Title: The Current Challenges in the North American Reprocessing Field and Proposed Solution!
1The Current Challenges in the North American
Reprocessing Field and Proposed Solution!
- By Colleen Landers
- Registered Nurse
- Consultant
2Objectives
- Define the major issues facing reprocessing
departments. - Provide options to solve these issues.
- Refer to the standards that reprocessing
departments require to solve these issues.
3Challenges
- Cleaning and sterilization instructions.
- Turn around demands.
- Sterilization extended cycles.
- Staff orientation, education and competency
testing. - Weight of Sets.
- Loaner Sets.
- Flash Sterilization
4Medical Procedures
- First Do no Harm
- New Millennium Age of the Internet and
Litigation!!(The Public are informed and
demanding more - of HealthCare).
5Cleaning and Sterilization of Medical Devices
- Medical Devices have become increasingly complex
as more sophisticated and less invasive surgical
procedures are developed. - The manufacturers validated reprocessing
instructions for these devices have not kept pace
with the requirements listed in the Canadian and
American standards.
6Cleaning and Sterilization of Medical Devices
- Along with this issue, the increased demand to do
more surgical procedures, requires faster turn
around of instruments to meet the surgical
demand. - Concern regarding the safety of patients
receiving healthcare in North America has
required the standards to be reviewed and
requirements increased.
7Cleaning of Medical Devices Who is responsible
for What??
- Manufacturers validate that instrument can be
reliably cleaned and sterilized/disinfected and
is therefore re-usable. - Users verify that cleaning sterilization
equipment is working properly and that
in-hospital cleaning sterilization methods are
consistently performed and verified.
8Medical Device Specific Instructions for
Reprocessing
9Manufacturer recommended cleaning that is not
validated!
- Ivory Snow!
- No detergent! (only sterile distilled water)
- Use regular hospital cleaning protocol!
10Crud Issue
Patient-used (no side-panel) un cleaned
Patient-used repeatedly side-panel removed
11Show Me the MONEY!!
- STERILE Crud???- Acetabular reamers- Electronic
drill Hand pieces - No infection risk ? ? What is the issue?
- Cost/Benefit What is realistic???
12Sterile Crud ???
- Foreign organic material- Endotoxin (LPS) from
dead bacteria- blood and other organic debris
from previous patient or from cleaning process
(e.g. enzymatic detergent residuals, water
organisms etc). - Adverse antigenic reaction to foreign organic
material (e.g. implants LPS). - Foreign body Nidus for endogenous infection.
Xing et al. Accumulation of LPS by polyethylene
particles decreases bone attachment to implants.
J Orthopedic Res 200624959-966
13Cleaning and Sterilization of Medical Devices
- Health Care Facilities need to make sound
decisions when purchasing a reusable medical
device. - Documents that guide you are
- ISO 17664 Information to be provided by the
manufacturer for reprocessing of a resterilizable
medical devices. (CSA Z17664) - AAMI ST81 Sterilization of medical devices
Information to be provided by the manufacturer
for the processing of resterilizable medical
devices.
14Comply with ISO 17664 AAMI ST81Validated
Cleaning Methods
Coiled reamer
shafts
Alternative solid shaft
Crud from between coils
15Information provided by manufacturers
- Adequate and appropriate cleaning and
sterilization. - Instruments with cannulation (lumens or holes)
these instructions need to be device specific
include ultrasonic. If the device retracts or
requires disassembly include pictures. - Instruments with crevices (i.e. nuts, bolts,
broaches, cutting blocks) device specific. - Instruments that are hinged (i.e. pliers,
cutters, forceps) instructions to include - Cleaning brush size and type
- Ultrasonic
- Lubrication if required type of lubricant that
is recommended.
16Information provided by manufacturers
- Flexible reamers Instructions need to be device
specific including pictures. - Manufacturers to give maximum number of
reprocessing times for implants (i.e. screws,
plates) how to identify when replacement is
required. - Manufacturers to review design. Cleaning tools
should be provided by manufacturers. - Ensure that when manufacturers update, revise
and/or change sterilization instructions for
medical devices the new procedures are
communicated to the current users in a systematic
documented fashion.
17WHY?
- Health Care facilities require a policy and
procedure requiring all information on
reprocessing from the manufacturer of the device
prior to reprocessing cleaning, decontamination,
wrapping and sterilization. - Required prior to purchase, lease, trial or
evaluation or purchase. - Why?
