Transducers - PowerPoint PPT Presentation

About This Presentation
Title:

Transducers

Description:

Transducers Ultrasound is produced and detected with a transducer, composed of one or more ceramic elements with electromechanical (piezoelectric) properties. – PowerPoint PPT presentation

Number of Views:654
Avg rating:3.0/5.0
Slides: 127
Provided by: facultyEt
Learn more at: http://faculty.etsu.edu
Category:

less

Transcript and Presenter's Notes

Title: Transducers


1
  • Transducers

2
  • Ultrasound is produced and detected with a
    transducer, composed of one or more ceramic
    elements with electromechanical (piezoelectric)
    properties.
  • The ceramic element converts electrical energy
    into mechanical energy to produce ultrasound and
    mechanical energy into electrical energy for
    ultrasound detection.

3
  • Over the past several decades, the transducer
    assembly has evolved considerably in design,
    function, and capability, from a single-element
    resonance crystal to a broadband transducer array
    of hundreds of individual elements.
  • A simple single-element, plane-piston source
    transducer has major components including the
  • piezoelectric material,
  • marching layer,
  • backing block,
  • acoustic absorber,
  • insulating cover,
  • sensor electrodes, and
  • transducer housing.

4
(No Transcript)
5
Piezoelectric Materials
  • A piezoelectric material (often a crystal or
    ceramic) is the functional component of the
    transducer.
  • It converts electrical energy into mechanical
    (sound) energy by physical deformation of the
    crystal structure.

6
  • ConverseIy, mechanical pressure applied to its
    surface creates electrical energy.
  • Piezoelectric materials are characterized by a
    well-defined molecular arrangement of electrical
    dipoles (Fig. 16-9).

7
  • An electrical dipole is a molecular entity
    containing positive and negative electric charges
    that has no net charge.
  • When mechanically compressed by an externally
    applied pressure, the alignment of the dipoles is
    disturbed from the equilibrium position to cause
    an imbalance of the charge distribution.

8
  • A potential difference (voltage) is created
    across the element with one surface maintaining a
    net positive charge and one surface a net
    negative charge.
  • Surface electrodes measure the voltage, which is
    proportional to the incident mechanical pressure
    amplitude.

9
  • Conversely, application of an external voltage
    through conductors attached to the surface
    electrodes induces the mechanical expansion and
    contraction of the transducer element.

10
  • There are natural and synthetic piezoelectric
    materials.
  • An example of a natural piezoelectric material is
    quartz crystal, commonly used in watches and
    other time pieces to provide a mechanical
    vibration source at 32.768 kHz for interval
    timing.
  • This is one of several oscillation frequencies of
    quartz, determined by the crystal cut and
    machining properties.

11
  • Ultrasound transducers for medical imaging
    applications employ a synthetic piezoelectric
    ceramic, most often lead-zirconate-titanate
    (PZT).
  • The piezoelectric attributes are attained after a
    process of
  • Molecular synthesis,
  • Heating,
  • Orientation of internal dipole structures with an
    applied external voltage,
  • Cooling to permanently maintain the dipole
    orientation, and
  • Cutting into a specific shape.

12
  • For PZT in its natural state, no piezoelectric
    properties are exhibited however, heating the
    material past its Curie temperature (i.e., 3280
    C to 3650 C) and applying an external voltage
    causes the dipoles to align in the ceramic.
  • The external voltage is maintained until the
    material has cooled to below its Curie
    temperature.
  • Once the material has cooled, the dipoles retain
    their alignment.

13
  • At equilibrium, there is no net charge on ceramic
    surfaces.
  • When compressed, an imbalance of charge produces
    a voltage between the surfaces.
  • Similarly, when a voltage is applied between
    electrodes attached to both surfaces, mechanical
    deformation occurs.

14
  • The piezoelectric element is composed of aligned
    molecular dipoles.

15
  • Under the influence of mechanical pressure from
    an adjacent medium (e.g., an ultrasound echo),
    the element thickness
  • Contracts (at the peak pressure amplitude),
  • Achieves equilibrium (with no pressure) or
  • Expands (at the peak rarefactional pressure),
  • This causes realignment of the electrical dipoles
    to produce positive and negative surface charge.

