Evaluation the relative emission probabilities for 56Co and 66Ga - PowerPoint PPT Presentation

1 / 28
About This Presentation
Title:

Evaluation the relative emission probabilities for 56Co and 66Ga

Description:

China Institute of Atomic Energy, P.O. Box 275(41), Beijing 102413, China ... G.Molnar calibrated the detector efficiency curve using 14N(n, )15N reaction, in ... – PowerPoint PPT presentation

Number of Views:23
Avg rating:3.0/5.0
Slides: 29
Provided by: nip4
Category:

less

Transcript and Presenter's Notes

Title: Evaluation the relative emission probabilities for 56Co and 66Ga


1
Evaluation the relative emission probabilities
for 56Co and 66Ga
  • Yu Weixiang Lu Hanlin Huang Xiaolong
  • China Nuclear Data Center
  • China Institute of Atomic Energy,
  • P.O. Box 275(41), Beijing 102413,
    China
  • e-mail huang_at_ciae.ac.cn

2
Measurement
  • A 136cm3 coaxial Ge(Li) detector connected to an
    ORTEC-919 data acquisition system operating on a
    PC is used in present work. Spectra were recorded
    by the front face of the detector 17.5cm far from
    the source

3
Full-energy peak efficiency
  • In the energy range from 100 to 2754keV, the
    primary standard radioactive sources with
    well-known activities were used to calibrate the
    absolute efficiencies of Ge(Li) detector. For
    example 24Na, 60Co, 54Mn, 65Zn, 137Cs and 133Ba
    were used to calibrate the efficiency curve in
    this region. At the same time, a mixed
    125Sb154Eu155Eu multi-energy source produced by
    National Institute of Standards Technology
    (NIST) with uncertainties in 0.6-1.3 was used.
  • The I? values used in the calibration were taken
    from the Table of Isotopes, Eighth Edition. The
    uncertainties (one standard deviation) of these
    sources are about 0.6-1.3. The uncertainties of
    full-energy peak efficiency curve are about 1-2
    in this region.

4
Full-energy peak efficiency
  • Above 2754keV, there are no radioactive sources
    suitable for efficiency calibration. Therefore,
    nuclear reactions are commonly used. In present
    work, the efficiency curve is obtained by the
    calculated results using the EGS4 M-C code above
    2754MeV, and calibrated at 6.13MeV using
    19F(p,??)16O reaction at resonance energy point
    Ep340keV.

5
Full-energy peak efficiency

6
Full-energy peak efficiency
7
Measurement
  • By using the efficiency curve measured above,the
    new relative intensities were determined for the
    emitted ?-rays of 56Co and 66Ga. The final
    results are presented in Table 2 and 3,
    respectively.
  • Its noted that our measurements are about 2
    lower than other new measurements in high energy
    range.

8
Measured relative ?-ray intensities for 56Co
9
Measured relative ?-ray intensities for 66Ga
10
Standards ?-ray emission probabilities for
14N(n,?)15N reaction
  • The main measurements of the ?-ray emission
    probabilities for 14N(n,?)15N reaction are
    T.J.Kennett et al., E.T.Jurney et al., H.Takayama
    and T.Belgya. The main difference among these
    measurements are from the level scheme. For
    example, the levels are 15, 19, 19 and 17,
    respectively and ?-rays are 28, 58, 64 and 55,
    respectively corresponding to these measurements.
  • Fig. 3 shows the comparison of the ?-ray emission
    probabilities. Its easy to find that the
    measurements of H.Takayama and E.T.Jurney et al.
    are in good agreement in 1.7-6MeV energy region.
    But the ratio of the ?-ray emission probabilities
    for Kennett/Jurney and Belgya/Jurney are exactly
    reverse(see Fig.4).

11
Standards ?-ray emission probabilities for
14N(n,?)15N reaction
12
Standards ?-ray emission probabilities for
14N(n,?)15N reaction
13
Standards ?-ray emission probabilities for
14N(n,?)15N reaction
  • From Fig.4 its easy to find that the
    measurements of Kennett are higher about 1.2 to
    3.7 than Jurneys in 2.54MeV and 68MeV range.
    To average of these two measurements simplely is
    not suitable for the standard for detector
    efficiency calibration.
  • About 58 ?-rays were observed in the measurements
    of Jurney, but only 30 ?-rays of Kennett. In
    order to keep the intensity balance(?I?(in-out)),
    the measured ?-ray emission probabilities of
    Kennett should be higher than Jurneys.
  • We think that the ?-ray emission probabilities of
    Jurney are better than Kennett because the level
    scheme of Jurney is reasonable and complete.

14
Correction
  • G.Molnar calibrated the detector efficiency curve
    using 14N(n,?)15N reaction, in which their ?-ray
    emission probabilities were from the measurements
    of T.J.Kennett et al.
  • S.Raman calibrated the detector efficiency curve
    using 14N(n,?)15N reaction, in which their ?-ray
    emission probabilities were from the average of
    measurements of T.J.Kennett et al. and E.T.Jurney
    et al.
  • So its necessary to correct the measurements of
    S.Raman and G.Molnar using the ?-ray emission
    probabilities of Jurney above 2.5MeV.
  • In present work the level scheme suggested by
    Jurney is adopted to correct the ?-rays
    intensities of 56Co and 66Ga.