- To protect the facility from error and
protect patients from risk.
18POLICY PROCEDURE FOR ORDERING REUSABLE MEDICAL
DEVICES
- Requires the following
- Purchase approved by the Reprocessing Manager and
Infection Prevention and Control prior to
evaluation, consignment, loaned/leased or
purchases. - Written manufacturers instructions for cleaning,
wrapping and sterilization reviewed and
attainable by Central Processing department
before purchase. - In-service for cleaning and sterilization shall
be provided to all staff, prior to the medical
device being placed in circulation for use on a
patient.
19POLICY PROCEDURE FOR ORDERING REUSABLE MEDICAL
DEVICES
- The decisions to make changes to cleaning
products, purchases of new equipment for surgical
procedures, etc., shall be performed by a
Product Evaluation Team endorsed by the OEM. - All medical devices that will be loaned/leased,
used on consignment, evaluated or purchased will
be reviewed and a decision made by the Product
Evaluation Team.
20Why?
- To determine if the medical device can be cleaned
and sterilized in the facility before evaluation,
consignment, loaned/leased or purchased in the
healthcare facility. - To ensure that the reprocessing will meet the
requirement for organism kill. - To prevent risk to patients and staff.
21Solution
- If healthcare facilities require staff to
reprocess medical devices then clearly written
instructions with pictures are required. - In service of all medical devices given to the
reprocessing staff on the cleaning, wrapping and
sterilization prior to the implementation of the
device in the health care facility. - Sound decisions, when purchasing the medical
device, to ensure it can reprocessed in the
facility to meet the best practices required in
the standards.
22Solution
- If increased numbers of surgical procedures are
required then increase in instrumentation to meet
these demands is required. - Short cuts in reprocessing are a risk to
patients! They are are not a means to meet the
demand. - All patients have the right to the same level of
care and the same level of reprocessed
instrument. - ST79 covers these requirements as does CSA Z314.3
and Z314.8.
23EXTENDED CYCLES
24Sterilization Cycles
- Typical steam sterilization method used in health
care are - the dynamic air removal cycles.
- Typical cycles are
- 3 Minutes at 135 degrees C (275F)
- 4 Minutes at 132 degrees C (270 F)
- Cycle is dependent upon specific sterilizer
manufacturer. - Hospital sterilizers are only validated for these
hospital - cycles.
-
- Gravity cycles rarely used today in healthcare
facilities.
25Extended Cycles
- Up to 10 different extended cycle times exist and
the number grow daily! - These cycles are greater than the standard
- hospital cycles.
- Extended cycles are now being recommended by
- various device manufacturers.
- e.g. Range of 5 to 40 minutes
26Sterilizer Manufacturer Validated Cycles
This table represents the variation in
sterilizer manufacturers recommendations for
exposure at different temperatures. For a
specific sterilizer, consult only that
manufacturers recommendations. Dry times have
not been indicated as this varies substantially
depending upon the sterilizer manufacturer,
season and/or other site specific issues. Note
Extracted from AAMI DS2/ST79/2005-02-07
27Problems
- Large sterilizers are designed to
- accommodate loads of multiple devices
- with similar requirements for sterilization.
-
- Having only a small number of sets in a large
sterilizer chamber increases the challenge to
sterilization and the likelihood that air will
not be completely removed from the set being
reprocessed. This may compromise sterilization. - It is inefficient to run a load with a single
device that requires a different sterilization
cycle. - Sterilizing items in multiple loads does not
allow - departments to meet the daily
sterilization demands.
28Problems
- Changing the settings on sterilizer
- equipment, for each load, can lead to
- human error resulting in the wrong cycle
- being selected.
- This can lead to devices being weakened,
destroyed or incorrectly sterilized in subsequent
loads.
29Problem
- Packaging for containment of devices or sets
- has not been validated for use in extended
- cycles e.g. reusable and single use wraps,
- pouches, rigid containers and their filters.
- May not maintain sterility .
- May not allow sterilization of all device
surfaces due to tightness of fitting, excessive
metal mass or layering of devices.
30Recommended Solution
- Standardize to a minimum number of cycles e.g.
- one extended cycle, the length of time to be
- determined, in addition to the 3 or 4 minute
cycle - routinely used.
