Cardiovascular%20Monitoring%20during%20Anesthesia - PowerPoint PPT Presentation

About This Presentation
Title:

Cardiovascular%20Monitoring%20during%20Anesthesia

Description:

Five-electrode system: one on each lead and one precordial lead (V5, along the ... Akinesia: absence of wall motion. Dyskinesia: paradoxical movement. 82 ... – PowerPoint PPT presentation

Number of Views:988
Avg rating:3.0/5.0
Slides: 85
Provided by: Hong100
Category:

less

Transcript and Presenter's Notes

Title: Cardiovascular%20Monitoring%20during%20Anesthesia


1
Cardiovascular Monitoring during Anesthesia
  • ???????
  • ???

2
Electrocardiogram (EKG)
  • One of the standardized monitors during any form
    of anesthesia.
  • For detection and diagnosis of
  • dysrhythmias
  • conduction defects
  • cardiac ischemia
  • electrolyte disturbance

3
Types of Monitoring
  • Three-electrode system right arm, left arm, left
    leg white is always right, black is on the left,
    red is even lower.
  • (Racial discrimination!!)

4
Types of Monitoring
  • Five-electrode system one on each lead and one
    precordial lead (V5, along the anterior axillary
    line in the fifth intercostal space ? for
    detection of anterior ischemia).

5
Electrocardiogram (EKG)
  • Kaplan et al. (Anesthesiology, 1976)
  • 90 of intraoperative cardiac ischemia will be
    detected by multiple EKG, especially V5. At least
    two leads should be simultaneously showed on the
    monitor.

6
Position of the Leads
  • The four limb leads should be placed on the back
    of shoulders and hips, where they will disturb
    the operative field the least.
  • ???????????
  • Every lead should be fixed and protected with
    tape to prevent dislodgement of leads during
    operation.

7
Artifact Source of EKG
  • loss of insulation
  • motion artifacts
  • crossing cables (especially the pulse oximeter
    cable, which transmits an amplified signal.)
  • Electrocauterization

8
Electronic Filtering System
  • 1. Monitoring mode 0.540 Hz, eliminates high
    and low frequency artifacts, but distorts the
    height of QRS and the degree of ST depression.
  • 2. Diagnostic mode 0.05100 Hz, does not filter
    high frequency artifacts.

9
Non-invasive Blood Pressure (NIBP) Monitoring
  • Drawbacks inaccuracy, intermittent data,
    requirement of a pulsatile flow.
  • Common methods of detection of change in flow
  • 1) Auscultation of Korotkoff sounds
  • 2) Microprocessor-assisted
    interpretation of
  • the oscillations.

10
(from Essential Noninvasive Monitoring in
Anesthesia. New York Grune and Stratton, 1980)
11
Non-invasive Blood Pressure (NIBP) Monitoring
  • Familiarity with various type of the automated
    monitors is of prime importance.

12
Cuff Size
  • Too small cuff will result in high blood pressure
    reading.
  • A loosely applied cuff will also produce a
    reading higher than it should be.
  • Too large cuff will severely distorted the
    reading obtained.

13
Cuff Size
  • from Barbara Bates A Guide to Physical
    Examination

14
Intravascular Pressure Measurement
  • Components include intravascular catheter,
    fluid-filled connector tubing, transducer,
    electronic analyzer, display system.
  • Rapid establishment of pressure-transducer system
    is essential in emergent conditions.

15
(No Transcript)
16
Definitions of Terms
  • Frequency Response (Amplitude ratio) Amplitude
    of input vs. output. Ideally 1.
  • Natural frequency (fn, or Resonant frequency)
    the frequency at which the monitor system itself
    resonates and amplifies the signal.
  • Common natural frequency of transducer system is
    1020 Hz.

17
Intravascular Monitoring
  • The natural frequency of the measuring system
    should be at least 10 times of the fundamental
    frequency (HR).
  • When heart rate is fast (more close to natural
    frequency of the system), wave can be abnormally
    amplified
  • ?Damping is necessary.

