Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses. - PowerPoint PPT Presentation

About This Presentation
Title:

Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses.

Description:

This prospective comparative study included patients with ... ophthalmologic examination besides senile cataract, and unsatisfactory correction with glasses. ... – PowerPoint PPT presentation

Number of Views:173
Avg rating:3.0/5.0

less

Transcript and Presenter's Notes

Title: Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses.


1
Comparison of visual function following piggyback
implantation of Acrysof ReSTOR intraocular lenses
with Tecnis multifocal ZM900 intraocular lenses.
Rodrigo Vaz, MD Leonardo Akaishi, MD Patrick F.
Tzelikis, MD
None of the authors has a financial or
proprietary interest in any material or method
mentioned.
2
Purpose
  • To compare the postoperative outcome of piggyback
    implantation using the AcrySof ReStor lens with
    Tecnis ZM900 multifocal intraocular lenses (IOL).

3
Material / Methods
  • This prospective comparative study included
    patients with age-related cataract, primary
    hyperopia or hyperopic astigmatism, a normal
    ophthalmologic examination besides senile
    cataract, and unsatisfactory correction with
    glasses.
  • Information collected included uncorrected visual
    acuity (UCVA) for near and distance, best
    corrected visual acuity (BCVA), and spherical
    equivalent (SE) before and after the surgery.
  • All patients were followed for 6 months

4
Material / Methods
  • The three lenses used in this study were TECNIS
    ZM900 multifocal IOL, AcrySof ReStor and
    CLARIFLEX.
  • The multifocal IOL (Tecnis ZM900 or AcrySof
    ReStor) was first implanted in the capsular bag
    and the Clariflex lens was implanted second in
    the ciliary sulcus.

5
Material / Methods
  • The TECNIS multifocal IOL (Model ZM900, AMO,
    Santa Ana, CA) is a 3-piece diffractive
    multifocal lens with 6.00 mm polysiloxane optic.
  • The AcrySof ReSTOR (Model SA60D3, Alcon
    Surgical Laboratories, Dallas, Texas) is an
    acrylic foldable single-piece IOL.
  • The CLARIFLEX (Model AMO, Advanced Medical
    Optics, Santa Ana, California) is 3-piece
    monofocal lens with 6.00 mm biconvex square-edged
    silicone optic and 10-degree angulated C
    polymethyl-methacrylate (PMMA) haptics.

6
Material / Methods
  • The Holladay formula was used to calculate the
    IOL power when there was a short axial length (lt
    22.0 mm) and the SRK/T formula was used when
    there was an average axial length ( 22.0 mm).
    The A-constant used was 119.8 for the TECNIS
    multifocal IOL, 118.6 for the AcrySof ReStor and
    118.4 for the CLARIFLEX IOL. Axial length was
    measured with the IOL Master (Carl Zeiss Meditec
    AG), and the targeted postoperative refractive
    error was 0.0

7
Results
Teste t Student (SPSS)
Preoperative Tecnis ZM900 (n20 eyes) ReStor (n21 eyes) P value
Axial length 20.80 mm (SD0.80) 21.68 mm (SD0.39) Plt0,001
Power IOL (D) 26.75 D (SD 1.95) 22.0 D (SD 1.75) Plt0.001
UCVA 0.08 (20/250) (SD 0.11) 0.25 (20/80) (SD 0.06) Plt0.001
BCVA 0.67 (20/30) (SD 0.08) 0.9 (20/22) (SD 0.17) P0.02
SE 6.25 D (SD 2.25) 4.25 D (SD 1.55) P0.002
8
Results
Teste t Student (SPSS)
Postoperative (6 months) Tecnis ZM900 (n20 eyes) ReStor (n21 eyes) P value
UCVA for distance 0.67 (20/30) (SD 0.15) 0.86 (20/23) (SD 0.16) P0.002
UCVA for near 33 cm J1 J1 P0.14
BCVA for distance 0.90 (20/22) (SD 0.10) 1.00 (20/20) (SD 0.05) P0.002
BCVA at 60 cm J4.5 J5 P0.08
SE 0.00 D (SD 0.45) -0.25 D (SD 0.40) P0.06
9
Results
10
Results
  • After a follow-up of 6 months, halos were
    reported as severe by 8.3 , moderate by 33.3
    and absent or mild by 58.4 of patients in the
    Tecnis group. In the ReStor group halos were
    reported as severe by 9.0, moderate by 27.3
    and absent or mild by 63.7.
  • No interlenticular opacification (ILO) or another
    postoperative complication related to piggyback
    lenses was observed in our series.

11
Conclusion
  • Overall, the use of a multifocal IOL in a
    piggyback fashion appears to be a safe and
    efficient procedure and a good refractive
    solution.
  • Implanting 2 piggyback IOLs was beneficial in
    eyes with short axial length. The AcrySof ReSTOR
    IOL gave better outcomes than the Tecnis
    multifocal IOL in UCVA and BCVA for distance,
    probably because in the Tecnis group the eyes had
    shorter axial length and required higher diopter
    lenses.