Title: Cardiovascular%20Monitoring%20during%20Anesthesia
1Cardiovascular Monitoring during Anesthesia
2Electrocardiogram (EKG)
- One of the standardized monitors during any form
of anesthesia. - For detection and diagnosis of
- dysrhythmias
- conduction defects
- cardiac ischemia
- electrolyte disturbance
3Types of Monitoring
- Three-electrode system right arm, left arm, left
leg white is always right, black is on the left,
red is even lower. - (Racial discrimination!!)
4Types of Monitoring
- Five-electrode system one on each lead and one
precordial lead (V5, along the anterior axillary
line in the fifth intercostal space ? for
detection of anterior ischemia).
5Electrocardiogram (EKG)
- Kaplan et al. (Anesthesiology, 1976)
- 90 of intraoperative cardiac ischemia will be
detected by multiple EKG, especially V5. At least
two leads should be simultaneously showed on the
monitor.
6Position of the Leads
- The four limb leads should be placed on the back
of shoulders and hips, where they will disturb
the operative field the least. - ???????????
- Every lead should be fixed and protected with
tape to prevent dislodgement of leads during
operation.
7Artifact Source of EKG
- loss of insulation
- motion artifacts
- crossing cables (especially the pulse oximeter
cable, which transmits an amplified signal.) - Electrocauterization
8Electronic Filtering System
- 1. Monitoring mode 0.540 Hz, eliminates high
and low frequency artifacts, but distorts the
height of QRS and the degree of ST depression. - 2. Diagnostic mode 0.05100 Hz, does not filter
high frequency artifacts.
9Non-invasive Blood Pressure (NIBP) Monitoring
- Drawbacks inaccuracy, intermittent data,
requirement of a pulsatile flow. - Common methods of detection of change in flow
- 1) Auscultation of Korotkoff sounds
- 2) Microprocessor-assisted
interpretation of - the oscillations.
10(from Essential Noninvasive Monitoring in
Anesthesia. New York Grune and Stratton, 1980)
11Non-invasive Blood Pressure (NIBP) Monitoring
- Familiarity with various type of the automated
monitors is of prime importance.
12Cuff Size
- Too small cuff will result in high blood pressure
reading. - A loosely applied cuff will also produce a
reading higher than it should be. - Too large cuff will severely distorted the
reading obtained.
13Cuff Size
- from Barbara Bates A Guide to Physical
Examination
14Intravascular Pressure Measurement
- Components include intravascular catheter,
fluid-filled connector tubing, transducer,
electronic analyzer, display system. - Rapid establishment of pressure-transducer system
is essential in emergent conditions.
15(No Transcript)
16Definitions of Terms
- Frequency Response (Amplitude ratio) Amplitude
of input vs. output. Ideally 1. - Natural frequency (fn, or Resonant frequency)
the frequency at which the monitor system itself
resonates and amplifies the signal. - Common natural frequency of transducer system is
1020 Hz.
17Intravascular Monitoring
- The natural frequency of the measuring system
should be at least 10 times of the fundamental
frequency (HR). - When heart rate is fast (more close to natural
frequency of the system), wave can be abnormally
amplified - ?Damping is necessary.
18(No Transcript)
19- D diameter L length of tubing
- ? density of the fluid ?P/?V compliance
20Damping Coefficient
- Damping coefficient the rate of dissipation of
the energy of a pressure wave. It can correct the
erroneous amplification of under-damped system.
21(No Transcript)
22(No Transcript)
23Flush Test
- The damping coefficient can be evaluated by
flush test(pop test). - After a high-pressure flush, an under-damping
system will continue to oscillate for a long
period of time. - ? Overestimation of SBP and under-estimation of
DBP.
24Flush Test
- An over-damped system will not oscillate after a
flush but will return to baseline slowly. - An optimally (or critically) damped system will
oscillate one or two times only and will
reproduce BP accurately.
25(No Transcript)
26Sources of Errors
- 1)Low-frequency responses Air in line or failure
in flush device with formation of partial clot in
catheter. ? Over-damping. - 2)Catheter whip Motion of catheter tip itself
produces a noticeable pressure swing not common
in A-line but common in PA catheter.
27Source of Errors
- 3)Resonance in peripheral vessels The systolic
pressure measured in a radial artery may be up to
2050 mmHg higher than in the central aorta. - 4)Change in electronic balance electronic
zeroing should be done periodically to preclude
baseline drift (for example due to change in
room temperature).
28(No Transcript)
29Source of Errors
- 5)Transducer position error ideally at right
atrium. (Midaxillary line in supine patients.)
