Variables that Influence Measured Concrete Compressive Strength - PowerPoint PPT Presentation

1 / 37
About This Presentation
Title:

Variables that Influence Measured Concrete Compressive Strength

Description:

Air pockets trapped underneath a sulfur cap can lead to a loss of strength up to 12 ... Insufficient humidity during initial curing can lower measured strength. ... – PowerPoint PPT presentation

Number of Views:185
Avg rating:3.0/5.0
Slides: 38
Provided by: heather150
Category:

less

Transcript and Presenter's Notes

Title: Variables that Influence Measured Concrete Compressive Strength


1
Variables that Influence Measured Concrete
Compressive Strength
  • Heather J. Sauter, E.I.T.
  • Ph.D. Candidate

2
Acknowledgement
  • The following paper was referenced throughout
    this presentation
  • Richardson, David N., Review of Variables that
    Influence Measured Concrete Compressive
    Strength, NAA Circular No. 132, NRMCA
    Publication No. 179, National Ready Mixed
    Concrete Association, Silver Spring, MD 20910.

3
Why concrete strength is tested
  • In research, physical laws or properties may be
    under investigation
  • During construction, an estimate of the in-place
    strength of concrete may be desired for
    determining the safe time to strip forms or to
    proceed with further work.
  • The adequacy of mix proportions may need to be
    verified.
  • Compressive strength data is necessary for
    quality control.

4
Testing of Cylinders
  • Hardened concrete is typically evaluated for
    acceptance using 6 in. (152 mm) by 12 in. (305
    mm) cylinders.
  • The measured results are dependent upon adhering
    strictly to standardized uniform procedures.
  • Most testing errors produce lower strength
    results.

5
Testing of Cylinders
  • The consequences of falsely low results can be
  • Unnecessary delays
  • Costly follow up testing
  • Wasteful overdesign
  • Possible rejection of good concrete

6
Variables
  • Sampling
  • Casting
  • Initial Curing
  • Transporting
  • Laboratory Curing
  • Capping
  • Testing
  • Reporting

7
Sampling
  • Proper batch representation by sampling is often
    not achieved.
  • ASTM C172 specifies that the sample should be
    taken from atleast two places in the middle
    portion of the load.
  • Remixing of the sample is also specified to
    ensure uniformity of the sample.
  • Maximum time interval between sampling and
    casting is also specified.
  • Make sure that the concrete for a set of
    cylinders comes from a single truck.

8
Casting
  • Consolidation
  • Mold Material
  • Specimen End Condition
  • Cylinder Uniformity

9
Consolidation
  • To get the proper consolidation, the method of
    consolidation should match the material being
    consolidated.
  • The proper number of layers, consolidation
    effort, rod type, and mold type are all
    important.
  • Insufficient consolidation can lead to a strength
    loss of as much as 61.

10
Mold Material
  • The type of mold material is important in terms
    of rigidity, water absorption, and expansion.
  • A more rigid mold will flex less during
    consolidation resulting in a more compact
    specimen.
  • Mold types
  • Cardboard molds with steel base plate
  • Steel
  • Plastic

11
Specimen End Condition
  • The finished condition of the cylinder ends are
    important to concrete strength.
  • A rough end can mean capping problems
  • Air pockets trapped underneath a sulfur cap can
    lead to a loss of strength up to 12.
  • The rough interface between the concrete and
    capping material will present a nonstandard
    stress distribution during testing

12
Specimen End Condition
  • Upon casting, the specimens should be left on a
    level, smooth surface.
  • Cutting the cylinder end evenly may be a better
    procedure but it becomes very costly.

13
Cylinder Uniformity
  • Unusual fracture types have been noted for
    cylinders when rod penetration has been
    insufficient, resulting in poor bond between
    layers.

14
Initial Curing
  • Temperature
  • Humidity
  • Specimen Disturbance during Initial Curing

15
Temperature
  • Lower than standard curing temperatures for 3 to
    7 days can cause as much as a 7 loss in
    strength.
  • One day of freezing followed by standard curing
    can lead to a loss up to 56.
  • Higher than standard curing temperatures may
    increase early strength, but later strengths will
    suffer.

16
Humidity
  • Insufficient humidity during initial curing can
    lower measured strength.
  • Proper humidity can be approached by covering the
    cylinders.
  • In general, the cylinders should be brought into
    the lab within 24 hours of casting.

