Title: NIST Traceable Flow Rate, Temperature and Pressure Calibration of FRM and Continuous Ambient Monitors
1NIST Traceable Flow Rate, Temperature and
Pressure Calibration of FRM and Continuous
Ambient Monitors
- 2005 National Air Quality Conference
- 13 February , 2005
- Quality of Air Means Quality of Life
2Introduction
- Federal, State and Tribal AQ Agencies Require
Flow Calibration and Audits on all Gravimetric
and Continuous PM Monitors - Commercial Claims of Primary Flow and NIST
Traceability are Confusing - A NIST Traceable Volumetric Flow Rate, Ambient
Temperature and Pressure Field Calibration Device
Has Been Developed and NIST Verified
3Investigation
- NIST Primary Standard vs. Flow Transfer Standard
- Requirements for a Field Calibrator
- Current Methods Available
4NIST Primary EPA Flow Transfer Standards
- NIST Primary Standard
- Colorado Engineering Experiment Station (CEESI)
- Titanium Vessel Microbalance
- Pressurized, Evacuated and Weighed for Primary
- Volume.
- NIST Certified Standard
- Three Stainless Steel Venturi Certified by NIST
- Annual Certification at CEESI
- ISO Laboratory Calibration Procedure
5CEESI - NIST Primary Laboratory
6BGI - NIST Certified Laboratory
7Requirements of a Field Calibrator
- NIST Transfer Standard. Accuracy to lt2to
Primary - Calibrator Field Usable to EPA FRM Specs for
- Atmospheric Conditions of Temperature
Pressure - Calibrator should not add artificial pressure to
- sampler to interfere with electronic flow
controller - Portable Battery Powered
- Direct Reading Records Flow Rate, Ambient
- Temperature Barometric Pressure
- RS232 Calibration Report-Digital Calibration
Report
8Flow Transfer Standard Calibrators
- Soap Bubble Meter (Excellent for Indoor Use)
- Gilibrator (Wet)
- Frictionless Piston (Not Field Hardened,
Piston Soiling) - DryCal/DC-Lite (Dry)
- Orifice - Quartz Tube (Sensitive to Wind, Adds
Pressure) - Streamline/Chinook (Dry)
- Venturi (Field Hardened with Gust Cap, Flow,
Temp, Press) - DeltaCal TriCal (Dry)
9DeltaCal FRM CPM Calibrator
- Measurements
- Barometric Pressure
- Ambient Temperature
- Filter Temperature
- Volumetric Flow Rate
- Digital Output for Recording Field Data
- Software to Print Summary report
10Schematic Diagram of DeltaCal
11DeltaCal NIST Traceable Specifications
- Operable Range 30 to 40C
- 2 to 20 aLPM Model DC 1
- 10 to 60 aLPM Model DC 100
- - 1 to NIST
- Temp Sensor Range 40 to 50C
- lt - 0.5C to NIST Mercury Thermometer
- Pressure Range 400 - 800mmHg
- lt - 2mmHg to NIST Mercury Barometer
12Certificate of Calibration
- Traceable to NIST
- Recommended Annual
- Calibration
- Venturi Can Be
- Cleaned by User
-
13Calibrator Software Display
- Historical Record
- User Configurable
- All Readings in one Display
- Auto Resolution
- ASCII File Export
- Flow Statistics
14Summary Printed Report
15TriCal Calibrator
- Operable Range -30 to 40C
- Flow Range 0.1 to 30.0 Lpm
- Venturi 1 0.1 - 1.2Lpm
- Venturi 2 1.2 -6.0Lpm
- Venturi 3 6.0 - 30Lpm
- Barometric Pressure
- Ambient Temperature
- (Option) Filter Temperature
16Calibrating a Continuous BAM Monitor
17One Step Auto-Calibration
- Connect Calibrator to FRM OMNI
- Connect RS232 between Calibrator OMNI
- Select Calibrate from Menu
- TriCal will Auto Calibrate OMNI Flow
- FRM OMNI Logs Calibration
- TriCal Prints Report
18Conclusions
- DeltaCal TriCal are Flow Transfer Standards
Directly Traceable to NIST Primary - Annual Recertification is Recommended
- Field Hardened Device with Low Pressure Drop and
DeltaCal Includes Wind Gust Cap - Cost Includes 4 Measurements, Rather than just
One (Price 1695.) - Re-Calibration by a Single Vendor (Price 150.)
- Exclusive for the PEP PM2.5 Network PM- Coarse
Method Development Program