A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout - PowerPoint PPT Presentation

About This Presentation
Title:

A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout

Description:

The surface between LSO slab and MCP glass was optically coupled with the optical grease. ... MCP/TL coupled to 1x1x10mm3 LSO crystal. ... – PowerPoint PPT presentation

Number of Views:188
Avg rating:3.0/5.0
Slides: 16
Provided by: psecUc
Category:

less

Transcript and Presenter's Notes

Title: A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout


1
A Design of PET detector using Microchannel Plate
PMT with Transmission Line Readout
  • Heejong Kim1, Chien-Min Kao1, Chin-Tu Chen1,
  • Jean-Francois Genat2, Fukun Tang2, Henry Frisch2,
  • Woong-Seng Choong3, William Moses3
  • 1. Department of Radiology, University of
    Chicago, IL
  • 2. Enrico Fermi Institute, University of Chicago,
    IL
  • 3. Lawrence Berkeley National Laboratory,
    Berkeley, CA

2
1. Introduction
  • The advantages of using Microchannel Plate(MCP)
    PMT.
  • Position sensitiveness.
  • Fast time response.
  • Compact size than conventional PMT.
  • LSO scintillator
  • High Light yield( 2500030000/MeV)
  • Fast decay time( 40ns)
  • Transmission Line readout scheme.
  • Readout both ends of the strip.
  • Position measurement by time difference
  • Efficient reduction of of readout channel( NxN
    -gt 2N)
  • Readout at both ends( Scintillator sandwiched by
    MCPs)
  • Possible to extract Depth of Interaction(DOI)
  • A PET detector design, using pixelated array of
    LSO scintillator with MCP PMT, has been
    investigated. Fast timing characteriscs of MCP
    combined with high sensitivity LSO makes this
    design suitable for TOF PET application. By
    design, DOI information is available by reading
    out the signals at both ends of scintillator. The
    preliminary results of Geant4 simulation study
    are presented here. The real tests to validate
    the simulation has been conducted with Photonis
    planacon MCP(XP85022) and the results are also
    shown.

3
MCP Transmission Line board
Fig.1 Photonis Planacon MCP(XP85022) with
1024(32x32) anodes(left) and Transmission
line(TL) baord with 32 microstrip (right). One
microstrip is connected to one raw of MCP
anode(32) and signals are readout at both ends of
a TL.
4
2.Material and Methods
A. Detector configuration
  • One detector module consists of 24x24 array of
    LSO scintillator and 2 MCP assemblies.
  • Two detector modules facing each other.
  • 5cm distance between them.
  • LSO pixel dimension 4x4x25mm3.
  • Crystal pitch 4.25mm
  • MCP assembly dimension 102x102x9.15mm3. It
    includes photocathode and TL structure. (MCP with
    8x8 area is under development.)?
  • MCP is coupled to LSO at both front and back ends.

5
Scintillators sandwiched by MCPs
Fig. 2 Simulation set-up with two detector
modules. Each module consist of 24x24 array of
pixelated LSO scintillators and two MCPs coupled
to the scintillators at both front and back side.
6
B. Simulation Setup
  • Optical Photon generation and transport was
    simulated by Geant4.
  • Two 511keV gammas are generated back to back at
    the middle of two detector modules and sent to
    the detector centers.
  • The reflective media was inserted between
    crystals.
  • The surface between LSO slab and MCP glass was
    optically coupled with the optical grease.
  • LSO characteristics( simulation input parameters)
  • Light yield 30,000/MeV
  • Decay time 40ns
  • Re?solution 10.4( FWHM)

7
Signal Readout Scheme
  • Electrical signal was formed based on the
    measured XP85022 characteristics combining with
    the Geant4 simulation outputs optical photons
    position and arrival time at photocathode.
  • For each individual photo electron, the measured
    single photo electron response was assigned.
    Convolute pulses due to all the photo-electron
    within the area of TL strip.
  • TL signal then propagates to both ends of TL.
  • In the forward MCP, 24 TL strips run vertically.
    By applying Anger logic to measured TL signals, X
    coordinate can be obtained.
  • TLs runs horizontally in the backward MCP to get
    Y coordinate in the same way.
  • The position also can be measured from the
    measured time difference at both ends of TL.

