LAT Integration and Test - PowerPoint PPT Presentation

About This Presentation
Title:

LAT Integration and Test

Description:

Option 4 TEM/TPS Removal (cont d) R&R of TEM/TPS - Basic Steps: Measure relative height to CAL base plate of TEM/TPS assembly in GRID. – PowerPoint PPT presentation

Number of Views:59
Avg rating:3.0/5.0
Slides: 24
Provided by: Pet1164
Category:
Tags: lat | basic | integration | test

less

Transcript and Presenter's Notes

Title: LAT Integration and Test


1
MRB NCR 562 Bay 0 and 4 TPS Removal and
Replacement
  • LAT Integration and Test
  • and Systems Engineering

2
Recommendation
  • Option 3 TPS Removal
  • Allows repair of TPS
  • Minimizes risk of damage to Flight Hardware
  • ELX has recalibration method for TEM/TPS
    telemetry
  • Accurate shim determination requires TPS or
    TPS/TEM/CAL measurement
  • TPS height and flatness data will be used to
    determine offset
  • TPS selection determined by best fit for shim,
    then ELX criteria
  • TEM or TEM/CAL could be removed if TPS data fails
    to meet criteria

3
Options
  • Options 1 through 5 are discussed in subsequent
    charts
  • Option 1 Use As Is
  • Option 2 TPS Removal and Rework
  • Option 3 TPS Removal
  • Option 4 TEM/TPS Removal
  • Option 5 CAL/TEM/TPS Removal

4
Option 1 Use-As-Is
  • IT Mechanical Pros
  • No demate of flight connections
  • No change to accurately measured CAL/TEM/TPS
    stack height
  • IT Mechanical Cons
  • None
  • IT Electrical Pros
  • No additional testing required
  • IT Electrical Cons
  • None
  • ELX Pros
  • No immediate change to TEM/TPS calibration or
    performance
  • ELX Cons
  • Power supply circuit design is marginally stable
    with temperature
  • 2 different TPS configurations within the LAT
  • Potential Risks
  • Power supply oscillation could negatively impact
    detector performance
  • Power supply oscillation could worsen over
    temperature and time
  • Rework later in the program could increase risk.
  • Discussion

5
Option 2 TPS Removal and Rework
  • IT Mechanical Pros
  • Demate TPS/TEM connection in Grid Low Risk
  • Demate TPS/PDU connection in Grid Low Risk, but
    connector jack screw may loosen
  • Maintains CAL/TEM/TPS stack height
  • IT Mechanical Cons
  • None
  • IT Electrical Pros
  • Maintains TEM/TPS telemetry calibration
  • Maintains calibration of Tracker and Calorimeter
    detectors
  • IT Electrical Cons
  • Need regression test definition to validate
    re-assembled TEM/TPS in the grid
  • ELX Pros
  • Permits rework of TPS for power supply
    oscillation
  • Maintains TEM/TPS telemetry accuracy
  • Maintains TEM/TPS Performance
  • ELX Cons
  • Testing of removed TPS required
  • Potential Risks
  • Impact to IT schedule while TPS is reworked and
    retested

6
Option 3 TPS Removal
  • IT Mechanical Pros
  • Good access to the interface between TEM and TPS
  • Shimming may not be required if TPS box height
    meets criteria
  • Demate TPS/TEM connection in Grid Low Risk
  • Demate TPS/PDU connection in Grid Low Risk, but
    connector jack screw may loosen
  • IT Mechanical Cons
  • Stack height measurement is invalidated and
    becomes inferred
  • If shimming is required it is a more complex
    operation to loosen and shim at the CAL interface
  • IT Electrical Pros
  • None
  • IT Electrical Cons
  • Requires recalibration of TEM/TPS telemetry
  • Need regression test definition to validate new
    TEM/TPS Assembly in the grid
  • Requires recalibration of Tracker and Calorimeter
    detectors
  • ELX Pros
  • Permits rework of TPS for power supply
    oscillation
  • ELX Cons
  • Degradation in TEM/TPS telemetry accuracy
  • TEM/TPS measured performance becomes inferred

7
Option 3 TPS Removal (contd)
  • Potential Risks
  • Unknown where or how shimming will be performed,
    if required
  • Shimming in the Grid has not been demonstrated.
  • Discussion
  • 18 TPS units have measured flatness variation
    within 1 mil
  • 17 TPS units have measured height variation
    within 4 mils
  • Height was not measured for TPS units currently
    installed in Bays 0 GLAT1749 sn 22 and 4
    GLAT1750 sn 03
  • GLAT1749 and GLAT1750 height and flatness would
    be measured after removal
  • GLAT1813 sn 11 and GLAT1828 sn 02 height and
    flatness would be measured prior to installation
  • Height variance of measured TPS chassis is 0.004
  • Flatness variance of TPS measured chassis is 0.001

