Title: Bilateral extrusion of Intracorneal ring segments 9 years post implantation


1
Bilateral extrusion of Intra-corneal ring
segments 9 years post implantation
  • Nisha R. Patel, MD, Barrie D. Soloway, MD,
  • P. Giordana Avila, OD
  • New York Eye and Ear Infirmary
  • Vista Alliance Eye Care Associates
  • New York, NY

2
Case Report
  • A 50 year-old Caucasian male presents with
    sudden-onset foreign body sensation after
    episodic eye rubbing. He reports having scratched
    his right eye while swimming 6 months prior, and
    felt that it never fully healed.
  • Ocular history was significant for Intacs OU 10
    years ago DV OD, NV OS. Medical, social and
    family history were all noncontributory and his
    medications included Genteal gel qAM OD and
    antibiotic drops OD qid, initiated by the
    referring physician, as well as a bandage contact
    lens, also initiated by the referring physician.
  • Examination revealed a visual acuity of OD 20/40,
    OS 20/150, brisk pupils with no RAPD in either
    eye. Motility was full and the patients
    refraction was OD 1.00-2.50 x 23, OS -1.75-0.50
    x 105.
  • Slit Lamp Exam 1 conjunctival injection OD, 50
    extrusion of the anterior surface of the temporal
    ICRS in the right eye with focal thinning and few
    SPK along its border. The other ICRS in the right
    eye, as well as both ICRS in the left eye, were
    in the anterior 25 of the cornea.
  • Both segments were removed OD. His contra lateral
    eye suffered the same extrusion 8 months later
    and both segments were removed from the left eye.

3
(No Transcript)
4
Case Report
  • Over a period of 9 months following the removal
    of the INTACS OD, the patient developed
    increasing with-the-rule astigmatism of 6.0D due
    to apparent ectasia, except at 12 and 6 oclock,
    where there were no segments.
  • After 11 months, his cornea stabilized and his
    astigmatism improved.
  • The left eye experienced a 3.0D increase in
    astigmatism which improved after 3 months.

5
Sequential Topographies OD
3 months post explantation Rx
-0.62 - 4.62 x 10
1 day prior to removal of Intacs OD Rx
1.00-2.50 x 23
6
Sequential Topographies OD
5 months post explantation Rx -1.62 - 4.50 x 7
6 months post explantation Rx -0.62 - 6.12 x 14
7
Sequential Topographies OD
11 months post explantation Rx -2.00 - 0.75 x 37
21 months post explantation Rx -3.75 - 0.87 x 14
8
Sequential Topographies OS
3 months prior to explantation Rx -0.75 - 1.25 x
131
3 days prior to explantation Rx 1.75-4.50 x 143
9
Sequential Topographies OS
2 months after removal of Intacs OS Rx -1.25 -
4.50 x 165
3 months post explantation Rx -2.25 - 3.37 x 170
10
Sequential Topographies OS
  • 13 months post explantation Rx -2.87 - 1.37 x
    145

11
Case Report
  • The patient was fit with rigid gas permeable
    lenses OU to achieve a BCVA of 20/20 in each eye
  • At the most recent visit, almost 2 years after
    the initial extrusion in the right eye, the
    patient was re-fit with soft contact lenses.

8 months after removal of Intacs OD
12
Conclusions
  • ICRS are approved as safe and effective for
    long-term control of myopia and keratoconus.
  • Anterior placement/ migration of these implants
    leads to overlying corneal tissue loss and
    extrusion.
  • We emphasize the need for proper depth of
    placement at the initial surgery to avoid the
    long-term post operative complication of
    extrusion.
  • Follow-up studies of patients beyond 3 months
    post-explantation may reveal further changes in
    refractive error.