    oil-based sample by causing immiscible droplets
  • Potential residue from previous sample can cause
    change of counts
  • Data from first run of series is often discarded

48
Sample Preparation
  • Contamination
  • Particles
  • Gases
  • Liquids

49
Sample Preparation
  • True Particle Sources
  • Residue on glassware and equipment
  • Ambient air
  • Paper dust
  • Glass
  • Diluent
  • Residue from previous sample
  • Colloidal suspensions
  • False Particle Sources
  • RF signals or other electronic interference
  • Bubbles from entrained gases

50
Sample Preparation
  • Work in controlled environment
  • Use particle-free gloves
  • Let water flow for 200 mL or more after opening a
    valve / tap
  • Opening / closing valve generates particles
  • Wipe the outside of containers before sampling
  • Particles on surface of vials or ampoules
  • Open vials and ampoules away from beaker or flask
  • Particles from activity can fall into open
    container
  • Wash outside of containers to reduce potential
    particle source

51
Degassing sample
  • Three common methods
  • Allowing to stand in ambient air
  • Risk of large particles settling
  • Sonification ultrasound
  • 80 to 120 watts USP lt788gt
  • 30 seconds USP lt788gt
  • Vacuum
  • Bell jar or dessicator
  • 0.6 0.8 atmospheres for 2 to 10 minutes

52
Sample Preparation
  • Possibility of decreasing true particle counts
  • Settling
  • Lack of agitation
  • Position of probe in sample vessel
  • Remaining material from previous sample run
  • Sample with lower counts
  • Blank

53
Sample Handling
  • Aggregation
  • Settling
  • Entrained gases

54
Sample Handling
  • Aggregation
  • Samples held over time or at extremes of
    temperature can develop aggregates of smaller
    particles
  • Exposure to light can also trigger this reaction
  • Sub-micron particles can thus add to particle
    counts above 10 µm
  • Suggestion
  • Profile counts under 10 µm e.g. at 2 or 5 µm in
    addition to standard count sizes at 10 and 25 µm

55
Sample Handling
  • Settling
  • Undercounting caused by
  • Gravitational settling
  • Failure to suspend particulate matter adhering to
    walls or stopper of container
  • Standards have recommended procedure for
    agitation
  • Multiple inversions of original container before
    decanting
  • Constant motion of liquid during sampling
  • Gently stir the contents of the container by
    hand-swirling or by mechanical means USP lt788gt

56
Optical Particle Counter
  • Instrument sensor
  • Must move fluid through sensor
  • Can quantify particles from 100 nm to 5000 µm
  • Counts particles individually (one at a time)
  • Cannot tell you composition
  • But results are immediate

57
Questions?
58
Patent Pending
59
New Hardware 9703
  • Key features
  • Auto stop for sensor elevator arm
  • Small vial holding clamp
  • Sample probe with reduced dead volume
  • Back-flush and forward flush from front panel
  • Supports MC-05 sensor (0.5 micron sensitivity)

60
Software PharmSpec 3
  • Key new features
  • Compendial test support continues
  • USP, EP, JP, KP looks same as previous PharmSpec
    versions
  • Uses same log on as for Windows
  • Improved Report format
  • Improved Error Detection and Display

61
Still the HIAC 9703 you know and trust only
better!
  • Syringes
  • 1 ml, 10 ml, 25 ml
  • Flow rate settings
  • 10 to 100 ml
  • Sensors
  • MC-05 is added
  • Sampling Probes
  • added shorter small-bore probe
  • Instrument size / shape
  • 50 of instruments are placed in laminar flow
    cabinets.
  • Smooth, curved surfaces create less turbulence
    for the air flow

62
Easier, Faster and Confident Sampling
  • Use less sample, save valuable time - protect
    your investment
  • Small vial clamp ensures that sample does not
    spill during testing
  • Probe needle safety switch prevents probe damage
  • New small needle probe with industrys smallest
    tare volume

63
Easier, Faster and Confident Sampling
  • Reduce uncertainty from data anomalies
  • Bubble alarm notification
  • Contamination alarm notification
  • Advanced notification when service or
    calibration is due
  • Invalid configuration notification

64
HIAC just got easier!
  • Less time needed for clean-up
  • Automated flushing and cleaning routines
  • Push a button, walk away and return to a clean
    sensor
  • Export your data with ease
  • Select one, several or all of your historical
    data records with our batch export utility
  • Select PDF, Word, Excel, or text files
  • Save time with electronic signature
  • Stricter interpretation of 21CFR Part 11
    electronic signature process. WITHOUT more
    manual inputs
  • Remembers user Login ID

65
HIAC 9703 Flexibility
  • Interchangeable sampling probes, syringes, and
    sensors
  • Ensure you have one instrument to manage all
    applications
  • Now supports MC05 sub-micron sensor
  • Change configuration with no impact to instrument
    validation
  • Customized reporting
  • Customize the number of reviewers and approvers
    for compendial test reports
  • Add company logo, user-defined descriptors
  • Customized test recipes
  • Procedure Builder enables the development of
    unique test recipes for your application
  • Enables testing to marketing license-specific
    applications

66
New Sampling Probe
  • 3 probes available
  • Tare
  • ¼ ID 1.2 ml tare volume
  • 1/16 ID 0.172 ml tare volume
  • New small / short probe
  • 1/16 ID 0.09 ml tare volume
  • Tests can be performed on 1 ml of product!

67
Small Vial Clamp
  • Small Vial Clamp Platform
  • Can be retrofitted
  • Can be removed
  • Ease of Use
  • Use one hand to compress lever arms
  • Use other hand to place sample in central
    location
  • Decompress hand
  • Clamp auto-centers and holds sample container

68
Docking Module
  • Docking Module
  • Enables removal/disabling of the stir bar
    mechanism
  • Enables field installation of small vial clamp
    outside of the laminar flow cabinet
  • Avoid potential of re-qualification that can
    occur if instrument is moved
  • Future developments to expand 9703 applications

69
Sampling Safety Switch
  • Sampling safety switch
  • Ensures the sampling probe does not crash (and
    bend or break) into the docking module
  • Ensures the probe does crash into or tip the
    sample container

70
Liquidborne Particle Counting using Light
Obscuration and Light Scattering Methods
Write a Comment
User Comments (0)
About PowerShow.com