Title: NIST Traceable Flow Rate, Temperature and Pressure Calibration of FRM and Continuous Ambient Monitors


1
NIST Traceable Flow Rate, Temperature and
Pressure Calibration of FRM and Continuous
Ambient Monitors
  • 2005 National Air Quality Conference
  • 13 February , 2005
  • Quality of Air Means Quality of Life

2
Introduction
  • Federal, State and Tribal AQ Agencies Require
    Flow Calibration and Audits on all Gravimetric
    and Continuous PM Monitors
  • Commercial Claims of Primary Flow and NIST
    Traceability are Confusing
  • A NIST Traceable Volumetric Flow Rate, Ambient
    Temperature and Pressure Field Calibration Device
    Has Been Developed and NIST Verified

3
Investigation
  • NIST Primary Standard vs. Flow Transfer Standard
  • Requirements for a Field Calibrator
  • Current Methods Available

4
NIST Primary EPA Flow Transfer Standards
  • NIST Primary Standard
  • Colorado Engineering Experiment Station (CEESI)
  • Titanium Vessel Microbalance
  • Pressurized, Evacuated and Weighed for Primary
  • Volume.
  • NIST Certified Standard
  • Three Stainless Steel Venturi Certified by NIST
  • Annual Certification at CEESI
  • ISO Laboratory Calibration Procedure

5
CEESI - NIST Primary Laboratory
6
BGI - NIST Certified Laboratory
7
Requirements of a Field Calibrator
  • NIST Transfer Standard. Accuracy to lt2to
    Primary
  • Calibrator Field Usable to EPA FRM Specs for
  • Atmospheric Conditions of Temperature
    Pressure
  • Calibrator should not add artificial pressure to
  • sampler to interfere with electronic flow
    controller
  • Portable Battery Powered
  • Direct Reading Records Flow Rate, Ambient
  • Temperature Barometric Pressure
  • RS232 Calibration Report-Digital Calibration
    Report

8
Flow Transfer Standard Calibrators
  • Soap Bubble Meter (Excellent for Indoor Use)
  • Gilibrator (Wet)
  • Frictionless Piston (Not Field Hardened,
    Piston Soiling)
  • DryCal/DC-Lite (Dry)
  • Orifice - Quartz Tube (Sensitive to Wind, Adds
    Pressure)
  • Streamline/Chinook (Dry)
  • Venturi (Field Hardened with Gust Cap, Flow,
    Temp, Press)
  • DeltaCal TriCal (Dry)

9
DeltaCal FRM CPM Calibrator
  • Measurements
  • Barometric Pressure
  • Ambient Temperature
  • Filter Temperature
  • Volumetric Flow Rate
  • Digital Output for Recording Field Data
  • Software to Print Summary report

10
Schematic Diagram of DeltaCal
11
DeltaCal NIST Traceable Specifications
  • Operable Range 30 to 40C
  • 2 to 20 aLPM Model DC 1
  • 10 to 60 aLPM Model DC 100
  • - 1 to NIST
  • Temp Sensor Range 40 to 50C
  • lt - 0.5C to NIST Mercury Thermometer
  • Pressure Range 400 - 800mmHg
  • lt - 2mmHg to NIST Mercury Barometer

12
Certificate of Calibration
  • Traceable to NIST
  • Recommended Annual
  • Calibration
  • Venturi Can Be
  • Cleaned by User

13
Calibrator Software Display
  • Historical Record
  • User Configurable
  • All Readings in one Display
  • Auto Resolution
  • ASCII File Export
  • Flow Statistics

14
Summary Printed Report
15
TriCal Calibrator
  • Operable Range -30 to 40C
  • Flow Range 0.1 to 30.0 Lpm
  • Venturi 1 0.1 - 1.2Lpm
  • Venturi 2 1.2 -6.0Lpm
  • Venturi 3 6.0 - 30Lpm
  • Barometric Pressure
  • Ambient Temperature
  • (Option) Filter Temperature

16
Calibrating a Continuous BAM Monitor
  • CPM Display

17
One Step Auto-Calibration
  • Connect Calibrator to FRM OMNI
  • Connect RS232 between Calibrator OMNI
  • Select Calibrate from Menu
  • TriCal will Auto Calibrate OMNI Flow
  • FRM OMNI Logs Calibration
  • TriCal Prints Report

18
Conclusions
  • DeltaCal TriCal are Flow Transfer Standards
    Directly Traceable to NIST Primary
  • Annual Recertification is Recommended
  • Field Hardened Device with Low Pressure Drop and
    DeltaCal Includes Wind Gust Cap
  • Cost Includes 4 Measurements, Rather than just
    One (Price 1695.)
  • Re-Calibration by a Single Vendor (Price 150.)
  • Exclusive for the PEP PM2.5 Network PM- Coarse
    Method Development Program

19
Thank You
  • BGI INSTRUMENTS
  • 58 GUINAN STREET
  • WALTHAM, MA 02451 USA
  • Tel 781 891-9380
  • Fax 781 891-8151
  • Electronic Mail info_at_bgiusa.com
  • Internet Catalogue http//www.bgiusa.com