Medical Imaging - PowerPoint PPT Presentation

1 / 47
About This Presentation
Title:

Medical Imaging

Description:

can't be discerned on X-ray image. Injecting contrast agents into blood vessels. iodine ... X-ray shadow image is obtained with the help of an image ... – PowerPoint PPT presentation

Number of Views:40
Avg rating:3.0/5.0
Slides: 48
Provided by: dml6
Category:
Tags: imaging | medical

less

Transcript and Presenter's Notes

Title: Medical Imaging


1
Medical Imaging
  • Dr. Der-Ming Liou
  • Health Informatics Group
  • Institute of Public Health
  • National Yang-Ming University

2
Medical Image
  • Generated by means of radiation
  • electromagnetic (EM)
  • ultrasound
  • electrons
  • Displayed for interpretation
  • the original image carrier
  • a processed film on which the image was formed
  • other carrier
  • photograph or computer display monitor

3
Computer Applied to Medical Image
  • Construct an image from measurements
  • Obtain an image reconstructed for optimal
    extraction of a particular feature from an image
  • Present images
  • Improve image quality by image processing
  • Store and retrieve images

4
Using Medical Image
  • Carriers
  • Modulation (by human body)
  • Detection
  • Demodulation
  • Display
  • Further Manipulations
  • Applications

5
Overview of the Use of Electromagnetic Radiation
6
Ultrasound
  • Produced by piezoelectric crystals
  • transform electrical energy to acoustic energy
  • vibrate with 2 10 MHz (audible sound 20 KHz)
  • pulsed sound waves are emitted
  • the energies and the arrival times of the
    received echo caused by the reflections are
    measured

short
l
2l
accuracy
Source
reflector
reflector
at least
2l
echo
distance
7
Echo Scanner (1)
  • Sound pulse
  • generated with a few MHz
  • absorbed, scattered, or reflected in the patient

different density
small
1020l
different velocity of sound
crystal
(diffuse)
different acoustic impedance
Reflected interface
water? soft tissue
bone
Soft tissue
air
small
scatter
object
8
Echo Scanner (2)
  • The resolution of echo scanner
  • the degree with which details located close
    together can still be distinguished
  • determined by the l of the sound wave and the
    duration of the emitted pulse
  • the smaller l the better the resolution
  • The attenuation coefficient a
  • sound frequency for soft tissue
  • square of the frequency for other types of tissue
  • the more the beam is attenuated, the more
    difficult it is to measure the reflections of
    deeper structures

9
Echo Scanner (3)
  • Resolution 8 penetration path
  • deeper structure can only be visualized with
    relative low frequency and with low resolution
  • Type of tissue influences the amount of
    absorption of the beam
  • air and bone are strong absorbers
  • muscle tissue and water hardly attenuate the beam
  • f 3 MHz (l 0.5 mm) depth up to 10 cm
  • f 513 MHz (l 0.250.075 mm) for eye
    examination

10
A-Mode
  • Amplitude mode
  • the energy of each echo is displayed as a
    function of the time interval between the pulse
    and the echo
  • vertical axis amplitude of the echo
  • horizontal axis time
  • the time interval between the pulse and the echo
    corresponds to the distance between the
    transducer and the reflecting tissue boundary
  • Provide one-dimensional information about the
    location of the reflecting boundaries

11
M-Mode
  • Motion mode (brightness mode)
  • similar to A-mode
  • one-dimensional information
  • the amplitude of each echo is represented as the
    brightness of the point located along the time
    axis
  • other axis is now used to display the echoes of
    subsequent pulses
  • possible to visualize the movement of objects
    that are crossed by the ultrasonic beam

12
C-Scan
  • Boundary not perpendicular to the beam axis
  • produce weak echoes
  • A-, B-, and M-mode have fixed beam directions
  • do not display all boundaries equally well
  • C-scan (compound-scan)
  • the crystal can be moved
  • the direction of the beam is changed
  • the crystal is connected to a flexible arm
  • linked to a fixed point of reference
  • disadvantage taking a few seconds to build up
    the image impossible to follow moving structures

