Title: Towards the automation of the test of photomultiplier tubes in Barcelona
1Towards the automation of the test of
photomultiplier tubes in Barcelona
- Rafael Ballabriga Suñé
- Xavier Xirgu Aleixandre
2Introduction
- Measures performed
- Instruments and Automation
- Summary
- Future Plans
- Open Questions
3Measures Performed
4Measures performed
- Absolute Gain
- Linearity
- Scintillation Pulses Analysis
- Crosstalk
5Gain Measurement
- Two Methods have been tested
- Fitting the spectrum of large number of
photoelectrons chosen - 2 measures spectrum/pedestal
- Easy to compute
6Gain Measurement (Setup)
7Gain Measurement (Results)
8Linearity measurement
- Double pulse method has been used.
- Deviation from linearity expressed as
9Linearity Measurement (setup)
10Linearity Measurement (results)
11Linearity as a function of High Voltage
- Linearity is a parameter dependent on the High
Voltage - The aim of this measure is to study this
dependence - Relative linearity has been found
- A small high frequency noise has been observed
when the LED driver is pulsed (400µV peak)
12Linearity as a function of High Voltage
13Linearity as a function of High Voltage
14Scintillator Pulse Analysis
- Objective To study the performance of the system
Scintillator-Fiber-PMT - A 12x12 pad has been used with 101 cmts WLS fiber
- Aiming at
- Mean number of photoelectrons/MIP
- Percentage of signal within 25 ns windows
15Scintillator Pulse Analysis (Setup I)
16Finding the fibers optimal position (setup)
17Scintillator Pulse Analysis (results)
Histogram of the number of photoelectrons per
signal
Time response waveform
18Scintillator Pulse Analysis (results)
Mean time waveform in lineal axis (time vs
voltage)
Mean time waveform in logarithmical axis (time vs
voltage) Slope-0.8768E-1 Statistics9568signals
19Scintillator Pulse Analysis (results)
20Scintillator Pulse Analysis (results)
21Crosstalk Measurement
- A measure of the system Scintillator-Fiber-PMT
- A system to position automatically the fibers in
the cathode has been designed - The mask used in this test places diagonally the
fibers in the fired pixel
22Crosstalk
- Crosstalk results of two studied channels.
- 2 diagonal WLS fibers are placed in close contact
with each channel.
23Crosstalk
- In the upper figure the fibers are placed in
close contact with the cathode. In the bottom the
fibers are placed 0.5 mm from the pixel.
24Instruments automation
25Instruments and automation
- 2x64 channels PMT H7546
- 2x1 channel PMT XP2262
- Light source (mimics scintillator pulses)
- 4 channel High Voltage Source
- 2x32 channel charge ADC (CAEN V792)
- PC with VME-MXI-2 interface (with LabVIEW
software) - 3 axis-motion stage ( X-Z motorized, Y manual)
- Oscilloscope TDS 380
- Timers, discriminators, logic units
26Some pictures of the setup
27Some pictures of the setup
28Some pictures of the setup
29Some pictures of the setup
The LED pulser
30Summary
31Summary
- The setup in Barcelona has been shown
- It is still under construction aiming at a fully
automated system - Some measures have been made in order to test the
performance of the PMT and calibrate the system
32Future plans
33Future plans
- Deadline October 2003
- Reduction of the noise of the LED pulse (1 month)
in parallel with signals analysis - Which tests are to be performed on each PMT?
- Absolute Gain of each channel
- Tests to be performed on some PMTs?
- Linearity
- Scintillator pulse analysis (Crosstalk)
34Open questions
35Open questions
- Define the tests to be performed on each tube
criteria to decide which cannels - Base study status
- Aging problem
- Performance under magnetic field?
- Possibility to test in the lab a complete set of
Scintillators-fibres-PMT