Introduction to Electron Microscopy Fundamental concepts in - PowerPoint PPT Presentation

1 / 41
About This Presentation
Title:

Introduction to Electron Microscopy Fundamental concepts in

Description:

Introduction to Electron Microscopy Fundamental concepts in electron microscopy The construction of transmission and scanning electron microscopes – PowerPoint PPT presentation

Number of Views:248
Avg rating:3.0/5.0
Slides: 42
Provided by: caurUgaE8
Category:

less

Transcript and Presenter's Notes

Title: Introduction to Electron Microscopy Fundamental concepts in


1
Introduction to Electron Microscopy
  • Fundamental concepts in electron microscopy
  • The construction of transmission and scanning
    electron microscopes
  • Sample examples of application

2
Electron Microscope vs. Optical Microscope
(first one built in 1931 by Ruska and Knoll)
(Leeuwenhoek in 17th century)
  • Electron vs. Photon
  • Electron charged, has rest mass, not visible
  • Photon neutral, has no rest mass, visible at the
    wavelength 400 nm-760 nm.

Because of these differences, the microscope
construction will also be different
What is the common property?
3
Comparison of EM and LM
  • a. Similarities (Arrangement and function of
    components are similar) 1) Illumination system
    produces required radiation and directs it onto
    the specimen. Consists of a source, which emits
    the radiation, and a condenser lens, which
    focuses the illuminating beam (allowing
    variations of intensity to be made) on the
    specimen. 2) Specimen stage situated between
    the illumination and imaging systems. 3)
    Imaging system Lenses which together produce the
    final magnified image of the specimen. Consists
    of i) an objective lens which focuses the beam
    after it passes through the specimen and forms an
    intermediate image of the specimen and ii) the
    projector lens(es) which magnifies a portion of
    the intermediate image to form the final image.
    4) Image recording system Converts the
    radiation into a permanent image (typically on a
    photographic emulsion) that can be viewed.

4
Comparison of EM and LM
  • b. Differences 1) Optical lenses are generally
    made of glass with fixed focal lengths whereas
    magnetic lenses are constructed with
    ferromagnetic materials and windings of copper
    wire producing a focal length which can be
    changed by varying the current through the coil.
    2) Magnification in the LM is generally changed
    by switching between different power objective
    lenses mounted on a rotating turret above the
    specimen. It can also be changed if oculars
    (eyepieces) of different power are used. In the
    TEM the magnification (focal length) of the
    objective remains fixed while the focal length of
    the projector lens is changed to vary
    magnification. 3) The LM has a small depth of
    field, thus different focal levels can be seen in
    the specimen. The large (relative) depth of field
    in the TEM means that the entire (thin) specimen
    is in focus simultaneously. 4) Mechanisms of
    image formation vary (phase and amplitude
    contrast). 5) TEMs are generally constructed
    with the radiation source at the top of the
    instrument the source is generally situated at
    the bottom of LMs. 6) TEM is operated at high
    vacuum (since the mean free path of electrons in
    air is very small) so most specimens (biological)
    must be dehydrated (i.e. dead !!). 7) TEM
    specimens (biological) are rapidly damaged by the
    electron beam. 8) TEMs can achieve higher
    magnification and better resolution than LMs.
    9) Price tag!!! (100x more than LM)

5
Resolution of a microscope
Where N.A. is the numerical aperture n(sina)
6
The resolution is proportional to the
wavelength!
Electron equivalent wavelength and accelerating
voltage
The dualism wave/particle is quantified by the De
Broglie equation ? h/p h/mv ? wavelength
h Planck constant p momentum
The energy of accelerate electrons is equal to
their kinetic energy E eV m0v2/2 V
acceleration voltage e / m0 / v charge / rest
mass / velocity of the electron These equations
can be combined to calculate the wave length of
an electron with a certain energy p m0v
(2m0eV)1/2 ? h / (2m0eV)1/2 ( 1.22 / V1/2 nm)
At the acceleration voltages used in TEM,
relativistic effects have to be taken into
account (e.g. Egt100 keV) ? h / 2m0eV (1
eV/2m0/c2)1/2
7

Wavelength and accelerating voltage
There are other factors that limit the resolution!
8
Types of Electron Microscope
  • Transmission Electron Microscope (TEM) uses a
    wide beam of electrons passing through a thin
    sliced specimen to form an image. This microscope
    is analogous to a standard upright or inverted
    light microscope
  • Scanning Electron Microscope (SEM) uses focused
    beam of electrons scanning over the surface of
    thick or thin specimens.. Images are produced one
    spot at a time in a grid-like raster pattern.
    (will be discussed in a later lecture)
  • Scanning Transmission Electron Microscope (STEM)
    uses a focused beam of electrons scanning through
    a thin sliced specimen to form an image. The STEM
    looks like a TEM but produces images as does an
    SEM (one spot at a time). It is most commonly
    used for elemental analysis of samples.

