Title: CRYOGENIC LTD


1
  • SQUID Magnetometer
  • Liquid helium cooled
  • Re-condensing
  • Cryogen Free

2
Possible approaches to SQUID Magnetometer
Concept 1 Liquid Helium cooled system with Low
loss cryostat Concept 2 Liquid helium with
re-condenser. Concept 3 Dry systems /
Cryogen Free Systems The magnet is cooled
directly and operates in vacuum. The sample is
cooled by a secondary helium circuit using 50
litres NTP of He gas.
3
Background
Cryogenic Ltd is a recognised leader for the
development and manufacture of advanced High
Field Measurement Systems for the study of the
electrical, magnetic or thermal properties of
materials. With proven experience in high field
superconducting magnets up to 21 Tesla and low
temperature equipment operating with and without
the use of liquid helium, Cryogenic can offer the
optimum solution for every type of research at
high field and low temperature.
4
Approaches to cool the SQUID Magnetometer
  • Concept 1 Conventional liquid helium or
    Nitrogen shielded system.
  • 50 litres reservoir for liquid helium or
    Nitrogen.
  • 2 to 4 litres per day of helium consumption.
  • 5 litres per day liquid Nitrogen.
  • Concept 2 Zero boil-off system with
    re-condenser.
  • 50 litres reservoir with Pulse Tube cooler for
    Zero boil-off system.
  • No liquid Nitrogen.
  • 6.5 kW compressor water cooled
  • Concept 3 Dry systems / Cryogen Free Systems
  • No liquid helium or nitrogen
  • The magnet and the sample space are cooled
    directly by a 1 Watt Pulse Tube cryocooler and
    operates in vacuum.
  • 6.5 kW compressor water cooled

5
The S700X System Structure
  • A Superconducting Magnet to 7 Tesla
  • Continuous operation from 320K down to 1.5K
  • A Variable Temperature Sample Space of 9mm with
    He-4 Gas
  • 10-8 EMU sensitivity for total moment
  • He-3 for temperatures down to 300mK
  • Oven option to 700K
  • Electronics Rack
  • LabVIEW operating software
  • Options
  • AC and DC measurements
  • Oscillator and extraction mode
  • MilliTesla field resolution
  • Full environmental Shielding built-in.
  • Flexible open LabVIEW software
  • Fast 20 bit data acquisition
  • Real data access during measurement
  • Transverse field

6
Zero boil-off Cryostat
  • Cooling via ultra quiet 1 Watt Pulse Tube
    Cryocooler
  • 45 litres of liquid helium reservoir.
  • The Cryocooler reliquifies the helium gas from a
    cylinder at the rate of 2 litres per day.
  • Cryostat required to be filled initially once
    only with 100 litres of liquid for pre-cool.
    Thereafter the liquid helium reservoir is filled
    and maintained by the Cryocooler.
  • Full magnetic shield built into the cryostat to
    protect the system from external influence.
  • Short- term fluctuation in heat load such as from
    ramping the magnet will not result in liquid
    helium loss as the system is supplied with a
    buffer tank full of reserve of helium gas.

7
Pulse Tube Cryocooler Stability Measurements
8
Cryogen Free Cryostat
  • Integrated Cryogen Free VTI with a sample space
    of 9mm
  • Single Pulse Tube cryocooler to provide cooling
    for both the magnet and the VTI
  • Samples top-loaded into the sample space via the
    top of the VTI
  • The PT cryocoolers provides up to 1.0 Watt of
    cooling power at 4K and more than 50W on the 60K
    stage.
  • System Cool-down time from start to room
    temperature takes typically 15 to 20 hours.

9
Cooling of the Cryogen Free Magnet System.
  • Cooling for the system is provided by a Pulse
    Tube (PT) cryocooler.
  • The PT cryocooler is quieter than the Gifford
    McMahon, has less vibration and has longer
    service intervals with lower servicing costs.
  • The PT cryocoolers can provide up to 1.0 Watt of
    cooling power at 4K and more than 50W on the 60K
    stage.
  • System Cool-down time from start to room
    temperature takes typically 15 to 20 hours.

