Title: Fluoroscopy Review Notes From CDPH RHB Syllabus
1Fluoroscopy Review NotesFrom CDPH RHB Syllabus
2Fluoroscopy Notes Ch 1
- Approximately 5 of the US population has a
fluoro procedure each year - The average number of fluoro exams per person is
1.3 - The average number of spot films is 4.6
3- The MC exam is a GI tract at 53
- A 2 minute UGI exam can produce an exposure
ranging from 5-15 rads, comparatively a KUB is
between 100-500 mrads. - Fluoro is defined as a rad exam utilizing
fluorescence for the observation of the transient
image.
4- Fluoro was first used as a dynamic procedure.
Second as a means of positioning for spot films. - Medical exposure accounts for about 20 of the
total radiation people receive. - Even though the percentage is small, for medical
exposure, it is the only exposure that is
controllable. - FLUOROSCOPY TO POSITION PATIENTS IS PROHIBITED.
5Notes Chapter 2Factors Directly Affect Exposure
- mA
- kVp
- Collimation
- Filtration
- Exposure time
- Total fluoro time
- Target to panel distance (TPD)
- Patient to II distance
- Sensitivity of the image receptor
- Essentially speed RSV
6The following will reduce exposure
- Collimating
- Last frame hold
- Shortest possible patient to II distance
- Highest possible kVp
- Pulsed fluoroscopy
- Using the largest II mode with collimation
7Factors indirectly influencing exposure
- Room illumination
- Image receptor quality
- Absorption of the table top
8mA
- 0.5 5 mA
- Usually 1 3 mA
- Spot films
- 100 mA or higher
- Output and dose are directly proportional to mA
9kVp
- Maximum photon energy
- Beam quality
- Penetrability of the beam
- Tube potential
10collimation
- Required by law
- Image quality improves as the beam is collimated
11Collimation
- Collimate tightly to the area of interest.
- Reduces the patients total entrance skin
exposure. - Improves image contrast.
- Scatter radiation to the operator will also
decrease.
12Factor affecting staff doses
FIELD SIZE DEPENDENCE
Scattered dose rate is higher when field size
increases
11x11 cm
17x17 cm
17x17 cm
100 kV
0.8 mGy/h
1.3 mGy/h
1 mA
0.6 mGy/h
1.1 mGy/h
0.3 mGy/h
0.7 mGy/h
1m patient distance
Patient thickness 18 cm
13filtration
- If the tube is operated above 125 kVp, 3 mm Al eq
is required. - Filtration reduces patient dose
14Source to table toptarget to panel distance
- Cannot be less than 12 and should be 18
- Mobiles are required to be at least 12
- Fixed units, 15
15Patient to II distance
- The closer the II, the lower the dose
- This is more pronounced with fixed units.
- Decreases the SID
16- Tabletop
- Less than 1 mm Al eq at 100 kVp
- Exposure switch
- Dead man type
17Primary protective barrier
- The II is the primary barrier and must have 2 mm
Pb eq for systems operating above 125 kVp - The II has to be in place for the tube to energize
18Protective Actions
- Bucky Slot Cover
- Automatically covered, 0.25 mm Pb eq
- Protective curtains
- 0.25 mm Pb eq
- Not required on c-arms
- Scatter at 1 foot can reach 500 mrad/hr
19- Allowable exposure rates
- Cannot exceed 5 rad/minutes
- Unless, ABC or image recording
- Cumulative timer
- Cannot exceed 5 minutes
- Illumination
20II Considerations
- Purpose
- The basic purpose of the II is to make the fluoro
image brighter - When the image is brighter it is easier to
visualize structures - Brightness Gain
- Minification gain multiplied by electronic (flux)
gain
21II Facts
- Input phosphor, cesium iodide
- Photcathode, danium antimony
- Output phosphor, zinc cadnium sulfide
22Image Quality Issues
- Quantum Mottle
- Caused by too few photons
- Contrast
- Subject
- Detector
- Image
- Resolution
23Effect of X ray Beam Penetration on Contrast,
Body Penetration, and Dose
24Dose vs. Noise
2 µR per frame
25- Distortion
- Size, shape, pincushion
- Lag
- Vignetting
- Less bright at the edges than center of image
- Magnification tubes, Multi-mode
- Variable FoV
26Closed Circuit TV Systems
- Camera
- MC is the vidicon
- Camera control unit
- Video amplifier
- Monitor
- CRT, etc.
