Quality Control in Serology - PowerPoint PPT Presentation

About This Presentation
Title:

Quality Control in Serology

Description:

Quality Control in Serology – PowerPoint PPT presentation

Number of Views:2793
Slides: 57
Provided by: doctorrao
Tags:

less

Transcript and Presenter's Notes

Title: Quality Control in Serology


1
Quality control in serology
  • Dr.T.V.Rao MD

2
What is Quality Control?
  • Process or system for monitoring the quality of
    laboratory testing, and the accuracy and
    precision of results
  • Routinely collect and analyze data from every
    test run or procedure
  • Allows for immediate corrective action

3
Definitions
  • Quality Control - QC refers to the measures that
    must be included during each assay run to verify
    that the test is working properly.
  • Quality Assurance - QA is defined as the overall
    program that ensures that the final results
    reported by the laboratory are correct.
  • The aim of quality control is simply to ensure
    that the results generated by the test are
    correct. However, quality assurance is concerned
    with much more that the right test is carried
    out on the right specimen, and that the right
    result and right interpretation is delivered to
    the right person at the right time

4
Definitions
  • Quality Assessment - quality assessment (also
    known as proficiency testing) is a means to
    determine the quality of the results generated by
    the laboratory. Quality assessment is a challenge
    to the effectiveness of the QA and QC programs.
  • Quality Assessment may be external or internal,
    examples of external programs

5
Designing a QC Program
  • Establish written policies and procedures
  • Corrective action procedures
  • Train all staff
  • Design forms
  • Assure complete documentation and review

6
The Quality System
Information Management
7
The Quality Assurance Cycle
Pre-Analytic
Patient/Client Prep Sample Collection
Personnel Competency Test Evaluations
Reporting
  • Data and Lab Management
  • Safety
  • Customer Service

Post-Analytic
Sample Receipt and Accessioning
Record Keeping
Sample Transport
Quality Control
Testing
Analytic
8
Need for Serological Tests
  • SEROLOGICAL TESTS are performed to demonstrate
    antigens in the serum, or the response of the
    human body to these infectious agent ( Antibodies
    ) to establish its contact with the immune
    system. Their diagnostic importance stems from
    demonstration of a rising titre of antibodies to
    the agent which inter alia indicates a
    progressive infection. In rare instances is the
    presence of antibody in a single sample
    indicative of infection and disease. Serological
    tests are of importance in epidemiological
    studies and to ascertain the response of the
    population to vaccines and other
    immunopotentiators.

9
Importance of Serological Tests
  • Serological tests are of importance in
    epidemiological studies and to ascertain the
    response of the population to vaccines and other
    immunopotentiators.

10
Serology performed mainly as vitro tests
  • Serological tests are also useful for the in
    vitro detection of microbial infections, and for
    the classification and sub classification of
    infectious agents (e.g. Salmonella, Shigella,
    Streptococcus, etc.).

11
COLLECTION OF SPECIMEN
  • There must be a system for the orderly and
    efficient requesting of tests collection and
    identification of specimens and transporting,
    preparation, and storage of specimens. Nothing is
    more important than having an adequate amount of
    an appropriate specimen in good condition for
    examination. If each specimen is not properly
    collected, labeled, and handled, or is not
    representative, the laboratory may do more harm
    than good by testing it.

12
Haemolysed specimens are not suitable for testing
  • Haemolysed blood specimens are not suitable for
    serological studies. It is always advisable to
    avoid factors which cause hemolysis (Table 14.1).
    Specimens containing precipitates should be
    centrifuged prior to testing.

13
Avoidable causes of hemolysis
  • Blood sampling through too small bore of a needle
  •       Forced suction of blood in the syringe
    during blood collection
  •       Vigorous shaking of blood from the syringe,
    especially through a needle
  •       Centrifuging blood sample at a high speed
    before clotting
  •       Freezing and thawing of blood
  •       Unclean tubes with residual detergents
  •      

14
Serology can Detect either antigen or antibody
  • Serological reactions detect either a specific
    antigen produced by the microorganism or a
    specific immune response of the human body.
    Serological tests may detect
  • 1 an immunological principle (antigen-antibody
    reaction ELISA, Widal)
  •  2 a non-specific reaction (VDRL test)
  •   3 a reaction mediated by complement (complement
    fixation test)

15
advantages with serological methods.
  •  Rapid identification of agent
  •      High specificity of detection of antigen
  •      Simplicity of performance
  •      Safe procedures
  •      Diagnostic aids
  •      Epidemiological tools
  •      Retrospective confirmation of diagnosis

16
List of tests added every day
  • A wide variety of serological tests are now
    available and every day new ones are added to an
    already impressive list. Every laboratory must
    define a policy for conducting these tests
    because some may be expensive, all require
    certain reagents (sera or antigens etc.) which
    have limited shelf life, and all require
    standardised techniques which must be documented
    in SOPM.