8
Option 3 TPS Removal (contd)
  • Discussion (contd)
  • Suggested re-calibration method for new assembled
    units
  • CAL and TKR HV Bias Currents
  • Maintain the slope constants for both CAL and TRK
    Bias currents.
  • Re-calibrate the intercept constants by taking
    current average of one thousand measurements
    before and after TPS replacement, and then adjust
    the new average measurement to old reading with
    new value of Intercept point.
  • For Total TEM/TPS current
  • Re-calibrate the slope and intercept constants by
    taking the average of one thousand measurements
    before and after replacement of TPS in
    conditions    - no TRK or CAL FE is powered   
    - CAL FE is powered and TRK FE is not    - TRK
    FE is powered and CAL FE is not    - TRK and CAL
    FE is poweredThese four points allow make it
    easy to make fit and find new calibration
    constants for new assembly.Telemetry points can
    be verified by recording external PS current
    reading for all 4 points before and after
    replacement.

9
Option 3 TPS Removal (contd)
  • RR of TPS Basic Steps
  • Measure relative height of TEM/TPS assembly to
    CAL base plate in the GRID.
  • Remove TPS from TEM on CAL in GRID.
  • Measure height of TPS removed from GRID on
    granite table.
  • Measure new power supply on granite table.
  • Determine new shim thickness.
  • Install new shims.
  • Install new TPS.
  • Verify relative height obtained in step 1.

10
Option 3 TPS Removal (contd)
  • TPS Replacement IT Risk Summary

Risk Description Weight Factor (1-5) Risk (1-3) Score Mitigation Plan
Demate and mate of new TPS to TEM connector. 4 2 8 Dry run assembly procedure
If shimming is required it is a more complex operation to loosen and shim at the CAL interface 4 3 12 Demonstrate with EM units in 1x4.
Regression Testing to Validate 2 2 4 Utilize existing single tower tests
Total 24
If shimming is not required the overall risk
score is reduced by 12 pts.
11
Option 4 TEM/TPS Removal
  • IT Mechanical Pros
  • Demate TPS/TEM connection outside Grid Low Risk
  • Demate TPS/PDU connection in Grid Low Risk, but
    connector jack screw may loosen
  • Exposes existing shimming interface between
    TEM/CAL
  • Shimming operation is cleaner and consistent with
    technique used in the metrology bay. Needs
    pathfinding.
  • Installation of thermistor cable is simplified
    with the box removed.
  • IT Mechanical Cons
  • Re-shimming required
  • Requires new in Grid shimming procedure
  • Demate 8 TEM/TKR flex cables in Grid Elevated
    Risk, EXTREMELY LOW Risk Tolerance
  • Demate 4 Calorimeter flex cables in Grid
    Elevated Risk
  • IT Electrical Pros
  • None
  • IT Electrical Cons
  • Need regression test definition to validate new
    TEM/TPS Assembly in the grid
  • Requires recalibration of Tracker and Calorimeter
    detectors

12
Option 4 TEM/TPS Removal (contd)
  • ELX Pros
  • Permits rework of TPS for power supply
    oscillation
  • No degradation in TEM/TPS telemetry accuracy
  • TEM/TPS performance is based on measurements
  • ELX Cons
  • None
  • Potential Risks
  • Shimming in the Grid has not been demonstrated
  • Requires demate of TEM/TKR flex cables that could
    require extensive TKR rework if damaged. There
    are no spares.
  • Requires demate of TEM/CAL cables.

13
Option 4 TEM/TPS Removal (contd)
  • RR of TEM/TPS - Basic Steps
  • Measure relative height to CAL base plate of
    TEM/TPS assembly in GRID.
  • De-mate CAL, TKR, TEM to GASU, and TPS to PDU
    cables from TEM/PSU assembly.
  • Remove TEM/TPS from CAL in GRID.
  • Measure height of TEM/TPS removed from GRID on
    granite table.
  • Measure new TEM/TPS on granite table.
  • Determine new shim thickness based on the height
    differences between the removed and replacement
    assemblies.
  • Install new TEM/TPS with shim stack up determined
    in previous step.
  • Verify relative height obtained in step 1.
  • Install thermistor cable.
  • Mate CAL, TKR, TEM to GASU, and TPS to PDU
    cables to the newly installed TEM/TPS assembly