27Cinefluoroscopy
- Synchronization
- Record with the x-ray pulses
- Framing frequency
- Division of 60
- The higher the rate the higher the dose
- F-number
- Video disk recording (electronic radiography)
- Exposure ends when image is formed
- Basically fluoro phototiming
- 95 dose reduction
28- Video tape
- Instant playback and no additional dose
- Spot films
- Conventional cassettes
- Photospot cameras
- ½ to 1/3 dose of convent. Cass.
- Lower image quality
29Accessories
- Gonadal shields
- Required when possible
- Grids
- Fluoro uses low ratio grids
- Cassettes
- Cine film
- Per frame basis 10 x the dose than fluoro
30Factors affecting an Increase in Scatter
- High kVp
- Large field size
- Thick body part
31Advantages of 3 phase and medium/high frequency
generators
- Relatively high mA
- Higher effective kVp
- Near constant potential
- Less ripple
32Notes for Chapter 3
- Fluoro Image Production
- Fluoro units have 2 basic components
- X-ray tube
- Image intensifier
33Fluoro X-ray Tube
- Regular rotating anode x-ray tube
- Runs at a lower mA
- Less than 5 mA
- Small focal spot
- Possible because of the low mA
34II
- The primary purpose is to increase the brightness
of the fluoro image - Components
- Glass envelope that provides a vacuum
- Input layer
- Converts x-ray photons to electrons
- Electronic (electostatic) lens
- Output layer
35- Input layer
- Converts x-ray photons to light photons
- Light photons then strike the photocathode and
convert into electrons - Electrons are then accelerated across the II
- Electrons strike the output phosphor and are
converted back into light photons
36The image intensifier (I.I.)
I.I. Input Screen
Electrode E1
Electrode E2
Electrode E3
Electrons Path
I.I.Output Screen
Photocathode
37Image intensifier component
- Input screen conversion of incident X Rays into
light photons (CsI) - 1 X Ray photon creates ? 3,000 light photons
- Photocathode conversion of light photons into
electrons - only 10 to 20 of light photons are converted
into photoelectrons - Electrodes focalization of electrons onto the
output screen - electrodes provide the electronic magnification
- Output screen conversion of accelerated
electrons into light photons
38Image Intensifier Magnification Modes
Same area
Output phosphor
Input Phosphor
9 inch field
6.5 inch field
39RELATIVE PATIENT ENTRANCE DOSE RATE FOR SOME UNITS
IMAGE INTENSIFIER
Active Field-of-View (FOV)
40Brightness Gain
- BG is the product of minification gain and
flux(electronic) gain - Minification Gain
- Input phosphor dia.2/output phosphor dia2
- Making the image smaller will make it brighter.
- The same number of photons are contained in a
smaller area - In most IIs the output phosphor is 1 inch.
41Flux Gain
- Caused by the conversion efficiency of the output
phosphor and the acceleration of the electrons
across the II - As the electrons accelerate they gain kinetic
energy - Flux gain is usually between 50 and 150.
42Measuring Brightness Gain
- The actual measurement is done by calculating the
conversion factor - Intensity of output phosphor (candelas)/mrads/sec
- Brightness gain will deteriorate 10 annually.
- This will ultimately decrease image contrast
43Beam Splitter Mirror
- 10 of the output light goes to the vidicon
(video camera) the remainder goes to the
photospot device. - NOTE not all units have a beam splitter
44Vignetting and Pincushion Distortion
- Pincushion
- The loss of shape at the edges of the fluoro
image - Vignetting
- Loss of brightness at the edge of the image
45Veiling Glare
- Occurs when the light from the output phosphor
reflects back into the II. - Remember, the photocathode is stimulated by
light, so light reflecting from the output
phosphor would also trigger electron production. - Decreased contrast results
46Automatic Brightness Stabilization
(ABS)Automatic Brightness Control (ABC)
- Keeps light output of the II constant.