18
(No Transcript)
19
  • D diameter L length of tubing
  • ? density of the fluid ?P/?V compliance

20
Damping Coefficient
  • Damping coefficient the rate of dissipation of
    the energy of a pressure wave. It can correct the
    erroneous amplification of under-damped system.

21
(No Transcript)
22
(No Transcript)
23
Flush Test
  • The damping coefficient can be evaluated by
    flush test(pop test).
  • After a high-pressure flush, an under-damping
    system will continue to oscillate for a long
    period of time.
  • ? Overestimation of SBP and under-estimation of
    DBP.

24
Flush Test
  • An over-damped system will not oscillate after a
    flush but will return to baseline slowly.
  • An optimally (or critically) damped system will
    oscillate one or two times only and will
    reproduce BP accurately.

25
(No Transcript)
26
Sources of Errors
  • 1)Low-frequency responses Air in line or failure
    in flush device with formation of partial clot in
    catheter. ? Over-damping.
  • 2)Catheter whip Motion of catheter tip itself
    produces a noticeable pressure swing not common
    in A-line but common in PA catheter.

27
Source of Errors
  • 3)Resonance in peripheral vessels The systolic
    pressure measured in a radial artery may be up to
    2050 mmHg higher than in the central aorta.
  • 4)Change in electronic balance electronic
    zeroing should be done periodically to preclude
    baseline drift (for example due to change in
    room temperature).

28
(No Transcript)
29
Source of Errors
  • 5)Transducer position error ideally at right
    atrium. (Midaxillary line in supine patients.)
    Changes may be less significant in blood pressure
    but significant in CVP or PA pressure/wedge
    pressure.

30
Air Bubbles in Line
  • Air bubbles can result in a lower frequency
    response and greater resonance response.
  • Small amount may augment systolic pressure
    reading while large amount cause an over-damped
    system.

31
Site for Arterial Cannulation
  • Radial artery Modified Allens test
  • Apply pressure over both radial and ulnar
    arteries simultaneously and ask the patient to
    squeeze his or her hand several times to promote
    exsanguination. Then release the pressure on the
    ulnar side and measure the time for refill of the
    nailbed of the capillary. If refill time is
    greater than 15 sec, it is a positive test,
    indicating inadequate collateral flow.

32
Site for Arterial Cannulation
  • Femoral artery excellent access to central
    arterial system. Insertion with a Seldinger
    method.
  • Brachial artery medial in antecubital fossa, but
    since it is an end artery, its cannulation is
    relatively contraindicated. However the incidence
    of thromboembolism is low.

33
Site for Arterial Cannulation
  • Dorsalis pedis and posterior tibialis artery
    distortion of the artery wave.
  • External temporary artery rarely used

34
Choice of Cannulation Site Special Considerations
  • Blalock-Taussig shunt contralateral to the shunt
  • Coarctation of Aorta right radial/brachial is
    preferred
  • Thoracic aortic aneurysm right radial
  • Ascending aorta surgery left radial
  • Liver transplantation two cannulations are
    required

35
Complications of Arterial Catheterization
  • Ischemia incidence is reportedly low.
  • Thrombosis incidence may be high but no adverse
    sequelae were reported.
  • Infection incidence is higher in femoral
    cannulation.
  • Bleeding

36
Complications of Arterial Catheterization
  • False lowering of radial artery pressure
    immediately after cardio-pulmonary bypass up to
    72 immediately after cardiopulmonary bypass the
    radial artery pressure is significantly lower
    than the aortic pressure. Mechanism
    Vasodilatation? Hypovolemia? AV shunting?? If an
    arterial trace is damped after cardiopulmonary
    bypass, a direct pressure measurement should be
    obtained from central site.