- to allow BI and CI manufacturers to develop
products specific to these cycles. - to decrease likelihood of human error.
- to increase productivity.
- to decrease likelihood of sterilization failure.
31Recommended Solutions
- When a device manufacturer develops extended
cycle recommendations, the type of appropriate
packaging to be used for the extended cycle
(reusable or single use wrappers, rigid
containers or pouches) should be included in the
validation process. - Packaging manufacturers should validate their
products for extended cycle times.
32AAMI TIR 31- Process Challenge Devices/test packs
used for healthcare facilities.
- The standard extended cycles has been addressed
in this AAMI TIR. - They include
- Pre Vacuum cycles 270 274 degrees at
- 10 or 20 minutes.
- Gravity 250 degrees 40 or 60 minutes.
- Recommend Solid Load PCD and Hollow
- Load PCD for each recommended cycle.
33Proposed Chart for Extended Cycles
Standard Extended Steam Sterilization Cycle Times and Temperature Standard Extended Steam Sterilization Cycle Times and Temperature Standard Extended Steam Sterilization Cycle Times and Temperature Standard Extended Steam Sterilization Cycle Times and Temperature
Cycle Type Temperature Time (minutes)
Dynamic-air removal 132C - 135C (270F - 275F) 10
Dynamic-air removal 132C - 135C (270F - 275F) 20
Gravity-displacement 121C (250F) 40
Gravity-displacement 121C (250F) 60
34Recommended Solution
- This will allow PCD manufacturers to develop PCDS
for these cycles. - Adopted by the ST 8 - Hospital Steam Sterilizers
so sterilizers validated for these cycles. - Adopted as hospital cycles in ST79 Comprehensive
Guide to steam sterilization and sterility
assurance in health care facilities. - Wrapping manufacturers can then test their wrap
at these cycles. - CSA in Canada has agreed to also adopt these
cycles recommended.
35Conclusion Extended Cycles
- Device manufacturers must work together, with
other manufacturers, to ensure that the
recommended extended cycle meets capabilities of
the sterilizer, the packaging, and process
monitors to be used. - Health Canada and the FDA, when reviewing pre
market applications for devices will require the
medical device manufacturer to provide
appropriate Validation - Packaging will be validated for these cycles.
- Process monitoring device will be available for
these cycles. - Healthcare sterilizers will be validated for all
of the hospital cycles recognized as healthcare
steam sterilization cycles.
36Staff Orientation, Training and Competency
Testing.
37Reprocessing Staff
- Since the 1990s with the advent of Minimal
Invasive Surgery, instruments became complex and
more difficult to clean. - Simple instruments like kockers and criles made
of stainless steel are no longer the norm. - These changes brought larger challenges for
cleaning, requiring knowledgeable staff that are
current and trained. - Staff now require greater education and critical
thinking abilities in order to do the cleaning
and reprocessing safely. - Reprocessing departments require a Quality System
- which requires monitoring and documentation.
38Quality Control
- Standards require
- Continuous supervision of personnel performance.
- Competency testing.
- Supervision of work practices.
- Ongoing verification of adherence to
- policies and procedures.
- Verification of reprocessing practices.
39Quality Systems
- Deviations from normal operating procedures are
identified, investigated and evaluated. - Corrective action required.
- Audits
- Internal audits by knowledgeable staff to verify
effectiveness of system. Audit reviews documented
and corrective action taken. - Periodic external reviews to measure progress are
recommended. - Contractors or other organizations that perform
- functions that could effect quality systems
then they - must demonstrate they will have adequate
control and - supervisions so as not to effect quality
system.
40This Requires Educated Staff with Critical
Thinking Skills!
- Challenge is to find these people and then to
maintain them in the field. - Orientation after they obtain a formal course
requires months of supervision before they can
function alone. - Due to the complexity of the instruments and vast
numbers of sets to master, this does not occur
overnight. - Ongoing education, competency testing of staff
yearly, in service on all new medical devices is
costly and time consuming. - Staffing compliments must address release time
for this to occur and is rarely considered when
financial allocations are given to reprocessing
areas.
41Proposed Solutions
- Require formal education as hiring practice.
- Ensure the staff have job satisfaction.
