Resuscitation Equipment
2
Manual Resuscitators
  • Are devices incorp. a selfinflating bag and a
    non-rebreathing valve mechanism
  • Provides manually applied positive pressure to
    the airways
  • Used in emergency situations, for suctioning, and
    for transport
  • Patient valve has a standard 22 - 15 mm connector
  • Sometimes called AMBU bag

3
Recommendations for a bag-mask unit
  • Self-refilling bag that is easily cleaned and
    sterilized (if reusable)
  • Non-sticking valve system at 15 L/M
  • Non-rebreathing valve system
  • Can have pop-off valve or not
  • Standard 15/22 mm adapter
  • Provide up to 100 O2 at high stroke volumes and
    rates
  • Available in adult, pediatric, infant

4
Hazards of manual resuscitators
  • Leaks around mask or tracheal tube may prevent
    proper lung inflation
  • Improperly functioning equipment due to missing
    parts, improper assembly
  • Poor ventilation technique by operator
  • Failure to recognize and properly handle tension
    pneumo., aspiration, or acute episodes of hypoxia

5
Gas - Powered Resuscitators
  • Usually pressure-limited units and may be used as
    demand valves
  • Can be time cycled or pressure cycled
  • Provide instantaneous flow rates of at least 100
    L/M
  • Have an insp. press. relief valve that opens at
    50 cm H2O
  • Provides 100 O2
  • Has standard 15/22 mm adapter