- Brightness of the image varies with changes to
kVp and mA. - Increase mA increase brightness direct
relationship - Increase kVp 10 double brightness
47Brightness Sensing
- II photocathode current
- Television camera signal sensing
- Lens coupled phototube sensing
48Types of ABS
- Variable mA, preset kVp
- Set the kVp and the unit adjusts mA
- Variable mA with kVp following
- If the mA range is exceeded the unit will
automatically adjust the kVp to compensate - Variable kVp, preset mA
- Set the mA and the unit adjusts kVp
- Variable kVp, variable mA
49Closed Circuit TV Systems
- Camera
- Camera control unit
- Power supply and video amplifier
- Monitor
50Cameras
- Vidicon
- MC, inexpensive, lag, 525 raster lines
- Plumbicon
- Cardiac cath labs
- Fixed gain (better contrast) and low lag
- Increased quantum mottle
- Image orthicon
- Not widely used
- CCD
- Solid state semiconductor
- Small, low power consumption, low price, long life
51Photoconductive camera tube
52Schematic structure of a charged couple device
(CCD)
53Monitor
- 525 lines 30 times per second
- Combined with the video camera improves image
contrast
54TV Image Quality
- Horizontal resolution
- Vertical resolution
- Contrast
- Brightness
- Lag
55- Horizontal resolution
- Bandwidth or bandpass
- Increase frequency bandwidth increase horizontal
resolution - Vertical resolution
- Determined by number of scan lines
- Kell factor
- Ratio of vertical resolution and scan lines
56- Contrast
- Brightness
- Adjust contrast first and brightness after
- Lag
- Occurs when the II is moved rapidly.
57Dynamic Image Recording
- Video Tape
- 2 advantages
- Instant replay
- No additional patient exposure
- Disadvantages
- Poor image quality, fixed frame rate,
58- Cinefluoroscopy
- 16 or 35 mm (MC high patient dose, better image)
- Synchronization
- Camera shutters open at the same rate as x-ray
pulses - Framing rate
- F-number
- The lower the number the more light hitting the
camera the lower the patient dose however, more
distortion at the edges
59- Framing
- Underframing should be avoided
- Exact framing, diameter of the II fits in the
shortest dimension of the film - Overframing, diameter of the II fits the largest
dimension of the film. Part of the image is lost - Total overframing, diameter of the II is equal to
the diagonal of the film
60Static Image Recording
- Video disk
- Last image freeze (hold) (sticky fluoroscopy)
- Electronic radiography, similar to AEC x-ray only
until image is made. - Decrease dose up to 95
- 1 to 30 frames per second
- Spot film
- Conventional cassettes
61- Spot film
- Photospot
- Image is taken from the II output phosphor
- Dose 20 50 X higher per frame than fluoro
because of higher mA - Currently, 70 mm roll, 105 roll, 100 mm chip
- 105 mm roll ½ the dose of cassette spots
- Cassette spots
- Slower frame rate
- Higher dose and better spatial resolution
62Digital Fluoroscopy
- Digital image is obtained from the output
phosphor. - A vidicon then a digital image processor
- Or, digital video camera
- Digital photospot
- Instant playback, possible image enlargement
63Pulsed Fluoroscopy
- Variable frame rates are possible with a
corresponding decrease in patient dose
64High level (boost) fluoroscopy
- Higher tube currents than normal
- 10-20 mA usual 40 mA potentially
- Increase patient dose 2-10 times reg. Fluoro
- 10-50 rads per minute
- Limited to 20 rad/minute unless recording the
image - Key points
- Special activation required, audible signal, dose
rate limited to 20 rad.minute
65(No Transcript)
66(No Transcript)
67Notes from Chapter 4Conducting the Fluoro Exam
- Operator dose is directly proportional to patient
dose - Image brightness is directly proportional to dose
rate at input phosphor
68Technical factors which directly influence dose
rate at the table top
- mA
- kVp
- Collimation
- Filtration
- Exposure time
- Target panel distance
69Technical factors which indirectly influence dose
by affecting technical factors
- Room lighting
- Image receptor quality
- tabletop
70Collimation
- A border needs to be visible when the II is 14
inches above the tabletop and the collimator is
fully opened - With an automatic collimator, a border should
always be visible - Image is not brighter with a less collimation
