Long-Term Quantitative Analysis of Posterior Capsule Opacification After Implantation of Dual-Optic Accommodating IOLs - PowerPoint PPT Presentation

About This Presentation
Title:

Long-Term Quantitative Analysis of Posterior Capsule Opacification After Implantation of Dual-Optic Accommodating IOLs

Description:

Subjective and objective evidence of accommodative amplitude 4,5,6 ... Based on texture analysis. Objective. Valid. Reliable. Galvis et al. Study Population. Synchrony ... – PowerPoint PPT presentation

Number of Views:34
Avg rating:3.0/5.0
Slides: 14
Provided by: ascrs2009A
Category:

less

Transcript and Presenter's Notes

Title: Long-Term Quantitative Analysis of Posterior Capsule Opacification After Implantation of Dual-Optic Accommodating IOLs


1
Long-Term Quantitative Analysis of Posterior
Capsule Opacification After Implantation of
Dual-Optic Accommodating IOLs
  • Andrea Galvis, MD 1 , 3
  • Ivan Ossma Gomez,MD MPH 1 , 2
  • 1 Fundación Valle del Lili Cali, Colombia
  • 2 Universidad Industrial de Santander
    Bucaramanga, Colombia
  • 3 Fundación Oftalmológica de Santander
    Bucaramanga, Colombia

Authors are clinical investigators for the
Synchrony Dual Optic Accommodating IOL
2
Synchrony Dual Optic Accommodating IOL
  • Fills the capsular bag
  • Designed to provide accommodation via forward
    movement of the anterior optic
  • Potential for movement could be impaired by
    capsule opacification
  • ACO
  • PCO

Galvis et al
3
Clinical Experience Synchrony
  • Good unaided distance, intermediate and near
    visual function 1,2,3
  • Subjective and objective evidence of
    accommodative amplitude 4,5,6
  • Stability of over 2 years 1,2
  • Low PCO rates 7
  • Functional range of vision exceeding that of
    multifocals 8

1 Alarcon et al. ESCRS London 2006 2 Ossma et al.
ASCRS San Francisco 2006 3 Beiko G. ESCRS London
2006 4 Ossma et al. ASCRS San Diego 2004 5 Galvis
et al. ASCRS Washington 2005 6 Bohorquez et al.
ESCRS London 2006 7 Galvis et al. ESCRS Berlin
2008 8 Ossma et al. ASCRS Chicago 2008
Galvis et al
4
Rabbit Studies
Synchrony
Plate Haptic IOL
  • Difference in PCO
  • Synchrony 0.714 0.751
  • Plate haptic IOL 3.142 0.899
  • Werner et al. JCRS 2004 1114-23

Galvis et al
5
Study Design
  • Prospective Observational Analytical Study
  • 3 Groups
  • Synchrony IOL Implantation (43 eyes)
  • Alcon SN60D3 (42 eyes)
  • Alcon SN60AT Implantation (54 eyes)

Galvis et al
6
Study Design
  • Digital retroillumination photographs (3,6,12 and
    24 months)
  • Evaluate incidence PCO
  • OSCA System
  • Evaluate incidence ACO
  • Photography grading one observer

Galvis et al
7
OSCA (Open Access Systematic Capsule Assessment)
  • Aslam et al. ¹-²-³
  • Based on texture analysis
  • Objective
  • Valid
  • Reliable

¹ Aslam et al. BMC Ophthalmology 2005, 59 ²
Aslam et al. BMC Ophthalmology 2006, 630 ³ Paton
et al. BJO 2002, 861181-87
Galvis et al
8
Study Population
Synchrony (n43) SN60D3 (n42) SN60AT (n54)
Age 68.8 5.4 68.5 5.4 70.1 5.6 p0.55
Residual Refraction Defocus Equivalent 0.32 0.153 0.22 0.184 0.20 0.157 p0.48
Kruskal Wallis Analysis of Variance
Galvis et al
9
Results
Alcon SN6AD3 21.3 months postop
Synchrony 20.7 months postop
Alcon SN60AT 21.1 months postop
Galvis et al
10
OSCA Score Evolution
p0.41
p0.83
OSCA Score
p0.27
p0.38
Mann-Whitney Test
Galvis et al
11
Anterior Capsule OpacificationMonth 12
Synchrony n43 SN60D3 n42 SN60AT n54
None - Mild 40 (93.1) 35 (83.3) 47 (87)
Moderate 3 (6.9) 7 (16.7) 6 (11.1)
Severe 0 0 1 (1.9)
Fishers Exact Test p0.37
Galvis et al
12
Conclusion
  • Elements in design of Synchrony Dual Optic
    Accommodating Intraocular Lens
  • Minimize ACO
  • Minimize PCO
  • Synchrony Dual Optic Accommodating IOL
  • PCO Rates comparable to those of acrylic
    monofocal and multifocal IOLs

Galvis et al
13
Thank you