Title: Integrating A Non-invasive Blood Pressure Monitor to the AYUSHMAN testbed
1Integrating A Non-invasive Blood Pressure Monitor
to the AYUSHMAN testbed
- Faculty Advisor
- Sandeep K. S. Gupta
- Associate Professor (CSE)
- Arizona State University
- Undergrad Team
- Tomas A. Velasquez - NIBP Team
- Waleed F. Masoom - NIBP Team
- Mohammad Lootah - Tiny-OS Team
- Koshin Mariano - Tiny-OS Team
- Graduate Mentor
- Valliappan Annamalai
2Project Overview
- Select and integrate with AYUSHMAN testbed a
Non-Invasive Blood Pressure (NIBP) monitoring
system that measures - Systolic Pressure
- Diastolic Pressure
- Arterial Pressure
- Heart Rate
- Broadcast procured data over a wireless network
to specified remote Host station and PDAs.
3NIBP Device Selection Criteria
- Measures Systolic, Diastolic, and Arterial
Pressure, and Heart Rate - Bi-Directional RS-232 Communication
- Meets Medical Standards
- Energy-Efficient
- Motion-Tolerant
4Final NIBP Device Candidates
Advantage Series Module 2.0
MJ-23 CNIBP Module
SunTech
Medwave
5Standards
- AAMI SP10 - 1992
- Association for the Advancement of Medical
Instrumentation - EN 60601-1-2-30
- International Standard for EMC
- MIL-STD 810 C, D, and E
- U.S. Department of Defense Signal transmission
6Device Selection Rating
Noninvasive Blood Pressure(NIBP) Noninvasive Blood Pressure(NIBP) Noninvasive Blood Pressure(NIBP) Noninvasive Blood Pressure(NIBP) Noninvasive Blood Pressure(NIBP) Noninvasive Blood Pressure(NIBP)
Selection Criteria MJ23 MJ23 ADVANGE ADVANGE
Weight Rating Weighted Score Rating Weighted Score
Communication Link 15 5 0.75 5 0.75
Compact 15 3 0.45 3 0.45
Operation Range 10 3 0.3 3 0.3
Alerts 5 3 0.15 3 0.15
Storage Conditions 5 3 0.15 3 0.15
Data Storage 5 3 0.15 3 0.15
Multifunctional 10 3 0.3 3 0.3
Battery Operated 15 0 0 5 0.75
Total Rank 2.75 2.75 3.5 3.5
7The Winner
Advantage Series Module 2.0
SunTech
8Battery Life
- Voltage Use
- The device operates between 6vdc and 14.5vdc. The
device will not operate when the voltage is below
5.3vdc. Starting with a voltage of 6vdc the
device will operate 20 times between battery
changes, the batteries would sustain the amps
required, but the voltage would drop to 5.3vdc.
When changed to 9vdc the last count of successful
operation was 45 times and counting. The team
used off the shelf plastic battery holder that
hold AA cells, with standard snap connection for
ease of change.
9NIBP Operational Data
START UP CURRENT
10NIBP Operational Data
PUMP OPERATION
VALVE OPERATION
?10 SEC
11MICA2 Motes
Mote Communication (916 MHz, 500 ft.)
PDA
BPControlBase
RS-232 Serial Communication
NIBP Device