Electron Microscopy - PowerPoint PPT Presentation

1 / 29
About This Presentation
Title:

Electron Microscopy

Description:

Cryo-Electron Microscopy. Methods. February 4, 2005. Phoebe L. Stewart ... phoebe.stewart_at_vanderbilt.edu. Suitable Samples for CryoEM. Saibil, Acta Cryst. 2000, ... – PowerPoint PPT presentation

Number of Views:250
Avg rating:3.0/5.0
Slides: 30
Provided by: phoebes1
Category:

less

Transcript and Presenter's Notes

Title: Electron Microscopy


1
Cryo-Electron Microscopy Methods February 4,
2005 Phoebe L. Stewart Molecular Physiology
Biophysics phoebe.stewart_at_vanderbilt.edu
2
Suitable Samples for CryoEM
Saibil, Acta Cryst. 2000, D561215
3
Single Particles - Suitable Samples (200 kDa
400 MDa)
Asymmetric monomeric proteins e.g. DNA-PKcs,
470 kDa
Protein/RNA or DNA complexes e.g. Ribosome, 2
MDa
Icosahedral viruses 60-fold symmetry e.g.
Adenovirus, 150 MDa
Detergent solubilized membrane proteins
e.g. Voltage-sensitive sodium channel 300 kDa,
Sato et al., Nature 2001
e.g. Platelet integrin aIIbb3 230 kDa, Adair
Yeager, PNAS 2002
4
Single Particle Method - Overview
1. Cryo-Plunge Samples
2. Collect Micrographs
3. Digitally Select Particle Images
4. 3D Image Processing
5. Structural Analysis Modeling
GOAL ? Use structural information to understand
biological function
Average 100s or 1000s of particle images
5
1. Cryo-Plunge Samples EM Sample Grids
  • For traditional electron microscopy, a copper
    mesh grid is covered with a thin carbon film.

6
  • For cryo-EM, a copper mesh grid is covered with a
    holey carbon film (A). Images are collected of
    the frozen sample suspended in holes of the
    carbon film (B). Any particles that are on the
    carbon support are not used in the image
    processing because they have a higher background
    signal.

7
Cryo-Plunging Device
  • The biological sample is applied to the grid,
    blotted to leave a thin film of water, and then
    plunge-frozen in ethane slush chilled by liquid
    nitrogen.
  • Cryo sample preparation methods were developed in
    the mid 1980s

8
Cryo Plunging and Grids
Home built Plunger
Vitrobot Reproducibility
Quantifoil Grids Automation of Data
Acquisition LEGINON System Scripps
Homemade Holey film
9
Three Forms of Ice
  • Plunge freezing of a thin (1,000Å) layer of
    water into a cryogen produces vitrified ice, or
    water in a glass-like state.
  • The diffraction pattern of vitrified ice shows no
    regular diffraction spacing, indicating a
    non-crystalline structure.
  • Hexagonal ice is the normal crystalline form of
    ice.
  • Cubic ice is formed when vitrified ice warms up
    above approx. -130C.
  • Cubic and hexagonal ice both have a greater
    volume than liquid water - expansion occurs
    during non-cryogenic freezing.
  • This expansion would distort the 3D structure of
    the biological sample.

10
2. Collect Cryo-Micrographs
Cryo Sample Holder
  • The frozen sample grid is normally kept at
    liquid nitrogen temperature (approximately
    -185C) while in the vacuum of the microscope by
    a cryo-holder.
  • Liquid helium microscopes allow the sample grid
    to be kept even colder (12 K, -261C) in the
    microscope.

11
Transmission Electron Microscope (TEM)
  • A high energy electron beam is focused by
    magnetic lenses, the column is under ultra-high
    vacuum, the sample must be dry or frozen.

12
Electro-Magnetic Spectrum
  • The electron wavelength is much shorter than 1
    Å (10-10 m) and does not limit the resolution in
    TEM.
  • Note, electrons are particles with a dual wave
    nature, rather than part of the traditional
    electromagnetic spectrum.

13
TEM Image Collection
  • A digital cryo-micrograph of adenovirus (shown
    artificially colored to enhance the image
    contrast)
  • The images are collected by
  • exposing and developing a piece of photographic
    film (the negative has to be scanned if you want
    to do image processing)
  • or by digitally collecting the image on a CCD
    (charge coupled device) camera

14
Radiation Damage
  • A low dose of electrons must be used to avoid
    radiation damage
  • A cryo-electron micrograph of adenovirus showing
    radiation damage. The virus particles have been
    destroyed by the electron beam.

15
Microscope and Camera
  • Both the type of microscope and camera can limit
    the resolution

120kV non-FEG Nitrogen
300kV FEG Nitrogen Helium High Coherence
4k x 4k camera 0.5 Å pixel High Resolution
1k x 1k camera 4 Å pixel
  • Expect 15-20 Å resolution
  • Expect 1-5 Å resolution

16
3. Digitally Select Particle Images
  • Individual particle images must be selected from
    digital cryo-electron micrographs. This has been
    done interactively (non-automatically) in the
    past. Software is being developed for automatic
    particle selection.

