Assessment of Anterior Chamber Angles Scanning Peripheral Anterior Chamber Depth Analyser SPAC compa - PowerPoint PPT Presentation

1 / 15
About This Presentation
Title:

Assessment of Anterior Chamber Angles Scanning Peripheral Anterior Chamber Depth Analyser SPAC compa

Description:

Scanning Peripheral Anterior Chamber Depth Analyser (SPAC) compared to Gonioscopy ... SPAC correlates well with the gradings of gonioscopy. ... – PowerPoint PPT presentation

Number of Views:857
Avg rating:3.0/5.0
Slides: 16
Provided by: Kevi113
Category:

less

Transcript and Presenter's Notes

Title: Assessment of Anterior Chamber Angles Scanning Peripheral Anterior Chamber Depth Analyser SPAC compa


1
Assessment of Anterior Chamber Angles Scanning
Peripheral Anterior Chamber Depth Analyser (SPAC)
compared to Gonioscopy 
  • Johnson CH TAN, MBBS, MRCSEd(Ophthalmology)
  • Colin SH TAN, MBBS, M Med (Ophthalmology), MRCSEd
    (Ophthalmology), FAMS
  • The authors have no financial interest in any of
    the products mentioned

2
Literature Review
  • A newly developed peripheral anterior chamber
    depth analysis system principle, accuracy, and
    reproducibility.
  • K Kashiwagi et al. Br. J. Ophthalmol.
    2004881030-1035
  • Computer installed program automatically evaluate
    the ACD, CT and CRC.
  • Dummy eye was used for investigation.
  • 3 examiners measured ACD in 10 normal eyes
    (shaffers grade 3 or 4) and inter-observer and
    intra-obserever reproducibility analysed.

3
Literature Review
  • Quantitative evaluation of changes in anterior
    segment biometry by peripheral laser iridotomy
    using newly developed scanning peripheral
    anterior chamber depth analyser.
  • K Kashiwagi et al. Br. J. Ophthalmol.
    2004881036-1041
  • Finding cases of angle closure glaucoma in clinic
    setting using a newly developed instrument.
  • Kashiwagi K et al. Eye. 2006 Mar20(3)319-24
  • No studies so far to compare the gradings of SPAC
    versus that of gonioscopic findings with
    shaffers grading.

4
Objective
  • To assess the correlation between the scanning
    peripheral anterior chamber depth analyser (SPAC)
    and gonioscopy in the assessment of the anterior
    chamber angles. 

5
Method
  • Prospective analytical study performed at Tan
    Tock Seng Hospital, Singapore.
  • The temporal anterior chamber angles of patients
    seen at the ophthalmology outpatient clinic were
    assessed by SPAC (graded from 1 to 12) and
    gonioscopy performed by a single masked
    investigator using the Shaffers grading (grades
    1 to 4).

6
Method
  • Exclusion criteria
  • Patients with peripheral corneal disease.
    obstructing the view of the angles
  • peripheral anterior synechiae.
  • patients whose pupils were pharmacologically
    dilated.

7
Results
8
Demographics
  • Age distribution
  • Mean 60.6 (SD 12.6), range 38-83

N 37
9
Comparison of SPAC vs Sex
P0.100
10
Comparison of SPAC vs Race
P0.101
11
Comparison of SPAC vs Gonioscopy
Plt0.001
12
Correlation of SPAC and Gonioscopy
  • Spearmans correlation coefficient
  • (r) 0.699
  • (plt0.001)

13
Discussions
  • Gonioscopy remains the gold standard in
    evaluating the angles.
  • However, subjected to inter-observer and
    intra-observer variability.
  • Limitations of SPAC unable to pick up
    peripheral anterior synechiae.

14
Conclusion
  • SPAC correlates well with the gradings of
    gonioscopy.
  • All gonioscopy gradings of 2 or less have SPAC
    grades of 6 or less.
  • But some with gonioscopy gradings of 3 or more
    have SPAC grades of 6 or less.
  • Performing gonioscopy on SPAC grades of 6 or less
    appears to pick up the narrow angles.

15
Thank you