17
Standard operating procedures are foundations in
all protocols
  • An important element in maintaining day-to-day
    uniformity in laboratory results is an
    established procedure manual (SOPM) which details
    all phases of the laboratorys operation
    (including safety precautions) and is used by all
    laboratory personnel. It should include
    instructions for collecting, transporting, and
    storing specimens, for preparing and storing
    reagents, and for performing tests. In addition,
    the controls and calibrators to be used should be
    listed along with directions for their use,
    expected results, and instructions for corrective
    measures if the expected results are not obtained.

18
Control Sera
  • Source
  • Some control sera are available commercially.
    Small volumes are generally available as
    components in kits but are intended to be used
    only with a single kit. A few may be available in
    larger quantities.
  • Preparation
  • Sera to be used as controls should be kept
    sterile to avoid deterioration. In general each
    procedure should have a normal control serum
    (negative), a strong positive control serum, and
    another positive control serum which is reactive
    at the critical concentration (borderline
    positive). With some tests, controls with a low
    concentration of analyze should be included.
    Controls recommended by the manufacturer of a
    particular test should always be used and
    additional control sera can be included if a test
    involves special problems.

19
Storage
  • Sera to be used as controls should be
    standardized against international reference
    materials when they are available. "Standards"
    included in commercial kits are not calibrated
    with each other and often are not
    interchangeable. These should be stored in
    aliquotes in frozen forms. Repeated freezing and
    thawing should be avoided.

20
Quality reagents give optimal results
  • Quality reagents are necessary for quality
    performance. A record should be kept of any
    changes in reagents in case the performance of a
    test changes. Before new reagents are introduced
    into a system they should be tested in parallel
    with the old reagents against a panel of
    appropriate reference sera to be sure that
    consistent reactions are obtained. The results
    obtained with the panel should reflect the
    sensitivity and specificity of the reagents being
    compared.

21
Label all the reagents
  • Reagents should be clearly labeled to indicate
    their identity, hazards involved in their use,
    recommended storage conditions, and preparation
    and expiration dates.

22
EQUIPMENT AND INSTRUMENTS
  • All glassware used in immunologic tests must be
    clean and free of detergent. Chipped or etched
    glassware should be discarded. Calibrated
    glassware should be checked for accuracy.
  • The users accuracy and precision requirements
    should be met or exceeded when equipment is
    tested under working conditions. The
    manufacturers specifications for performance
    should be checked and met. Instruments and
    equipment should be monitored routinely. The
    temperature of water baths, incubators,
    refrigerators, and freezers should be checked
    periodically and records maintained. Maintenance
    should be performed and records kept on a regular
    basis by individuals who are trained and are
    familiar with the equipment.

23
Quality control of instruments
  • Instruments used for measurements including
    spectro-photometers, spectrometers, dilutors, and
    automatic pipettes should be calibrated on a
    regular basis.

24
Selecting a procedure or a protocol
25
Choose the appropriate test to your laboratory
  • As new tests and methods are developed for
    various analytes (antibodies or antigens), the
    most appropriate must be chosen for each
    laboratorys needs. A number of factors must be
    considered, including bias, specificity,
    sensitivity, precision, cost and ease of
    performance. Bias, specificity and sensitivity
    may be related. Frequently the more sensitive a
    test, the less specific it is. Bias may result
    from low specificity or sensitivity.

26
How to reduce the presence of Bias
  • To determine the presence of bias, the proposed
    method should be compared with other reliable
    methods, preferably with a standard method or
    clinical data. The same specimens should be run
    with both methods in the same laboratory and the
    results compared, although interlaboratory
    comparisons are also useful. If the results from
    the different methods do not agree, one must
    determine the reason for the difference and then
    decide which result is more useful.

27
What is clinical specificity
  • The clinical specificity of a method is evaluated
    by testing negative samples and samples
    containing substances which might cause
    interference. Closely related or cross-reacting
    substances frequently found in clinical specimens
    should be included.