16
(No Transcript)
17
  • Surface electrodes (not shown) measure the
    voltage as a function of time.

18
  • An external voltage source applied to the element
    surfaces causes compression or expansion from
    equilibrium by realignment of the dipoles in
    response to the electrical attraction or
    repulsion force.

19
(No Transcript)
20
Resonance Transducers
  • Resonance transducers for pulse echo ultrasound
    imaging are manufactured to operate in a
    resonance mode, whereby a voItage (commonly 150
    V) of very short duration (a voltage spike of ?1
    msec) is applied, causing the piezoelectric
    material to initially contract, and subsequently
    vibrate at a natural resonance frequency.
  • This frequency is selected by the thickness
    cut, due to the preferential emission of
    ultrasound waves whose wavelength is twice the
    thickness of the piezoelectric material.

21
  • The operating frequency is determined from the
    speed of sound in, and the thickness of, the
    piezoelectric material.
  • For example, a 5-MHz transducer will have a
    wavelength in PZT (speed of sound in PZT is ?
    4,000 m/sec) of

22
  • A short duration voltage spike causes the
    resonance piezoelectric element to vibrate at its
    natural frequency, fo, which is determined by the
    thickness of the transducer equal to 1/A.

23
  • To achieve the 5-MHz resonance frequency, a
    transducer element thickness of ½ X 0.8 mm 0.4
    mm is required.
  • Higher frequencies are achieved with thinner
    elements, and lower frequencies with thicker
    elements.
  • Resonance transducers transmit and receive
    preferentially at a single center frequency.

24
Damping Block
  • The damping block, layered on the back of the
    piezoelectric element, absorbs the backward
    directed ultrasound energy and attenuates stray
    ultrasound signals from the housing.
  • This component also dampens he transducer
    vibration in create an ultrasound pulse width a
    short spatial pulse length, which is necessary to
    preserve detail along he beam axis (axial
    resolution).

25
(No Transcript)
26
(No Transcript)
27
(No Transcript)
28
  • Dampening of the vibration (also known as
    ring-down) lessens the purity of the resonance
    frequency and introduces a broadband frequency
    spectrum.
  • With ring-down, an increase in he bandwidth
    (range of frequencies) of he ultrasound pulse
    occurs by introducing higher and lower
    frequencies above and below the center
    (resonance) frequency.

29
  • The Q factor describes the bandwidth of the
    sound emanating from a transducer as
  • where fo is the center frequency and the
    bandwidth is the width of the frequency
    distribution.

30
  • A high Q transducer has a narrow bandwidth
    (i.e., very little damping) and a corresponding
    long spatial pulse length.
  • A low Q transducer has a wide bandwidth and
    short spatial pulse length.

31
  • Imaging applications require a broad bandwidth
    transducer in order to achieve high spatial
    resolution along the direction of beam travel.
  • Blood velocity measurements by Doppler
    instrumentation require a relatively narrow-band
    transducer response in order to preserve velocity
    information encoded by changes in the echo
    frequency relative to the incident frequency.

32
  • Continuous-wave ultrasound transducers have a
    very high Q characteristic.
  • While the Q factor is derived from the term
    quality factor, a transducer with a low Q does
    not imply poor quality in the signal.

33
Matching Layer
  • The matching layer provides the interface between
    the transducer element and the tissue and
    minimizes the acoustic impedance differences
    between the transducer and the patient.
  • It consists of layers of materials with acoustic
    impedances that are intermediate to those of soft
    tissue and the transducer material.
  • The thickness of each layer is equal to
    one-fourth the wavelength, determined from the
    center operating frequency of the transducer and
    speed of sound in the matching layer.

34
  • For example, the wavelength of sound in a
    matching layer with a speed of sound of 2,000
    m/sec for a 5-MHz ultrasound beam is 0.4 mm.
  • The optimal matching layer thickness is equal to
    ¼l ¼ x 0.4 mm 0. 1 mm.
  • In addition to the matching layers, acoustic
    coupling gel (with acoustic impedance similar to
    soft tissue) is used between the transducer and
    the skin of the patient to eliminate air pockets
    that could attenuate and reflect the ultrasound
    beam.

