Meter Testing - Traditional - PowerPoint PPT Presentation

1 / 19
About This Presentation
Title:

Meter Testing - Traditional

Description:

Dave Elmore, Don Bullock (ret), Les Rosenau, Mike Coit, Warren Germer, Joe Provost ... Kh - Watt-hours per (equivalent) disk revolution ... – PowerPoint PPT presentation

Number of Views:300
Avg rating:3.0/5.0
Slides: 20
Provided by: tjwol
Category:

less

Transcript and Presenter's Notes

Title: Meter Testing - Traditional


1
Meter Testing - Traditional New Techniques
GE Meter School
  • Presented by
  • T Jeffrey Wolters
  • GE Meter, Somersworth, NH
  • The following people designed, developed or
    provided the substance for the material
    presented
  • Dave Elmore, Don Bullock (ret), Les Rosenau, Mike
    Coit, Warren Germer, Joe Provost

2
Some K Definitions
  • Kh - Watt-hours per (equivalent) disk revolution
  • Ke - Energy value per pulse (used commonly for
    PI)
  • K - Register programming watt-hour constant (for
    EV, Kh/12 K)
  • Kt - Test constant Watt-hours (or VArH) per
    pulse
  • Kr - Dial Multiplier
  • Inside the kV, K values have no real meaning or
    bearing on energy calculations. All energy values
    are supplied from the DSP in engineering units.

3
GE kV Meter Test LED
Test LED
4
GE kV Meter Functional Diagram
  • DSP blinks the Optocom LED every Kt watt-hours
    (or VArh)
  • Through programming, one can set the calibration
    pulse rate (Kt) to just about any practical
    value. This can
  • allow one to adjust testing speed (within the
    confines of other, more restrictive parameters)
  • standardize pulse values regardless of meter
    type.
  • Remember the kV DSP outputs (and the register
    stores) actual energy values rather than pulses
    which can be converted to energy values.

5
Kt Value Selections
  • The kV meter offers a great deal of flexibility
    in specifying Kt (test pulse) and PI (Pulse
    Initiator) values.
  • To obtain the most accurate results, it is
    important to understand more about how the kV
    calculates its values.
  • There is a number stored in the kVs firmware
    which is responsible for making the math work
    based on meter class.
  • Rev3 below Rev 04 above
  • Meter Class kV divisor kV divisor
  • 20 0.0006 0.0005
  • 200 0.006 0.005
  • 320 0.009 0.0075

6
Kt Value Selections
  • Rev3 below Rev 04 above
  • Meter Class kV divisor kV divisor
  • 20 0.0006 0.0005
  • 200 0.006 0.005
  • 320 0.009 0.0075
  • The use of Kt (and PI) values that are evenly
    divisible by these numbers will prevent small
    errors from being introduced due to truncation of
    repeating or long decimal values.
  • Most traditional Kt (and PI) value selections
    provide error-free operation, or contain
    insignificant errors.

7
Kt Value Selections
  • To calculate the magnitude of any possible error,
    use the formula above.
  • Example Use a Kt value of 1.0 on a CL 20 meter,
    Firmware 3
  • 1-((Integer(1/0.0006))/(1/0.0006))x100
  • 1-(1666/1666.666)x100
  • 0.04 error

8
Kt Value Selections
  • Now Use a Kt value of 1.0 on a CL 20 meter,
    Firmware 4
  • 1-((Integer(1/0.0005))/(1/0.0005))x100
  • 1-(2000/2000)x100
  • 0.00 error

9
Some Factors Affecting Measurement Uncertainty
  • Total meter performance is a function of many
    variables.
  • The methods we use to verify the meters
    performance are also subject to many variables.
  • In general, any variable which affects the meter
    and the standard differently will lead to
    UNCERTAINTY.

10
Some Factors Affecting Measurement Uncertainty
  • In digital meters, the more obvious factors
    affecting uncertainty include
  • Repeatability limits of the test equipment being
    used
  • Random systematic errors including response
    time and characteristics of the meter calibration
    LED circuit and the test equipment optotransistor
    circuit used in the pickup device
  • Digitizing error (remember we reduced this by
    careful selection of sampling rate)
  • Inherent measurement differences between filtered
    and unfiltered methods of measuring AC power
  • Temperature, frequency, stray magnetic fields,
    current and voltage non-linearities...

