Radiation Protection in Radiotherapy - PowerPoint PPT Presentation

1 / 71
About This Presentation
Title:

Radiation Protection in Radiotherapy

Description:

IAEA Training Material on Radiation Protection in Radiotherapy Radiation Protection in Radiotherapy Part 10 Good Practice in EBT Lecture 1 (cont.): Equipment design – PowerPoint PPT presentation

Number of Views:844
Avg rating:3.0/5.0
Slides: 72
Provided by: iaea
Category:

less

Transcript and Presenter's Notes

Title: Radiation Protection in Radiotherapy


1
Radiation Protection inRadiotherapy
IAEA Training Material on Radiation Protection in
Radiotherapy
  • Part 10
  • Good Practice in EBT
  • Lecture 1 (cont.) Equipment design

2
2. Features of safe design in practice
  • A General considerations
  • B Kilovoltage radiation units
  • C Telecurie units
  • D Megavoltage units
  • E Other irradiation units

3
A. General Safety Requirements
  • Radiation Protection Measures include
  • Protection of the patient during treatment
  • Equipment shielding
  • Collimation system
  • Patient comfort and control
  • Protection of others
  • Room shielding (this was covered in part 7)

4
Equipment shielding
  • Part of dose reduction strategy for patients
  • Dose to patient other than target due to scatter
    and leakage

5
Equipment shielding
  • X Ray equipment - only needed when machine is on
  • protects the patient during treatment
  • Telecobalt units - shielding needed all the time
  • protects patient and staff during set-up

General design limit - leakage should be
less than 0.1 of the primary radiation
6
Testing of shielding integrity of a linac head
using film
About 2t of lead
7
Collimation
  • Creates outlines of the radiation field which
    should conform to the target
  • Can be done by a variety of different measures
    depending on the treatment unit type
  • Always includes some leakage through the
    collimation - typically lt2 of the primary beam

8
Collimation
Customized blocks or prefabricated blocks in
geometric shapes
  • Aim to limit field to the target only

9
Collimation
  • Applicators
  • electron beams
  • superficial beams
  • Movable jaws
  • Lead blocks
  • fixed shapes
  • customized
  • Multileaf collimator

10
Custom shielding may reduce the dose to critical
organs
  • e.g. scrotal shields to reduce dose to scrotum
    due to scattered radiation

11
Patient comfort and control
  • The best collimation does not help if the patient
    is not stable
  • need good immobilization devices
  • need to put patient in a reasonably comfortable
    position (this is often difficult with very sick
    patients)
  • need to make them feel comfortable

12
Immobilization/set-up devices
  • There are innumerable systems - many of them home
    built and designed
  • A good mould room is essential - they are
    responsible for both,
  • immobilization and
  • block making

13
Immobilization/set-up devices
  • Head rests

14
Head and Neck Immobilization
Head rests to fit
Prone head rest
All MedTec
15
Lateral Head position
16
Immobilization/set-up devices
  • The more accuracy is required, the more effort
    one must make e.g.
  • Stereotactic head frame with repositioning
    accuracy better than 2mm

17
Immobilization/set-up devices
  • Immobilization shells for head
  • Vacuum bag for body immobilization

18
Various body immobilisation devices
Body fix with external markers for set-up
All MedTec
19
Belly board for prone position
  • Allows belly to move into space
  • Some of the bowel can be moved out of the field

20
Vacuum bags
Customized for every patient
All MedTec
21
Immobilization/set-up devices
  • Board for set-up of breast patients

Arm rest to get arm out of the treatment field
Head rest
Slope to straighten sternum in order to minimize
lung dose
Leg rest
22
sometimes movement is difficult to control...
  • e.g. rectal and bladder filling in prostate
    treatment
  • determine location of the prostate prior to each
    treatment fraction using ultrasound

23
sometimes movement is difficult to control...
  • e.g. lung motion due to breathing
  • determine motion and gate radiation beam

External markers on the patient which can be
tracked by a video system
24
Low cost solutions
  • Ask patients to
  • hold still
  • have reproducible bladder filling (e.g. always
    full or always empty)
  • provide dietary advise
  • breath shallow
  • Make patients feel comfortable and secure

