bmd Crossvent 3
2
Modes Touch Screen
  • Adult to pediatric range
  • Assist/Control Volume Control
  • Assist/Control Pressure Control
  • Sighs
  • SIMV
  • Pressure Support
  • CPAP

3
Turn knobs for setting
  • Maximum pressure
  • PEEP
  • Flow

4
LCD Touch ScreenSelect Mode then press MAIN menu
  • Then press Rate TV
  • adjust with up/down arrows
  • IT, ET IE Ratio displayed
  • Flow, PEEP Max Press limit
  • adjust with dials

5
On the left side
External Power Supply
O2 Sensor
ON/OFF
Alarm Reset
LED Charging Light
Pt Circuit
Pneumotach/Flow Sensor Input
Airway Pressure Connector
Exhalation Valve Connector
6
bmd CrossVent 3
  • powered by a rechargeable NiMH battery
  • lasts 6 hours
  • BioMed 16vDC 3A charger/converter recharges
    internal battery and allows for AC operation
  • LCD screen has LCD charging indicator and
    visually large LOW BATTERY alarm

7
Flow Sensor at proximal airway
8
Flow Sensor
9
On the right side
Gas Entrainment ON/OFF
Bleed Exhaust
Gas Supply Inlet (31 - 75 psig or 45 - 75
when using blender)
10
Gas Entrainment
  • When connected to 100 oxygen source gas with the
    Gas Entrainment ON, the oxygen concentration is
    diluted to 50.
  • There will be some decrease in delivered flow
    rate with Entrainment ON under high load
    conditions

11
The rear panel
  • Maximum pressure relief valve (120 cm H2O)
  • Negative pressure relief valve opens at - 4 cm
    H2O with machine shutdown

12
SIMV settings
13
Press arrows to increase or decrease Rate or
Tidal Volume
14
Touch screen to highlight TV Adjust with
up/down arrows
15
CPAP Mode
16
ALARMS
  • ALARM 1
  • Peak Pressure
  • Respiratory Rate
  • Exhaled Tidal Volume (can be switched off)
  • Exhaled Minute Volume (can be switched off)
  • ALARM 2
  • PEEP
  • Mean Pressure (can be switched off)
  • Oxygen (can be switched off)

17
To set Alarms Press ALARM 1
  • Press High or Low Limits to Change
  • Adjust with up/down arrows

18
Repeat for ALARM 2 settings
19
3rd Level Alarms
Low Gas Supply Pressure Low Battery Ventilator
Failure
20
With Alarm Condition
the value for the exceeded limit flashes red
-1
21
When power off, press and hold ALARM RESET
Write a Comment