Changes may be less significant in blood pressure
but significant in CVP or PA pressure/wedge
pressure.
30Air Bubbles in Line
- Air bubbles can result in a lower frequency
response and greater resonance response. - Small amount may augment systolic pressure
reading while large amount cause an over-damped
system.
31Site for Arterial Cannulation
- Radial artery Modified Allens test
- Apply pressure over both radial and ulnar
arteries simultaneously and ask the patient to
squeeze his or her hand several times to promote
exsanguination. Then release the pressure on the
ulnar side and measure the time for refill of the
nailbed of the capillary. If refill time is
greater than 15 sec, it is a positive test,
indicating inadequate collateral flow.
32Site for Arterial Cannulation
- Femoral artery excellent access to central
arterial system. Insertion with a Seldinger
method. - Brachial artery medial in antecubital fossa, but
since it is an end artery, its cannulation is
relatively contraindicated. However the incidence
of thromboembolism is low.
33Site for Arterial Cannulation
- Dorsalis pedis and posterior tibialis artery
distortion of the artery wave. - External temporary artery rarely used
34Choice of Cannulation Site Special Considerations
- Blalock-Taussig shunt contralateral to the shunt
- Coarctation of Aorta right radial/brachial is
preferred - Thoracic aortic aneurysm right radial
- Ascending aorta surgery left radial
- Liver transplantation two cannulations are
required
35Complications of Arterial Catheterization
- Ischemia incidence is reportedly low.
- Thrombosis incidence may be high but no adverse
sequelae were reported. - Infection incidence is higher in femoral
cannulation. - Bleeding
36Complications of Arterial Catheterization
- False lowering of radial artery pressure
immediately after cardio-pulmonary bypass up to
72 immediately after cardiopulmonary bypass the
radial artery pressure is significantly lower
than the aortic pressure. Mechanism
Vasodilatation? Hypovolemia? AV shunting?? If an
arterial trace is damped after cardiopulmonary
bypass, a direct pressure measurement should be
obtained from central site.
37(No Transcript)
38Central Venous Pressure (CVP)
- measures right atrial pressure
- A wave with P wave, atrial contraction
- C wave QRS complex, tricuspid bulging into RA
- X descent when tricuspid valve is pulling down
during the latter stage of ventricular systole
indicating emptying of blood from ventricle.
39Central Venous Pressure (CVP)
- V wave after T wave, RA filling before tricuspid
opening. - Y descent when tricuspid valve opens and atrium
begins to empty. - TR giant V wave replacing C wave
- af absence of A wave
40Central Venous Pressure (CVP)
- AV dissociation frequent cannon A wave (atrial
contraction over a closed tricuspid valve) - Rapid and exaggerated x, y descent in
constrictive pericarditis.
41CVP values
- a direct measurement of RV filling pressure, CVP
is good measurement of LV filing pressure only in
the absence of pulmonary hypertension or mitral
disease.
42Site of Insertion
- Internal jugular vein, external jugular vein,
subclavian vein, femoral vein - PICC peripheral inserted central catheter,
inserted from antecubital vein (or basilic,
cephalic vein).
43- Measurement should be done at end-expiration
phase.
44CVP and PEEP
- The effect of PEEP is rarely significant when
PEEP is less than 7.5 cmH2O. - PEEP may cause increased CVP value while
decreases cardiac output. - Discontinuation of PEEP for measurement of CVP is
not recommended.
45Pulmonary Artery Catheter
- It can measure
- PA pressure
- Pulmonary capillary wedged pressure (PCWP) a
balloon at catheter tip (volume 1.5 ml), when the
balloon is inflated and the vessel is wedged, a
valveless hydrostatic column exists between the
distal port and LA.
46Pulmonary Artery Catheter
- CVP a port for CVP measurement is located at 30
cm from the tip - Cardiac output measurement of RV output
- Blood temperature
- Derived hemodynamic data
- Mixed Venous O2 saturation (SvO2)
47Pulmonary Artery Catheter
- PA Port YELLOW
- CVP Port BLUE
- PA balloon Port PINK
48Waveform during Insertion
49Length of Insertion
- Usual conditions (just for ease to memorize)
- 35 cm RV
- 45cm PA
- 55 cm wedge
- But the actual length may vary greatly between
patients!