8
3. Experimental Tests
The test set-up was built using a XP85022 MCP and
TL board to measure the characteristics of the
MCP. The measured single photo-electron
response(SER) was fed to the simulation for the
electrical signal.
  • XP85022
  • Chevron type, 10um pore
  • Textronix DPO7354 Digital Oscilloscope recorded
    the waveform of TL at 10GS/s.
  • The charge of pulse are
    obtained by integrating TL waveform.

Fig. 3 MCP/TL assembled for the real test. 4 TL
channels were connected through SMA to the
DPO7354 Oscilloscope. A LSO crystal with
1x1x10mm3 was placed on top of the XP85022 MCP
surface.
9
A. Single Electron Response(SER)
  • SER was measured using the pulsed LED as a light
    source.
  • The rise time of SER was measured 560ps.
  • The SER signal was spread in 5 TL.
  • The XP85022 gain at HV -2300V 1.5 x 106

mV
Fig. 4 Integrated charge of SER waveforms(left).
Averaged waveform of SER the maximum TL signal
only (middle). XP85022 MCP gain as a function of
HV
10
B. Responses to 511keV photon
  • MCP/TL coupled to 1x1x10mm3 LSO crystal.
  • Hamamatsu R9800PMT with 6.2x6.2x25mm3 LSO for
    coincidence
  • Use Na22 for positron source.?
  • Waveform recorded by Tektronix DPO7354 scope

E resolution 13.8 fwhm
Fig.5 Test set-up for 511keV gamma coincidence
(left). Energy distribution of R9800PMT(right).
11
Energy( real test)?
  • Charge sum of 3 TL signal only left side of TL.
  • Compton 511keV peak structure is clearly found.
  • Discrepancy between the real test and simulation.
  • E resolution 22.3 vs 15.8 ( at 511keV peak)
  • Shape of compton continuum.
  • Due to simplified simulation setup( gamma
    direction).

Test set-up simulation
Real Test
15.8 fwhm
22.3 fwhm
Fig. 6 Energy sum of 3 TL signal by 511keV
photon.
12
Coincidence Timing ( real test)?
  • Event selection requirement for the coincidence
    timing.
  • R9800PMT 400 lt E lt 600 keV
  • MCP 3TL Sum 35 lt Int. Charge lt 60pC
  • Coincidence timing resolution 416ps( FWHM)?
  • contribution from R9800PMT side 200ps
    (FWHM)?

Real Test 416ps
Test set-up simulation 398ps
Fig. 7 coincidence time distribution.
13
4. Results Energy Timing
  • Sum of 5 TL signals around the maximum amplitude.
  • Energy resolution 11
  • Use the measured XP85022 SER for the TL signal.
  • The event time was extracted by Leading Edge(LE)
    to the maximum TL signal. ( Threshold 3mV)?
  • Energy window 450, 600 keV required for
    coincidence event.
  • The detection efficiency 40( 63 for one
    module).
  • Coincidence timing resolution 323 ps.

323ps(fwhm)
E_res 11(fwhm)
Fig 8. Energy (left) and Coincidence timing
distribution (right)
14
Results - DOI
  • 511keV gamma injected from side of detector with
    1mm step along Z axis.
  • Energy asymmetry and time difference of front and
    back due to different interaction depth.
  • (EFront EBack)/(EFront EBack)
  • Clear correlations were found.

Fig. 9 Energy asymmetry( left) and time
difference( right) measured at both front and
back MCP as a function of depth of interaction.
15
5. Summary
  • A PET detector design using pixelated array of
    LSO scintillator and MCP PMT with Transmission
    Line readout was studied.
  • Geant4 was used for optical photon simulation.
  • Real test setup using XP85022 MCP and TL board
    was built to measure SER of MCP. The measurement
    from the test set-up was fed to the simulation
    for TL signal forming.
  • The preliminary results from the study show
    promising results.
  • Energy resolution11 at 511keV was obtained.
  • The coincidence time resolution 323ps with 40
    detection efficiency were estimated.
  • Readout at both ends of scintillator makes it
    possible to extract the DOI information.
Write a Comment
User Comments (0)
About PowerShow.com