(bigger field size)
71Filtration
- Total filtration
- 2.5 mm Al eq lt 125 kVp
- 3.0 mm Al eq at 125 kVp and above
- Total filtration includes inherent and added
- Exposure rate should be less than 2.2 rads/min at
80 kVp - HVL
72Allowable exposure rates
- Limited to 5 rads per minute
- If the unit has ABC/ABS then 10 rads/minute is
allowed - However, if the unit has boost then the limit
is 5rads/minute - ABC/ABS units have to be checked by a physicist
annually - Also have to have weekly fluoro checks of mA and
kVp - No ABS 3 year check by physicist
73TPD
- TPD increases from 12 to 18 inches
- Pt. Dose decreases by 30
- II as close as possible to the patient
74Room Lighting
- Affects visual acuity
- Photopic acuity is 10 X better than scotopic
acuity - Day versus night vision
- Normal viewing distance is 12 15 inches
- Image recognition in 0.2 seconds
75Gonadal Shields
- 0.5 mm Pb eq
- 97 effective at 100 kVp and 3 mm Al filtration
76Notes from Chapter 5Basic Operational Procedures
- Minimize exposure time by utilizing short looks
- Use the cumulative timer
- Use the highest applicable kVp
- Collimate
- Use mag and boost only when necessary
- Use last image hold
77- Use a photospot instead of cassettes
- Use video tape
- Use II with good contrast
- Monitor the TV monitor for brightness and
contrast - Minimize the pt. II distance
- Position the II prior to exposure
78- Prevent pt. Motion with instructions
- Use gonadal shielding when possible
- Use compression devices
79Factor affecting staff doses
Scattered dose rate is higher near the area into
which the X-ray beam enters the patient
ANGLE DEPENDENCE
100 kV
0.9 mGy/h
1 mA
0.6 mGy/h
11x11 cm
0.3 mGy/h
1m patient distance
patient thickness 18 cm
80Factor affecting staff doses
DISTANCE VARIATION
mGy/h at 1m
mGy/h at 0.5m
Scattered dose rate is lower when distance to the
patient increases
100 kV
1 mA
11x11 cm
81Factor affecting staff doses
Tube undercouch position reduces, in general,
high dose rates to the specialists eye lens
X-Ray tube
mGy/h
100 kV
2.2 (100)
1 m
2.0 (91)
20x20 cm
1.3 (59)
mGy/h
1 Gy/h
(17mGy/min)
1.2 (55)
1.2 (55)
1.2 (55)
1 Gy/h
1m patient distance
(17 mGy/min)
1.3 (59)
20x20 cm
2.2 (100)
100 kV
1m patient distance
1 m
X-Ray tube
82Mobile Concerns
- Audible indicator
- Cumulative timer visible on the monitor
- Video storage
- Last frame hold
- Longest possible TPD
83Notes for Chapter 6Pediatric Fluoroscopy
- Motion
- Sedation
- Mechanical devices
- Personnel and Parental Protection
- Everyone in the room needs lead
- Gonadal shielding
- Artifacts
84- ABS/ABC
- Watch putting the II directly over large
concentrations of contrast - Distance
- Collimation
- Photospots instead of cassette spots if possible
85Notes for Chapter 7Mobile Fluoroscopic Equipment
- Primary beam is intercepted by the II
- If the II is used routinely in a single location
it needs to have secondary shielding - SSD has to be at least 12 inches
- Must have an II
- Collimation has to be used or unit will not
energize - Unit cannot energize unless the II is in the
primary beam
86- Maximum dose rate of 5 rads/minute
- Personnel monitoring in required for all persons
operating fluoro equipment - Protective aprons are required if exposed to more
than 5mrads/hour - Boost mode should only be used after areas of
interest have been localized
87Notes for Chapter 8Responsibilities of X-ray
Supervisor
- Chief Radiologist or designees are responsible
- Licentiates who can use and supervise fluoro
- Radiology supervisor and Operator
- Fluoro supervisor and Operator
- Techs with Fluoro permits can only use fluoro
when supervised by above
88Specific requirements
- Establish a fluoro procedures manual
- Annual review of manual
- Assure that techs dont practice medicine
- Observe tech performance
- Assure techs are offered training
- Assure equipment monitoring is adhered to
89Personnel Protection
- Operator is adequately protected from scatter
- Individuals in the room need to wear aprons and
film badges - Use protective devices as applicable
90Protection Devices
SCREEN AND GOGGLES
CURTAIN
THYROID
91Restrictions
- Techs can only use fluoro equip. under the
supervision of a supervisor operator - Techs cannot interpret films
- Techs cannot use a title implying the right to