13
Sector Scan
  • Parallel scan
  • the movement of the transducer can be sped up by
    mechanical means
  • each successive beam makes a small angle with the
    previous one
  • two dimensional images are obtained
  • to use a stationary probe containing a linear
    array of about 100 crystals
  • electrically activated one after one
  • distance between certain boundaries can be
    obtained

14
Sector Scan by Ultrasound of part of the Aorta
15
Dopper Effect
  • The measurement of flow velocities
  • f ? when the target approach to the sound source
  • the shift in frequency is a the frequency of the
    incident beam and to the velocity of the target
  • everything that moves inside the beam contributes
    to the Doppler signal
  • by repositioning the sample volume of the
    transducer is a systematic way, a two dimensional
    velocity distribution can be determined
  • the flow map can be superimposed on the echo image

Target
sound
wave
velocity
transducer
16
Advantages of Ultrasound
  • No harmful side effect
  • for examining pregnant women and young children
  • Applications
  • Determination of heart function
  • examination of the brain
  • obstetric examinations
  • eye examinations
  • determination of the perfusion of tissues
  • detection of tumors and cysts

17
Radiology
  • X-rays
  • discovery in 1895
  • simple applications
  • computed radiology
  • digital subtraction angiography (DSA)
  • complex applications
  • computed tomography (CT)
  • Magnetic resonance image (MRI)

18
Principle of an X-ray
19
Principle of DSA
  • Blood vessels often absorb as much as radiation
    as the surrounding tissues
  • cant be discerned on X-ray image
  • Injecting contrast agents into blood vessels
  • iodine
  • Due to the presence of bone
  • small contrast medium in the blood vessels are
    difficult to distinguish
  • the eye is not able to detect contrast
    differences of less than 3

20
Procedure of DSA
  • An initial X-ray shadow image is obtained with
    the help of an image intensifier
  • signal coming from the television camera is
    digitized and stored in the computer
  • The contrast medium is injected into the veins
  • The precontrast image (mask) is subtracted from
    the subsequent images
  • the resulting images only contain the information
    about the location of the contrast medium
  • since the contrast medium fills the vessels, the
    resulting images only show the vessels

21
Procedure of DSA (cont.)
  • bone structures that disturb the image have been
    removed by the subtraction
  • The differences in intensity can be amplified in
    such a way that the eye is able to perceive the
    blood vessels in the image
  • The quality of the image deteriorates when
    patient move
  • corrected by using suitable software
  • the contrast medium is injected in the form of
    bolus
  • the proximal part of vessels is visible on the
    first image
  • the distal part is visible on the later
    subtraction images

22
Principle of Contrast Enhancement
23
Example DSA of the Bifurcation of the Aorta
24
Two Dimension Image of X-ray
  • X-ray image are projection (shadow) images
  • can not reveal the real geometric distribution of
    organs
  • organ are situated behind each other
  • superimposed in the image so that a three
    dimensional volume is projected in two dimension
  • To obtain a three-dimension impression
  • obtain a number of images from different angles

25
Computed Tomography (1)
  • A cross section with a thickness about 1mm
  • divide the cross section into a large number of
    small squares
  • each with an area of about 1 mm2
  • when a narrow X-ray beam (pencil beam) pass
    through the slice (the beam attenuate)
  • the attenuation is determined by the molecular
    composition and the density of the tissue present
    in the square
  • an X-ray tube and a detector can be both shifted
    along a line and rotate

26
Computed Tomography (2)
  • To obtain an anatomical image
  • by displaying the attenuation coefficients
  • to determine the attenuation coefficients of each
    square
  • determination of the attenuation of each square
    in the cross-section is the purpose of the
    procedure
  • Translate the beam over a distance width (row)
  • take into account the attenuation coefficient of
    the square of the squares located on a neighbor
    row
  • repeated by translating the beam and measuring
    the transmitted intensity
  • until cover the total cross section

27
Computed Tomography (3)
  • Its not possible to get determine the individual
    attenuation coefficient of each square
  • an intensity profile of the transmitted beam as
    a function of the position of the beam
  • a measure of total attenuation
  • each point in the profile indicates how strongly
    the incident beam was attenuated by the row of
    squares that was passed by the beam in that
    position
  • Repeat the procedure outline above for various
    angles of the beam
  • possible to compute the attenuation per square