25
A note on intercom systems
  • Need to be able to see the patient - is he/she
    comfortable? Is she/he moving?
  • Need to be able to talk to the patient
  • Need to be able to hear if the patient is in
    distress

26
B. Kilovoltage Equipment (10 - 150 kV)
  • Dose rate is approximately proportional to the
    nth power of the accelerating potential as kVn
    where 2 lt n lt 3
  • Dose rate is approximately proportional to
    current (mA)
  • Therefore important that kV and mA are stable.
  • It is obviously important that the timer is
    accurate and stable

27
Kilovoltage Equipment (10 150 kV)
  • Dose control is achieved by a dual timer system
    as it is usually not practical to use a
    transmission ionization chamber
  • Interlocks should be present to prevent incorrect
    combinations of kV, mA, and filtration

28
Quick Question
  • What are the fluctuations of the mains voltage in
    your hospital? What would be the consequence in
    dose if these would not be filtered out before
    generating the high voltage for the X Ray tube?

29
Answer
  • A /- 10 voltage variation is not uncommon due
    to loading of the net at different times of the
    day or heavy occasional uses on the same mains
    (e.g. a lift)
  • This translates into 40 dose variation which is
    unacceptable
  • Mains stabilization is a MUST

30
Kilovoltage Equipment (10 - 150 kV)
  • Leakage from the tube housing, the Air Kerma Rate
    (AKR) shall not exceed
  • 10 mGy h-1 at 1 metre from focus
  • 300 mGy h-1 at 5 cm from housing or accessory
    equipment
  • if the tube is designed to operate in the range
    10 - 50 kV then a special housing is required
    with a maximum leakage of 1 mGy h-1
  • Testing for hot spots should be carried out using
    film-wrap techniques

31
Patient shielding
  • May be done on the skin using lead sheets cut
    into customized shapes
  • Special shields may be used - e.g. eye shields

32
Kilovoltage Equipment (150 - 400 kV)Orthovoltage
irradiation units
  • It is practical to use a transmission ionization
    chamber with this equipment and the primary dose
    control system should be an integrating
    dosemeter.
  • The backup (secondary) dose control system can be
    either an independent integrating dosemeter or a
    timer

33
Kilovoltage Equipment (150 - 400 kV)
  • Leakage from the tube housing, the Air Kerma Rate
    (AKR) shall not exceed
  • 10 mGy h-1 at 1 metre from focus
  • 300 mGy h-1 at 5 cm from housing or accessory
    equipment (including the beam collimation system
    such as cones)
  • Testing for hot spots should be carried out using
    film-wrap techniques

34
C. Telecurie units
  • 137-Cs or more importantly 60-Co
  • High activity in treatment head
  • Termination of exposure is usually by dual
    independent timers

35
Timers
  • Need two completely independent timers
  • One should count time up, one down

36
Gamma-ray equipment
  • The source should be sealed such that the
    container can withstand temperatures likely to be
    obtained in building fires.
  • Wipe tests should be carried out initially at
    installation and at regular intervals to check
    for surface contamination. This test need not be
    carried out directly on the source surface and
    can be carried out on a surface which comes into
    contact with the source during normal operation
    of the equipment.

37
Cobalt unit designs
38
Gamma-ray equipment
  • At commissioning, cross-sectional drawings of the
    head should be examined to identify possible
    locations where radiation leakage could be a
    problem.
  • Film wrap techniques can be used to identify
    positions of hot spots.
  • Accurate integrated ionization chamber readings
    should be made at the location of any hot spots
    and also in a regular pattern around the head.