35
Nonresonance (Broad-Bandwidth) Multifrequency
Transducers
  • Modern transducer design coupled with digital
    signal processing enables multifrequency or
    multihertz transducer operation, whereby rhe
    center frequency can be adjusted in he transmit
    mode.
  • Unlike the resonance transducer design, the
    piezoelectric element is intricately machined
    into a large number of small rods, and then
    filled with an epoxy resin to create a smooth
    surface.

36
(No Transcript)
37
(No Transcript)
38
  • The acoustic properties are closer to issue than
    a pure PZT material, and thus provide a greater
    transmission efficiency of the ultrasound beam
    without resorting to multiple matching layers.
  • Multifrequency transducers have bandwidths that
    exceed 80 of the center frequency.

39
  • Excitation of the multifrequency transducer is
    accomplished with a short square wave burst of
    150 V with one to three cycles, unlike the
    voltage spike used for resonance transducers.
  • This allows the center frequency to be selected
    within the limits of the transducer bandwidth.

40
  • Likewise, the broad bandwidth response permits
    the reception of echoes within a wide range of
    frequencies.
  • For instance, ultrasound pulses can be produced
    at a low frequency, and the echoes received at
    higher frequency.

41
  • Harmonic imaging is a recently introduced
    technique that uses this ability
  • lower frequency ultrasound is transmitted into
    the patient, and the higher frequency harmonics
    (e.g., two times the transmitted center
    frequency) created from the interaction with
    contrast agents and tissues, are received as
    echoes.

42
  • Native tissue harmonic imaging has certain
    advantages including greater depth of
    penetration, noise and clutter removal, and
    improved lateral spatial resolution.

43
Transducer Arrays
  • The majority of ultrasound systems employ
    transducers with many individual rectangular
    piezoelectric elements arranged in linear or
    curvilinear arrays.
  • Typically, 128 to 512 individual rectangular
    elements compose the transducer assembly.
  • Each element has a width typically less than half
    the wavelength and a length of several
    millimeters.

44
  • Two modes of activation are used to produce a
    beam.
  • These are the linear (sequential) and phased
    activation/receive modes.

45
Linear Arrays
  • Linear array transducers typically contain 256 to
    512 elements physically these are the largest
    transducer assemblies.

46
  • In operation, the simultaneous firing of a small
    group of ? 20 adjacent elements produces the
    ultrasound beam.
  • The simultaneous activation produces a synthetic
    aperture (effetive transducer width) defined by
    the number of active elements.

47
  • Echoes are detected in the receive mode by
    acquiring signals from most of the transducer
    elements.
  • Subsequent A-line acquisition occurs by firing
    another group of transducer elements displaced by
    one or two elements.

48
  • A rectangular field of view is produced with this
    transducer arrangement.
  • For a curvilinear array, a trapezoidal field of
    view is produced.

49
Phased Arrays
  • A phased-array transducer is usually composed of
    64 to 128 individual elements in a smaller
    package than a linear array transducer.
  • All transducer elements are activated nearly (but
    not exactly) simultaneously to produce a single
    ultrasound beam.

50
  • By using time delays in the electrical activarion
    of the discrete elements across the face of the
    transducer, the ultrasound beam can be steered
    and focused electronically without moving the
    transducer.
  • During ultrasound signal reception, all of the
    transducer elements detect the returning echoes
    from the beam path, and sophisticated algorithms
    synthesize the image from the detected data.

51
BEAM PROPERTIES
  • The ultrasound beam propagates as a longitudinal
    wave from the transducer surface into the
    propagation medium, and exhibits two distinct
    beam patterns
  • a slightly converging beam out to a distance
    specified by the geometry and frequency of the
    transducer (the near field), and
  • a diverging beam beyond that point (the far
    field).

52
  • For an unfocused, single-element transducer, the
    length of the near field is determined by the
    transducer diameter and the frequency of the
    transmitted sound.

