Title: X-ray powder
1X-ray powder diffractometer
2X-ray powder diffractometer
radiation counter
3X-ray powder diffractometer
4Powder diffractometer (see Jenkins Snyder,
chap. 7)
5Powder diffractometer (see Jenkins Snyder,
chap. 7)
Bragg-Brentano geometry
6X-ray powder diffractometer
7X-ray powder diffractometer
8X-ray powder diffractometer
9X-ray powder diffractometer
10X-ray powder diffractometer
11X-ray powder diffractometer
12X-ray powder diffractometer
13X-ray powder diffractometer
14X-ray powder diffractometer
15X-ray powder diffractometer
16X-ray powder diffractometer
17X-ray powder diffractometer
18X-ray powder diffractometer
To maintain the focusing geometry, when counter
is rotated at rate of 2?/min, specimen is also
rotated at rate of ?/min
??/min
?/min
19Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here, is
main rotation axis - asymmetric broadening
toward low 2? - negative 2? error up to 90
- positive 2? error beyond 90
20Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here, is
main rotation axis
21Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here is
main rotation axis each section of Soller
slit produces its own Debye ring from a
specific region in the specimen divergence
angle typically 2.5
22Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Axial divergence - axis, here is
main rotation axis each section of Soller
slit produces its own Debye ring from a
specific region in the specimen divergence
angle typically 2.5
23Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Flat specimen peak shifts to
lower 2? asymmetric broadening depends on
beam divergence 2?, plus specimen shape
24Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Specimen transparency
25Powder diffractometer
Bragg-Brentano instrument - errors (see Klug
Alexander) Specimen displacement Usually
largest error in expt'l data Neg. peak shift -
0.01 2?/15??displacement Asymmetric
broadening toward lower 2?
26Powder diffractometer
Bragg-Brentano instrument Divergence slit
choice
27Powder diffractometer
Bragg-Brentano instrument Divergence slit
choice Or, theta-compensating slit -
divergence slit opens as theta increases
28Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice
29Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice lowers intensity - little
improvement in resolution poor
resolution - little improvement in
intensity
30Powder diffractometer
Bragg-Brentano instrument Receiving slit
choice correct slit size
0.15
31Powder diffractometer
Alignment steps
32Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle
33Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
34Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
35Powder diffractometer
Alignment steps Align to focal spot Set
take-off angle Zero alignment 21
alignment
36Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard NIST 640c (Si)
37Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard
38Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard Watch for Intensity
loss Angle shifts Resolution loss
39Powder diffractometer
Alignment checks
40Powder diffractometer
Alignment check "Five fingers of
quartz"
41Powder diffractometer
Alignment check "Five fingers of
quartz" Measure position intensity of
(203)?1 H1/H2
42Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard NIST 1976 (Al2O3)
43Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard
44Powder diffractometer
Alignment checks Run calibration pattern - use
suitable standard