Title: Phase Referencing Using More Than One Calibrator Ed Fomalont NRAO
1Phase Referencing Using More Than One
CalibratorEd Fomalont (NRAO)
2Summary
- Nodding between Calibrator and Target
- - Lessens short-term temporal phase
fluctuations - VERA two simultaneous beams
- Imaging of weak sources possible
- Position of target wrt calibrator
obtained - - Persistent systematic phase differences
between source and target produce position shifts
and large side-lobes. - - The smaller the Calibrator-Target
separation the better. - In-beam calibrator(s) best, but only
lt2 GHz - .
- Using more than One Calibrator
- - Removes most of angular phase dependence.
- - Higher quality images and positions (8 GHz,
0.8deg separation) - Dynamic range 201 ? 601
- Position accuracy 30 ? lt10 microarcsec
3PR Experiment Used for Jupiter Bending
- - Cycle among three sources for 10 hours on five
days - - 4.5-min cycling time
- - Measure change of position of J0842 on Sept 8,
2002 - - Parameters
- J0842 - J0839 Sep 48
- J0842 - J0854 Sep 193
- J0842 0.3 Jy (peak)
- J0854 1.5 Jy (OJ287)
- J0839 0.08 Jy
- - Good test of PR with one or two sources.
4Angular Phase Dependence of VLBI Observations
- Phase for each source using calib with self-cal
image - Angular dependence of phase separation
- Ph(J0842) Ph(J0854)
- 4xPh(J0842) Ph(J0839)
- (low elevation problems)
- Consistent with phase wedge in sky
- Main contribution is ionosphere/troposphere plus
geometric residuals - (calibrator position errors later)
- PR of J0842 using
- Ph(J0839)
- is not a good as using
- 0.8xPh(J0839) 0.2xPh(J0854)
5More Examples of Phase Wedge Behavior
6Phase Referencing Schemes
- Self-calibration Images
- J0842 image, position using PR J0839 (0.8 d)
- J0842 image, position using PR J0854 (3.2 d)
- J0842 image, position using
- PR (0.8 x J0839 0.2 x J0854)
- Compare results among five days to determine
- image quality and position
stability
7IMAGE TESTS
Self-cal 0.8 d PR 3.2 d PR
PR-2
VLBA Effelsberg 8.4 GHz, 64 MHz BW 10-hour
integrations Data editing during times of poor
stability 20 x 12 mas field of view
Sep 4
Sep 8
Lowest contour level 3 mJy 6
12 3
Peak in image 333 mJy 290
200 325
Most negative SL -2 mJy -14
-26 -6
8Grav bending sensitivity Sep 8 AVG 2 /- 4
e/w --5 /- 9 n/s (Correlator model assumes
GR is correct)
unitsmJy
unitsmicroarcsec
- COMPARISON OF 1-PR to 2-PR RESULTS
- --Significantly higher coherence with 2-PR. 97
versus 86 - -Negative side-lobes lower with 2-PR 1.5 versus
5 - -RMS scatter of positions 9 microasec for 2-PR
versus 35 microarcsec for 1-PR - -Systematic position change lt5 microarcsec in
R.A., 20 microarcsec in decl.
9(No Transcript)
10Phase Referencing Properties at 8 GHz
Self 2 PR 1 PR 1 PR 1 PR
Cal 0.8,3.2d 0.8d 3.2d 4.1d
Coher. ( of Peak) Min Side-lobe Level Position
RMS Position offset (micro-arcset)
11Solving for the Phase Wedge
Two calibrators within 10d of linearity (calibra
tor position errors in a few slides)
General solution to Determine phase
wedge (algorithm on request)
12If Phase Error is Entirely Caused by Troposphere
Source structure and position must be
accurately known Phase difference between
calibrators gives tropospheric zenith path
length correction, L Use C2 for frequent
calibration Use C1 less occasionally to
determine angular phase offset
13CALIBRATOR PROBLEMS
- Position Offsets
- Introduces position offset to target to
linear sum - used for combining calibrator phases.
- BUT harmful if assuming atmosphere is main
- cause of phase error (includes large
astrometric - errors also).
- Non-stationarity of Calibrators
- BIG problem for long-term, high frequency
obs. - Will introduce errors in long-term position
- stability.
- Using more than one calibrator can help
sort - out more unstable calibrators.
14SUMMARY
- If source is weak (lt201 dynamic range)
- Simple phase referening is okay. Image
- and position limited by SNR.
- Two or three calibrators within about
- four deg will increase image fidelity
and - positional accuracy even though less
time - is spent on target and cycling time
is a - bit longer.
- With care lt10 microarcsec relative position
- accuracy is obtainable with the VLBA.
- 4. If the calibrators are well-chosen, more
simpler - calibration schemes can take out
troposphere - if this is dominant error.
- 5. Over long-term projects, position errors are
- affected by calibrator changes. Try
to - register calibrator(s) as well as
possible!!