Instrumentation noise equivalent exposure INEE for routine quality assurance: INEE measurements on a - PowerPoint PPT Presentation

1 / 27
About This Presentation
Title:

Instrumentation noise equivalent exposure INEE for routine quality assurance: INEE measurements on a

Description:

Typical Result for Digital Angiography (DA) Mode. Typical Result for Digital Subtraction Angiography (DSA) mode. Comparison of DA and DSA Modes ... – PowerPoint PPT presentation

Number of Views:99
Avg rating:3.0/5.0
Slides: 28
Provided by: Timo179
Category:

less

Transcript and Presenter's Notes

Title: Instrumentation noise equivalent exposure INEE for routine quality assurance: INEE measurements on a


1
Instrumentation noise equivalent exposure (INEE)
for routine quality assurance INEE measurements
on a clinical flat panel detector
  • T. Szczykutowicz, A. Kuhls-Gilcrist,
  • D. Bednarek, S. Rudin

2
Instrumentation Noise Equivalent Exposure (INEE)
  • INEE provides a measure of total instrumentation
    noise in terms of detector entrance exposure.
  • INEE model N2 k(Exposure INEE)
  • Noise (N2) is the variance in pixel gray values
    from flat-field images
  • k is a proportionality constant relating pixel
    gray values to units of exposure (µR)
  • If Exposure gt INEE ? Quantum Limited
  • If Exposure lt INEE ? Instrumentation Noise
    Limited

3
Instrumentation Noise Equivalent Exposure (INEE)
  • INEE provides a measure of total instrumentation
    noise in terms of detector entrance exposure.
  • INEE model N2 k(Exposure INEE)
  • Noise (N2) is the variance in pixel gray values
    from flat-field images
  • k is a proportionality constant relating pixel
    gray values to units of exposure (µR)
  • If Exposure gt INEE ? Quantum Limited
  • If Exposure lt INEE ? Instrumentation Noise
    Limited

4
Instrumentation Noise Equivalent Exposure (INEE)
  • INEE provides a measure of total instrumentation
    noise in terms of detector entrance exposure.
  • INEE model N2 k(Exposure INEE)
  • Noise (N2) is the variance in pixel gray values
    from flat-field images
  • k is a proportionality constant relating pixel
    gray values to units of exposure (µR)
  • If Exposure gt INEE ? Quantum Limited
  • If Exposure lt INEE ? Instrumentation Noise
    Limited

5
Instrumentation Noise Equivalent Exposure (INEE)
  • INEE provides a measure of total instrumentation
    noise in terms of detector entrance exposure.
  • INEE model N2 k(Exposure INEE)
  • Noise (N2) is the variance in pixel gray values
    from flat-field images
  • k is a proportionality constant relating pixel
    gray values to units of exposure (µR)
  • If Exposure gt INEE ? Quantum Limited
  • If Exposure lt INEE ? Instrumentation Noise
    Limited

6
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • If grid cannot be removed, the grid transmission
    factor can be used to acquire the detector
    entrance exposure
  • Removing the grid eliminates grid line noise
    patterns

7
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)

8
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)
  • Measure detector entrance exposure for range of
    mAs values

9
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)
  • Measure detector entrance exposure for range of
    mAs values
  • Disable image processing

10
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)
  • Measure detector entrance exposure for range of
    mAs values
  • Disable image processing
  • Acquire image sequence

11
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)
  • Measure detector entrance exposure for range of
    mAs values
  • Disable image processing
  • Acquire image sequence
  • Calculate the variance for each image sequence
    for each mAs value used and plot versus detector
    entrance exposure

12
Procedure for Quality Assurance
  • Remove anti-scatter grid
  • Add filtration to get a standardized spectrum
  • (e.g. RQA IEC 62220-1)
  • Measure detector entrance exposure for range of
    mAs values
  • Disable image processing
  • Acquire image sequence
  • Calculate the variance for each image sequence
    for each mAs value used and plot versus detector
    entrance exposure
  • Fit plot using linear regression and obtain the
    slope and intercept INEE Intercept / Slope

13
Varian PaxScan 2020 FPD
14
Varian PaxScan 2020 FPD
  • 1 frame / sec was used to minimize temporal
    effects.
  • Spectrum Used 70kV, 51.2 mm of AL (includes
    inherent tube filtration)

15
Typical Result for Digital Angiography (DA) Mode
16
Typical Result for Digital Subtraction
Angiography (DSA) mode
17
Comparison of DA and DSA Modes
DA
DSA
18
Comparison of DA and DSA Modes
Integration Capacitance 0.5 pF
DA
Integration Capacitance 4 pF
DSA
19
INEE Stability
  • Over a four month period, our system was tested
    three times and showed maximum deviations of 5
    and 2 for DA and DSA modes.

20
Effects of Image processingDA Mode
  • In DA mode, non linear image processing resulted
    in unusual variance behavior.

21
Effects of Image processingDA Mode
  • Even with linearization of the mean signal to
    detector entrance exposure and recalculation of
    the variance, the resulting trend was still found
    to be non linear.
  • DA mode, with (non-linear) image processing
    enabled, yielded no INEE.

22
Effects of Image processingDSA Mode
  • The un-subtracted images were used to calculate
    the mean signal and variance.
  • In DSA mode, no linearization was required (i.e.
    the image processing was linear).
  • INEE 39.2 0.4 µR.

23
Summary
  • INEE was determined for DA and DSA modes their
    differences could be due to integration
    capacitance differences.
  • DA and DSA modes implemented different image
    processing. In DA mode, the imaging processing
    was non-linear and the INEE could not be
    measured. In DSA mode, the image processing was
    linear and had no effect on the INEE.

24
Conclusions
  • The INEE metric was applied to a clinical FPD to
    obtain the minimal exposure at which the detector
    was quantum noise limited.
  • If, Exposure lt INEE detector performance is
    instrumentation noise limited
  • If, Exposure gt INEE detector performance is
    quantum noise limited
  • The INEE is a simple and direct measure,
    utilizing the variance in pixel gray values of a
    flat image over a range of exposures and
    requiring no special phantoms, and can be readily
    implemented in a clinical setting.

25
Future Work Incorporating Spatial Frequency
  • Use NPS(u,v) Instead of Signal Variance

Spatial Frequency
A Kuhls-Gilcrist et al., Instrumentation Noise
Equivalent Exposure (INEE) An Investigation of
Spatial Frequency Effects, AAPM 2008, SU-DD-A4-6
26
Future Work Incorporating Spatial Frequency
  • Relating to Frequency Independent Case

Varian Paxscan 2020
Quantum Limited Performance _at_ Exposures gt 3 µR
for all Spatial Frequencies (up to Nyquist)
A Kuhls-Gilcrist et al., Instrumentation Noise
Equivalent Exposure (INEE) An Investigation of
Spatial Frequency Effects, AAPM 2008, SU-DD-A4-6
27
Acknowledgements
  • NIH Grant R01-NS43924
  • NIH Grant R01-EB002873
  • Toshiba Medical Systems Corporation
Write a Comment
User Comments (0)
About PowerShow.com