39
Gamma-ray equipment
  • Leakage from the head with the source in the Off
    position the Air Kerma Rate (AKR) shall not
    exceed
  • 10 ?Gy h-1 at 1 metre from source
  • 200 ?Gy h-1 at 5 cm from housing or accessory
    equipment

40
Gamma-ray equipment
  • Leakage from the head with the source in the On
    position the Air Kerma Rate (AKR) shall not
    exceed
  • 10 mGy h-1 at 1 metre from source or
  • 0.1 of the useful beam AKR
  • whichever is the greater

41
Gamma-ray equipment
  • The beam control mechanism shall be of the fail
    to safety type and will return to the Off
    position in the event of
  • end of normal exposure
  • any breakdown situation
  • interruption of the force holding the beam
    control mechanism in the On position, for example
    failure of electrical power or compressed air
    supply

42
Gamma-ray equipment
Mechanical source position indicator
  • In case of failure of the automatic source return
    section of the beam control mechanism, it shall
    be possible to interrupt the exposure by other
    means, for example, a manual return system
  • It shall be possible to unload or repair the
    treatment head without exceeding the dose limit
    for occupational exposure recommended by
    regulation

43
Gamma-ray equipment
  • Collimation, patient immobilisation and blocking
    as described in first section of part 10 and the
    case of linacs.
  • Two particularities
  • No commercial MLC available (but several home
    built systems)
  • Due to large source size and wide penumbra
    penumbra trimmers (collimation close to the
    patient can be employed)

44
Specific design for Co units
  • Penumbra trimmers - collimation close to patient
    reduces penumbra width

45
Beam stopper
  • Metal disk at the exit side
  • reduces primary beam shielding requirements
  • may make set-up of patients more cumbersome

46
D Megavoltage units
  • Electron linear accelerators - linacs
  • Capable of X Ray (4 to 25MV) and electron (4 to
    25MeV) irradiation

47
Linacs
  • Radiation exposure is usually controlled by two
    independent integrating transmission ionization
    chamber systems.
  • One of these is designated as the primary system
    and should terminate the exposure at the correct
    number of monitor units
  • The other system is termed the secondary system
    and is usually set to terminate the exposure
    after an additional dose, typically set around
    0.25 Gy
  • Most modern accelerators also have a timer which
    will terminate the exposure if both ionization
    chamber systems fail

48
Linacs
  • Modern accelerators have a lot of treatment
    options as discussed in part 6, for example
  • X Rays or electrons (dual mode)
  • multiple energies
  • 2 X Ray energies
  • 5 or more electron energies
  • wedges
  • 3 or more fixed wedges
  • auto-wedge
  • dynamic wedge

49
Linacs
  • With such a large number of possible settings it
    is essential that interlocks be provided to
    prevent inappropriate combinations from being
    selected
  • It is also essential that the control console
    provide a clear indication of what functions have
    been set

50
A linac control example
Active selection
Parameter display
Varian
51
Linacs
  • Verification systems
  • All accelerator manufacturers now produce
    computer controlled verification systems which
    provide an additional check that the settings on
    the accelerator console are correct for
  • proper accelerator function and
  • correspond exactly with the parameters determined
    for the individual patient during the treatment
    planning process

52
Linacs - a note on MLCs
  • X Ray Collimators may be
  • rectangular (conventional)
  • Multi-Leaf collimators (MLC)
  • the transmission through the collimators should
    be less than 2 of the primary (open) beam
  • The transmission between the leaves should be
    checked to ensure that it is less than the
    manufacturers specification - this can be done
    using radiographic film

53
Linacs - electrons
  • Electron applicators, these may be
  • open sided for modern accelerators using double
    scattering foils or scanned beams
  • enclosed for older accelerators using single
    scattering foils
  • should be checked for leakage
  • adjacent to the open beam
  • on the sides of the applicators

Cut-out at the end of the applicator
54
Electron collimation
  • Done at the end of the applicator using
    customized cut-outs

Pour LMA around it
Cut-out foam where field should be
55
IEC 601.2
  • Limit values at different locations around the
    useful field

56
Electron Accelerators
  • Head leakage
  • the Air Kerma Rate (AKR) due to leakage radiation
    at any point outside the maximum useful beam, but
    inside a plane circular area with a radius of 2
    metres centred around, and perpendicular to, the
    central axis of the beam at the normal distance
    of treatment shall not exceed 0.2 of the AKR at
    the central axis of the open beam. The
    measurement shall be done with a thick shielding
    block covering the open beam.

