Transcript and Presenter's Notes

Title: Digital and multicolor flow cytometry


1
Digital and multicolor flow cytometry
  • Holden Maecker

2
An LSRII is not a Calibur with more colors
  • Optics fiber optics and octagons/trigons
  • Fluidics optimized for high flow rates
  • Electronics digital signal processing

3
PMT Octagon and Trigon
PMT
Bandpass filter
Longpass filter
4
Filter Nomenclature Conventions
  • Longpass filter allows light above a certain
    wavelength to pass, reflects shorter wavelengths
  • Example 505LP 505 nm longpass
  • Bandpass filter allows light within a certain
    range of wavelengths to pass (above and below a
    specified midpoint)
  • Example 530/30BP 515-545 nm bandpass

5
Effect of flow rate
  • Higher flow rates mean a broader sample stream
    (less precise focusing)
  • Less precise focusing means less accurate
    fluorescence measurement of dim populations
    (population spreading)
  • Higher flow rates also increase the frequency of
    coincident events (can be gated out based on FSC
    area vs. height)
  • In practice, flow rates up to 8000-12,000 events
    per second are acceptable on the LSRII (vs.
    2000-3000 events per second on a FACS Calibur)

6
Digital Signal Processing
  • Generates high resolution fluorescence values
    that can include negative numbers
  • No compression of populations at the low end of
    the fluorescence scale
  • More accurate representation of dim populations
  • Allows compensation to be performed in the
    software at any time
  • Uncompensated data and any associated
    compensation matrix are both stored
  • Compensation can be changed at any time
  • Peak area and peak height can both be recorded
    for all parameters

7
Biexponential display of digital data
Antibody capture beads stained with 3 levels
of an APC reagent The transformed display shows
aligned populations In the APC-Cy7 dimension
APC-Cy7 Area
All populations align correctly
APC Area
8
Spillover affects resolution senstivity
9
Conclusions
  • LSRII optics (using octagons and trigons) mean
    more efficient light collection and flexibility
    in use of detectors and filters
  • LSRII fluidics allow running at higher flow rates
    with minimal compromise to the data
  • Digital signal processing provides more accurate
    representation of dim populations, and more
    accurate and flexible compensationbut
    logarithmic data display may not be appropriate
  • More colors mean more spillover issues, with loss
    of resolution sensitivity in affected detectors