- Provide ongoing education opportunities for
staff. - Have internal audits be opportunities for
improvement. - Staff that know why they are required to do
something do a better job. - Ensure you have detailed procedure cards for all
cleaning and reprocessing practices for all
medical devices that are current and have
pictures to facilitate processes. - Provide opportunities for open houses in the
department so users understand the complex tasks
the reprocessing staff deal with on a daily
basis.
42Proposed Solutions!
- Offer opportunities for user staff to spend time
in reprocessing. - Offer opportunities for reprocessing staff to
view surgical procedures so they understand the
use of medical devices. - Celebrate success!
- Teach staff to understand that an error is a
learning opportunity as long as corrective action
results. - Do customer service surveys and involve staff in
the results and solutions. - Liaison staff between the operating room and
reprocessing is a requirement, in order to solve
issues rather then let them become issues that
break relationships.
43Proposed Solutions!
- Reprocessing budgets must include release time
for training and ongoing education. - Auditing must be scheduled documented!
- Recommendations must be documented, and
addressed to improve process and action taken! -
44Why knowledge is Important?
45Weight of Sets!
Imagine working in a Reprocessing Department or
Operating Room where there is a maximum weight
limit on instrument trays! Imagine that this
maximum weight limit policy is routinely
enforced! Picture your staff no longer
complaining about having to repeatedly lift heavy
trays! Imagine if you did not have to deal with
wet pack problems or other sterilization
concerns due to heavy trays! This vision can
come true by just saying NO to instrument
trays that are too heavy!
46WEIGHT OF SETS
47Heavy Sets!
- No matter what the reason for an instrument set
being on the heavy side, it can be a problem when
it comes to the sterilization process. -
- Overweight sets and sets that do not have the
metal mass evenly distributed can create concerns
with adequate steam contact with all surfaces. - In addition, a problem with drying may result in
the potential for wet packs. - The Association of Preoperative Registered Nurses
(A O R N s) 2006 recommended practice for
Packaging Systems states rigid container systems
should be used according to manufacturers
instructions. The approved sterilization method
and cycle exposure times for each rigid container
system should be provided in the manufacturers
data and instructions. - Your sterilization standards state weight of a
wrapped instruments set should be no more then 16
pounds! - The sterilizers are validated for this weight.
48Heavy Sets
- Why? The more mass metal the more steam forms
water thus drying issues wet packs. - The new CSA Z314.3 states the total weight with
container cannot be over 10 kg/22 pounds. - Sterilization cannot occur unless the steriliant
can reach all surfaces of the instrument. With
multi layer sets users must verify that the
steriliant will sterilize all medical devices in
all areas. - Testing with BIs has been done and failure has
occurred. (User Alert- Problems with process
monitors for extended cycles)
49Instrument sets excessive weight mass
- If failure of any BIs report to medical device
manufacturer - Break down set to smaller tray sets and re-test
50Ergonomic Concerns with Heavy Sets
- In the hospital setting we see many injuries of
staff due to repeated lifting of heavy instrument
containers. - The health and safety of healthcare workers
should be protected through reasonable weight
restrictions, especially when these trays must be
lifted an average of 12 to 15 times each time the
set is used and reprocessed. - Instrument sets used for a surgical procedure are
lifted on and off case carts, on and off surgical
areas, in and out of instrument washers, onto and
off of processing tables, lifted to be wrapped or
containerized, into and out of the sterilizer, on
and off storage shelves, etc. - This repetitive lifting can be an ergonomic
nightmare, particularly when the set weight
exceeds 10kg/22 pounds or if there are several
heavy sets. - Studies have shown that repetitive lifting
greater then 9kg will cause injury to staff such
as shoulder, arm, back, neck. - The statistics are showing that these heavy sets
are making patients of our reprocessing and
operating room staff.
51Solution Say No!
- Write policy statement, no instrument set,
including its package is to exceed 20 pounds.
Once we had the policy in place we could enforce
it. - SPD purchased a scale for the instrument room,
and the weighed all of the sets that we owned
that seemed heavy. - We found about 25 sets that exceeded the 20-pound
weight limit. The majority of hefty sets were
orthopedic, and we also found some urology,
neurology, cardiovascular and retractor sets that
exceeded our weight limit. - We put together a list and then asked the OR
service leader for each area to help us eliminate
some items in each set or to separate the sets.
52Solution Say No!