53
  • For multiple transducer element arrays, an
    effective transducer diameter is determined by
    the excitation of a group of transducer
    elements.
  • Because of the interactions of each of the
    individual beams and the ability to focus and
    steer the overall beam, the formulas for a
    single-element, unfocused transducer are not
    directly applicable.

54
The Near Field
  • The near field, also known as the Fresnel zone,
    is adjacent to the transducer face and has a
    converging beam profile.
  • Beam convergence in the near field occurs because
    of multiple constructive and destructive
    interference patterns of the ultrasound waves
    from the transducer surface.

55
  • Huygens principle describes a large transducer
    surface as an infinite number of point sources of
    sound energy where each point is characterized as
    a radial emitter.
  • By analogy, a pebble dropped in a quiet pond
    creates a radial wave pattern.

56
  • As individual wave patterns interact, the peaks
    and troughs from adjacent sources constructively
    and destructively interfere, causing the beam
    profile to be tightly collimated in the near
    field.

57
  • The ultrasound beam path is thus largely confined
    to the dimensions of the active portion of the
    transducer surface, with the beam diameter
    converging to approximately half the transducer
    diameter at the end of the near field.

58
  • The near field length is dependent on the
    transducer frequency and diameter
  • where d is the transducer diameter, r is the
    transducer radius, and l is the wavelength of
    ultrasound in the propagation medium.

59
  • In soft tissue, l 1.54mm/f(MHz), and the near
    field length can be expressed as a function of
    frequency

60
  • A higher transducer frequency (shorter
    wavelength) will result in a longer near field,
    as will a larger diameter element.

61
  • For a 10-mm-diameter transducer, the near field
    extends 5.7 cm at 3.5 MHz and 16.2 cm at 10 MHz
    in soft tissue.
  • For a 15-mm-diameter transducer, the
    corresponding near field lengths are 12.8 and
    36.4 cm, respectively.

62
  • Lateral resolution (the ability of the system to
    resolve objects in a direction perpendicular to
    the beam direction) is dependent on the beam
    diameter and is best at the end of the near field
    for a single-element transducer.
  • Lateral resolution is worst in areas close to and
    far from the transducer surface.

63
  • Pressure amplitude characteristics in the near
    field are very complex, caused by the
    constructive and destructive interference wave
    patterns of the ultrasound beam.
  • Peak ultrasound pressure occurs at the end of the
    near field, corresponding to the minimum beam
    diameter for a single-element transducer.

64
  • Pressures vary rapidly from peak compression to
    peak rarefaction several times during transit
    through the near field.
  • Only when the far field is reached do the
    ultrasound pressure variations decrease
    continuously.

65
  • The far field is also known as the Fraunhofer
    zone, and is where the beam diverges.
  • For a large-area single-element transducer, the
    angle of ultrasound beam divergence, 0, for the
    far field is given by
  • where d is the effective diameter of the
    transducer and l is the wavelength both must
    have the same units of distance.

66
  • Less beam divergence occurs with high-frequency,
    large-diameter transducers.
  • Unlike the near field, where beam intensity
    varies from maximum to minimum to maximum in a
    converging beam, ultrasound intensity in the far
    field decreases monotonically with distance.

67
Focused Transducers
  • Single-element transducers are focused by using a
    curved piezoelectric element or a curved acoustic
    lens to reduce the beam profile.
  • The focal distance, the length from the
    transducer to the narrowest beam width, is
    shorter than the focal length of a non-focused
    transducer and is fixed.

68
  • The focal zone is defined as the region over
    which the width of the beam is less than two
    times the width at the focal distance
  • Thus, the transducer frequency and dimensions
    should be chosen to match the depth requirements
    of the clinical situation.

69
Transducer Array Beam Formation and Focusing
  • In a transducer array, the narrow piezoelectric
    element width (typically less than one
    wavelength) produces a diverging beam at a
    distance very close to the transducer face.
  • Formation and convergence of the ultrasound beam
    occurs with the operation of several or all of
    the transducer elements at the same time.

