Diagnostic Instrumentation for the Fermilab
Vertical Cavity Test Facility C.M. Ginsburg , R.
Carcagno, M. Champion, N. Dhanaraj, A. Lunin,
W.-D. Moeller, A. Mukherjee, R. Nehring, D.
Orris, J. Ozelis, V. Poloubotko, D.A.
SergatskovAll Fermilab except (DESY)
  • The Fermilab vertical cavity test facility
    (VCTF), for CW RF vertical testing of bare ILC
    1.3 GHz 9-cell SRF cavities, was completed in
    July 2007.
  • Primary purpose of VCTF is to assess cavity
    performance
  • Study cavity production and processing
  • Perform acceptance test prior to insertion in a
    cryomodule
  • VCTF diagnostic instrumentation and measurements
    are required to understand cavity performance
  • Quench Location on AES01 Cavity
  • Mode measurements indicate the poorly performing
    cell is 3 or 7
  • Initial Fast thermometry test (8 sensors on
    each equator) implicates 45o region of cell 7
    near equator
  • Comprehensive Fast thermometry test pinpoints
    quench on AES01 cell 7
  • AES01 quench location
    using Fast thermometry
  • Eight Cernox sensors
  • Temperature (K) vs. time (sec) over 4
    quench-recovery cycles input power constant
  • Temperature rise 0.1 K over 2 sec seen in
    sensors 3 and 4 before quench
  • Quench seen in all 8 sensors
  • Fast thermometry - hot spot detection
  • Cernox RTD (resistive thermometry device) sensors
  • Calibrated, absolute temperature measurement
  • Flexible sensor placement, depending on test
    requirements
  • Wired in series with 1 mA current source and
    shunt resistor for measuring current
  • G10 band or nylon thread, plus Apiezon grease,
    used for mounting
  • Fast 10 kHz sampling rate suitable for quench
    detection and temperature rises
  • Status well tested and reliable system
  • Successfully and repeatedly used in other test
    systems since 2002
  • Used to pinpoint quench in AES01 cell 7

AES01 cell 7
3
4
  • Mode measurements determine cavity performance
    limiting cell(s)
  • Determine limitation to within one (cell 5) or
    two cells
  • Measure frequencies of 8 out of 9 TM010 passband
    modes on network analyzer (the p/9 was not
    measurable)
  • Determine total cavity stored energy at maximum
    input power just before quench for each
    measurable passband mode
  • Maximum gradient for each cell extrapolated from
    measurement using Superfish simulation by Haipeng
    Wang (JLab)
  • Cell 3 7 had the lowest maximum gradient, and
    cause the performance limitation
  • Field emission observed for 7p/9 mode only
  • 9-cell thermometry - hot spot detection
  • System requirements
  • Measure temperature rise, not absolute
    temperature, in a comprehensive pattern around
    each of the 9 cells, all 9 cells tested in a
    single step
  • Fast installation use for every 9 cell test
  • Design
  • Diode sensors 1mm x 1mm sense area, 1cm
    spacing
  • Diodes allow multiplexing without active
    multiplexer
  • 1N4148 diode in a SOD523 package
  • 960 diodes per cell 60 cards x 16 diodes/card
  • 8640 diodes for a 9-cell cavity
  • Cards G10 boards, Kapton printed circuit, Kapton
    flex loop provides spring to hold diodes against
    surface
  • Cage Cards installed once on two half-shells
    which bolt together
  • Status Single-cell prototype being built test
    CY07


R
Completed card
z
G10 boards
Kapton printed circuit
  • Variable RF input coupler
  • Keeps cavity critically coupled
  • Facilitates mode measurements useful for cavity
    diagnostics
  • Mounts to standard input coupler port
  • Movement range DR15 mm (equiv. 2109 . Qext .
    41010)
  • Status Parts arriving test CY07

Diode sensor
Flex shown in compressed state as if contacting
cavity