15
Previous evaluation I? for 56Co
  • The decay data of 56Co were evaluated based on 33
    measured data sets from 1965 to 2002 years by
    C.M.Baglin et al. in 2004.
  • Firstly several measured values are statistical
    outliers(about 11) according to the Chauvenet
    criterion. Secondly the measured data relied on
    linear extrapolations of the efficiency on a
    log-log plot above 3MeV were excluded. The rest
    measured data were processed by several
    evaluation methods weight average(WM),
    limitation of relative statistical weight
    average(LWM), normalized residuals method(NR) and
    Rajeval method(RA). At the same time, the whole
    measured data, the whole measured data except the
    exceed Chauvenet criterion value, were also
    processed by these evaluation methods.
  • The recommended values are obtained from one of
    these processed values according to authors
    judgments. Its easy to find that the final
    recommended data are evaluated based on the whole
    measured data below 2598keV and 8 data sets above
    2598keV.

16
Present evaluation I? for 56Co
  • Fig.5 shows the comparison of the ratio of newly
    measurements of S.Raman et al.,G.Molnar et al.
    and present work above 2.5MeV, present evaluation
    below 2.5MeV to the evaluations of C.M.Baglin for
    the relative ?-ray emission probabilities of
    56Co. Below 2.5MeV the present evaluation is in
    good agreement with Baglins within 0.4. But
    above 2.5MeV, the evaluation of Baglin is larger
    than the newly three measurements.
  • The similar trend can be found for 66Ga (see the
    next section in detail). We think that the
    different detector efficiency curve may cause
    this discrepancy.

17
Present evaluation I? for 56Co
18
Present evaluation I? for 56Co
  • The modified values of S.Raman and G.Molnar using
    the ?-ray emission probabilities of Jurney for
    56Co are listed in table 2. Its noted from table
    2 that the systematic deviation among the
    measurements of S.Raman, G.Molnar and present
    work is not existed. Present measurements are in
    good agreement with the modified values of
    S.Raman and G.Molnar within 1. Fig. 6 shows the
    comparison of present evaluations to the modified
    measurements of S.Raman and G.Molnar, and
    evaluations of C.M.Baglin.
  • The present evaluation is obtained from the
    average of 11 measured data sets including
    present measurements using LWM below 2.5MeV
    energy region. Above 2.5MeV energy region, the
    present evaluation is obtained from the
    unweighted average of the measurements of S.Raman
    et al.,G.Molnar et al. and present work(the
    measurements before 2000 are rejected when
    evaluated due to present knowledge of detector
    efficiency curve).

19
Present evaluation I? for 56Co
  • Table2

20
Present evaluation I? for 56Co
21
Temporary evaluation I? for 66Ga
  • The newly recommended relative ?-ray emission
    probabilities for 66Ga were evaluated by E.Browne
    et al. based on the measurements of S.Raman et
    al.(quoted data from Budapest), G.Molnar et
    al.(Budapest) and C.M.Baglin et al.(Berkeley) in
    2004. Weve finished the measurements of relative
    ?-ray emission probabilities for 66Ga. A
    temporary evaluation of the relative ?-ray
    emission probabilities for 66Ga was given based
    on the measurements of S.Raman et al., G.Molnar
    et al.(Budapest), C.M.Baglin et al.(Berkeley) and
    present work.
  • This two evaluations are in agreement with each
    other within 1, but the deviation between the
    evaluations and newly measurements is up to 2.5,
    which is shown in Fig. 7.

22
Temporary evaluation I? for 66Ga
23
Present evaluation I? for 66Ga
  • We corrected the measurements of S.Raman and
    G.Molnar using the ?-ray emission probabilities
    of 14N(n,?)15N reaction from Jurney above 2.5MeV.
    We also corrected the measurements of C.M.Baglin
    using present evaluated ?-ray emission
    probabilities of 56Co. The modified values are
    given in table 3 and shown in Fig.8.

24
Present evaluation I? for 66Ga
25
Present evaluation I? for 66Ga
  • Present final evaluations are recommended based
    on present measurements and modified measurements
    of S.Raman, G.Molnar and C.M.Baglin shown in
    Fig.9. The present evaluations are in good
    agreement with the modified measured values
    within 1.3.

26
Present evaluation I? for 66Ga
27
Present evaluation I? for 66Ga
28
Conclusion
  • The newly recommended relative ?-ray emission
    probabilities for 56Co and 66Ga evaluated by
    C.M.Baglin et al., E.Browne et al. and present
    work show that the old standard for detector
    efficiency calibration existes systematic errors
    (up to 30) in high energies range. But present
    evaluation is lower than the evaluation of Browne
    and Baglin above 2.5MeV. The deviation at 3.4MeV
    is up to 2.7 and this deviation is consistent
    with the difference shown in Fig.6. The
    rationality of present evaluation and corrected
    method will be dependent upon new measurements,
    and more precise standard data are desirable in
    further.
Write a Comment
User Comments (0)
About PowerShow.com