Title: Morphology, Accuracy, and Histology of Corneal Flap Cuts Using a 200 kHz Femtosecond Laser


1
Morphology, Accuracy, and Histology of Corneal
Flap Cuts Using a 200 kHz Femtosecond Laser
R. Khoramnia C.P. Lohmann J. Salgado C.
Winkler von Mohrenfels
ASCRS 2010
  • Augenklinik, Klinikum rechts der Isar, Technische
    Universität München, Munich, Germany
  • All authors have no financial interest in the
    subject matter of this poster.
  • The laser system used is not FDA approved and
    does not yet have a CE mark.

2
Purpose
  • - LASIK requires precise corneal flap cutting.
  • Especially the creation of thin flaps has
    recently gained importance for Sub-Bowman
    keratomileusis (SBK), which requires precise
    cutting.
  • There is a tendency towards faster femtosecond
    lasers that produce flaps more quickly.

We analyzed flaps created with a prototype 200
kHz femtosecond laser (UltraFlap WaveLight AG,
Erlangen, Germany) concerning their cut accuracy,
morphology, and histology.
3
Material and methods
Preparation 36 fresh porcine eyes Epithelial
abrasion
Surgery Femtosecond laser flap cutting using the
200 kHz femtosecond laser -gt Three thickness
groups (100 µm, 130 µm, or 180 µm) -gt Flap
diameters were varied ranging from 8.0 to 9.5 mm
Examination 1.) Flap diameters were measured with
a sliding caliper. 2.) Flap thickness was
measured with a micrometer gauge (Extramess
2000 Mahr, Esslingen, Germany). Accuracy of
this device 0.3 µm 3.) Histologic examination
4
Results Flap thickness
Box-and-whisker plot
  • Accuracy - Statistical analysis
  • Significant difference between the 130- and
    180-µm group (Mann-Whitney-U test p lt 0.05)
  • All other comparisons No significant
    differences (Mann-Whitney-U test p gt 0.05).

Mean values 96.337.45
134.674.96 174.599.35 µm
5
Results Flap diameter
Box-and-whisker plot
  • Accuracy - Statistical analysis
  • The deviation between measured and attempted
    diameters was not significantly different between
    the four diameters (Kruskal-Wallis test p gt
    0.05).

Mean values 8.030.15 8.030.15
9.090.10 9.540.15 mm mm
6
Results Histological examination
  • (A) A 180-µm flap in the region of the side cut.
  • (B) Stromal bed in the region of the side cut
    (flap thickness 100 µm).
  • A 130-µm flap and the stromal bed.
  • HE staining original magnification, x20.

Histology showed a smooth cut with very little to
almost no changes in the corneal structure.
7
Conclusions
  • Flap creation was performed easily without any
    complications.
  • Morphology and accuracy of the cuts were very
    reliable and precise.
  • Histology showed a smooth cut with very little to
    almost no changes in the corneal structure.
  • -gt Therefore, the 200 kHz femtosecond laser
    seems to be a promising new device in refractive
    surgery.

8
CONTACT
Klinik und Poliklinik für Augenheilkunde
Klinik und Poliklinik für Augenheilkunde am
Klinikum rechts der Isar Ismaninger Str.
22 81675 Munich GERMANY SECRETARIAT Tel. 49 89
41 40-23 20 Fax 49 89 41 40-48 58 E-Mail
augenklinik_at_lrz.tum.de Internet
www.augenklinik-mri.de
8