10
Benefits of the Cryogen-Free Platform
  • Low Operating Costs - No liquid helium or
    Nitrogen required for start-up or operation. No
    costs associated with storage, transport or
    liquids and safety issues.
  • Simple to use - as there are no cryogen transfers
    to make. Just switch on the cooler and wait for
    it to reach operating temperature within 24
    hours. Can be run to cooldown overnight.
  • Independent and convenient operation Only
    mains electrical power is required to operate the
    system.
  • Compact absence of liquid helium and nitrogen
    reservoirs means that this system is more compact
    and allows easy access to field centre.

11
Competitive Benefits of the Cryogenic S700X
measurement system
  • Ultra light design of the sample chamber
    allowing fast temperature changes and quick
    stability.
  • Any temperature within the wide range of 1.6K to
    320K can be kept continuously, as long as there
    is liquid helium in the main reservoir (i.e for
    months at a time).
  • Temperatures down to 300mK with a He-3 insert
  • AC susceptibility can be cross calibrated with
    DC magnetic moment.
  • Sample exchange via airlock in 10 to 15 minutes
    including cooling to 2K.
  • Change of measurement options / probes in 10 to
    15 minutes.
  • Removal of Remanent Field by magnet quench or
    heating in less than 30 minutes.

12
Competitive Benefits of the Cryogenic S700X
measurement system
  • Faster measurement than any other commercial
    susceptometer (makes more measurements per scan
    and reads in both directions)
  • Full access to the operational functions
  • Upgradeable to allow further measurement
    additions at a later stage
  • Automated measurement sequences allowing
    unattended operation.
  • High quality electronics from well-established
    producers, such as the Lakeshore Temperature
    controller, allowing high reliability and
    replacement or recalibration of measurement
    electronics, as well independent warranty and
    support
  • Full environmental shielding (both magnetic and
    electromagnetic)

13
Cryogen Free Variable Temperature Sample Space
  • Simultaneous controlled cooling of the sample
    and magnet using a single cryocooler.
  • Integrated VTI built into the bore of the
    magnet.
  • Temperature range 1.6K to 300K. Optionally, a
    sample heater is used to extend the temperature
    range up to 700K.

14
Automation
  • All Cryogenic Measurement systems are supplied
    with complete control, monitoring and data
    acquisition software using the LabVIEW operating
    environment.
  • This allows 
  • Control of magnetic field
  • Control of individual instruments (magnet power
    supply, temperature controller, lock-in
    amplifier, etc.)
  • Combine functions into measurement sequences.
  • Fail-safe shutdown even with power failure.
  • Customisation of measurement sequences as the
    software is fully open-source.

15
(No Transcript)
16
He3 Insert For Measurements Below 0.3K
  • The He3 unit can be directly exchanged for the
    standard VTI to allow extension of the
    temperature range from 1.6 to 325K (VTI) to 0.26K
    100K.
  • The Cryogen Free He3 units are constructed as
    stand-alone items with their own cryocooler or
    fitted permanently into a cryogen free system.

17
He3 Insert For Measurements Below 0.3K
18
High Temperature Option
  • To perform measurements above room temperature,
    furnace or heated sample probe insert is offered
    to increase the temperature range to 700K.
  • A series of measurements can be often be
    performed over the whole range from 1.6 to 700K.

19
Quench Test
Magnet quenched at full field as the compressor
was stopped imitating a power failure. After the
compressor was restarted, the magnet was fully
operational in 27 minutes.
20
Magnetisation of typical Paramagnetic Salt of at
base temperature
21
6 Al impurity in a Ti alloy. Thermally conducted
with Si grease.
22
(No Transcript)
23
Measurements for the SQUID Magnetometer
  • DC Magnetisation of G - Fe2-O3 particles in
    silica
  • The measurement consists of two parts 1.
    Measurement from 2K to 100K at 1 Tesla after zero
    field cooling
  • (lower curve)2. Measurement from 100K to 2 K at
    1Tesla (upper curve)

24
Measurements for the SQUID Magnetometer
  • Relaxation Measurement
  • The sample was cooled down to 2K in zero field,
    the first measurement was made after a field
    change of 1T. The intervals in time between
    measurements increase in 'pseudo logarithmic'
    steps. This example demonstrates continuous
    operation at temperatures below 4.2K.

25
Measurements for the SQUID Magnetometer
  • Magnetic Viscosity as a Function of Temperature
    of G - Fe2-O3 particles in silica

26
Measurements for the SQUID Magnetometer
  • Magnetic Moment of the of G - Fe2-O3 particles in
    silica Sample