Title: Comparison of Automated and Manual Measurements of Horizontal Corneal Diameter in Contact Lens Weare


1
Comparison of Automated and Manual Measurements
of Horizontal Corneal Diameter in Contact
LensWearers Using Orbscan Eye Metrics Function
and Orbscan 11z Corneal Topographer
  • Dr Venkataraman Arvind MD , MRCOphth.
  • M.Krithiga B.S.(Optom)
  • Vasan Eye Care Hospital,
  • Chennai, India

2
Financial disclosure
  • The authors have no financial interest in this
    study or presentation

3
AIM
  • Comparison of automated and manual measurements
    of horizontal corneal diameter in contact lens
    wearers using Orbscan Eye Metrics Function and
    Orbscan 11z corneal topographer.

4
METHODS
  • Automated White to White (WTW) measurement was
    obtained from the Orbscan II.
  • Using the Eye metrics function the WTW was
    measured manually by two observers along the five
    point corneal reflections on the Orbscan real
    image.
  • The data was analysed using Statistica software
    for correlation ,reliability and Inter-rater
    repeatability.

5
Orbscan WTW
6
WTW WITH EYEMETRICS
5 point corneal reflections through which WTW is
measured
7
Results
  • 73 eyes of 37 patients were included in the
    study.
  • 25 females and 12 males.
  • Average duration of contact lens usage was 34
    months (range 16-52 months).
  • All of them were using soft contact lenses

8
Results
9
Results
Obs A
Obs B
Orbscan
10
RESULTS
  • Manual measurements (Observer AB) using eye
    metrics was statistically different compared to
    the automated measurements of Orbscan (plt0.001).
  • Standard deviation difference
  • Orbscan Vs OB-A is 0.19
  • Orbscan Vs OB-B is 0.10

11
Results
Correlation R 0.0009(pgt0.05). Slope -0.008
(Pgt0.05). Intercept -0.079 (P 0.05).
12
Results
Observer A Observer B
Average of measurements of Observer A B
Correlation R 0.0043(pgt0.05). Slope -0.0336
(Pgt0.05). Intercept 0.468 (P 0.05).
13
Results
  • Horizontal corneal diameter by manual Eye metrics
    measurements are longer than Orbscan derived
    measurements.
  • Manual measurement shows higher
  • inter-rater variability and test retest
    variability.
  • Variations tend to decrease with repeat
    measurements.

14
Conclusion
  • Orbscan IIz corneal topographer underestimates
    the WTW especially in long term contact lens
    wearers. This error can be reduced by manual
    measurements using Eye metrics software.
  • Manual corneal measurement can be very useful in
    patients with pannus and other peripheral corneal
    changes.
  • This method will be very useful adjuvant to
    automated measurement of corneal diameter in
    selection of Phakic intraocular lenses ( ICLs)
    and ACIOLs.

15
Thank you