Title: Quickvue In-Line rapid strep a testing


1
Quickvue In-Line rapid strep a testing
Click here to move on
WellOne Primary Medical and Dental Care
  • For Medical Clinical Staff

2
WARNING
  • Conducting this test exposes the operator to
    potentially
  • infectious material. Standard precautions,
    including glove
  • use, are required for this procedure.
  • this test exposes the operator to potentially
    infectious material. Standard precautions,
    including glove use are required for this
    procedure.
  • Controls are potentially infectious and should be
    handled in the same manner as a patient specimen,
    using all appropriate precautions.

Click here to move on
3
Rapid Strep A Testing Overview
  • Pharyngitis, can be caused by streptococci from
    serogroups other than Group A, as well as other
    pathogens. The rapid strep test will not detect
    these other groups and pathogens.
  • The Quick Vue Strep test will not
    differentiate asymptomatic carriers of Group A
    Streptococcus from those exhibiting streptococcal
    infection.
  • The test is designated as a screening test at
    WellOne since, if clinical signs and symptoms are
    not consistent with test results, follow-up
    laboratory performed culture procedure should be
    performed.

Click here to move on
4
Question (True or False) A negative Rapid Strep
test is proof positive that the patient does not
have strep throat.
Click here for this answer choice
TRUE
Click here for this answer choice
FALSE
Click here to go back and review previous slide
5
The correct answer is FALSE
  • The rapid strep test is 92 sensitive (meaning
    that 8 out of 100 persons with strep throat will
    have a negative rapid strep test)
  • If clinical signs and symptoms are not consistent
    with test results, follow-up laboratory performed
    culture procedure should be performed.

Click here to move on
Click here to go back and review previous slide
6
Rapid Strep A Testing Storage and Stability
  • Reagents should be stored at room temperature, in
    the laboratory 59 ?F -86?F (15 ?C -30?C) and out
    of direct sunlight. The laboratory room
    temperature must be checked daily and action
    taken for out of range temperatures.
  • Kit contents are stable until the expiration date
    printed on the outer box.
  • Swab specimens should be processed as soon as
    possible after collection and the test kit not
    opened until just prior to testing.
  • The Test Cassette must remain sealed in the
    protective foil pouch until just prior to use.

Click here to move on
Click here to go back and review previous slide
7
Rapid Strep A Testing Initial Control Testing
  • Each test kit comes with two control swabs (one
    positive and one negative) that must be checked
    prior to use of a new box of tests
  • Be sure that controls have been run and are
    acceptable for the test kit that you are using
    prior to patient testing
  • If results are not as expected (ie. Positive
    control is negative and/or negative control is
    positive) DO NOT USE BOX. Segregate box of tests
    and notify supervisor immediately.
  • Document the results of control testing on the
    strep testing log form
  • Use a new log form for each new box of tests
    opened

Click here to move on
Click here to go back and review previous slide
8
Steps in Testing General
  • Identify patient using two patient identifiers
    (name and DOB)
  • Explain procedure to patient in culturally
    appropriate, age appropriate terms.
  • Gather all necessary equipment and supplies prior
    to start of the procedure.
  • Inspect all equipment and supplies for damage,
    general integrity and expiration date- do not use
    any damaged or expired materials
  • Perform hand hygiene before and after patient
    contact before and after donning gloves.

Click here to move on
Click here to go back and review previous slide
9
Before conducting a strep test, the tester
must
  • Ensure that the lot number on the test kits
    begins with 1003
  • External controls have been run on the test kit
    and the control results were acceptable
  • The green lettering on the box has not faded

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
10
The correct answer is
  • External controls have been run on the test kit
    and the control results were acceptable.
    External controls must be run on each test kit
    prior to use and recorded on the strep testing
    log.

B
Click here to move on
Click here to go back and review previous slide
11
Steps in Testing Collecting the Specimen
  • Collect the specimen Use ONLY the swabs
    supplied with the test kit for Strep testing
    (these swabs have a green handle)

Avoid the uvula as stimulation may result in
gagging
Stand off to the side of the patient to collect
the specimen (helpful if the patient coughs or
gags)
Take the sample from both sides of the throat at
the tonsillar areas
Click here to move on
Click here to go back and review previous slide
12
The specimen should be collected from
  • The uvula
  • The posterior
    gyros
  • Both sides of
    the tonsillar areas

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
13
The correct answer is both sides of the tonsillar
areas
Avoid the uvula as stimulation may result in
gagging
Take the sample from both sides of the throat at
the tonsillar areas
Click here to move on
Click here to go back and review previous slide
14
The proper test swabs to use for Rapid Strep A
testing are
  • Wooden with a
    cotton tip
  • Supplied with
    the kit and have a green handle
  • Supplied with
    the kit and have a wooden handle

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
15
The correct answer is
Supplied with the kit and have a green handle
Click here to go back and review previous slide
Click here to move on
16
Strep Testing Standing Orders for MA Staff
  • Collect and test specimen on all patients
    presenting with sore throat or patients stating
    that they have recently been exposed to strep and
    are ill.

Click here to go back and review previous slide
Click here to move on
17
Steps in Testing Testing the sample
  • Label with patients name (or initials) and date
    of birth
  • Open a test cassette and place on a clean, level
    surface.
  •  Insert the swab completely into the swab
    chamber.