17
4. Three-Dimensional Image Processing
  • Each particle image represents a 2D projection of
    the 3D object

3D object (a duck)
2D projections in different views
  • The difficult step in 3D image processing is to
    determine the orientational angles (Euler angles)
    for each projection image

18
How many projection images are needed to generate
a 3D reconstruction?
  • A single projection image is plainly insufficient
    to infer the structure of an object.
  • (Originally from The New Yorker Magazine, 1991)

19
If the particle has no symmetry, projection
images are needed for essentially all possible
views
  • This Euler sphere representation (which is like a
    flat map of the globe) shows each projection
    image in a cryoEM data set as a dot
  • The entire Euler sphere should be spanned for an
    asymmetric particle.

20
If the particle has symmetry, fewer projection
images are needed
  • For a highly symmetric object, such as an
    icosahedral (60-fold symmetric) virus, projection
    images only need to be collected within the
    asymmetric unit (1/60th of the surface).
  • The number of projection images also affect the
    resolution that is attainable

21
3D Image Processing
Particle images (with determined Euler angles)
3D Reconstruction
Reprojections (should match particle images)
22
(No Transcript)
23
Resolution Assessment
  • FOURIER SHELL CORRELATION (FSC) METHOD - The data
    set is split into two halves and two independent
    half reconstructions are calculated
  • A correlation coefficient is then calculated
    between the two half reconstructions and plotted
    as a function of resolution
  • The resolution estimate is where the FSC
    correlation falls to 0.5

24
(No Transcript)
25
5. Structural Analysis Modeling
Structural Analysis at 20 Å Resolution
  • If 20 Å resolution is reached, approaches such
    as peptide tagging, antibody labeling, and
    fitting the cryoEM density with atomic structures
    of components all help to interpret the cryoEM
    density.

Hepatitis B capsid with an 8 residue peptide tag
(red) added to the N-terminus of the recombinant
capsid protein. Conway et al., PNAS 1998 9514622
CryoEM reconstruction of adenovirus (just one
capsid protein is shown in yellow) complexed with
a monoclonal antibody (red) analyzed with the
atomic structure of the major capsid protein
(magenta, green, blue). Varghese et al., J.
Virol. 2004 7812320
26
Structural Analysis at 10 Å Resolution
  • If 10 Å resolution is reached, it is possible to
    resolve alpha helices in the cryoEM density
  • It is also possible to determine if a predicted
    homology model for a protein domain agrees with
    the shape of the cryoEM density

Jiang et al., 2001, J. Mol. Biol. Helixhunter
Rice Dwarf Virus Zhou et al. 2001, Nat. Struct.
Biol.
27
Structural Analysis at 1 Å Resolution
  • If 1 to 4.5 Å resolution is reached, it should
    be possible to determine atomic resolution
    structures (similar to x-ray crystallography)
  • The best published resolution by cryoEM single
    particle reconstruction so far is 7 Å
  • Other cryoEM methods (2D crystals and helical
    filaments) have reached 3 to 4.5 Å resolution, so
    similar resolutions should be possible by single
    particle reconstruction
  • It is predicted that 1 million particle images
    will be needed to reach 3 Å resolution
  • The number of particle images may be divided by
    the symmetry. So for an icosahedral virus, 1
    million/60 16,000

X-ray crystallography density (red) and atomic
model (yellow) Terwilliger, LANL,
website www.solve.lanl.gov/index.html
28
Summary CryoEM Methods
  • No crystals are needed
  • Cryo freezing preserves the 3D structure of the
    sample. (Traditional sample staining usually
    produces structural distortions)
  • The electron dose must be kept low to avoid
    damaging the frozen sample
  • There is low image contrast in cryo-electron
    micrographs
  • Many particle images (projection images) must be
    averaged to generate a 3D reconstruction
  • The difficult part is determining the Euler
    (orientational) angles for each projection image
  • Biological complexes in the range of 300kDa to
    400 MDa may be imaged (100 - 1,500 Å in diameter)
  • Typical resolutions after 3D reconstruction are
    7-25 Å
  • 3-4.5 Å resolution may be feasible with high-end
    cryo-electron microscopes and by averaging of 1
    million particle images

29
CryoEM - Applications
On Monday (2/7) we will discuss two cryoEM
papers Z. Hong Zhou et al., Electron
cryomicroscopy and bioinformatics suggest protein
fold models for rice dwarf virus. (2001) Nat.
Struct. Biol. 8868 Brian D. Adair and Mark
Yeager, Three-dimensional model of the human
platelet integrin IIb3 based on electron
cryomicroscopy and x-ray crystallography. (2002)
PNAS 9914059
Write a Comment
User Comments (0)
About PowerShow.com