X-ray powder - PowerPoint PPT Presentation

About This Presentation
Title:

X-ray powder

Description:

X-ray powder diffractometer Can use calibration data to correct peak positions Measure periodically to see how diffractometer is performing Note axial divergence and ... – PowerPoint PPT presentation

Number of Views:58
Avg rating:3.0/5.0
Slides: 45
Provided by: Earle9
Category:
Tags: powder | ray

less

Transcript and Presenter's Notes

Title: X-ray powder


1
X-ray powder diffractometer
2
X-ray powder diffractometer
radiation counter
3
X-ray powder diffractometer
4
Powder diffractometer (see Jenkins Snyder,
chap. 7)

5
Powder diffractometer (see Jenkins Snyder,
chap. 7)
Bragg-Brentano geometry
6
X-ray powder diffractometer
7
X-ray powder diffractometer
8
X-ray powder diffractometer
9
X-ray powder diffractometer
10
X-ray powder diffractometer
11
X-ray powder diffractometer
12
X-ray powder diffractometer
13
X-ray powder diffractometer
14
X-ray powder diffractometer
15
X-ray powder diffractometer
16
X-ray powder diffractometer
17
X-ray powder diffractometer
18
X-ray powder diffractometer
To maintain the focusing geometry, when counter
is rotated at rate of 2?/min, specimen is also
rotated at rate of ?/min
??/min
?/min
19
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here, is
main rotation axis - asymmetric broadening
toward low 2? - negative 2? error up to 90
- positive 2? error beyond 90
20
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here, is
main rotation axis
21
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here is
main rotation axis each section of Soller
slit produces its own Debye ring from a
specific region in the specimen divergence
angle typically 2.5
22
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here is
main rotation axis each section of Soller
slit produces its own Debye ring from a
specific region in the specimen divergence
angle typically 2.5
23
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Flat specimen peak shifts to
lower 2? asymmetric broadening depends on
beam divergence 2?, plus specimen shape
24
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Specimen transparency
25
Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Specimen displacement Usually
largest error in expt'l data Neg. peak shift -
0.01 2?/15??displacement Asymmetric
broadening toward lower 2?
26
Powder diffractometer
Bragg-Brentano instrument Divergence slit
choice
27
Powder diffractometer
Bragg-Brentano instrument Divergence slit
choice Or, theta-compensating slit -
divergence slit opens as theta increases
28
Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice
29
Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice lowers intensity - little
improvement in resolution poor
resolution - little improvement in
intensity
30
Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice correct slit size
0.15
31
Powder diffractometer
Alignment steps
32
Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle
33
Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
34
Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
35
Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
36
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard NIST 640c (Si)
37
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard
38
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard Watch for Intensity
loss Angle shifts Resolution loss
39
Powder diffractometer
Alignment checks
40
Powder diffractometer
Alignment check "Five fingers of
quartz"
41
Powder diffractometer
Alignment check "Five fingers of
quartz" Measure position intensity of
(203)?1 H1/H2
42
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard NIST 1976 (Al2O3)
43
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard
44
Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard
Write a Comment
User Comments (0)
About PowerShow.com