Title: PRELIMINARY RESULTS from PIE 2005 the PAN Intercomparison Exercise F' Flocke, G' Tyndall, E' Apel NC


1
PRELIMINARY RESULTSfromPIE 2005the PAN
Intercomparison ExerciseF. Flocke, G. Tyndall,
E. Apel (NCAR)E. Williams (NOAA)
2
PIE 2005
  • 6/28-7/8/2005
  • Enchanted Mesa
  • eight different instruments
  • from Canada, Germany, Japan, the UK, and the US
  • five different methods
  • ambient and standards measurements

3
results arePRELIMINARY
4
Instruments
  • Hiroshi Tanimoto, NIES, Japan GC/MS
  • Graham Mills, UEA, UK, GC/ECD
  • Stefan Gilge, DWD, Germany, GC/ECD
  • Bernhard Rappenglueck and Rainer Schmitt, U.
    Houston and Metcon, Germany, GC/ECD
  • Jason OBrien, Environment Canada, GC/PDD
  • Ron Cohen, Paul Wooldridge, Anne Perring,
    UCBerkeley, TD-LIF
  • Jim Roberts, NOAA, GC/ECD
  • Aaron Swanson, NCAR and Saewung Kim, GaTech, CIMS
  • (Frank Flocke, NCAR, Dual-GC/ECD)

5
(No Transcript)
6
Other Instruments
  • Meteorological parameters (NOAA)
  • NO, NO2, O3 (NOAA)
  • NOy reference instrument in referee trailer (NCAR)

7
Enchanted Mesa
8
The forbidden zone
Flow 1220 slpm
standards
ambient
GC5
Internet Access, email
9
(No Transcript)
10
(No Transcript)
11
(No Transcript)
12
(No Transcript)
13
(No Transcript)
14
UEA GC/ECD
  • Restek MXT-200 Column
  • 0.53 ID, 10 m, 1.0µ film
  • 22 sccm
  • 17 ºC isothermal
  • cycle time 10 minutes
  • PAN, PPN

15
(No Transcript)
16
Env. CAN GC/PDD
  • Restek MXT-200 Column
  • 0.53 ID, 8 m, 1.0µ film
  • 35 sccm
  • 15 ºC isothermal
  • cycle time 10 minutes
  • PAN, PPN

17
(No Transcript)
18
NOAA GC/ECD
  • Restek Rtx-200 Column
  • 0.53 ID, 7.5 m, 1.0µ film
  • 10 sccm
  • 15 ºC isothermal
  • cycle time 5 minutes
  • PAN, PPN, APAN, MPAN, PiBN

19
(No Transcript)
20
NIES GC/MS
  • DB-1 MS Column
  • 0.25 ID, 10 m, 0.25µ film
  • 2 sccm
  • cycle time 10 minutes
  • PAN, PPN, APAN, MPAN, PiBN, PnBN

21
(No Transcript)
22
UH/Metcon GC/ECDDWD GC/ECD
  • DB1/DB-5 Column combination
  • 0.53 ID, 10 m total
  • 20 sccm
  • 15 ºC isothermal
  • PAN, PPN
  • cycle time 10 minutes
  • DWD GC has modified flow control (PAN only)

23
(No Transcript)
24
UC Berkeley TD-LIF
  • measures NO2 by LIF
  • has 3 thermal decomposition stages held at 150,
    300 and 550 degrees decomposing all PANs, ANs,
    NOy into NO2
  • cycle time 10 seconds
  • measures the sum of all PANs

25
(No Transcript)
26
NCAR/GaTech CIGAR
  • measures PANs by CIMS
  • PAN, PPN, PBN, APAN, MPAN, PBzN, MoPAN, PPeN, ...
  • cycle time 2 seconds
  • scans should show additional PANs unless they are
    very sticky

27
(No Transcript)
28
NCAR 2 Channel GC/ECD(reference instrument)
  • 2 RTX-200 Columns
  • 7.5m, 0.53 ID, 1.0µ film
  • 20 sccm
  • 20 ºC isothermal
  • cycle time 3 minutes
  • PAN, PPN, APAN, MPAN, PiBN

29
(No Transcript)
30
Measurements
  • all measurements from common manifold
  • ambient air
  • PAN and PPN standards in dry zero-air from
    manifold
  • PAN and PPN standards in humidified zero air
  • i-Propyl nitrate (interferent) in zero-air

31
PRELIMINARY
32
Ambient Measurements
33
6/30/05 trace
34
6/30/05 trace
35
7/4/05 trace
36
7/4/05 trace
37
All ambient data
38
All ambient data
39
All ambient data
40
All ambient data
41
i-Propyl nitrate
42
PPN ambient
43
PPN ambient
44
PPN ambient
45
Standards
46
Standards
47
Standards
48
Standards
49
Standards
50
PRELIMINARY
51
LIF and the sum of PANs
52
6/30 ambient trace
53
7/1 ambient trace
54
7/5 ambient trace
55
Standard comparison with LIF
56
Conclusions (for now)
  • very good agreement for ambient PAN measurements,
    considering its field data
  • acceptable agreement for ambient PPN measurements
    (might improve)
  • LIF measures consistently higher than sum of
    individual PANs (expected?)
  • extra PANs that GCs and CIMS miss?
  • no indication of LIF counting ANs as PANs
  • LIF NO2 background subtraction problem?
  • more work has to be done to produce final data
    and merge measurements together
  • compare MPAN, APAN, PBN