17
Specimen Disturbance during Initial Curing
  • Rough treatment of the cylinder while it is
    undergoing setting and initial hardening can
    damage the specimen.
  • Disturbance can be in the form of gross
    disturbance or vibration.
  • For traffic induced vibrations, wet mixes can
    lose as much as 5 strength through the
    segregation mechanism, while stiff mixes can gain
    4 strength by virtue of improved consolidation.

18
Transporting
  • Rolling and bumping in the back of a pickup truck
    and result in a 7 strength loss.
  • Dropping cylinders from waist level can lower
    strength by atleast 5.
  • Cylinders should be cushioned during transport
    and handled gently at all times.

19
Laboratory Curing
  • Upon arrival to the lab, cylinders should
    immediately be stripped, logged in, and placed in
    a proper curing environment until the time of
    test.
  • Temperature needs to be maintained at 73 /- 3 F
    either in a fog room or a lime-saturated water
    tank.
  • The purpose of wet curing is to maximize
    hydration of the cement.

20
Capping
  • End Conditions
  • Capping Materials

21
End Conditions
  • Four basic specifications per ASTM C617
  • The capped ends must be flat within 0.002 in.
    (0.05 mm) in 6 inches.
  • The ends must be approximately perpendicular to
    the specimen axis.
  • The ends must be approximately parallel to each
    other.
  • The caps must not be excessively thick.

22
Capping Materials
  • Any capping material used should not induce a
    stress distribution in the cylinder that would
    cause inaccurate measured strength results.

23
Testing
  • Moisture Condition and Temperature
  • Loading Rate
  • Specimen Misalignment
  • Loading Platens
  • Specimen Behavior
  • Seating Behavior
  • Machine Calibration
  • Frame Stiffness
  • Postfailure Inspection

24
Moisture Condition and Temperature
  • The cylinder must not be allowed to dry out after
    capping. A dry cylinder will exhibit 2-30
    higher strength than a saturated one.
  • A recent study showed that the moisture gradient
    is important and that the surface drying causes
    shrinkage and lateral biaxial compression, which
    increases measured strength.

25
Loading Rate
  • Compressive strength increases with increased
    loading rate the relationship between strength
    and rate of loading is logarithmic.
  • Choose a loading rate per ASTM C39 and repeat it
    to reduce variability.

26
Specimen Misalignment
  • A 12 loss in strength can result from eccentric
    loading.

27
Loading Platens
  • The degree of stiffness of the steel loading
    platens can affect the stress distribution within
    the test cylinder.
  • Radius of the spherically seated head is
    important to provide sufficient restraint during
    loading and obtain good initial alignment.

28
Loading Platens
29
Specimen Behavior
  • There are three basic types of specimen behavior
    during loading
  • both ends pinned
  • both ends fixed
  • one end fixed, one end pinned

30
Specimen Behavior
  • Once significant loads are applied, any tilting
    will induce bending and nonuniform straining
    which will result in a loss of strength.
  • The fixed end situation is desirable so that
    there is no tilting during loading.

31
Seating Behavior
  • Seating behavior is a function of lubricant type,
    sphere-seat contact area, contact type,
    sphere-seat surface finish, and platen geometry.
  • It is recommended that high-pressure grease not
    be used.
  • A thin coat of petroleum jelly has proven good
    performance. ASTM C39 allows the use of motor
    oil also.

32
Machine Calibration
  • A significant percentage of testing machines in
    use can fail to be within calibration standards
    giving strength differences of as much as 19.
  • Accuracy of a machine can change over a
    relatively short time
  • Care must be taken during calibration not to
    accidentally damage the platens.

33
Frame Stiffness
  • Machine frame stiffness can be important,
    especially for high strength concrete.
  • A less rigid frame will stretch during loading
    storing more elastic energy.
  • The less rigid machine will more closely follow
    the rapid deformation of the specimen and failure
    will be more explosive.
  • It is not unanimous whether this lowers
    compressive strength.

34
Postfailure Inspection
  • Visual inspection of the break can reveal
  • poor bonding to coarse aggregate
  • segregation
  • a porous and fragile matrix
  • high air content
  • soft coarse aggregate

35
Conclusions
36
(No Transcript)
37
  • Thank you
  • Questions?
Write a Comment
User Comments (0)
About PowerShow.com