practice medicine
92Notification Requirements
- Immediate notification (prompt phone call and
timely letter) - Total dose 25 rems
- Eye dose 75 rems
- Skin or extremity dose of 250 rems
- 24 hour notification ( call within 24 hours and a
letter follow-up) - Total dose 5 rems
- Eye dose 15 rems
- Skin or extremity 50 rems
93Notes on Chapter 9Supervision of Techs w/Fluoro
permits
- Clear the room of unnecessary personnel
- Collimate
- Use shields
- Use correct technical factors
- Position the patient correctly
- Avoid patient motion
94- Direct supervision
- Use equipment only as trained
- Indirect supervision
- Spot filming and video taping
95Notes on Chapter 10Health Effects of Low Level
Radiation dose
- Somatic dose indicators
- Injuries to superficial tissue
- Induction of cancer
- Cataracts, fertility issues, life-span shortening
- Injuries to developing fetus
- Based on dose at specific locations or points
96- High marrow dose exams
- BE, UGI, abdominal angio
- Genetic dose indicators
- 50 rads temp male sterility
- 30 rads temp female sterility
- Genetically significiant dose
- Number of future kids
- X-ray exam rate
- Mean gonadal dose/exam
97Notes on Chapter 11Biological Effects and
Significance of Dose
- Effects appear to follow a linear non-threshold
dose curve - Dose rate to tissue
- Total dose
- Type of cell exposed
98Radiobiological injury
- Cellular amplification
- Gross cellular effects
- MC effect is the cessation of cell division
- Latent Period
- Short term, weeks or less
- Immediate or early effects
- Long term, years or longer
- Delayed or late effects
99Dose relationship curves
Non-threshold linear
Threshold Non-linear
Regulations are based on non-threshold linear
curves
100Variations in Cell Sensitivity
- Bergonie and Tribondeau
- Number of undifferentiated cells
- Degree of mitotic activity
- Duration of active proliferation
- Radiation induced mitotic delay is usually
reversable
101Cell Sensitivity
- Lymphocytes or white blood cells
- RBCs
- Epithelial
- Endothelial
- Connective tissue cells
- Bone
- Nerve
- Brain
- Muscle
102Short Term Effects
- 25 rads or less demonstrate no effects
103Long Term Effects
- No specific effect associated with radiation
exposure - Somatic damage
- Increased incidence of cancer
- Embryological effects
- Cataracts
- Life span shortening
- Genetic mutations
104Carcinogenic Effects
- Human evidence
- Early radiologists and dentists
- Radium dial painters
- Uranium miners
- Survivors of Hiroshima and Nagasaki
105Radiation Induced Cancers
- Female breast
- Thyroid
- Hemopoitic tissue
- Lungs
- GI tract
- Bones
106Embryological Effects
- As little as 10 rads demonstrates effects in
animal models - 50 rads can cause spontaneous abortion
107Notes on Chapter 12Personnel Radiation Protection
- ALARA
- Basis for radiation protection requirements
- Stochastic effects
- Probability of an event occurring, ie cancer
- Non-stochastic effects (deterministic)
- Severity of the effects varies with exposure
108Operator Exposure
- Distance
- Apparel
- Aprons, 0.25 mm Pb 97 effective
- 0.5 mm Pb, 99.9 effective
- Should be placed on hangers when not in use
- Aprons cover 80 of the bone marrow
109Notes on Chapter 13Personnel Monitoring
- Record exposure
- Measure accumulated exposure
- Indicate type of exposure
- Provide a record of exposure
110Types
- Film badge
- 10 mrad to 700 rads
- /- 25 accuracy
- TLD
- Lithium fluoride
- /- 9, cannot be reread
- Others
- Pocket dosimeter
- Audible device
111Maximum Permissible DoseFor adults over 18 y/o
- Whole body - head, trunk, arms above the elbow,
and legs above the knee - 5 Rem
- Skin and extremities
- 50 rem
- Lens
- 15 Rem
- Occupational dose for people under 18 y/o 10 of
adult dose
112Occupational Exposure Limit
Whole Body 5rem/year Extremities
50rem/year Eye 15rem/year Pregnant workers
0.5rem/gestation period
General public Limited to 0.1 rem/year
(Addition to the background radiation) 0.002 Rem
(2 mrem) per hour
113Who must be monitored?
- Persons in high radiation area
- 0.1 Rem per hour at 30 cm
- Fluoro rooms
- Persons operating mobile x-ray equipment
- Radiation Area
- 0.005 Rem per hour at 30 cm
114Typical Exposures
5000
4000
mrem/yr
3000
2000
1000
0
X-Ray Tech
Pain Mgnt.
MPD
Cardio.