Title: Reduction in Exposure Rates during Realtime Computerized Axial Tomography Interventional Procedures
1Reduction in Exposure Rates during Real-time
Computerized Axial Tomography Interventional
Procedures
- Dave Tripp, Ph.D.
- Karen Langley, M.S.
- William Orrison, M.D.
- Department of Radiology
- University of Utah Health Sciences Center
- Salt Lake City, UT
- Radiological Health Department
- University of Utah
- Salt Lake City, UT
2Introduction
- CT fluoroscopy or fluoro CT provides real-time
reconstruction and display of CT images. - This modality has advantages over conventional CT
in performing certain types of interventional
procedures such as chest biopsies. - Primarily by reducing the time required to
perform complex interventional procedures. - However, fluoro CT can produce large increases in
primary radiation dose to patients and secondary
radiation (scatter) to attending staff by a
factor of 10-20 times that of conventional
fluoroscopy.
3Fluoro - CT Entrance and Scatter Exposure Rate
Measurements
4Possible Remedies
- By reducing imaging times
- By reducing x-ray technique factors such as kVp
and mA. - By using a portable lead shield adjacent to the
imaging plane to shield attending staff from
scattered radiation.
5Response
- Imaging time is a function of difficulty of
procedure, anatomy of patient, and radiologist
technique. In our Department, imaging time has
been cut to less than 3 min. - The fluoroscopic technique factors cannot be
lowered below certain values in order to maintain
minimum image quality. At the lowest technique
used at our institution, the dose rates is still
10 times conventional fluoroscopy. - The use of a portable lead shield creates
problems with maintaining sterile conditions, and
blocking physicians access to the biopsy site.
6Alternate Solution?
- Place a sterile drape over the biopsy site.
- Use a surgical drape that is composed of a
relatively high Z material -Bismuth in our case
(which is easily disposed of)- to shield
attending staff from scattered radiation and
partially shield the patient from primary
radiation. - Cut a small opening in the surgical drape for the
purpose of biopsy needle placement.
7Bismuth Shield on Anthropomorphic Phantom with
Biopsy Needle in Place
8Purpose of the Investigation
- With the use of an anthropomorphic chest phantom
determine the patient entrance dose rates, and
attending staff scatter dose rates. - To use the bismuth surgical drapes with various
lead equivalencies to determine their
effectiveness in - reducing patient entrance exposure
- reducing exposure to attending staff to scatter,
and - in determining the degree of associated image
degradation.
9Experimental Methods
- Patient entrance dose was determined using TLDs
on top of and underneath the drapes. - Attending staff exposure to scatter was measured
using an ion chamber placed at the approximate
position of the physicians face. - 18-gauge biopsy needle placed in phantom lung
field was imaged to evaluate the degree of
distortion.
10Results
- Patient entrance dose
- skin entrance dose decreased 65-75 in the region
covered by the drape using the lowest CT
technique with the optimal surgical drape. - Obviously, the remainder of the body not covered
by drape would not receive the the benefit of the
skin entrance dose reduction.
11Results (continued)
- Scattered radiation to attending staff
- scatter rate also significantly reduced
- scatter rate not significantly reduced beyond 0.2
mm lead equivalency. - Scatter from the phantom other than that
attenuated by drape seemed to dominate beyond 0.2
mm lead equivalency.
12CT Fluoro Scatter Rate and Reduction as a
Function of Shielding (lead equivalency)
13Results (continued)
- Image degradation
- bismuth drapes with up to 0.4 mm lead equivalency
produced acceptable images - image degradation in area of needle puncture site
was minimal - image of needle in the region of interest (biopsy
site) was of sufficient quality so as not to
impede the required precision in placing the
needle in the lesion.
14Axial CT Image of Chest Phantom with Simulated
Needle Biopsy - No Shielding
15With 0.273 mm Lead Equivalency Shielding
16With 0.359 mm Lead Equivalency Shielding
17With 0.430 mm Lead Equivalency Shielding
18Discussion
- The advent of CT fluoro has provided a more rapid
and easy means of performing interventional
procedures when a detailed and real-time 3
dimensional image is of benefit. - Even when used under optimum conditions, the
exposure rates to patient and staff are
significantly greater than either conventional CT
or most routine fluoroscopic procedures. - The use of sterile surgical drapes loaded with
bismuth is a simple way to reduce staff exposure
by as much as 50 without significant degradation
of the patient image.
19Discussion (continued)
- The problem associated with placing a shield on
the patient to prevent scatter during
conventional fluoroscopy is the significant
degradation of the image as well as an increase
in skin entrance dose caused by the auto
brightness control associated with a fluoroscopy
unit. - However, with CT fluoro the image is not acquired
using x-rays emanating from a single direction. - In fact, the CT fluoro image is not reconstructed
for each rotation, but rather every 60o . Thus
the remaining views make it possible to obtain
adequate image quality in the area of interest.
20Conclusion
- The use of CT fluoro has definite benefits that
may include significant savings in procedural
time, ease of access, and in some cases, even
procedural viability. - The trade-off is the potential for excessive
exposure of patient and staff to increased
amounts of radiation. - Our investigation shows that with the use of
shielding materials, such as bismuth -containing
sterile surgical drapes can lower dose to the
patient, and provides significantly less
occupational exposure to attending staff while
maintaining a sterile field and allowing the
radiologist easy access to the biopsy site.