70
  • Transducer elements in a linear array that are
    fired simultaneously produce an effective
    transducer width equal to the sum of the widths
    of the individual elements.
  • Individual beams interact via constructive and
    destructive interference to produce a collimated
    beam that has properties similar to the
    properties of a single transducer of the same
    size.

71
  • With a phased-array transducer, the beam is
    formed by interaction of the individual wave
    fronts from each transducer, each with a slight
    difference in excitation time.
  • Minor phase differences of adjacent beams form
    constructive and destructive wave summations that
    steer or focus the beam profile.

72
Transmit Focus
  • For a single transducer or group of
    simultaneously fired elements in a linear array,
  • The focal distance is a function of the
    transducer diameter (or the width of the group of
    simultaneously fired elements),
  • The center operating frequency, and
  • The presence of any acoustic lenses attached to
    the element surface.

73
  • Phased array transducers and many linear array
    transducers allow a selectable focal distance by
    applying specific timing delays between
    transducer elements that cause the beam to
    converge at a specified distance.

74
  • A shallow focal zone (close to the transducer
    surface) is produced by firing outer transducers
    in the array before the inner transducers in a
    symmetrical pattern.

75
  • Greater focal distances are achieved by reducing
    the delay time differences among the transducer
    elements, resulting in more distal beam
    convergence.
  • Multiple transmit focal zones are created by
    repeatedly acquiring data over the same volume,
    but with different phase timing of the transducer
    array elements.

76
Receive Focus
  • In a phased array transducer, the echoes received
    by all of the individual transducer elements are
    summed together to create the ultrasound signal
    from a given depth.
  • Echoes received at the edge of the element array
    travel a slightly longer distance than those
    received at the center of the array, particularly
    at shallow depths.

77
  • Signals from individual transducer elements
    therefore must be rephased to avoid a loss of
    resolution when the individual signals are
    synthesized into an image.
  • Dynamic receive focusing is a method to rephase
    the signals by dynamically introducing electronic
    delays as a function of depth (time).

78
  • At shallow depths, rephasing delays between
    adjacent transducer elements are greatest.

79
  • With greater depth, there is less phase shift, so
    the phase delay circuitry for the receiver varies
    as a function of echo listening time.

80
  • In addition to phased array transducers, many
    linear array transducers permit dynamic receive
    focusing among the active element group.

81
Dynamic Aperture
  • The lateral spatial resolution of the linear
    array beam varies with depth, dependent on the
    total width of the simultaneously fired elements
    (aperture).
  • A process termed dynamic aperture increases the
    number of active receiving elements in the array
    with reflector depth so that the lateral
    resolution does not degrade with depth of
    propagation.

82
  • Side Lobes and Grating Lobes

83
  • Side lobes are unwanted emissions of ultrasound
    energy directed away from the main pulse, caused
    by the radial expansion and contraction of the
    transducer element during thickness contraction
    and expansion.
  • In the receive mode of transducer operation,
    echoes generated from the side lobes are
    unavoidably remapped along the main beam, which
    can introduce artifacts in the image.

84
  • In continuous mode operation, the narrow
    frequency bandwidth of the transducer (high Q)
    causes the side lobe energy to be a significant
    fraction of the total beam.
  • In pulsed mode operation. the low Q broadband
    ultrasound beam produces a spectrum of acoustic
    wavelengths chat reduces the emission of side
    lobe energy.

85
  • For multielement arrays, side lobe emission
    occurs in a forward direction along the main
    beam.

86
  • By keeping the individual transducer element
    widths small (less than half the wavelength) the
    side lobe emissions are reduced.
  • Another method to minimize side lobes with array
    transducers is to reduce the amplitude of the
    peripheral transducer element excitations
    relative to the central element excitations.

87
  • Grating lobes result when ultrasound energy is
    emitted far off-axis by multielement arrays, and
    are a consequence of the noncontinuous transducer
    surface of the discrete elements.
  • The grating lobe effect is equivalent to placing
    a grating in front of a continuous transducer
    element, producing coherent waves directed at a
    large angle away from the main beam.

