1
Basic Microscopy An Overview
  • October 2005
  • Protistology Course
  • MBL, Woods Hole, MA

2
  • Brief History of the Microscope
  • Whats a microscope?
  • Definition of Magnification
  • Conventional Viewing Distance
  • Leeuwenhoek gt Compound gt Stereo Microscope
  • The Telescope, a simple detour
  • How to make the specimen visible Contrast!
  • Definition of Contrast
  • Techniques
  • Brightfield
  • Phase
  • Darkfield
  • Pol
  • DIC (Differential Interference Contrast)
  • Fluorescence
  • Optical Sectioning an expansion of Fluorescence
  • Setting up the Microscope (Lab)
  • Koehler Illumination
  • Resolution Empty Magnification

Agenda
3
(No Transcript)
4
(No Transcript)
5
(No Transcript)
6
Objects appear to the eye at different
magnifications, depending on their distance from
the eye. Accommodation (lens) will make it
possible.
What is Magnification?
MB 2x MA
A
B
7
Conventional Viewing Distance
?
1x
250 mm
8
Magnification 1x
1x
f 250 mm
1x
9
Magnification via Single Lens
1x
f 250 mm
Magnifying Glass (Loupe)
5x
Example f50mm
10
The simple microscope
Leeuwenhoek Microscope
11
The current ?-corrected Compound
Microscope
Eyepiece
Tube lens (Zeiss f164.5mm)
Objective
12
Q What happens if we take the objective away?
Eyepiece
8
Tube lens (Zeiss f164.5mm)
8
Objective
250mm
f


M
Tube


f
250mm
Eyepiece
Answer We have created a Telescope
13
AxioImager Upright Research Microscope
14
Axiovert 200 Inverted Research
Microscope
15
The basic light microscope types
Upright microscope .
Inverted microscope
16
Illumination via Transmitted Light
The specimen must be transparent !
17
Upright microscope .
Inverted microscope
18
Illumination via Reflected (Incident) Light
Eg. Fluorescence, Opaque Samples
19
Upright microscope .
Inverted microscope
20
Mixed Illumination
Upright microscope .
Inverted microscope
21
Other Ways to Illuminate
Reflectors Ring Lights Fiber Optics LEDs Etc.
22
  • Couldnt one build a microscope for both
    eyes, and thereby generate spatial images?
  • Question addressed to Ernst Abbe in 1896
  • by Horatio S. Greenough

23
1897 the first Stereo Microscope in the world,
built by Zeiss, according to the Greenough
principle
1896 Drawing by Horatio S. Greenough
24
Greenough Type
Telescope Type
Introduced first by Zeiss - 1946
Introduced first by Zeiss - 1897
25
Greenough Stereo Microscopes
SteMi DV4
26
Greenough Stereo Microscopes
SteMi 2000 (2000-C, 2000-CS)
27
Research Stereo Microscopes
SteREO Discovery V12
SteREO Lumar V12
28
  • How to make the specimen visible
  • CONTRAST!
  • Definition of Contrast
  • Techniques
  • Brightfield
  • Phase
  • Darkfield
  • Pol
  • DIC (Differential Interference Contrast)
  • Fluorescence
  • Optical Sectioning an expansion of Fluorescence

Agenda
29
C ONTRAST
50 Units
0 Units
100 Units
50
50
50 Units
50 100 / 50 100 -0.33
50 0 / 50 0 1
50 50 / 50 50 0
30
Common Illumination Techniques
  • Brightfield
  • Darkfield
  • Phase Contrast
  • Polarized Light
  • DIC (Differential Interference Contrast)
  • Fluorescence (and related techniques)

31
(No Transcript)
32
Brightfield
  • For naturally absorbing or stained samples
  • True Color Representation
  • Proper Technique for Measurements
  • Spectral
  • Dimensional

33
(No Transcript)
34
(No Transcript)
35
Bacillaria
36
Paramecium bursaria
Condenser diaphragm open
Condenser Diaphragm almost closed
37
Paramecium bursaria
Different Staining Techniques
Indian Ink Staining
Feulgen Staining
Silver Staining
38
(No Transcript)
39
Phase Contrast (Frits Zernike 1934)
  • - Halo effect gt Reduced resolution
  • No staining necessary
  • Good Depth of Field
  • Easy alignment
  • Orientation independent
  • Repeatable setup
  • Works with plastic dishes

40
  • Required Components for Phase Contrast
  • Objective with built-in Phase Annulus
  • Condenser or Slider with Centerable Phase Ring
    for illumination (Ph0, 1, 2 or 3)

Required Adjustment Superimpose Phase Ring of
condenser over (dark) phase plate of objective
(after Koehler Illumination)
41
  • Illumination bypasses Specimen gt no phase shift
  • Illumination passes through thin part of Specimen
    gt small phase retardation
  • Illumination passes through thick part of
    Specimen gt larger phase retardation

Phase Shifts Cells have higher n than water.
Light moves slower in higher n, consequently
resulting in a phase retardation Phase shift
depends on n and on thickness of specimen detail
42
  • Non-diffracted and diffracted light are focused
    via tube lens ?into intermediate image and
    interfere with each other ¼¼ ½ wave
    shift causes destructive interference i.e.
    Specimen detail appears dark ?
  • Affected rays from specimen, expressed by the
    higher diffraction orders, do not pass through
    phase ring of objective gt¼ wave
    retarded ?

Tube Lens
  • Objective Phase Ring
    a) attenuates the non-diffracted 0th Order
    b) shifts it ¼ wave forward ?