28
Making Suitable Dilutions
100 ul serum in tube 1
Mix and Transfer
Discard
100ul diluent in each tube
Each tube is a 12 dilution of the previous tube
29
Selecting a Suitable Sample Dilution
Serial Dilutions on Abbott AxSYM HIV-1/HIV-2 MEIA
20
18
16
14
12
S/Co Ratio
10
8
6
Pos Cont 3.3
4
Cut Off 1.0
2
Neg Cont 0.38
0
Doubling Dilutions
30
What is clinical Sensitivity
  • The clinical sensitivity of a method being
    evaluated should be compared to that of other
    methods, but the purpose of the test must also be
    considered. In general, a definitive test need
    not be as sensitive as a screening test. The test
    should distinguish between normal and abnormal
    levels of analyze.

31
Evaluation on precision
  • The precision of a quantitative or
    Semiquantitative test must be evaluated in light
    of the precision required for the clinical
    application of the test results. Many factors
    affect precision, but one that is frequently
    overlooked in serologic tests is the size of the
    dilution increments. If all other variables are
    held constant, serologic tests tend to become
    less precise as the size of the dilution
    increment increases. For example, it should be
    expected that a test based on a four fold
    dilution would be less precise than the same test
    with a two-fold dilution.

32
Errors in measurement
  • True value - this is an ideal concept which
    cannot be achieved.
  • Accepted true value - the value approximating the
    true value, the difference between the two values
    is negligible.
  • Error - the discrepancy between the result of a
    measurement and the true (or accepted true
    value).

33
When you need a highly sensitive test
  • A test with maximum possible sensitivity is
    desirable when a disease is serious and its
    diagnosis should not be missed when the disease
    is treatable, and when false-positive results do
    not lead to serious problems. Similarly a test
    with maximum specificity is desirable when a
    disease is serious but is not treatable, the
    knowledge that the disease is absent has
    psychological or public health value, and
    false-positive results can lead to serious
    problems. A high predictive value of a positive
    test result is desirable when treatment of a
    false positive might have serious consequences.

34
Quality control of tests detecting antibodies
  • The performance of tests is monitored with
    controls. Antigenic serum panels as well as sera
    with known quantities of antibodies are available
    and should be routinely used. Correct performance
    of reagents is reflected by the expected reaction
    in tubes which lack one or more of the components
    necessary for the reaction. For example, the
    presence of anti-streptolysin O reagent is
    demonstrated by haemolysis in the tube containing
    the reagent buffer and cells but no antibody to
    inhibit haemolysis.

35
Antibody test
  • Flocculation test(RPR)
  • control procedures required
  • Nonreactive serum controlWeakly reactive serum
    controlReactive serum control
  • Expected results
  • No clumpingClumping of graded activityClumping
    of graded activity

36
Antibody test
  • Antibody test
  • Latex agglutination test(ASO)
  • Control procedures required
  • Negative control serumPositive control serum
  • Expected results
  • No clumpingClumping

37
Antibody test
  • Antibody test
  • Direct agglutination(Widal test, STA for
    Brucellosis)
  • Control procedures required
  • Antigen controlNegative control serumPositive
    control serum
  • Expected results
  • No clumpingNo clumpingClumping

38
Antibody test
  • Antibody test
  • Passive haemagglutination (ASO)
  • Control procedures required
  • Streptolysin controlRed cell control
  • Expected results
  • HemolysisNo hemolysis

39
Antigen test
  • Antigen test
  • Coagglutination test(Haemolytic streptococci
    meningitis antigens)
  • Control material
  • Group A,B,C streptococciN.meningitidis
  • Expected result
  • Agglutination with corresponding serum,

40
Quality control procedures for tests detecting
antigens
  • Antigen test
  • capsular Quelling reaction(Omni serum,
    H.influenzae type b)
  • Control material
  • PneumococciHaemolytic streptococciH.influenzae
    type bAcinetobacter anitratum
  • Expected result
  • Capsular swellingNo reactionCapsular
    swellingNo reaction

41
REPORTING AND RECORD KEEPING
  • Complete and accurate records must be maintained
    in a good quality assurance programme. These
    records should include personnel information
    details of equipment, preventive maintenance,
    service, and repair copies of reports to
    physicians or other clients accession records
    records of reagents and materials used records
    of observations made concurrently with the
    performance of each step in the examination of
    specimens proficiency testing results and
    internal quality control results.
  •  

42
Accuracy and Precision
  • The degree of fluctuation in the measurements is
    indicative of the precision of the assay.
  • The closeness of measurements to the true value
    is indicative of the accuracy of the assay.
  • Quality Control is used to monitor both the
    precision and the accuracy of the assay in order
    to provide reliable results.