50Hemodynamic Measurements--Normal Range
- RA pressure 26 mmHg
- RV pressure systolic 1525 mmHg diastolic 04
mmHg - PA pressure systolic 1525 mmHg diastolic 816
mmHg - Mean PAP 1020 mmHg
- Pulmonary Capillary Wedge Pressure (PCWP) 612
mmHg
51Derived Hemodynamic Profiles
- Systemic Vascular Resistance (SVR) 80 x
(MAP-CVP)/CO 8001200 dyne-sec-cm-5 - Pulmonary Vascular Resistance (PVR) 80 x
(PAP-PCWP)/CO 20130 dyne-sec-cm-5
52Derived Hemodynamic Profiles
- Cardiac Output thermodilution method 48 L/min
- Cardiac Index CO/BSA 2.54.2 L/min/m2
53(No Transcript)
54Guidelines for Injection
- Injection port CVP
- Solution Normal Saline
- Volume 5 or 10 ml, usually 10 ml, set up in the
machine before injection - Injection time better within 4 sec. Prolonged
injection will cause low measurement. - Consecutive 3 measurements are required.
55(No Transcript)
56Continuous Cardiac Output
- Continuous Cardiac Output (CCO) measurement can
be achieved by a electric coil attached on the
tip of PA catheter. It automatically measures CO
every 3 min.
57Mixed Venous Oxygen Saturation (SvO2)
- Mixed by blood from both SVC and IVC, sampled at
PA - O2 consumption SaO2-SvO2
- ?SvO2SaO2 - (VO2/Q x Hb x 13)
58SvO2SaO2 - (VO2/Q x Hb x 13)
- Causes for decreasing SvO2
- Hypoxemia
- Increased Metabolic Rate
- Anemia (Blood loss)
- Low Cardiac Output
- ????????, SvO2???CO???.
59(No Transcript)
60(No Transcript)
61Intravascular Volume
Blood Pressure
Vascular Compliance
62TEE (Transesophageal Echocardiography)
- A vibrating piezoelectric crystal at the probe
tip both produces and receives the ultrasound
wave. The time required for the wave to travel
through biological structures, together with the
intensity of reflected waves, provides
information about the size and the intensity of
the structures.
63TEE (Transesophageal Echocardiography)
- Waves can be processed to analyze the frequency
shift (Doppler effect) between the emitted and
reflected waves to provide the direction and the
velocity of blood flow.
64Modes of Imaging
- M-mode one-dimensional view
- 2D mode multiplanar view
- Doppler echocardiography
65Doppler Echocardiography
- Pulse-wave (PW) Doppler repetitive bursts of
ultrasound focused at a precise location,
drawback limitation for high-velocity flow - Continuous-wave (CW) Doppler two transducers,
one emits one receives. It measures velocity
accurately but sampling location is unknown.
66Doppler Echocardiography
- A frequency shift is produced when sound waves
are reflected by a moving target. - V ?fc/2f0 x cos ?
- It can detect the velocity of blood flow,
producing both visual and audible signals
67Doppler Echocardiography
- Color flow mapping
- Flow toward the probe red
- Flow away from the probe blue
68Doppler Echocardiography
- Turbulent flow contains RBCs of different
velocities. Doppler flow showed a shaggy pattern,
known as spectral broadening.
69TEE Contraindications
- Absolute
- Esophageal disease
- Active upper GI bleeding
- Severe C-spine instability
70(No Transcript)
71(No Transcript)
72Pressure Gradient
- Peak pressure gradient can be calculated by a
modified Bernoulli equation - ?P 4 x V2
73Pressure Half Time
- The degree of mitral stenosis can be assessed by
pressure half time (T 1/2). - Mitral Area 220/ T 1/2
74(No Transcript)
75TEE Contraindications
- Relative
- Recent gastroesophageal operation
- Esophageal varices
- Severe cervical arthritis
- Dysphagia or odynophagia
76Complications and Risks
- Major complications are unusual (0.180.5).
- Esophageal tear/hemorrhage
- Transient vocal cord paralysis
- Sore throat with pharyngitis
77Basic Views of TEE
78Basic Views of TEE
79Basic Views of TEE
80Regional Wall Motion Abnormality (RWMA)
- A more sensitive indicator of myocardial ischemia
than ECG changes. - A transient RWMA is consistent with ischemia a
persistent change is consistent with infarction.
81Regional Wall Motion Abnormality (RWMA)
- Normal contraction greater than 30 of
shortening of the radius from the center of the
ventricle to the endocardial border. - Mild hypokinesia shortening 1030
- Severe hypokinesia less than 10
- Akinesia absence of wall motion
- Dyskinesia paradoxical movement
82Regional Wall Motion Abnormality (RWMA)
83Duomo, Milan, Italy
84Suggested Reading
- Paul L. Marino The ICU book
- Anesthesia Secrets, J. Duke and ST Rosenberg,
Mosby - A Practical Approach to Cardiac Anesthesia, FA
Hensley Jr., DE Martin, Little Brown