Title: Morphology, Accuracy, and Histology of Corneal Flap Cuts Using a 200 kHz Femtosecond Laser
1Morphology, Accuracy, and Histology of Corneal
Flap Cuts Using a 200 kHz Femtosecond Laser
R. Khoramnia C.P. Lohmann J. Salgado C.
Winkler von Mohrenfels
ASCRS 2010
- Augenklinik, Klinikum rechts der Isar, Technische
Universität München, Munich, Germany - All authors have no financial interest in the
subject matter of this poster. - The laser system used is not FDA approved and
does not yet have a CE mark.
2Purpose
- - LASIK requires precise corneal flap cutting.
- Especially the creation of thin flaps has
recently gained importance for Sub-Bowman
keratomileusis (SBK), which requires precise
cutting. - There is a tendency towards faster femtosecond
lasers that produce flaps more quickly.
We analyzed flaps created with a prototype 200
kHz femtosecond laser (UltraFlap WaveLight AG,
Erlangen, Germany) concerning their cut accuracy,
morphology, and histology.
3Material and methods
Preparation 36 fresh porcine eyes Epithelial
abrasion
Surgery Femtosecond laser flap cutting using the
200 kHz femtosecond laser -gt Three thickness
groups (100 µm, 130 µm, or 180 µm) -gt Flap
diameters were varied ranging from 8.0 to 9.5 mm
Examination 1.) Flap diameters were measured with
a sliding caliper. 2.) Flap thickness was
measured with a micrometer gauge (Extramess
2000 Mahr, Esslingen, Germany). Accuracy of
this device 0.3 µm 3.) Histologic examination
4Results Flap thickness
Box-and-whisker plot
- Accuracy - Statistical analysis
- Significant difference between the 130- and
180-µm group (Mann-Whitney-U test p lt 0.05) -
- All other comparisons No significant
differences (Mann-Whitney-U test p gt 0.05).
Mean values 96.337.45
134.674.96 174.599.35 µm
5Results Flap diameter
Box-and-whisker plot
- Accuracy - Statistical analysis
- The deviation between measured and attempted
diameters was not significantly different between
the four diameters (Kruskal-Wallis test p gt
0.05).
Mean values 8.030.15 8.030.15
9.090.10 9.540.15 mm mm
6Results Histological examination
- (A) A 180-µm flap in the region of the side cut.
- (B) Stromal bed in the region of the side cut
(flap thickness 100 µm). - A 130-µm flap and the stromal bed.
- HE staining original magnification, x20.
Histology showed a smooth cut with very little to
almost no changes in the corneal structure.
7Conclusions
- Flap creation was performed easily without any
complications. - Morphology and accuracy of the cuts were very
reliable and precise. - Histology showed a smooth cut with very little to
almost no changes in the corneal structure. - -gt Therefore, the 200 kHz femtosecond laser
seems to be a promising new device in refractive
surgery.
8CONTACT
Klinik und Poliklinik für Augenheilkunde
Klinik und Poliklinik für Augenheilkunde am
Klinikum rechts der Isar Ismaninger Str.
22 81675 Munich GERMANY SECRETARIAT Tel. 49 89
41 40-23 20 Fax 49 89 41 40-48 58 E-Mail
augenklinik_at_lrz.tum.de Internet
www.augenklinik-mri.de
8