Title: RAPID MEASUREMENT OF GASEOUS VOC MIXTURES WITH A NEW MASS SPECTROMETRY TECHNIQUE Robert Walters Moll
1RAPID MEASUREMENT OF GASEOUS VOC MIXTURES WITH A
NEW MASS SPECTROMETRY TECHNIQUERobert
WaltersMolly GribbJerome ImonigieBoise State
University Geoenvironmental Research
Labcoen.boisestate.edu/gerlPresented at the
2004 INRA Environmental Subsurface Science
Symposium
2Rapid VOC Measurement
- The Big Picture
- Our Approach
- The Payoff
- Future Directions
3Rapid VOC Measurement
- Our team is developing a subsurface probe that
- detects and identifies gaseous volatile organic
compounds (VOCs) in the unsaturated zone, - is inexpensive and minimally-invasive,
-
- provides real-time and long-term sampling,
- is deployed underground in networked arrays,
-
- performs wireless reporting and allows remote
access.
4Rapid VOC Measurement
- The high-resolution ion mobility spectrometer
(HRIMS) probe
5Rapid VOC Measurement
The HRIMS probe is first validated in the lab in
a large soil column using comparison measurements
with the GC/MS.
Once fully validated, the HRIMS probe will be
deployed in the field using a direct-push truck.
Approximately 65 cm in length, the HRIMS probe
can be deployed to a depth of up to 100 feet for
over six months
6Rapid VOC Measurement
- The Approach
- Cyclic Valving
- Continuous sampling from soil column effluent
- Maintain packet integrity from soil column to
detector - Limit peak spreading in GC column
- GC/MS in Selective Ion Monitoring (SIM) mode
- Isothermal chromatographic column conditions
- Resolve co-eluting peaks
- Compare mixtures with single components
7Rapid VOC Measurement
An external timer is programmed to cycle the
sampling valve at a specific interval. The
sample passes through the GC column and into the
MS for quantitation.
8The flow controller and a small soil column
during a test.
9Rapid VOC Measurement
Traditional GC/MS method
The new SIM-CV method
10Rapid VOC Measurement
INTENSITY
RETENTION TIME min
11Rapid VOC Measurement
- CYCLIC VALVING
- Continuous sampling using external timer
- Packet integrity maintained
- Timer interval and measured interval are equal
- No peak spreading
- No skewness or flattening of peaks was observed
when comparing traditional to SIM-CV
12(No Transcript)
13Rapid VOC Measurement
- GC/MS in SIM MODE
- Thermal effects
- Pronounced skewing at temps below 300 ºC
- GC column temp gt 300 ºC
- Co-eluting peaks
- Look at secondary and tertiary ions
- Mixtures vs. single components
- Examine toluene in mixture and as single specie
14Rapid VOC Measurement
- BTEX mixtures produce six ions of interest
- m/z 77.05 Benzene 78.05, 77.05
- 78.05 Toluene 91.10, 92.05
- 91.10 E.benzene 91.10, 106.1
- 92.05 m-Xylene 91.10, 106.1, 105.1
- 105.1 o-Xylene 91.10, 106.1, 105.1
- 106.1 p-Xylene 91.10, 106.1, 105.1
15(No Transcript)
16(No Transcript)
17Rapid VOC Measurement
18(No Transcript)
19(No Transcript)
20Rapid VOC Measurement
- Why Use SIM-CV?
- Utilizes existing equipment and configuration
- A high measurement rate allowing breakthrough
tests at higher interstitial velocities - A lower detection limit than GC/FID for low
concentration/residual studies - High selectivity, allowing simultaneous, multiple
compound testing
21Rapid VOC Measurement
- SIM-CV is suitable for gaseous transport studies.
However, the SIM-CV method can be further
refined by - Extend the deconvolution method to capture all
components - Characterization of a wider range of volatiles
(aromatics, alcohols, etc.) - Investigation of multiple, competitive
interactions (sorption/desorption and reactions)
22Partners
- Funded by the United States EPA under Contract
no. X97031101-0 - Shimadzu Scientific Instruments