- We then added a deadline of three weeks to
complete this project, knowing that if we set a
deadline it was more likely to be completed in a
timely manner. - In some instances we were able to remove
infrequently or unused instruments that were
added to the set in case they were needed. We put
these additional items in peel packs or small
wrappers. - In other cases, such as cardiovascular (CV) sets,
we separated the string instruments and the
retractor and put them in separate smaller pans.
53Solution!
- The number of pans is irrelevant to the outcome
of the procedure, as long as the correct items
are readily available. Moreover, in addition to
being readily available, the sterility assurance
of each device had been verified, which is
essential for safe patient care. - SPD gets asked to add an instrument to a set, we
ask if we can remove anything first. - Once the changes had been made, and the
sterilization process was verified, the surgeons
preference lists were updated to reflect the
changes and the need to pick both parts of a set,
since some were now in more than one package
(i.e., trays are labeled A and B). In addition,
we made sure the pick list made note of just
in case items to be picked that were taken out
of sets and wrapped or peel packed separately. - We realize that even though surgeons may say they
want everything in one pan, all they really want
is to make sure they have what they need, when
they need it. It doesnt matter if it is in one
or more pans.
54Weighing Sets and Limiting Weight
Instrument set being weighed
Fukushima Day Retractor set with one level
wrapped separately
.
55Solution!
- Work with the manufacturers as users to educate
them that the weight is important and to aim for
10kg/22 pound weight. - Users when making purchases start standing up for
the weight requirement to achieve sterility,
protect workers as our sterilizers are validated
with 16 pounds of mass metal. - Health care committees when making purchasing
decisions this must be taken into account. - If we dont say NO to overweight sets, you will
all be an occupational injury before you reach
retirement age. - Remember we are humans not robots!
- Our body will react to abuse by breaking down!
- Why should we make reprocessing and operating
staff suffer injury when we can divide sets into
reasonable weights?
56Loaner Specialty Equipment Sets
57Loaner Specialty Equipment Sets
- Inventory issues!
- Receipt on site 48 hours prior to scheduled
surgical procedure. - - Cleaning- Sterilization
- Quality Control for sets!
58Loaner Sets
- Healthcare facilities use loaned and shared
medical devices for a number of procedures for
potential cost savings from not purchasing
expensive devices in their inventory. - Loaned and shared devices allow facilities to
save on maintenance costs and keep up with
changing technology. - The biggest concern comes from those instrument
sets that we do not own. - Loaner sets come and go without written
instructions, pictures, in service, etc. all not
meeting ISO 17664, AAMI ST81, and CSA Z314.22
Management of Loaned, Shared, Leased Medical
Devices. -
- If you write your policy for your facility
meeting these standards you will eliminate the
problems.
59Problems!
- Sets coming with no list of items.
- No pictures.
- No cleaning or sterilization instructions.
- No in service.
- Heavy sets.
- Sets received in complete or soiled from previous
user. - Come with no time to check and reprocess the sets
properly. - Sets received and not sure what is supposed to be
in the set. - All of these issues place the patient at risk of
injury and infection! - If manufacturers want your business then they
will follow your policy. - When doing the evaluation of which vendor you
will deal with use the standards to set policy to
get all your concerns addressed.
60Write Policy as a Solution!
- Policy must address
- Identify responsibilities as sender, receiver,
qualifications and training of personnel, risk
management, quality management and insurance. - Have a policy to cover emergencies related to
loaned and shared devices including the necessary
signatures to designate that situation as an
emergency. - Not accept Medical Devices unless they are
licensed by Health Canada and FDA, and conform to
codes and standards. - Have procedures in place for sending,
transporting and receiving of loaned and shared
devices. - Have in place mutually accepted written
documentation, outlining the responsibilities of
individuals during the transactions between
senders and receivers.
61Write Policy as a Solution!
- In written form (e.g. in binders, manuals or
monographs) and/or electronic format. - In a central location so they are readily
accessible and shall be updated as required. - Documented and reviewed on a regular basis by the
facilitys Infection Prevention and Control or
Risk Management personnel, or their designates. - Policies required for
- Quality system!
- Documentation!
62Transportation
- Check environmental and handling conditions
- required and transport accordingly.
- Routinely clean and maintain the trucks.
- Prevent physical damage, tampering and loss.
- Maintain records of devices transported.