43
Precision and Accuracy
  • Precise and inaccurate
  • Precise and accurate

44
Designing a QC Program
  • Establish written policies and procedures
  • Corrective action procedures
  • Train all staff
  • Design forms
  • Assure complete documentation and review

45
Qualitative QC
  • Quality control is performed for both, system is
    somewhat different
  • Controls available
  • Blood Bank/Serology/Micro
  • RPR/TPHA
  • Dipstick technology
  • Pregnancy

46
Establishing Control Ranges
  • Select appropriate controls
  • Assay them repeatedly over time
  • at least 20 data points
  • Make sure any procedural variation is
    represented
  • different operators
  • different times of day
  • Determine the degree of variability in the data
    to establish acceptable range

47
Measurement of Variability
  • A certain amount of variability will naturally
    occur when a control is tested repeatedly.
  • Variability is affected by operator technique,
    environmental conditions, and the performance
    characteristics of the assay method.
  • The goal is to differentiate between variability
    due to chance from that due to error.

48
Sources of error
  • Input data required - such as standards used,
    calibration values, and values of physical
    constants.
  • Inherent characteristics of the quantity being
    measured - e.g. CFT and HAI titer.
  • Instruments used - accuracy, repeatability.
  • Observer fallibility - reading errors, blunders,
    equipment selection, analysis and computation
    errors.
  • Environment - any external influences affecting
    the measurement.
  • Theory assumed - validity of mathematical methods
    and approximations.

49
Avoiding the errors
  • The accessioning and reporting system should
    minimize the possibility of clerical errors.
    Precautions should be taken to prevent reporting
    results on the wrong specimen and transposing
    digits in reporting quantitative data. The system
    should be so designed that the history associated
    with a sample can be reconstructed in detail if
    necessary. Who performed which tests, what
    reagents and lot numbers they used, what the
    control results were for that run, and how and
    when the results were reported should also be
    documented

50
Random Error
  • An error which varies in an unpredictable manner,
    in magnitude and sign, when a large number of
    measurements of the same quantity are made under
    effectively identical conditions.
  • Random errors create a characteristic spread of
    results for any test method and cannot be
    accounted for by applying corrections. Random
    errors are difficult to eliminate but repetition
    reduces the influences of random errors.
  • Examples of random errors include errors in
    pipetting and changes in incubation period.
    Random errors can be minimized by training,
    supervision and adherence to standard operating
    procedures.

51
Sources of error
  • Input data required - such as standards used,
    calibration values, and values of physical
    constants.
  • Inherent characteristics of the quantity being
    measured - e.g. CFT and HAI titer.
  • Instruments used - accuracy, repeatability.
  • Observer fallibility - reading errors, blunders,
    equipment selection, analysis and computation
    errors.
  • Environment - any external influences affecting
    the measurement.
  • Theory assumed - validity of mathematical methods
    and approximations.

52
How to implement a QC program?
  • Establish written policies and procedures
  • Assign responsibility for monitoring and
    reviewing
  • Train staff
  • Obtain control materials
  • Collect data
  • Set target values (mean, SD)
  • Establish Levey-Jennings charts
  • Routinely plot control data
  • Establish and implement troubleshooting and
    corrective action protocols
  • Establish and maintain system for documentation

53
Monitoring QC Data
  • Use Levey-Jennings chart
  • Plot control values each run, make decision
    regarding acceptability of run
  • Monitor over time to evaluate the precision and
    accuracy of repeated measurements
  • Review charts at defined intervals, take
    necessary action, and document

54
Internal Quality Control Program for Serological
Testing
  • An internal quality control program depend on the
    use of internal quality control (IQC) specimens,
    Shewhart Control Charts, and the use of
    statistical methods for interpretation.
  • Internal Quality Control Specimens
  • IQC specimens comprises either (1) in-house
    patient sera (single or pooled clinical samples),
    or (2) international serum standards with values
    within each clinically significant ranges.

55
For Articles of Interest on microbiology follow
me on
56
  • Created by Dr.T.V.Rao MD for e learning
    resources for Microbiologists in Developing World
  • Email
  • doctortvrao_at_gmail.com
Write a Comment
User Comments (0)
About PowerShow.com