9
(No Transcript)
10
Leo 982 SEM
For SEM a fine probe (beam spot) is formed by
condenser lens and its size determines the
resolution (this differs from the TEM which is
diffraction limited)
11
Electron Gun
FEG
W hairpin
LaB6 crystal
12
Thermionic Sources
Increasing the filament current will increase the
beam current but only to the point of saturation
at which point an increase in the filament
current will only shorten the life of the emitter
13
Beam spot image at different stage of heating
14
(No Transcript)
15
(No Transcript)
16
Electromagnetic lens
17
Condenser-lens system
C1 controls the spot size
The condenser aperture must be centered
C2 changes the convergence of the beam
18
TEM imaging modes
19
STEM image
Bright and dark field STEM image of Au particles
on a carbon film
What is the differences between this dark field
and the previous one?
20
Magnification in TEM
Mob Mint Mproj Total Mag
Depends on the magnification, some lens may not
be used
21
Electron-specimen interaction
22
  • Backscattered Electrons
  • Formation
  • Caused by an incident electron colliding with
    an atom in the specimen which is nearly normal to
    the incident's path. The incident electron is
    then scattered "backward" 180 degrees.
    Utilization
  • The production of backscattered electrons
    varies directly with the specimen's atomic
    number. This differing production rates causes
    higher atomic number elements to appear brighter
    than lower atomic number elements. This
    interaction is utilized to differentiate parts of
    the specimen that have different average atomic
    number.

23
Backscatter Detector
The most common design is a four quadrant solid
state detector that is positioned directly above
the specimen
24
Gold particles on E. coli appear as bright white
dots due to the higher percentage of
backscattered electrons compared to the low
atomic weight elements in the specimen
25
Backscatter image of Nickel in a leaf
26
  • Secondary Electrons
  • Source
  • Caused by an incident electron passing "near" an
    atom in the specimen, near enough to impart some
    of its energy to a lower energy electron (usually
    in the K-shell). This causes a slight energy loss
    and path change in the incident electron and the
    ionization of the electron in the specimen atom.
    This ionized electron then leaves the atom with a
    very small kinetic energy (5eV) and is then
    termed a "secondary electron". Each incident
    electron can produce several secondary electrons.
  • Utilization
  • Production of secondary electrons is very
    topography related. Due to their low energy, 5eV,
    only secondaries that are very near the surface
    (lt 10 nm) can exit the sample and be examined.
    Any changes in topography in the sample that are
    larger than this sampling depth will change the
    yield of secondaries due to collection
    efficiencies. Collection of these electrons is
    aided by using a "collector" in conjunction with
    the secondary electron detector. The collector is
    a grid or mesh with a 100V potential applied to
    it which is placed in front of the detector,
    attracting the negatively charged secondary
    electrons to it which then pass through the
    grid-holes and into the detector to be counted.

27
A conventional secondary electron detector is
positioned off to the side of the specimen. A
faraday cage (kept at a positive bias) draws in
the low energy secondary electrons. The electrons
are then accelerated towards a scintillator which
is kept at a very high bias in order to
accelerate them into the phosphor.
28
The position of the secondary electron detector
also affects signal collection and shadow. An
in-lens detector within the column is more
efficient at collecting secondary electrons that
are generated close to the final lens (i.e. short
working distance).
29
Secondary Electron Detector
Side Mounted In-Lens
What are the differences between these two images?
30
  • Auger Electrons
  • Source
  • Caused by the de-energization of the specimen
    atom after a secondary electron is produced.
    Since a lower (usually K-shell) electron was
    emitted from the atom during the secondary
    electron process an inner (lower energy) shell
    now has a vacancy. A higher energy electron from
    the same atom can "fall" to a lower energy,
    filling the vacancy. This creates and energy
    surplus in the atom which can be corrected by
    emitting an outer (lower energy) electron an
    Auger Electron.
  • Utilization
  • Auger Electrons have a characteristic energy,
    unique to each element from which it was emitted
    from. These electrons are collected and sorted
    according to energy to give compositional
    information about the specimen. Since Auger
    Electrons have relatively low energy they are
    only emitted from the bulk specimen from a depth
    of lt 3 nm