14
Option 4 TEM/TPS Removal (contd)
  • TEM/TPS Replacement IT Risk Summary

Risk Description Weight Factor (1-5) Risk (1-3) Score Mitigation Plan
Requires breaking flight connections to CAL, TKR, GASU and PDU connections (13 Connectors). 5 3 15 Proceed cautiously. Do not use excessive force for mate or demate.
Re-shimming Required 4 2 8 Demonstrate with EM units in 1x4.
Regression Testing to Validate 2 1 2 Utilize existing single tower tests
Total 25
15
Option 5 CAL/TEM/TPS Removal
  • IT Mechanical Pros
  • Demate TPS/TEM connection outside Grid Low Risk
  • Demate TPS/PDU connection in Grid Low Risk, but
    connector jack screw may loosen
  • Permits accurate CAL/TEM/TPS shimming in
    metrology bay
  • IT Mechanical Cons
  • Demate 8 TEM/TKR flex cables in Grid Elevated
    Risk, EXTREMELY LOW Risk Tolerance
  • Demate 4 Calorimeter flex cables outside Grid
  • Removal of installed shear plates for Bays 0 and
    4 may not be desirable with existing nut failure
    (NCR 579)
  • IT Electrical Pros
  • TEM/TPS regression testing could be performed
    outside of Grid
  • Permits TEM/CAL re-calibration in metrology bay
  • IT Electrical Cons
  • Requires re-calibration of TKR detector in Grid
  • ELX Pros
  • Permits rework of TPS for power supply
    oscillation
  • No degradation in TEM/TPS telemetry accuracy
  • TEM/TPS performance is based on measurements
  • ELX Cons
  • None

16
Reference
  • Completed Action Items From Previous MRB
  • COMPLETE - QA identify which TPS chassis s/n are
    installed and to be installed.
  • TPS TPS/TEM ASSY TPS CHASSIS Height
  • GLAT1749 GLAT1752 LAT-DS-00995-51-22 Not
    measured
  • GLAT1750 GLAT1753 LAT-DS-00995-51-03 N
    ot measured
  • GLAT1813 GLAT1832 LAT-DS-00995-51-11 3
    .0618
  • GLAT1828 GLAT1845 LAT-DS-00995-51-02
    3.0597
  • COMPLETE - SE Assess impact of TPS swap based on
    chassis dimensions with inputs from IT and M.
    Nordby.
  • See Recommendation
  • Only 19 of the 23 TPS chassis were measured.
  • Flatness data variance of 19 boxes is .001
  • Height data variance of 18 boxes is .004
  • IT and M. Nordby will attend MRB.
  • COMPLETE QA How were TPS chassis heights
    measured?
  • The measured TPS units were measured on the CMM
    (Coordinate Measurement Machine). The height
    measurement of 3.060 0.005 was measured by the
    CMM's gaussian best fit plane. The part was
    constrained to the granite table per drawing
    instructions for parallelism. The TPS units were
    measured for a 0.010" parallelism free state
    condition, there were a few close to the limit,
    but not out of tolerance. The parallelism
    measurement was performed with an indicator and
    height gage.
  • COMPLETE (See presentation) - IT assess
    feasibility of performing height measurements in
    Grid for all options.
  • COMPLETE (See presentation) - IT assess
    shimming in Grid shimming for all options.
  • COMPLETE - ELX identify candidate TPS for
    installation in to Bays 0 and 4
  • Suggestion is to use for TPS  replacement on 
    tower A and B the following units1. GLAT1813
    TPS from Assembly GLAT1832 should replace TPS 
    GLAT1749 from  assembly GLAT1752 2. GLAT1828 TPS
    from Assembly GLAT1845 should replace TPS 
    GLAT1750 from  assembly GLAT1753.

17
Reference
  • COMPLETE - ELX identify candidate TPS for
    installation in to Bays 0 and 4
  • Suggestion is to use for TPS  replacement on 
    tower A and B the following units
  • 1. GLAT1813 TPS from Assembly GLAT1832 should
    replace TPS  GLAT1749 from  assembly GLAT1752
  • 2. GLAT1828 TPS from Assembly GLAT1845 should
    replace TPS  GLAT1750 from  assembly
    GLAT1753.Selection of these units are based on
    similarity of HV calibration constants

18
Reference
  • Bay 0 GLAT1813 -gt GLAT17491752 (TPS GLAT1749)

Calibration Date Slope Intercept
CAL Bias Current 2/2/05 0.930382662178 -0.0014032780581
TKR Bias Current 2/2/05 0.935558213381 0.00234355041672
TEM/TPS Current 2/2/05 66.973583 -1.518328
  • 1832 (TPS GLAT1813)