- Deliver the device to the next organization or
- facility at the contractually agreed time.
63Solution to Loaner Set Weights
- How can you enforce the weight limit on those
sets that we - borrowed or that the company brought in for us to
use as - long as we purchased their implants?
- The answer is in policy. The weight limit
statement into your loaner policy then shared
with the MDM representatives that had surgical
instrument sets in our facility. - Ask the MDMs to work with you to get their set
weights down to 20 pounds. Some were very helpful
and made suggestions on how we could divide the
sets, while others took a little convincing and
negotiating to get them to see that if they
wanted us to sterilize their sets, they had to
follow our policy. - Purchase additional instrument container or two
to house the items removed from the
vendor-owned sets. - (MDM medical device manufacturer)
64Solution to Loaner Set Weights
- Validated for sterilization by the MDM, and even
lighter by placing them in two containers, we
could use its sterilization process
recommendation. - Follow manufacturers instructions and conduct
your own sterilization verification testing per
AAMIs recommendation on product testing. - When it comes to MDMs sterilization validation
testing, never take the representatives verbal
word for it. Always ask for that information in
writing directly from the MDM.
65FLASH STERILIZATION
66What is flash sterilization?
Z314.13-01(CSA) Emergency Sterilization
Which now will be replaced by CSA Z314.3
Why? Flash sterilization - a special steam
sterilization process designed and used for the
emergency sterilization of surgical goods when
proper sterilization cannot be done.
67Flash Sterilization
Flash sterilization, if used by facility policy,
shall be used only for emergency sterilization
(for example when a dropped instrument requires
immediate sterilization). Flash sterilization
shall not be used to compensate for inventory
shortages or scheduling problems. Flash
sterilizers will be biologically monitored
minimally daily and all items traceable to the
patient.
68Current practice!
Flashing total sets and implants is
routine. Quick rinse in a sub-sterile room or
scrub-sink area -Uniform or scrub suit sink is
now contaminated Excess flashing due to lack of
inventory and scheduling practices. Flash
Sterilizers with no preventative maintenance No
verification of sterility of items being
sterilized.
69Where do we need to be?
- Instrument sets should not be flash sterilized.
Flash sterilizers are only validated for 1-2
instruments. - The following conditions shall be met if you are
flashing - a) there is an urgent unplanned need
- b) there are documented procedures for this
practice - the physical layout and documented procedures
assure direct delivery of the sterilized item to
the point of immediate use. - d) The flash sterilizers has routine preventative
maintenance that is documented. - e) Staff trained in reprocessing practices and
yearly competency testing.
70Where we need to be contd
d. Proper decontamination of instruments
Performed in a separate room/environment
by people who are properly attired with PPE
(personal protective equipment). CSA Z314.8
e. Load documentation monitoring system is
in place. f. All flashed items tracked to
the patient .
71 Measures to Avoid Flash Sterilization
- Measure to avoid flash sterilization
- maintenance of adequate inventories of medical
devices - coordinating device reprocessing with surgical
schedules so that properly processed devices are
available when needed - adherence to policies and procedures for loaned
and shared instrumentation (see Z314.22) so that
devices arrive in time for proper reprocessing
and - education of medical staff and materials managers
in the risks associated with flash sterilization - Computerized booking systems that cannot be
converted to override inventory.
72 Why not Flash Sterilize?
- Sterilization process itself can be similar to
sterilization process performed in reprocessing
department but the lack of pre- and post-
sterilization precautions introduces additional
risks to the use of flashed devices. - The facility has a responsibility to ensure that
every patient receives the same level of care
therefore should endeavour to avoid the
additional risks to patient safety associated
with flash sterilization. - Remember that flash sterilizers are only
validated for 1-2 instruments, and should only be
used for life and death situations or loss of
limb not routine for scheduling and lack of
instrumentation. - CSA Standard Z314.3 Flash sterilization shall
not be used on - complete sets or trays of instruments.
73What can be done to change?
- Develop policies based on CSA standards that both
- SPD and OR staff could follow and live with!
- How do we do that?
- Research the CSA documents which are best
practices in reprocessing - Invite an expert to speak to the Peri-operative
Policy and Procedure Committee about the CSA
Standards. - ORNAC Standards are based on CSA.
74CSA Standards and ORNAC
- Why are they written if hospitals cant follow?