88
  • This misdirected energy of relatively low
    amplitude results in the appearance of highly
    reflective, off-axis objects in the main beam.

89
  • Spatial Resolution

90
  • In ultrasound, the major factor that limits the
    spatial resolution and visibility of detail is
    the volume of the acoustic pulse.

91
  • The axial, lateral, and elevational (slice
    thickness) dimensions determine the minimal
    volume element.

92
  • Each dimension has an effect on the resolvability
    of objects in the image.

93
Axial Resolution
  • Axial resolution (also known as linear, range,
    longitudinal, or depth resolution) refers to the
    ability to discern two closely spaced objects in
    the direction of the beam.
  • Achieving good axial resolution requires that the
    returning echoes be distinct without overlap.

94
  • The minimal required separation distance between
    two reflectors is one-half of the spatial pulse
    length (SPL) to avoid the overlap of returning
    echoes, as the distance traveled between two
    reflectors is twice the separation distance.

95
  • Objects spaced closer than ½ SPL will not be
    resolved.

96
  • The SPL is the number of cycles emitted per pulse
    by the transducer multiplied by the wavelength.
  • Shorter pulses, producing better axial
    resolution, can be achieved with greater damping
    of the transducer element (to reduce the pulse
    duration and number of cycles) or with higher
    frequency (to reduce wavelength).

97
  • For imaging applications, the ultrasound pulse
    typically consists of three cycles.
  • At 5 MHz (wavelength of 0.31 mm), the SPL is
    about 3 x 0.31 0.93 mm, which provides an axial
    resolution of /2(0.93 mm) 0.47 mm.

98
  • At a given frequency, shorter pulse lengths
    require heavy damping and low Q, broad-bandwidth
    operation.
  • For a constant damping factor, higher frequencies
    (shorter wavelengths) give better axial
    resolution, but the imaging depth is reduced.
  • Axial resolution remains constant with depth.

99
Lateral Resolution
  • Lateral resolution, also known as azimuthal
    resolution, refers to the ability to discern as
    separate two closely spaced objects perpendicular
    to the beam direction.

100
  • For both single element transducers and
    multielement array transducers, the beam diameter
    determines the lateral resolution.

101
  • Since the beam diameter varies with the distance
    from the transducer in the near and far field,
    the lateral resolution is depth dependent.
  • The best lateral resolution occurs at the near
    fieldfar field face.

102
  • At this depth, the effective beam diameter is
    approximately equal to half the transducer
    diameter.
  • In the far field, the beam diverges and
    substantially reduces the lateral resolution.

103
  • The typical lateral resolution for an unfocused
    transducer is approximately 2 to 5 mm.
  • A focused transducer uses an acoustic lens (a
    curved acoustic material analogous to an optical
    lens) to decrease the beam diameter at a
    specified distance from the transducer.

104
  • With an acoustic lens, lateral resolution at the
    near field-far field interface is traded for
    better lateral resolution at a shorter depth, but
    the far field beam divergence is substantially
    increased.
  • The lateral resolution of linear and curvilinear
    array transducers can be varied.

105
  • The number of elements simultaneously activated
    in a group defines an effective transducer
    width that has similar behavior to a single
    transducer element of the same width.
  • Transmit and receive focusing can produce focal
    at varying depths along each line.

106
  • For the phased array transducer, focusing to a
    specific depth is achieved by both beam steering
    and transmit/receive focusing to reduce the
    effective beam width and improve lateral
    resolution, especially in the near field.

107
  • Multiple transmit/receive focal zones can be
    implemented to maintain Iateral resolution as a
    function of depth.

108
  • Each focal zone requires separate pulse echo
    sequences to acquire data.

109
  • One way to accomplish this is to acquire data
    along one beam line multiple times (depending on
    the number of transmit focal zones), and accept
    only the echoes within each focal zone, building
    up a single line of in-focus zones.
  • Increasing the number of focal zones improves
    overall lateral resolution, but the amount of
    time required to produce an image increases and
    reduces the frame rate and/or number of scan
    lines per image.