37
(No Transcript)
38
Central Venous Pressure (CVP)
  • measures right atrial pressure
  • A wave with P wave, atrial contraction
  • C wave QRS complex, tricuspid bulging into RA
  • X descent when tricuspid valve is pulling down
    during the latter stage of ventricular systole
    indicating emptying of blood from ventricle.

39
Central Venous Pressure (CVP)
  • V wave after T wave, RA filling before tricuspid
    opening.
  • Y descent when tricuspid valve opens and atrium
    begins to empty.
  • TR giant V wave replacing C wave
  • af absence of A wave

40
Central Venous Pressure (CVP)
  • AV dissociation frequent cannon A wave (atrial
    contraction over a closed tricuspid valve)
  • Rapid and exaggerated x, y descent in
    constrictive pericarditis.

41
CVP values
  • a direct measurement of RV filling pressure, CVP
    is good measurement of LV filing pressure only in
    the absence of pulmonary hypertension or mitral
    disease.

42
Site of Insertion
  • Internal jugular vein, external jugular vein,
    subclavian vein, femoral vein
  • PICC peripheral inserted central catheter,
    inserted from antecubital vein (or basilic,
    cephalic vein).

43
  • Measurement should be done at end-expiration
    phase.

44
CVP and PEEP
  • The effect of PEEP is rarely significant when
    PEEP is less than 7.5 cmH2O.
  • PEEP may cause increased CVP value while
    decreases cardiac output.
  • Discontinuation of PEEP for measurement of CVP is
    not recommended.

45
Pulmonary Artery Catheter
  • It can measure
  • PA pressure
  • Pulmonary capillary wedged pressure (PCWP) a
    balloon at catheter tip (volume 1.5 ml), when the
    balloon is inflated and the vessel is wedged, a
    valveless hydrostatic column exists between the
    distal port and LA.

46
Pulmonary Artery Catheter
  • CVP a port for CVP measurement is located at 30
    cm from the tip
  • Cardiac output measurement of RV output
  • Blood temperature
  • Derived hemodynamic data
  • Mixed Venous O2 saturation (SvO2)

47
Pulmonary Artery Catheter
  • PA Port YELLOW
  • CVP Port BLUE
  • PA balloon Port PINK

48
Waveform during Insertion
49
Length of Insertion
  • Usual conditions (just for ease to memorize)
  • 35 cm RV
  • 45cm PA
  • 55 cm wedge
  • But the actual length may vary greatly between
    patients!

50
Hemodynamic Measurements--Normal Range
  • RA pressure 26 mmHg
  • RV pressure systolic 1525 mmHg diastolic 04
    mmHg
  • PA pressure systolic 1525 mmHg diastolic 816
    mmHg
  • Mean PAP 1020 mmHg
  • Pulmonary Capillary Wedge Pressure (PCWP) 612
    mmHg

51
Derived Hemodynamic Profiles
  • Systemic Vascular Resistance (SVR) 80 x
    (MAP-CVP)/CO 8001200 dyne-sec-cm-5
  • Pulmonary Vascular Resistance (PVR) 80 x
    (PAP-PCWP)/CO 20130 dyne-sec-cm-5

52
Derived Hemodynamic Profiles
  • Cardiac Output thermodilution method 48 L/min
  • Cardiac Index CO/BSA 2.54.2 L/min/m2

53
(No Transcript)
54
Guidelines for Injection
  • Injection port CVP
  • Solution Normal Saline
  • Volume 5 or 10 ml, usually 10 ml, set up in the
    machine before injection
  • Injection time better within 4 sec. Prolonged
    injection will cause low measurement.
  • Consecutive 3 measurements are required.

55
(No Transcript)
56
Continuous Cardiac Output
  • Continuous Cardiac Output (CCO) measurement can
    be achieved by a electric coil attached on the
    tip of PA catheter. It automatically measures CO
    every 3 min.