Click here to move on
Click here to go back and review previous slide
18
Steps in Testing Testing the sample
  • Inspect the extraction solution bottle and do not
    use if the solution is not clear. DO NOT USE
    EXTRACTION SOLUTION IF IT IS GREEN PRIOR TO
    BREAKING AMPULE.
  • Squeeze carefully to crush the glass ampule
    inside the extraction solution bottle.

Click here to move on
Click here to go back and review previous slide
19
Steps in Testing Testing the sample
  • Vigorously shake the bottle 5 times to mix
    solution. Solution should turn green after the
    ampule is broken (do not use if green color is
    not present).
  • Solution must be used immediately.

Click here to move on
Click here to go back and review previous slide
20
The extraction solution bottle should be crushed
and __________ prior to use.
  • shaken
    vigorously 5 times
  • warmed
  • cooled
  • frozen

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
21
The correct answer is shaken 5 times
SHAKE 5 TIMES
SHAKE 5 TIMES
SHAKE 5 TIMES
SHAKE 5 TIMES
SHAKE 5 TIMES
Click here to go back and review previous slide
Click here to move on
22
The extraction solution before crushing should
  • Be clear green
  • Be pink
  • Contain a green hue
  • Be Clear

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
23
The correct answer is be clear..
  • If the solution is GREEN before crushing, the
    ampule has already been crushed before you
    handled it.
  • Since the solution must be used immediately after
    mixing, the extraction bottle must be discarded.

Click here to move on
Click here to go back and review previous slide
24
The extraction solution (after crushing) should
  • Not be used if it is not
    clear green
  • Be pink
  • Contain a black hue
  • Be orange

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
25
The correct answer is clear and green
  • Before crushing and mixing the ampule, the
    extraction solution should be
  • After crushing and mixing the ampule, the
    extraction solution should be clear (not cloudy)
    and

CLEAR
GREEN
Click here to move on
Click here to go back and review previous slide
26
Steps in Testing Testing the sample
  • Remove the cap from the bottle and quickly fill
    the chamber to the rim.
  • Begin timing (5 minutes)
  • If liquid has not moved across the Result Window
    in 1 minute, completely remove the swab and
    re-insert. If liquid still does not move across,
    retest with a new specimen, Test Cassette and
    Extraction Solution Bottle.
  • The Test Cassette should not be moved until the
    assay is complete.

Click here to move on
Click here to go back and review previous slide
27
Steps in Testing Testing the sample
Read and record results at 5 minutes verbally
alert provider of positive results Discard
test Record information on Strep testing
log Check procedure on encounter form
Click here to move on
Click here to go back and review previous slide
28
Steps in Testing Interpreting Results
  • Positive Result
  • The appearance of any pink - to purple line next
    to the letter T in the Result window, along
    with the blue control Line next to the letter C
    means that the test is positive for group A
    Streptococcus.
  • C for control Even a light line
    at the T is positive
  • T for Test

Click here to move on
Click here to go back and review previous slide
29
Steps in Testing Interpreting Results
The Control line is vivid blue- indicating an
acceptable test
The Test line is light pink indicating a positive
test
Click here to go back and review previous slide
Click here to move on
30
Steps in Testing Interpreting Results
  • Negative Result
  • The appearance of only the blue control Line next
    to letter C in the Result Window means that the
    test is negative. A negative QuickVue result
    means that the swab is presumptive negative for
    group A Streptococcus. A line at the C
    acceptable test
  • No line at the T
    negative test

Click here to go back and review previous slide
Click here to move on
31
Steps in Testing Interpreting Results
No line at the C Test is not acceptable- Discard
and repeat If second test does show a control
line, place the box out of use and notify
supervisor
Click here to move on
Click here to go back and review previous slide
32

A blue control line next to the letter C means
that

The test is positive This was a good (or
reliable) test The patient has strep
throat The results are not valid
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
33
The correct answer is This was a good (or
reliable) test
  • For more on controls, place mouse here

Click here to move on
Click here to go back and review previous slide
34

Any shade of a light pink to purple line at the
T means

The test is positive This was a good (or
reliable) test The patient has mono The
results are not valid
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
35
The correct answer is
  • The test is positive

Click here to move on
Click here to go back and review previous slide
36
Steps in Testing Interpreting Results
  • A negative background control is provided by the
    clearing of background color in the result window
    and indicates that there were no immunological
    interfering substances in the specimen. This
    area should be white to light pink within 5
    minutes.

Click here to move on
Click here to go back and review previous slide
37
Rapid strep results should be read at
  • 20 minutes
    5 minutes
  • 8 minutes
  •  

Click here for this answer choice
Click here for this answer choice
Click here for this answer choice
Click here to go back and review previous slide
Click here to move on
38
Steps in Testing Testing the sample
Read and record results at 5 minutes
Click here to move on
Click here to go back and review previous slide
39
You have reached the end of this program..
  • Click below to view a video on
  • QuickVue Rapid A Strep Testing
  • http//www.quidel.com/assets/swf/StrepATrainingvid
    eoHQ.htm
  • Once you are confident that you are comfortable
    with all the materials presented, proceed to the
    QuickVue Rapid A Strep Testing post test at
    http//www.classmarker.com/professional/

Start Over
Click here to go back and review previous slide
Exit