Programming - PowerPoint PPT Presentation

1 / 15
About This Presentation
Title:

Programming

Description:

A-dec 500/300 * Programming, Deluxe Touchpad (2006 Present) * The Handpiece Cop The system allows more than one handpiece to be out of its holder at a time ... – PowerPoint PPT presentation

Number of Views:91
Avg rating:3.0/5.0
Slides: 16
Provided by: Trai55
Category:

less

Transcript and Presenter's Notes

Title: Programming


1
Programming
  • A-dec 500/300

2
Learning Objectives
  • As a result of this training you will be able to
  • Identify an A-dec 500 touchpad from an A-dec 300
    touchpad
  • Configure handpieces on an A-dec 500 and A-dec
    300 unit

3
A-dec 500 Touchpads
  • Standard Touchpad, Assistants or Doctors
  • Deluxe Touchpad (2004 2006)
  • Endo Deluxe (May 2006)

4
Standard Touchpad Pg 52
  • System Status LED
  • Dental Light Control with 3-Way Switching
  • Manual Chair Positioning
  • Program Button
  • Presets
  • Cup Fill / Bowl Rinse
  • Only if a Cuspidor is installed
  • Auxiliary Buttons
  • A-dec relay functionality only if a Cuspidor is
    not installed

5
Chair Programming Pg 53
  • Position Programming
  • Position the chair to the desired operating
    position, using the manual controls (arrows) on
    the touchpad or foot switch.
  • Press and release Program button (one beep).
  • Within four seconds, press a position button
    (three beeps).
  • Position 3 (X-ray/Rinse)
  • Press and hold Program and Position 3at the same
    time for three seconds. One beep confirms that
    the button 3 is now fully programmable.
  • Note Disabling the x-ray/rinse function is a
    global function for Dr. A and B.

6
Cuspidor Functions Pg 54
  • Cuspidor Factory Presets
  • Cup fill 2.5 seconds
  • Bowl rinse 30 seconds

7
Stand-alone Cuspidor
  • After March, 2008
  • Hold BOTH cup fill and bowl rinse buttons on the
    cuspidor tower down for 3 seconds. Release BOTH
    buttons.
  • Select cup fill OR bowl rinse to program for the
    desired (program) time.
  • When the desired program time (amount of water)
    has been achieved, release the button. Three
    beeps indicates the cup fill or bowl rinse is now
    programmed.

8
Handpiece Configuration
2
3
  • Overview

1
0
4
9
Handpiece Configuration
  • Press Program and Doctor A/B buttons

10
Deluxe Touchpad Pg 56
  • Allows switching between Standard and Endo use of
    the EA51LT Electric Motor
  • Mode change
  • Select with E button
  • 4 memory settings
  • M12,000 RPM
  • M210,000 RPM
  • M320,000 RPM
  • M440,000RPM

11
Deluxe Touchpad Pg 57
  • Dual Operator Support
  • Auxiliary Buttons
  • Cup Fill / Bowl Rinse
  • Only if a Cuspidor is installed
  • If a Cuspidor is not installed,buttons are
    inactive
  • Water Coolant / Air Coolant
  • Electric Handpiece / Ancillary Controls

12
Programming Pg 58-60
13
Deluxe Touchpad Displays
  • Turbine selected
  • More than one of theBig Three selected
  • Scaler
  • Electric Handpiece
  • Pneumatic Handpiece
  • No Holdback Air
  • The power is on but there isnt anyair to the
    unit

14
Touchpad Programming
  • Factory Preset Functions
  • Disable/Enable the Footswitch or Touchpad
  • Place the spare jumper in the EN/DIS TP/FS
    position of the test points header.
  • Push the buttons to be disabled/enabled (PRGM,
    PRGM 0, PRGM 1, PRGM 2, PRGM 3).
  • One beep means the button is disabled.
  • Three beeps means the button is enabled.
  • Place the spare jumper back in the spare position.

15
A-dec 300 Deluxe Touchpad
  • Programmable chair positions
  • X-ray/rinse button
  • Cupfill/bowl rinse
  • Auto Light
  • Electric handpiece operation
  • Endodontics

16
Deluxe Touchpad Setup
17
Thank you!
  • What questions do you have?
Write a Comment
User Comments (0)
About PowerShow.com