31
  • X-rays
  • Source
  • Caused by the de-energization of the specimen
    atom after a secondary electron is produced.
    Since a lower (usually K-shell) electron was
    emitted from the atom during the secondary
    electron process an inner (lower energy) shell
    now has a vacancy. A higher energy electron can
    "fall" into the lower energy shell, filling the
    vacancy. As the electron "falls" it emits energy,
    usually X-rays to balance the total energy of the
    atom so it.
  • Utilization
  • X-rays or Light emitted from the atom will have
    a characteristic energy which is unique to the
    element from which it originated.
  • (will be discussed in a separate lecture)

32
  • Unscattered Electrons
  • Source
  • Incident electrons which are transmitted through
    the thin specimen without any interaction
    occurring inside the specimen.
  • Utilization
  • The transmission of unscattered electrons is
    inversely proportional to the specimen thickness.
    Areas of the specimen that are thicker will have
    fewer transmitted unscattered electrons and so
    will appear darker, conversely the thinner areas
    will have more transmitted and thus will appear
    lighter.

33
The overlapping areas appear darker
34
  • Elasticity Scattered electrons
  • Source
  • Incident electrons that are scattered (deflected
    from their original path) by atoms in the
    specimen in an elastic fashion (no loss of
    energy). These scattered electrons are then
    transmitted through the remaining portions of the
    specimen.
  • Utilization
  • All electrons follow Bragg's Law and thus are
    scattered according to ml2dsin q (angle of
    scattering). All incident electrons have the same
    energy(thus wavelength) and enter the specimen
    normal to its surface. All incidents that are
    scattered by the same atomic spacing will be
    scattered by the same angle. These "similar
    angle" scattered electrons can be collated using
    magnetic lenses to form a pattern of spots each
    spot corresponding to a specific atomic spacing
    (a plane). This pattern can then yield
    information about the orientation, atomic
    arrangements and phases present in the area being
    examined.

35
The diffraction pattern is highly dependable on
the structure of the specimen
Dr. Schroeder will have a lecture dedicated to
diffraction
36
  • Inelastically Scattered Electrons
  • Source
  • Incident electrons that interact with specimen
    atoms in a inelastic fashion, loosing energy
    during the interaction. These electrons are then
    transmitted trough the rest of the specimen
  • Utilization
  • Inelastically scattered electrons can be
    utilized two ways
  • Electron Energy Loss Spectroscopy (EELS) The
    inelastic loss of energy by the incident
    electrons is characteristic of the elements that
    were interacted with. These energies are unique
    to each bonding state of each element and thus
    can be used to extract both compositional and
    bonding (i.e. oxidation state) information on the
    specimen region being examined.
  • Kikuchi Bands Bands of alternating light and
    dark lines that are formed by inelastic
    scattering interactions that are related to
    atomic spacings in the specimen. These bands can
    be either measured (their width is inversely
    proportional to atomic spacing) or "followed"
    like a roadmap to the "real" elasticity scattered
    electron pattern.

37
EELS of NiO
Combine with STEM, can do element mapping in TEM
compare to EDX, EELS is better for lighter
elements
Unlike diffraction pattern blinks on and off, the
Kikuchi line pattern rotates when one tilts the
crystal. Thus is helpful in orientating a crystal
to certain zone axis.
38
Specimen interaction volume
  • Atomic number of the material being examined
    higher atomic number materials absorb or stop
    more electrons and so have a smaller interaction
    volume.
  • Accelerating voltage being used higher
    voltages penetrate farther into the sample and
    generate larger interaction volumes
  • Angle of incidence for the electron beam the
    greater the angle (further from normal) the
    smaller the volume

39
Monte Carlo simulation of the interaction volume
Try the online simulation at
http//www.matter.org.uk/tem/electron_scattering.h
tm
40
High-Resolution Electron Microscopy
Wave functions for elastically-scattered, forward
electrons
  • For a weak phase object, the observable image
    intensity is
  • I1 - 2??(r) ? FFTT(H).
  • represents convolution.
  • This shows a pure phase-contrast image.

Illustration of electron wave passing through
under phase object approximation. The phase
changes (contrast) is imaged.
41
High resolution Images
Discovery of the carbon nanotube
Pt nanoparticles
S. Iijima, Nature 354, 56 (1991).
Write a Comment
User Comments (0)
About PowerShow.com