Why write a policy based on CSA Standard that
cant be upheld due to resources available?
- Do we bend the Standard to fit resources of the
hospital?
75What can be done to change practices?
- Develop policies AND..
- Standardize practices throughout Canada for OR
SPD - Identify staff competencies
- Develop an education plan
- Decrease flash sterilization
- Share our resourcesÂ
- Be a leader!
- Provide every patient with the same level of
care
76Policies and Procedures
- The requirements for policies, procedures,
documentation and quality - system are required.
- If a flash sterilized medical device is used in a
procedure, an incident - report shall be created that includes the
following information - patients name
- physicians name
- date and time the of procedure using the flash
sterilized device - procedure it was used for
- reason that flash sterilization was needed
- results of physical monitoring and biological /
chemical indicator monitoring of the sterilizer
and load. - Periodic review incident reports to help identify
ways to reduce the need for flash sterilization.
77Policies
- Flash sterilization process is composed of
several phases that - are critical to achieving and maintaining
sterility. - These phases include
- transport of contaminated devices to
decontamination area - disassembly and decontamination
- preparation
- loading the sterilizer
- sterilization procedures
- monitoring and recording of cycle parameters
- unloading and aseptic transfer of items to the
sterile field.
78Personnel Requirements
- 1. All aspects of emergency steam sterilization
require - supervised staff that are knowledgeable,
trained - personnel wearing appropriate attire,
including personal - protective equipment.
- 2. Personnel must have demonstrated expertise in
medical - device reprocessing and in the use and
application of the - steam sterilization process.
- 3. Control of bio burden and containment of
contaminants are essential to the sterilization
process.
79Personnel Requirements
- 4. The health care facility administration to
ensure that operators of all steam sterilizers
used for flash sterilization in the facility are
trained. - 5. Education documented with ongoing competency
in the operation of steam sterilizers. - 6. Staff are aware of the hazards associated with
steam sterilizers. - 7. Monitoring and maintenance of sterilizers
required.
80Flash Sterilization Personnel
- Flash sterilization personnel must be responsible
for verifying exposure time and temperature, and
for aseptically transferring the sterilized items
to the point of use. - Personnel engaged in the processing of items for
flash sterilization require an initial
orientation and on-the-job training, including
instruction in the following -
- a) sterilizer operation and monitoring
(sterility assurance) - b) the parameters of steam sterilization
- c) basic microbiological principles
- d) the health care facility's infection
control and prevention - policies and procedures
- e) decontamination of medical devices
-
81Flash Sterilization Personnel
- f) monitoring of sterilization
- g) preparation of the incident report
- h) action to be taken if there is a failure
during reprocessing - i) unloading and transfer to the sterile
field. - The facility shall assign specific responsibility
to a staff - member or position for the maintenance of quality
system - documentation relating to flash sterilization,
including - documentation of staff training, and the
collection, review - and retention of flash sterilization incident
reports.
82Work Area for Flash Sterilization
- The steam sterilizer used for flash sterilization
shall be - located in a restricted-access area
immediately adjacent to the area where the
sterilized items will be used in patient care. - The sterilizer shall not be located in the
operating room theatre, and shall not be near any
potential source of contamination, such as sinks,
hoppers, linen, or trash disposal areas. - Sterile storage shall not be located immediately
adjacent to - the sterilizer used for flash sterilization.
- Flash sterilization must be carried out in a
clean environment. - Devices processed by this method must be
transferred and handled as little as possible as
may not be protected by packaging before or
immediately after the sterilization process.
83Work Area
- At least one flash sterilizer in the OR suite
shall be on the emergency power system. - The facility shall have written validation from
the sterilizer manufacturer that it is suitable
for flash sterilization and specifying the size
of loads for which it has been validated. - All instruments shall be cleaned and
decontaminated prior to sterilization by trained
staff wearing appropriate PPE. Refer to Z314.8.
Saline shall not be used. Before they are
cleaned, general operating instruments shall be
separated from delicate instruments or devices
requiring special handling.
84Work Area Decontame
- Soiled items shall be contained during the
transport from the point of use to the
decontamination area. - They shall be cleaned immediately.
- Gloves and appropriate PPE shall be worn when
handling soiled items. - Containment may be accomplished by any means that
adequately prevents inadvertent personnel contact
with or exposure to the soiled items during the
transfer. - Reusable transportation containers shall be
cleaned and decontaminated after each use.