110
Elevational Resolution
  • The elevational or slice-thickness dimension of
    the ultrasound beam is perpendicular to the image
    plane.
  • Slice thickness plays a significant part in image
    resolution, particularly with respect to volume
    averaging of acoustic details in the regions dose
    to the transducer and in the far field beyond the
    focal zone.

111
  • Elevational resolution is dependent on the
    transducer element height in much the same way
    that the lateral resolution is dependent on the
    transducer element width.

112
  • Slice thickness is typically the worst measure of
    resolution for array transducers.
  • Use of a fixed focaI length lens across the
    entire surface of the array provides improved
    elevational resolution at the focal distance.

113
  • Unfortunately, this compromises resolution due to
    partial volume averaging before and after the
    elevational focal zone (elevational resolution
    quality control phantom image shows the effects
    of variable resolution with depth.

114
  • Multiple linear array transducers with five to
    seven rows, known as 1.5-dimensional (1.5-D)
    transducer arrays, have the ability to steer and
    focus the beam in the elevational dimension.

115
  • Elevational focusing is implemented with phased
    excitation of the outer to inner arrays to
    minimize the slice thickness dimension at a given
    depth (Fig. 16-25).

116
  • By using subsequent excitations with different
    focusing distances, multiple transmit focusing
    can produce smaller slice thickness over a range
    of tissue depths.
  • A disadvantage of elevational focusing is a frame
    rate reduction penalty required for multiple
    excitations to build one image.

117
  • The increased width of the transducer array also
    limits positioning flexibility.
  • Extension to full 2D transducer arrays with
    enhancements in computational power will allow 3D
    imaging with uniform resolution throughout the
    image volume.

118
IMAGE DATA ACQUISITION
  • Understanding ultrasonic image formation requires
    knowledge of ultrasound production, propagation,
    and interactions.
  • Images are created using a pulse echo method of
    ultrasound production and detection.

119
  • Each pulse transmits directionally into the
    patient, and then experiences partial reflections
    from tissue interfaces that create echoes, which
    return to the transducer.

120
  • Image formation using the pulse echo approach
    requires a number of hardware components
  • the beam former,
  • pulser,
  • receiver,
  • amplifier,
  • scan converter/image memory, and
  • display system.

121
  • Ultrasound equipment is rapidly evolving toward
    digital electronics and processing, and current
    state-of-the-art systems use various
    combin-ations of analog and digital electronics.

122
Beam Formers
  • The beam former is responsible for generating the
    electronic delays for individual transducer
    elements in an array to achieve transmit and
    receive focusing and, in phased arrays, beam
    steering.
  • Most modern, high-end ultrasound equipment
    incorporates a digital beam former and digital
    electronics for both transmit and receive
    functions.

123
  • A digital beam former controls application-specifi
    c integrated circuits (ASICs) that provide
    transmit/receive switches, digital-to-analog and
    analog-to-digital converters, and
    preamplification and time gain compensation
    circuitry for each of the transducer elements in
    the array.

124
  • Major advantages of digital acquisition and
    processing include the flexibility to introduce
    new ultrasound capabilities by programmable
    software algorithms and to enhance control of the
    acoustic beam.

125
Pulser
  • The pulser (also known as the transmitter)
    provides the electrical voltage for exciting the
    piezoelectric transducer elcnwnts, and controls
    the output transmit power by adjustment of the
    applied voltage.
  • In digital beam-former systems, a digital-to
    analog-converter determines the amplitude of the
    voltage. An increase in transmit amplitude
    creates higher intensity sound and improves echo
    detection from weaker reflectors.

126
  • A direct consequence is higher signal-to-noise
    ratio in the images, but also higher power
    deposition to the patient. User controls of the
    output power are labeled output, power, dB,
    or transmit by the manufacturer. In some
    systems, a low power setting for obstetric
    imaging is available to reduce power deposition
    to the fetus. A method for indicating output
    power in terms of a thermal index (TI) and
    mechanical index (MI) is usually provided (see
    section 16.1 1).
Write a Comment
User Comments (0)
About PowerShow.com