Reduction in Exposure Rates during Realtime Computerized Axial Tomography Interventional Procedures - PowerPoint PPT Presentation

1 / 20
About This Presentation
Title:

Reduction in Exposure Rates during Realtime Computerized Axial Tomography Interventional Procedures

Description:

Bismuth Shield on Anthropomorphic Phantom with Biopsy Needle in Place ... To use the bismuth surgical drapes with various lead equivalencies to determine ... – PowerPoint PPT presentation

Number of Views:64
Avg rating:3.0/5.0
Slides: 21
Provided by: oir3
Category:

less

Transcript and Presenter's Notes

Title: Reduction in Exposure Rates during Realtime Computerized Axial Tomography Interventional Procedures


1
Reduction in Exposure Rates during Real-time
Computerized Axial Tomography Interventional
Procedures
  • Dave Tripp, Ph.D.
  • Karen Langley, M.S.
  • William Orrison, M.D.
  • Department of Radiology
  • University of Utah Health Sciences Center
  • Salt Lake City, UT
  • Radiological Health Department
  • University of Utah
  • Salt Lake City, UT

2
Introduction
  • CT fluoroscopy or fluoro CT provides real-time
    reconstruction and display of CT images.
  • This modality has advantages over conventional CT
    in performing certain types of interventional
    procedures such as chest biopsies.
  • Primarily by reducing the time required to
    perform complex interventional procedures.
  • However, fluoro CT can produce large increases in
    primary radiation dose to patients and secondary
    radiation (scatter) to attending staff by a
    factor of 10-20 times that of conventional
    fluoroscopy.

3
Fluoro - CT Entrance and Scatter Exposure Rate
Measurements
4
Possible Remedies
  • By reducing imaging times
  • By reducing x-ray technique factors such as kVp
    and mA.
  • By using a portable lead shield adjacent to the
    imaging plane to shield attending staff from
    scattered radiation.

5
Response
  • Imaging time is a function of difficulty of
    procedure, anatomy of patient, and radiologist
    technique. In our Department, imaging time has
    been cut to less than 3 min.
  • The fluoroscopic technique factors cannot be
    lowered below certain values in order to maintain
    minimum image quality. At the lowest technique
    used at our institution, the dose rates is still
    10 times conventional fluoroscopy.
  • The use of a portable lead shield creates
    problems with maintaining sterile conditions, and
    blocking physicians access to the biopsy site.

6
Alternate Solution?
  • Place a sterile drape over the biopsy site.
  • Use a surgical drape that is composed of a
    relatively high Z material -Bismuth in our case
    (which is easily disposed of)- to shield
    attending staff from scattered radiation and
    partially shield the patient from primary
    radiation.
  • Cut a small opening in the surgical drape for the
    purpose of biopsy needle placement.

7
Bismuth Shield on Anthropomorphic Phantom with
Biopsy Needle in Place
8
Purpose of the Investigation
  • With the use of an anthropomorphic chest phantom
    determine the patient entrance dose rates, and
    attending staff scatter dose rates.
  • To use the bismuth surgical drapes with various
    lead equivalencies to determine their
    effectiveness in
  • reducing patient entrance exposure
  • reducing exposure to attending staff to scatter,
    and
  • in determining the degree of associated image
    degradation.

9
Experimental Methods
  • Patient entrance dose was determined using TLDs
    on top of and underneath the drapes.
  • Attending staff exposure to scatter was measured
    using an ion chamber placed at the approximate
    position of the physicians face.
  • 18-gauge biopsy needle placed in phantom lung
    field was imaged to evaluate the degree of
    distortion.

10
Results
  • Patient entrance dose
  • skin entrance dose decreased 65-75 in the region
    covered by the drape using the lowest CT
    technique with the optimal surgical drape.
  • Obviously, the remainder of the body not covered
    by drape would not receive the the benefit of the
    skin entrance dose reduction.

11
Results (continued)
  • Scattered radiation to attending staff
  • scatter rate also significantly reduced
  • scatter rate not significantly reduced beyond 0.2
    mm lead equivalency.
  • Scatter from the phantom other than that
    attenuated by drape seemed to dominate beyond 0.2
    mm lead equivalency.

12
CT Fluoro Scatter Rate and Reduction as a
Function of Shielding (lead equivalency)
13
Results (continued)
  • Image degradation
  • bismuth drapes with up to 0.4 mm lead equivalency
    produced acceptable images
  • image degradation in area of needle puncture site
    was minimal
  • image of needle in the region of interest (biopsy
    site) was of sufficient quality so as not to
    impede the required precision in placing the
    needle in the lesion.

14
Axial CT Image of Chest Phantom with Simulated
Needle Biopsy - No Shielding
15
With 0.273 mm Lead Equivalency Shielding
16
With 0.359 mm Lead Equivalency Shielding
17
With 0.430 mm Lead Equivalency Shielding
18
Discussion
  • The advent of CT fluoro has provided a more rapid
    and easy means of performing interventional
    procedures when a detailed and real-time 3
    dimensional image is of benefit.
  • Even when used under optimum conditions, the
    exposure rates to patient and staff are
    significantly greater than either conventional CT
    or most routine fluoroscopic procedures.
  • The use of sterile surgical drapes loaded with
    bismuth is a simple way to reduce staff exposure
    by as much as 50 without significant degradation
    of the patient image.

19
Discussion (continued)
  • The problem associated with placing a shield on
    the patient to prevent scatter during
    conventional fluoroscopy is the significant
    degradation of the image as well as an increase
    in skin entrance dose caused by the auto
    brightness control associated with a fluoroscopy
    unit.
  • However, with CT fluoro the image is not acquired
    using x-rays emanating from a single direction.
  • In fact, the CT fluoro image is not reconstructed
    for each rotation, but rather every 60o . Thus
    the remaining views make it possible to obtain
    adequate image quality in the area of interest.

20
Conclusion
  • The use of CT fluoro has definite benefits that
    may include significant savings in procedural
    time, ease of access, and in some cases, even
    procedural viability.
  • The trade-off is the potential for excessive
    exposure of patient and staff to increased
    amounts of radiation.
  • Our investigation shows that with the use of
    shielding materials, such as bismuth -containing
    sterile surgical drapes can lower dose to the
    patient, and provides significantly less
    occupational exposure to attending staff while
    maintaining a sterile field and allowing the
    radiologist easy access to the biopsy site.
Write a Comment
User Comments (0)
About PowerShow.com