57
Mixed Venous Oxygen Saturation (SvO2)
  • Mixed by blood from both SVC and IVC, sampled at
    PA
  • O2 consumption SaO2-SvO2
  • ?SvO2SaO2 - (VO2/Q x Hb x 13)

58
SvO2SaO2 - (VO2/Q x Hb x 13)
  • Causes for decreasing SvO2
  • Hypoxemia
  • Increased Metabolic Rate
  • Anemia (Blood loss)
  • Low Cardiac Output
  • ????????, SvO2???CO???.

59
(No Transcript)
60
(No Transcript)
61
Intravascular Volume
Blood Pressure
Vascular Compliance
62
TEE (Transesophageal Echocardiography)
  • A vibrating piezoelectric crystal at the probe
    tip both produces and receives the ultrasound
    wave. The time required for the wave to travel
    through biological structures, together with the
    intensity of reflected waves, provides
    information about the size and the intensity of
    the structures.

63
TEE (Transesophageal Echocardiography)
  • Waves can be processed to analyze the frequency
    shift (Doppler effect) between the emitted and
    reflected waves to provide the direction and the
    velocity of blood flow.

64
Modes of Imaging
  • M-mode one-dimensional view
  • 2D mode multiplanar view
  • Doppler echocardiography

65
Doppler Echocardiography
  • Pulse-wave (PW) Doppler repetitive bursts of
    ultrasound focused at a precise location,
    drawback limitation for high-velocity flow
  • Continuous-wave (CW) Doppler two transducers,
    one emits one receives. It measures velocity
    accurately but sampling location is unknown.

66
Doppler Echocardiography
  • A frequency shift is produced when sound waves
    are reflected by a moving target.
  • V ?fc/2f0 x cos ?
  • It can detect the velocity of blood flow,
    producing both visual and audible signals

67
Doppler Echocardiography
  • Color flow mapping
  • Flow toward the probe red
  • Flow away from the probe blue

68
Doppler Echocardiography
  • Turbulent flow contains RBCs of different
    velocities. Doppler flow showed a shaggy pattern,
    known as spectral broadening.

69
TEE Contraindications
  • Absolute
  • Esophageal disease
  • Active upper GI bleeding
  • Severe C-spine instability

70
(No Transcript)
71
(No Transcript)
72
Pressure Gradient
  • Peak pressure gradient can be calculated by a
    modified Bernoulli equation
  • ?P 4 x V2

73
Pressure Half Time
  • The degree of mitral stenosis can be assessed by
    pressure half time (T 1/2).
  • Mitral Area 220/ T 1/2

74
(No Transcript)
75
TEE Contraindications
  • Relative
  • Recent gastroesophageal operation
  • Esophageal varices
  • Severe cervical arthritis
  • Dysphagia or odynophagia

76
Complications and Risks
  • Major complications are unusual (0.180.5).
  • Esophageal tear/hemorrhage
  • Transient vocal cord paralysis
  • Sore throat with pharyngitis

77
Basic Views of TEE
78
Basic Views of TEE
  • Mid-papillary view

79
Basic Views of TEE
  • Aortic short-axis view

80
Regional Wall Motion Abnormality (RWMA)
  • A more sensitive indicator of myocardial ischemia
    than ECG changes.
  • A transient RWMA is consistent with ischemia a
    persistent change is consistent with infarction.

81
Regional Wall Motion Abnormality (RWMA)
  • Normal contraction greater than 30 of
    shortening of the radius from the center of the
    ventricle to the endocardial border.
  • Mild hypokinesia shortening 1030
  • Severe hypokinesia less than 10
  • Akinesia absence of wall motion
  • Dyskinesia paradoxical movement

82
Regional Wall Motion Abnormality (RWMA)
83
Duomo, Milan, Italy
84
Suggested Reading
  • Paul L. Marino The ICU book
  • Anesthesia Secrets, J. Duke and ST Rosenberg,
    Mosby
  • A Practical Approach to Cardiac Anesthesia, FA
    Hensley Jr., DE Martin, Little Brown
Write a Comment
User Comments (0)
About PowerShow.com