A Rapid Analysis of Actinides In Urine Samples
Using Eichroms Resin Cartridges
  • Anil H.Thakkar
  • Eichrom Industries, Inc..
  • Darien, IL

2
Objective
  • To perform analysis for Am, Pu and U in urine
    samples using Eichroms new chromatographic
    configuration of UTEVA and TRU Resin

3
Current Eichrom Method
  • UTEVA and TRU Resin Column
  • Industry Standard
  • Flow Rate Limitation

4
Precedent Actinides in Water
  • 1 liter water samples
  • Coprecipitate with calcium phosphate
  • Dissolve in load solution
  • Load on UTEVA and TRU columns in tandem

5
Tandem Column Arrangement
Load
Discard
6
(No Transcript)
7
Eichrom TRU Resin Column Separation
Load 3 M HNO3, 1.0M Al(NO3)3
to waste
9m HCl 4M HCl
Am
0.1M ammonium bioxalate
Pu
8
Decontamination Factors (DF)
  • To determine the ruggedness of the method
  • To determine the degree of separation of the
    analyte from potentially interfering matrix
    components

9
Experiment for DF
  • Approximately 100 pCi (3.7 Bq) of each actinide
    was added to a load solution and run through the
    method.
  • DF in uranium, plutonium and americium fractions
    were calculated.

10
Equation for Decontamination Factor
  • DF Ao
  • Af
  • Where
  • Ao Original activity of the analyte added
  • Af Final activity of analyte in the fraction

11
Criteria for Acceptance
  • For actinides, a DF value of gt 1000 is
    acceptable
  • Tracer Recovery 70-105

12
Decontamination Factors (with column)
13
New ConfigurationResin Cartridge
  • Packed with 2 mL of resin
  • Cartridges with Luer lock
  • Ease of tandem arrangement for sequential
    analysis

14
Eichrom Vacuum Box System
  • 24 Hole clear polycarbonate vacuum box
  • Inexpensive cartridge connection
  • Yellow outer tip provides support
  • White inner tip provides excellent seal
  • Vacuum gauge
  • Plastic rack for twenty four 50 mL C Tubes
  • Eichrom has limited stock on hand

15
Decontamination Factors (with Cartridge)
16
Sample Preparation
  • 1.2 liter aliquots of acidified urine samples
    were spiked with Am, Pu and U
  • Samples were heated to near boiling for 2 hours
  • A Ca3(PO4)2 precipitation was performed
  • Decant supernatant, centrifuge and collect the
    precipitate
  • Perform a wet digestion on the precipitate until
    white ash

17
Load Solution for UTEVA and TRU Resin Columns
  • Dissolve precipitate in 3M HNO3/1M Al(NO3)3
    solution
  • Add ferrous sulfamate and ascorbic acid
  • Load on UTEVA and TRU Resin cartridge in tandem

18
Am/Cm, Pu and U Analyses
Calcium phosphate precipitation Wet ash
digestion Eichrom U/TEVA Resin U
Eichrom TRU Resin
Am/Cm, Pu Preparation for counting Alpha
spectrometry
19
Flow rate
  • Sample load Typically 1 mL/minute
  • Rinse solution 2-3 mL/minute
  • Strip solution 1 mL/minute
  • Achievable with vacuum pump

20
Recoveries for Direct Spikes
21
Recoveries in DI Water
22
Recoveries in Simulated Hanford Ground Water
23
Recoveries in Synthetic Urine Samples
24
(No Transcript)
25
Recoveries in Real Urine Samples
26
(No Transcript)
27
(No Transcript)
28
PROCORADFrench Intercomparison Samples
29
(No Transcript)
30
Summary
  • Eichrom Resin cartridges provide fast
    separation for actinides
  • Good, reproducible and consistent recoveries
    achievable
  • Rapid turn-around possible

31
Eichrom Cartridges
  • Available in TRU, TEVA, UTEVA and Sr
  • Other Resins available by special order
  • 2 mL Resin volume
  • Standard S grade, 50-100 micron beads
  • Can be stacked using Luer connections
  • Standard syringe barrel used for reservoir
  • Individually labeled w/ lot number

32
Future Objectives
  • Evaluate effect of faster flow rate on chemical
    recovery
  • To test other matrices such as fecal, soil and
    vegetation for actinides using Eichrom Resin
    cartridges
  • To evaluate separation of Sr using Eichrom Sr
    Resin cartridges