85Sterilization
- High-temperature saturated steam must come into
direct contact with all surfaces of all items. - Air removal, steam contact, and drainage of
condensate are enhanced by proper positioning and
by the use of perforated or mesh bottom trays.
Items to be sterilized may be placed in - (a) protective organizing cases
- (b) open trays
- (c) rigid sterilization container systems
- (d) single wrapped trays
- Closed container system is preferred, to minimize
the chance - of contamination occurring between sterilizer and
patient.
86STERILIZATION
- The health care facility shall follow the
device manufacturers instructions, including
those for - time and temperature
- load size (including any necessary adjustments to
time or temperature with relation to load size) - maximum loads (by weight or by number of
instruments) - load configuration
- Each flash sterilization cycle shall be verified
by the health care facility through actual
testing of loads configured as specified in the
manufacturers instructions.
87Sterilization
- If flashing will occur for life, limb or organ
emergency disaster then - after cleaning, items shall be placed in a
perforated or mesh bottom - instrument tray and arranged in the following
manner - (a) instruments and devices with concave surfaces
shall be positioned with the open side down - (b) all hinged instruments shall be opened,
without engaging the ratchet, and shall be
placed on racks or stringers as needed - (c) items with easily removable parts shall be
disassembled - (e) for complex instruments (e.g., air-powered
instruments), the instructions of the device
manufacturer shall be followed - (f) lumens of tubing, suction devices, and
needles shall be moistened with pyrogen-free
water immediately prior to steam sterilization.
88Do Health Care facilities meet these requirements?
- Auditing of healthcare facilities across the
nation will show that most - do not meet these requirements.
- Most
- Wash instruments in scrub sinks or small sink in
between OR theatres. - Decontame areas not under negative pressure
Z314.8 - Maintenance on these sterilizers not done,
sterilizers not cleaned, records not kept. Some
so old should be replaced. - Staff not trained to reprocess and do not have
competency testing or required education. -
- It goes on and
on!
89Notice from Ministry Ontario of Health re Flash
Sterilization
90Do you meet all these requirements?
- IF NOT YOUR FACILITY NEEDS
- TO TAKE ON A PROJECT TO
- IMPROVE PROCESS!
91Implement to improve patient care!
92Set up a Quality Improvement Project
- Joint Quality Assurance/Accreditation Project
- Sterile Processing Department and Surgical
Suites - Called The Flash Project.
- Must have buy in from all parties!
-
- Why do it? Best Practices are provided in the CSA
standards and every facility should be aiming
for these practices. - Accreditation will now be examining flashing
practices as a part of their review. -
93Reprocessing Challenges!
- I have touched on a few challenges but there are
many more. - Reprocessing is complex, always changing, and a
very demanding profession! - It requires ongoing education, and knowledge that
is current. - Staff with critical thinking skills,
knowledgeable is required. - Reprocessing staff must be the patient advocates!
94Change can be accomplished!
- Working together for the betterment of the
patient we can all meet best practices in
reprocessing as - Remember!
- You could be that patient having a surgical
procedure and - would you want a totally processed instrument or
a flashed - one?
- I know what I would want and so do you!
95 Remember!We
never know when we will require treatment in the
facility we work at! Always advocate for all
patients!When any reprocessing staff or their
family require care, we want to know that the
instruments are the best quality and meet
standards!
96What Can We do?
- Raise the bar in reprocessing!
- Ensure all patients are not put at risk during
their hospital stay! - PARTNERSHIP
- OPTIMAL PATIENT CARE
- REQUIRES WORKING
- TOGETHER!
97References
- AAMI ST 79 Comprehensive guide to steam
sterilization and sterility assurance in a health
care facilities - AAMI TIR 31 Process Challenge devices/test packs
for use in health care facilities - CSA Z314.3 Effective Sterilization in Health Care
Facilities by the Steam Process - CSA Z314.8 Decontamination of Reusable Medical
Devices - CSA Z314.22 Management of Loaned, Shared and
Leased Medical Devices - Just Say No by Rose Seavey RN, MBA, CNOR, ACSP
- Best Practices for Cleaning, Disinfection and
Sterilization in Health Care Facilities
98Lets all provide every patient with the care
that we all would like to receive!
Questions!