Setup errors using daily CT and CBCT for imageguided stereotactic body radiation therapy SBRT - PowerPoint PPT Presentation

1 / 1
About This Presentation
Title:

Setup errors using daily CT and CBCT for imageguided stereotactic body radiation therapy SBRT

Description:

Setup errors using daily CT and CBCT for image-guided stereotactic body radiation therapy (SBRT) ... Necessary positional setup shifts were established by ... – PowerPoint PPT presentation

Number of Views:476
Avg rating:5.0/5.0
Slides: 2
Provided by: wujz
Category:

less

Transcript and Presenter's Notes

Title: Setup errors using daily CT and CBCT for imageguided stereotactic body radiation therapy SBRT


1
Setup errors using daily CT and CBCT for
image-guided stereotactic body radiation therapy
(SBRT)
Jianzhou Wu1, Martin Fuss1, and Warren
DSouza1,2 1Department of Radiation Medicine,
Oregon Health Science University, Portland,
OR 2Department of Radiation Oncology, University
of Maryland School of Medicine, Baltimore, MD
Purpose
Absolute alignment errors over 30 setups
  • Immobilization devices, used for patients
    underwent stereotactic body radiation therapy
    (SBRT) improve patient alignment and
    repositioning accuracy
  • The BodyFIX system (BodyFIX, Medical
    Intelligence, Schwabmuenchen, Germany) has been
    shown to perform favorably compared with other
    body immobilization systems
  • In this study, we retrospectively compared
    patient alignments using daily simulation CT with
    patients immobilized within BodyFIX and
    alignments using on-board cone-beam CT (CBCT)

Materials and Methods
error shift from daily CT alignment shift
from CBCT alignment
  • Ten patients who underwent SBRT of lung cancers
    were immobilized using BodyFIX system. For
    image-guidance, all patients were immobilized and
    scanned in the simulation suite. An
    image-guidance control CT was acquired on the
    dedicated simulation CT scanner (Philips
    Brilliance Big Bore 16, Philips Medical Systems,
    Cleveland, OH) before each fraction of treatment.
    Necessary positional setup shifts were
    established by comparison between original
    simulation CT and the respective control CT.
  • Patients were then transferred onto a stretcher,
    and moved onto the linear accelerator treatment
    table within their BodyFIX for radiotherapy.
  • Prior to treatment initiation, patients were
    scanned again using CBCT for alignment
    confirmation.
  • This study investigated the capability of patient
    alignments using a daily simulation CT. Owing to
    the fact that patients were imaged in the
    simulation suite, and then moved to the treatment
    vault, we assigned the CBCT based alignment as
    the reference for accurate treatment setup.
  • Alignment variations between modalities were
    evaluated by a two-step image registration, which
    registered both the image-guidance CT and the
    CBCT to the original simulation CT.
  • First a rigid registration was performed to align
    the image-guidance CT (and CBCT) to the
    simulation CT. After rigid registration,
    patients were aligned with match in bony
    structures.
  • Next a deformable registration was performed,
    which allows soft tissue deformation so that a
    better local match was achieve. After this step,
    patients were aligned with tumor-to-tumor match.
  • After image registration, tumor contours
    delineated on the simulation CT were warped to
    the daily image-guidance CT (and CBCT) using the
    transformation fields saved from image
    registration.
  • The center-of-mass (COM) of the mapped contours
    was calculated. The difference between the COM
    after deformable registration and that after
    rigid registration stands for the shift after
    soft tissue based alignment with respect to bony
    structure based alignment
  • The setup optimization shifts derived from
    image-guidance CT based alignments were compared
    with the corresponding shifts from CBCT based
    alignments for the evaluation of setup accuracy.
    Particular interested was directed towards the
    question of detection of patient setup variations
    owing to transfer from the imaging system in the
    simulation suite to the treatment system in a
    different room.

Alignment deviations between modalities (30
setups in 10 patients)
A t-test was performed for alignments using
daily CT and CBCT over 30 setups. The
corresponding p-value was shown in the above table
Discussion
  • Daily simulation CT based alignments were
    compared with CBCT image-guidance based
    alignments for patients immobilized using BodyFIX
    system and transferred between 2 rooms
  • On average the alignment mismatch between the two
    modalities was less than 2 mm along the three
    principle axes
  • A t-test demonstrated that alignments using the
    two methods were indistinguishable along the
    principle axes, except in the left-right
    direction
  • This discrepancy may be related to one or two
    distinguishable observations. (1) The transfer of
    patient and immobilization system from CT table
    onto the stretcher and then again onto the linear
    accelerator table was always a lateral transfer.
    This particular form of transfer may be a cause
    the observed setup inconsistency.
  • (2) Abrupt lateral motion of the linear
    accelerator treatment table related to moving the
    tabletop into a safe, collision free position for
    CBCT acquisition. The impact of the created
    momentum on stability of patient setup in an
    immobilization system is subject to ongoing
    testing.
  • The averaged alignment discrepancy was well
    within the 5 mm margin, which was added to the
    ITV when creating the PTV to account for setup
    uncertainty.
  • Therefore daily simulation CT based
    image-guidance for SBRT with patients
    immobilization and transfer using BodyFIX system
    provides for accurate and safe patient setup
    assessment.

Demonstration of evaluation process
Alignment using CBCT
Rigid registration
Deformable registration
Contours mapped after deformable registration
Contours mapped after rigid registration
Alignment using daily CT
Conclusion
Contour on simulation CT
  • Patient alignments using daily simulation CT was
    consistent with the alignments using CBCT
  • Alignment using image-guidance CT with patient
    immobilized using BodyFIX system provides a
    solution for image- guided radiation therapy on
    facilities without the on-board imaging capability

shift COM of mapped contour after deformable
registration COM of contour after rigid
registration
Write a Comment
User Comments (0)
About PowerShow.com