Title: A Step-By-Step Guide to Field Calibration


1
A Step-By-Step Guide to Field Calibration
  • for Gilian Air Sampling Pump

2
Items required
  • Sampling Pump
  • Collection Media
  • Gilibrator Calibrator

3
Field Calibration
  • A Field Calibration Consists of Two Parts
  • The Pre-Calibration (Pre-Cal) is done before the
    sample is taken
  • The Post-Calibration (Post Cal) is done after the
    sample is taken

4
The Pre-Cal
  • The Pre-Cal is used to set the desired sample
    flow rate, and to verify it on an air flow
    calibrator.

5
Conducting the Pre-Calibration
6
The Pre-Cal
  • Set the Flow Rate
  • Observe the Flow Rate
  • Record the Flow Rate
  • An average of 6 to 10 Gilibrator readings is
    recommended

7
Run The Sample
  • Pump is placed on the belt
  • Sample media is placed on the lapel (Breathing
    Zone) and pointed downward
  • Record the Start Time
  • Record the End Time

8
The Post-Cal
  • The Post-Cal is used to obtain the final flow
    rate and to verify that the flow rate stayed
    within /- 5 of the set flow.

9
The Post-Cal
  • Observe the Final Flow Rate
  • Record the Final Flow Rate
  • An average of 6 to 10 Gilibrator readings is
    recommended

10
Average Flow Rate
  • The Pre-Cal flow rate and the Post-Cal flow rate
    may be averaged to obtain an average flow rate
    for the sample.

11
Determining the Average Flow Rate
  • Pre-Cal Flow Rate Post Cal Flow Rate Avg Flow
    Rate 2

12
Cyclone Calibration
  • The Gilian 10mm nylon cyclone has only an exit
    fitting and must be placed inside a calibration
    jar to accept the Gilibrator.

13
Low Flow Modules
  • Calibrating with low flow modules

14
Constant Flow Modules
  • Set pump flow to 1500cc/min
  • Turn module switch on
  • Connect sample tubing to pump
  • Adjust desired flow rate at module
  • Read flow rate at Gilibrator

15
Multi-Flow Modules
  • Set pump to 1500 cc/min
  • Turn module switch on
  • Connect sample tubing to pump
  • Adjust flow rate at holder
  • Adjust each tube twice
  • Read the flow rates

16
Further Information
  • For detailed information on any of the Gilian
    products, please consult the product operation
    manuals.
  • Manuals may be accessed through our web site at
    www.Sensidyne.com

17
Thank You
  • Sensidyne, LP.
  • Clearwater, FL, USA
  • www.Sensidyne.com
  • Fax 727-539-0550
  • Phone 1 727-530-3602 / 800-451-9444