Objective
Specimen
  • Illumination from Condenser Phase Ring ?(0
    Order) gt meets phase ring ? of objective

Condenser
43
Paramecium bursaria
Phase Contrast
44
Rhipidodendron
Phase Contrast
45
Cochliopodium
Phase Contrast
46
Lyngbya Bacteria
Phase Contrast
47
(No Transcript)
48
Darkfield
  • No staining necessary
  • Detection of sub-resolution details possible
  • Excellent, reversed contrast
  • Central Darkfield via hollow cone
  • Oblique Darkfield via Illumination from the side
  • Not useful for Measurements (sizes exaggerated)

49
Required conditions for DarkfieldIllumination
Aperture must be larger than objective aperture
I.e. direct light must bypass observer
Low NA Objective
High NA Objective
Iris Diaphragm
50
Paramecium bursaria
Polarized Light
Darkfield
51
Polarized Light
  • Specimen is placed between 2 crossed polarizers.
  • Only light produced by birefringent particles
    (e.g. crystals) or coming from the edges of
    particles (edge birefringence) is visible.
  • Looks sometimes like Darkfield
  • Orientation-specific (linear Pol)
  • Linear / circular Polarized Light

52
Birefringent Material
Color of sample and background modified by wave
plate
Background
Brightfield
Polarized Light
Pol Red I
53
Polarized Light
When Polarizers are crossed, only items that
rotate the plane of polarization reach the
detector. Wave plate adds color
Polarizer 2 (Analyzer)
Specimen
Polarizer 1
54
Required / Recommended Components
  • Polarizer (fixed or rotatable)
  • Analyzer (fixed or rotatable)
  • Strain-free Condenser and Objective
  • Rotating, centerable Stage
  • Wave plate and/or Compensator
  • Crossline Eyepiece

55
(No Transcript)
56
DIC (Differential Interference Contrast after
Nomarski)
  • High Contrast and high resolution
  • Control of condenser aperture for optimum
    contrast
  • Changes GRADIENTS into brightness differences
  • 3-D Image appearance
  • Color DIC by adding a wave plate
  • Best contrast / resolution via different DIC
    sliders
  • Orientation-specific gt orient fine details
    perpendicular to DIC prism

57
DIC
Observing local differences in phase retardation
58
  • 9 Image
  • 8 Tube lens
  • 7 Analyzer (7a with Wave Plate)
  • 6 Wollaston Prism behind objective
  • 5 Objective
  • 4 Specimen
  • 3 Condenser with receptacle for prisms

59
Wollaston Prism
Birefringence (Different refractive index for
different polarization orientations) Polarized
beam, under 45 to prism, gets split into
ordinary and extraordinary beam
60
Required Components for DIC
  • Nosepiece with DIC receptacles
  • Polarizer (or Sénarmont Polarizer)
  • Low Strain Condenser and Objective
  • DIC Prisms for Condenser ( I or II or III)
  • Appropriate DIC Slider for each objective
  • Analyzer (or Sénarmont Analyzer)
  • Not needed for New Plas-DIC (up to 40x)

61
Paramecium bursaria
DIC
Interference
62
Fluorescence
  • Easy to set up gt Objective Condenser
  • Highly specific technique, wide selection of
    markers
  • Detection and Identification of Proteins,
    Bacteria, Viruses
  • Basics for
  • Special Techniques eg. TIRF, FRET, FRAP etc.
  • 3-D imaging
  • Deconvolution
  • Structured Illumination
  • Confocal Techniques

63
Epi - Fluorescence
Observation port
Excitation Filter
Emission Filter
FL Light Source
Dichromatic Mirror



Example Specimen containing green fluorescing
Fluorochrome
64
Epi - Fluorescence Filter Sets
Example
Curve for a typical GFP filter set
65
Epi - Fluorescence
(Specimen containing green fluorescing
Fluorochrome)
Observation port
Excitation Filter
Emission Filter
FL Light Source
Dichromatic Mirror



Specimen containing green fluorescing Fluorochrome
66
Paramecium bursaria
Fluorescence
67
How to improve Fluorescence Imaging in a major
way
  • Optical Sectioning

68
Optical sectioning increased contrast and
sharpness
69
Overview of Optical
sectioning Methods
  • Confocal and Multi-photon
    Laser Scanning Microscopy
  • Pinhole prevents out-of-focus light getting to
    the sensor(s) (PMT - Photomultiplier) (30 70
    µm)
  • Multi Photon does not require pinhole (90 500
    µm)
  • Spinning disk systems
  • A large number of pinholes (used for excitation
    and emission) is used to prevent out-of-focus
    light getting to the camera
  • E.g. Perkin Elmer, Solamere ( up to 30 µm)
  • Structured Illumination
  • Moving grid represents the reference for in-focus
    information
  • Zeiss Apotome (10-50 µm)

70
Overview of Optical
sectioning Methods- contd -
  • Total Internal Reflection Fluorescence (TIRF)
  • High NA Objective projects beam at angle which
    exceeds critical angle.
  • Area touching cover slip (evanescent field) is
    typically smaller than 200 nm
  • Deconvolution
  • Point-Spread function (PSF) information is used
    to calculate light back to its origin
  • Post processing of an image stack

71
Limited Depth of Field With Standard Microcopy
  • Amber fossil (Chironomide)
  • Thickness app. 300 µm
  • Conventional incident light

72
Optical Sectioning Extended Focus Software
  • Amber fossil (Chironomide)
  • Thickness app. 300 µm
  • Conventional incident light
  • 3D reconstruction