Calibration Date Slope Intercept
CAL Bias Current 5/9/05 0.917697111398 0.000884641517652
TKR Bias Current 5/9/05 0.937436903933 0.0211388061165
TEM/TPS Current 5/9/05 66.993715 -0.837730
19
Reference
  • Bay 4 GLAT1828 -gt GLAT1750 
  • 1753 (TPS GLAT1750)

Calibration Date Slope Intercept
CAL Bias Current 1/31/05 1.05784628111 0.00439458111717
TKR Bias Current 1/31/05 0.930910460477 -0.0122075240151
TEM/TPS Current 1/31/05 66.510847 0.064750
  • 1845  (TPS GLAT1828)

Calibration Date Slope Intercept
CAL Bias Current 5/9/05 1.00367085354         -0.00551484961688
TKR Bias Current 5/9/05 0.956402741661 0.00380408724678
TEM/TPS Current 5/9/05 66.388597 0.089839
20
Reference
  • NCR 562 - TPS CCAs on GLAT 1752 and GLAT 1753
    TEM/TPS assys have not been reworked per NCR 397
    to correct observed oscillation of CAL 3.3V
    digital. Assemblies were integrated onto the
    Grid before the problem was observed in TEM/TPS
    test data
  • From Parent NCR 397
  • Description of Non Conformance
  • Oscillation of CAL-3.3V Digital when TEM/TPS
    tester board is connected. Amplitude was found
    around 100mV, 1-2 KHz frequency.
  • Oscillation is triggered at power-on of
    calorimeter and sustained. No oscillation is
    present when turning the power on at cold
    temperatures (-40C). At room-temperature
    oscillation might or might not be triggered by
    power-on. At hot temperatures (50C) oscillation
    was always triggered at power-on. Whether or not
    the oscillation is triggered at power-up, is
    dependant of the temperature at power-up. If the
    temperature of the module is changed after power
    up, the presence of oscillation (or not) is
    maintained and is independent of whether the
    temperature is changed.
  • Test showed that the supply only oscillates with
    loads drawing around/more than 1.2A. It does not
    oscillate at loads of around 600mA (the CAL has
    lt1A). In addition it appears to only oscillate
    when capacitive loads are added in parallel to
    the resistive loads.
  • No oscillation was observed at the TPS-only test.
    That tester includes tests at 150 of the cal
    nominal load, but does not have capacitive loads
    in parallel. It is not clear whether the problem
    is present when CAL AFEE boards are connected. It
    is known that on tower A and B the TPS system is
    stable at room-temperature.
  • Dispositions
  • (Cullinan, 4/5/05) Per MRB discussion 4/1/05 and
    4/4/05, the following rework on the qual TPS
    board is underway on AIDS 974 to resolve the
    oscillation issue observed
  • 1. Conformal coating on board in areas adjacent
    to CAL resistors R53, R58, TKR resistors R122,
    R142, R61, TEM resistor R20, R50 has been removed
    using wooden stick in preparation for attaching
    capacitors to be connected in parallel with these
    resistors, per direction from G. Haller. Caps
    will use 30 gauge wire pigtails to connect to on
    board resistors. Caps CDR33BX473AKUS will be
    used in this application. Caps will be bonded to
    board using Hysol EA9394.