11
Measurement Uncertainty Example
  • Inherent measurement differences between filtered
    and unfiltered methods of measuring AC power
  • Take two meters having equal accuracy
  • The first, a heavily filtered meter (e.g.,
    RadianTM standard). It integrates average power
    with respect to time
  • The second, an unfiltered meter (e.g., GE kV
    meter). It integrates instantaneous power with
    respect to time.

12
Measurement Uncertainty Example
Inherent measurement differences between filtered
and unfiltered methods of measuring AC power
  • Compare the meters power calculation during the
    time interval from a to b in the figure.
  • The unfiltered meter will record energy
    proportional to the horizontally shaded area plus
    the cross-shaded area.
  • The filtered meter will record energy
    proportional to the cross-shaded area.

13
Measurement Uncertainty Example
Inherent measurement differences between filtered
and unfiltered methods of measuring AC power
  • In this example, the Instantaneous power
    measurement is obviously larger than the Average
    power measurement. For other small time
    differences (a and b), Average power could be
    larger or the same.
  • As test time increases, the effect of this source
    of uncertainty disappears. Also, if you could
    test over an exact integral number of cycles,
    this uncertainty would not exist.

14
GE kV Flash Calibration
  • Flash Calibration Objective
  • Calibrate voltage gain, voltage phase, current
    gain and current phase
  • Accomplish multiple measurements simultaneously
  • Minimize calibration time
  • Minimize number of conditions necessary to assure
    overall meter accuracy

15
GE kV Flash Calibration
  • How Flash Calibration Works
  • Meter is put into Flash Cal Mode via Optocom
  • Meter blinks Optocom LED and starts
    accumulation
  • After specified time, meter stops accumulation
    and blinks LED
  • Accumulated quantities are read and compared to
    standard meters
  • New correction factors are calculated and
    programmed into the meter as required
  • Rerun until all results are acceptable

16
GE kV Flash Calibration
  • How Flash Calibration Works (cont)
  • Factory has independent, per-phase equipment
  • Meter Flash Calibrated at FL and LAG
  • Meter verified using both Flash Calibrate mode
    AND pulse mode
  • Total Flash Cal time is just a few seconds
  • How can we do that? ... Remember, if you choose
    an amount of time that is an exact integral
    number of power cycles, there is no uncertainty
    error between instantaneous and average power
    measurements.

17
9S vs 8S Testing Conditions
  • Why doesnt the 9S kV meter seem to test
    correctly on my calibration panel set up for an
    8S meter?
  • The socket terminal connections are the same.
  • The actual (in)SERVICE phasors presented will be
    read by either meter. In fact, since the 9S meter
    is a Blondel solution, it will properly meter
    unbalanced conditions, where the 8S will not.
  • The problem is that most test panels do not (by
    default) apply the voltages the way they would be
    presented in service. They treat the 8S as a two
    voltage element device, applying one voltage A-B
    and the other C-N. For the 8S, this is good
    enough to test, but for the 9S kV, there is no
    real A-N voltage reference since the A-B voltage
    is subjected to a voltage divider between kVs
    A-N and B-N voltage sensing resistors.
  • In service, this is not a problem because the
    line to neutral volts are all independently
    regulated by the distribution transformers.

18
9S vs 8S Testing Conditions
  • Some test panels can be configured to present a
    more realistic 4 wire delta input to the meter
    under test.
  • This can also be used on 8S meters.

19
kV Test Voltages
  • Typically, apply Test Voltage (TV nameplate).
    120v.
  • Other voltage(s) may be used. Wide range meter is
    120-480v Line to Line (not Line to Neutral).
  • WHEN TESTING, DO NOT APPLY 480 L-N.
  • IN POLYPHASE, THIS WILL EXCEED THE 575V LIMIT ON
    THE POWER SUPPLY
  • WITH REVENUE GUARD, THIS WILL EXCEED THE 575V
    LIMIT ON THE POWER SUPPLY AND THE REVENUE GUARD
    OPTION BOARD.
  • Be aware of test board power supply limitations
  • some boards were not designed with newer meter
    switch-mode power supplies in mind.
  • Test board error magnitudes typically increase
    with test voltage increase.
Write a Comment
User Comments (0)
About PowerShow.com