28
Principle of the CT
29
Intensity Profile
30
Example of Cross-sections
31
Back Projection (1)
  • Used in practice to obtain the attenuation
    coefficients mi
  • Can be used when intensity profiles that cover
    the total cross section under various angles are
    available
  • In an individual profile
  • each point represents the amount of attenuation
    by the pixels transmitted by the beam
  • Each profile show a dip at the location where the
    beam passed

32
Back Projection (2)
33
Back Projection (3)
  • If we have only one intensity profile
  • we cant determine where on the path the pixel
    was located
  • we cant even decide whether the absorption was
    due to a single pixel or was due to the
    attenuating medium that was present over the
    whole path
  • the only inference is that the attenuating medium
    was present only along one line in the cross
    section
  • the intensity profile showed a dip at only one
    point
  • The back-projection method assume
  • the absorption medium is uniformly distributed
    over the line

34
Back Projection (4)
  • The error from the assumption can be corrected
  • if we have several intensity profiles obtained at
    different angles
  • The reconstruction has a star-like distribution
  • The intensity in the center will increase much
    faster than the intensity at the periphery
  • With the use of more angles
  • the back-projection image become similar to the
    actual one only it is less sharp
  • instead of an image showing one attenuating
    pixel, the neighboring pixels are visible in the
    reconstructed image as well

35
Back Projection (5)
  • This blurring effect can be corrected to a
    certain extent by using appropriate filtering
    techniques
  • A real cross section can be considered a union of
    cross section with each one containing only one
    attenuation pixel
  • The back projection technique can be used
  • MRI (magnetic resonance imaging)
  • SPECT (single photon emission computed tomography)

36
Attenuation Coefficients
37
CT Equipment
  • Second generation scanners
  • the single detector is replaced by a frame of
    multiple detectors
  • the X-ray tube produces a fan-shaped beam
  • translation and rotation remain necessary
  • Third generation
  • the detector array is so large
  • transmission through the complete cross section
    of the patient can be measured simultaneously
  • translation is no longer necessary, rotation is
    necessary
  • Fourth generation
  • a stationary detector array covers 360o
  • only X-ray tube rotate

38
Principle of MRI (1)
  • Certain atomic nuclei behave like a spinning top
  • behave like small magnets
  • Under normal circumstances
  • the body is not magnetic
  • the hydrogen nuclei within the body point into
    all directions randomly
  • the net magnetic field strength (magnetization)
    0
  • When we place an ensemble of nuclei with spin in
    a strong magnetic field
  • the nuclei tend to align themselves with the
    magnetic field

39
Principle of MRI (2)
  • This alignment occurs
  • the nuclei prefer to be in a state with the
    lowest energy
  • 00 K ?all nuclei align themselves to the external
    magnetic field
  • At room temperature
  • the nuclei also possess thermal energy
  • external magnetic field
  • 0.1 tesla excess 1/106
  • 1 ml H20 3 x 1022 molecules 1017 hydrogen
    atoms aligning parallel to the magnetic field

40
Spin Alignment
41
EM Radiation
  • While the nuclei are under influence of the
    external magnetic field
  • pulse of electromagnetic radiation are beamed
    into the tissue
  • EM radiation is characterized by
  • an electric and a magnetic component
  • the magnetic component of the EM radiation exerts
    a force on the magnetic nuclei
  • When the magnetic component of the EM radiation
    has a direction perpendicular to the external
    magnetic field
  • cause the magnetization to precess around the
    direction of external field

42
Larmor Frequency
  • in such a way
  • the angle between the direction of the
    magnetization and the external field will
    increase linearly with time
  • only happen when the EM radiation has a certain
    frequency
  • the frequency is proportional to the strength of
    the external magnetic field
  • gyromagnetic ratio
  • characteristic for the element (isotope)
  • the range of radio frequencies 2 to 50 MHz

43
Precession of Magnetization
44
Principle of Gamma Camera
45
A Scintigram of the Lungs
46
Principle of ECG-gated Scintigraphy
47
Rotating Gamma Camera
Write a Comment
User Comments (0)
About PowerShow.com