1
Nellcor Puritan Bennett Sublingual PCO2 Monitor
B. cepacia
  • Joanna Weitershausen
  • Food and Drug Administration
  • Center for Devices and Radiological Health
  • Office of Compliance

1
2
Device Description
  • The probe is placed under the tongue and the
    device is intended to monitor sublingual PCO2.
  • The device is shipped non-sterile in a saline
    solution to preserve sensor integrity.

2
3
Case Events
  • Non-significant risk study
  • Cluster of Burkholderia cepacia colonization
  • No MDRs received
  • CDC and Health Department involvement
  • CapnoProbe suspected as causing infections
  • Initial notification to the public

3
4
Testing for B. cepacia
Electron Micrograph of Burkholderia cepacia.
Credit CDC.
  • Culture unopened devices
  • Testing
  • Manufacturer
  • Health Department

B. cepacia expressing CF mucous-binding Credit
Richard Goldstein and J Bacteriol
19951771039-52.
4
5
Manufacturer DH Results
Dendrogram for pulsed-field gel electrophoresis
patterns of Burkholderia cepacia isolates
analyzed by Joey Stringer at the Texas Dept. of
Health Laboratory
5
6
FDA Actions
  • Manufacturer voluntary recall
  • Press releases
  • FDA Press release
  • Manufacturers press release
  • MMWR - CDC
  • Health Hazard Evaluation (HHE)
  • Class I recall
  • MDRs

6
7
Timeline Recap
  • Aug 16 Hospital contacts mfr mfr starts
    testing
  • Aug 18 CDC contacts FDA
  • Aug 19 CDC post information on website
  • Aug 20 Texas HD starts testing
  • Aug 24 Mfr identifies multiple types of
    bacteria, initiates device recall
  • Aug 25 Texas DNA match
  • Aug 26, 27 Press releases
  • Aug 27 FDA HHE, classifies recall

7