57
Electron Accelerators
  • Head leakage
  • Except in the area defined in the previous slide
    the Air Kerma Rate (AKR) due to leakage radiation
    (excluding neutrons) at any point 1 metre from
    the path of the electrons between their origin
    and the target or electron window shall not
    exceed 0.5

58
IEC standard 601.2
  • Leakage in from linac head particularly of
    concern if the radiation can reach the patient

59
Guidance on leakage levelsin different parts of
the field
60
Also consider
  • Treatment in different patient positions e.g.
    sitting or standing next to the linac for
    treatment of a hand

61
Linacs - a note on neutrons
  • Neutrons will only be a problem if the X Ray
    energy is greater than 10 MV - in practice
    consideration MUST be given to neutrons if the
    energy is greater than or equal to 15MV
  • The rate of equivalent dose of the neutrons
    should not exceed 1 of the dose-equivalent rate
    of the X Rays - measured in sievert
  • The radiation weighting factor for the neutrons
    should be taken as 20. The above limit means that
    the neutron absorbed dose rate is always less
    than the X Ray absorbed dose rate

62
Accidents due to equipment design...
  • An operator of an accelerator quickly selected X
    Ray mode and quickly changed to electron mode
    before the machine was able to complete the first
    request (to operate in X Ray mode) and it
    operated with hybrid instructions. The same
    accident occurred in six different hospitals and
    two patients died due to doses as high as about
    160-180 Gy

63
This should not have happened...
  • Contributing factors
  • The computer controlled accelerators were not
    tested for the extreme conditions that occurred
    in practice at the hospitals.
  • It took too long for the manufacturer to identify
    the problem and disseminate the information and
    by then six hospitals had experienced the same
    failure and two patients had died from radiation

64
E Other irradiation units
  • Diagnostic units in radiotherapy
  • CT scanner
  • Simulator
  • Other therapy irradiation units
  • heavy particles

65
Diagnostic units in radiotherapy
  • Essential and often integral part of a modern
    radiotherapy department
  • Essential for adequate target definition -
    therefore important also for optimization of
    medical exposure from a radiation protection
    point of view
  • Includes not only X Ray equipment but may be MRI,
    ultrasound and nuclear medicine
  • Beyond the scope of this course - however,
    covered in separate courses on diagnostic
    radiology and nuclear medicine

66
A note on simulators
  • The simulator should be capable of reproducing
    all motions and X Ray exposure types (not
    radiation energy and dose though) of the
    treatment units

67
Simulator control
Patient clearly visible through large lead glass
window
Fluoroscopy screen
Varian Medical Systems
Control screen similar to linac
68
Simulators and other diagnostic equipment
  • Often the most important aspect of design is to
    ensure that the simulator patient set-up can be
    transferred without any modifications to the
    treatment unit. This includes imperfections of
    the systems such as couch sag under patients
    weight.

69
Heavy Particles
  • These could include
  • Neutrons
  • Protons
  • Helium nuclei (alpha particles)
  • Other heavy nuclei (Carbon nuclei)
  • Negative pi mesons
  • Protons are most common and increasing in their
    use

70
Heavy Particles treatment facilities
  • These are very specialized installations
  • shielding with high neutron fluxes can be
    extensive and complex
  • neutrons require hydrogen rich materials for good
    energy absorption for example wood and or
    plastics
  • many neutron interactions produce high energy
    gamma rays requiring large thicknesses of
    concrete , or steel to absorb them

71
Additional note on heavy particles
  • Many of the points covered for electron
    accelerators are also applicable for these
    installations
  • Specialized systems for positioning patients may
    be required
  • The charged particle accelerators are often
    multipurpose facilities which also serve research
    objectives (e.g. material research). These
    applications may require entirely different beam
    parameters (e.g. high particle flux) than medical
    treatment. More care has to be taken to ensure
    that only the correct beam can reach the patient.
  • There may be several treatment rooms for one
    accelerator
Write a Comment
User Comments (0)
About PowerShow.com