21
Reference
  • NCR 562 (contd)
  • 2. On board CAL resistors R28, R25, R51, R52
    will be removed and replaced with 2.74KOhm
    resistors M55342K06B2E74R per G. Haller
    direction.
  • 3. On board Compensation Capacitors C531, C532
    (CAL bias) and C631, C632, C633 (TKR bias) have
    present value of 10pF. Remove and replace these
    caps with 100pF caps (on TPS CCA parts list), P/N
    CDR31BP101BKUS.
  • Dispo 2 (Cullinan, 4/7/05) Rework was completed
    4/6/05 and the qual TPS board was tested on the
    TPS tester and connected to the AFEE/MCMs and
    tested up to 75 deg. C. Per G. Haller,
    oscillation was not observed with simulated load
    of greater than 150 of expected TKR load.
    Oscillations were noted on high voltage when
    simulated load was approx. 250 of expected TKR
    load. MRB was held 4/7/05. In attendance were
    Haller, Sapozhnikov, Hascall, Bright, Klaisner,
    Cullinan, Bloom, J. Lohr, B. Graf, N. Johnson to
    discuss test results. MRB concurred that testing
    performed to date was sufficient to validate new
    qual TPS board configuration and the elimination
    of the oscillation within TPS performance
    requirements.
  • Additional TPS board testing will be completed
    4/7/05. All testing will be performed to AIDS
    974, with data packages attached to this AIDS.
    Qual TPS board will be reinstalled into TPS base.
    Thermal compound CV-2946 will not be used (SLAC
    does not have material on hand) between board and
    TPS base. Analysis by DAQ showed temperature
    rise of TPS board was approximately 3-4 deg C
    without the thermal compound. Also, areas where
    conformal coating was removed and newly added
    pigtailed capacitors will not be covered with
    Arathane conformal coating at this time. In
    addition, all screws securing the TPS board into
    the TPS base, securing the TPS lid, and securing
    the TPS to the TEM will not be staked with Hysol
    0151. These materials are not immediately
    available at SLAC. Need for qual TEM/TPS for
    testing at NRL is urgent and the qual TPS board
    can be used for NRL testing in this
    configuration. When Qual TEM/TPS is returned to
    SLAC, the TPS will be reassembled to include
    these materials to return the unit to meet the
    assembly drawing.
  • MRB also discussed the plan to document these
    configuration changes. Relevant TPS assembly
    drawings, schematics and BOMs will need to be
    revised. Report list of documents that will be
    revised to this NCR once list is compiled.

22
Reference
  • NCR 562 (contd)
  • MRB also concurred to rework remaining TPS boards
    at SLAC prior to pre-conformal coat performance
    testing and thermal cycle testing. The CCAs will
    arrive from General Technology and will be
    reworked per this NCR and AIDS 994. The
    redlined schematic used to document configuration
    changes for the qual TPS CCA will be used for
    reworking the other TPS boards. The TPS CCA
    schematic diagram, assembly drawing and BOM/parts
    list will be revised to capture redline changes.
    GSFC inspection of reworked CCAs will be
    performed by Tracy Shepherd. Following rework,
    TPS CCAs will complete performance and thermal
    cycle testing and returned to General Technology
    for final assembly.
  • Dispo 3 (Cullinan, 4/12/05) Rework (including
    capacitor/jumper wire staking with EA9394),
    inspection and Performance testing has been
    completed on TPS CCAs GT104, 105, 106, 108, 109,
    110. GT104, 105, 106 and 108 to be shipped to
    GT GT109, 110 still require performance and
    thermal cycle testing following rework.
    Photographs of individually staked
    capacitors/jumper wires and overall board views
    were taken for all boards and are stored on V
    drive under Ancillary file in Electronics folder
    (under GLAST folder on V drive).
  • All remaining boards received from GT will be
    reworked per this NCR and AIDS 994.
  • Dispo 4 (cullinan, 4/20/05) TPS boards GT109
    and GT110 successfully completed EICIT
    (LAT-TD-04850-1), SVT (LAT-TD-04849-1),
    performance testing (LAT-TD-01652-4), thermal
    cycling (5 cycles) and post thermal cycle
    performance testing (LAT-TD-01652-4) on AIDS 872
    following rework.
  • All test results have been retained in the work
    order binders generated for each CCA and archived
    in bonded stores.
  • The following drawings have been revised to
    document this TPS CCA rework and are currently in
    sign off cycle
  • 1. TPS CCA assembly LAT-DS-02388, update to rev.
    58
  • 2. TPS CCA Schematic, LAT-DS-02390, updated to
    rev. 57
  • 3. TPS CCA BOM, LAT-DS-02391, updated to rev. 57
  • Per MRB 4/14/05, DAQ Electronics group to provide
    an advisory memo to Systems Engineering
    documenting this anomaly and directing test and
    service operations of the TEM/TPS to avoid any
    condition that would risk initiating oscillation.

23
Reference
  • NCR 562 (contd)
  • MRB concluded that no additional EMI/EMC testing
    or workmanship vibration testing is required for
    the reworked qual TEM/TPS module.
  • Plan is to close this NCR with MRB approval, and
    open new NCR to document required rework for TPS
    CCAs currently installed in TPS boxes of GLAT
    1752 and GLAT 1753 flight modules currently
    integrated onto Towers 1 and 2. Plan is to
    remove these TPS boxes and rework/replace with
    reworked TPS CCAs.
  • (Cullinan, 7-14-05) NCR 562 created per MRB to
    direct disposition of GLAT 1752 and GLAT 1753
    TEM/TPS for reworking the TPS CCAs of these two
    assemblies.
Write a Comment
User Comments (0)
About PowerShow.com