1
Team 6 Barbara Tuday
Nathan Schreppler
Michael Nolan Lauren
McIlwain
Position Sensing Glove
Sponsors David Cadogan Scott
DavidsonAdvisor Dr. Michael Keefe
Final Concept
Proof of Concept Prototype
Background
Glove with Model Sensors and Securing Devices
Circuit Diagram and Test Setup
LabView DAQ Program
During a hand therapy session, physical
therapists must manually measure the angle of
each joint in a patients hand and wrist.
Problem Statement
Our goal is to design, develop, produce, and test
a prototypic, low-cost, and functional glove that
will sense key finger, thumb, and wrist positions
using compliant advanced technologies.
Testing
  • Attachment of sensors
  • Effect of backing thickness
  • Need for electrical connection stiffener
  • Behavior of sensors
  • Tension Test
  • Bend Radius Test
  • Pressure Test
  • Twist Test

Concept Selection
Glove with Sensors
Top Ten Wants
  • Key Customers
  • ILC Dover
  • Physical Therapists
  • Patients

Results
Key Metrics Target Values
Accuracy /- 2.5 degrees
Cost 200.00
Cycles to Failure 50,000
Resistance 35,000 ohms
  • Optimal backing is the medium-thickness backing
    (three layers of packing tape).
  • No effect from pressure.
  • Minimal effects from tension and twist.
  • Bend radius has no effect on voltage readings.
  • An electrical connection stiffener is necessary
    for consistent results.

Features
Subsystems
  • Spectra Symbol flex sensors glued to a nylon
    knit comfort glove.
  • Sensors wired with 24 gauge wire to an eight-
    channel DAQ board.
  • Data collected using LabView, which will convert
    voltage readings to angular measurements.
  • Stiffeners made of polypropylene stabilize and
    protect the electrical connections on the sensors.

Project Transition
Selected Concepts
  • Sensors Peratech QTC material
  • Wiring Calmont brand wire
  • Glove Knit Comfort glove
  • DAQ LabView
  • Used in Prototype
  • Sensors Spectra Symbol flex sensors
  • Wiring Copper wire
  • Replace Sensors with Peratech QTC Technology
  • Rewire with Calmont brand wire
  • Improve DAQ system (USB Port Input)
  • Improve LabView Program