Title: BESTemc


1
BESTemc
Compact Tester of Schaffner for
Conducted immunity test system
Best_emc.ppt 23.05.2001
2
BESTemc
  • DimensionHeight 177.8 mmWidth 448.8
    mmDepth 590.0 mmWeight 40 kg
  • DUT InputAC 80 - 260V 16A maxDC 0 -
    65V 10AConnector 16A


WARNING This instrument requires two people to
lift
2
3
DUT connection
  • gt95 of all DUT connectors are 10A types with
    differentcountry specific connectors.
  • The BESTemc has a16A type DUT connector.The
    user has to adapt themselves.

95 use 10Aconnectors
BESTemc
4
BESTemc control is not user-friendly
  • Voltage setting Speed is 10 seconds per kV
  • The button needs to be pressed during the entire
    time
  • Setting of phase angle is too fast to stop at the
    desired phase angle
  • Test time setting in Seconds, Minutes, Hours,
    Counts, or Continuous- An easy setting like
    230 Min Second is not possible

5
BESTemc Standard Routine Settings
  • BESTemc offers a lot of defined Standard
    settings. The offered settings are the max.
    voltage level in both polarities for EN 50082-1/
    -2. The variation of the phase angles and some of
    the couplings must be set manually. Tests with
    Level 1 and Level4 are normally not in the menu.
  • The green fields show all pre-defined
    settings of the BESTemc.

Level 4
Level 3
Level 2
Level 1
L-N, L-PE, N-PE
L, N, PE, LN, all
Air Contact
50Hz 60Hz
Tests with capacitive coupling clamp are not in
this table.
0º 90º 180 270
Burst Surge
ESD Power fail
6
Burst
  • Burst frequency 1kHz...100kHz with steps of 1kHz
    max. 75 pulses / Burst2.5kHz at 4kV is not
    possible according to the data sheet.
  • Continuous Burst Endless with max 75 pulses /
    Burst with 100ms Repetition time.

Continuous Mode
Constant Burst
t Burst
t Rep
BESTemc can not test with Constant Burst
7
Surge Standard Test Settings
  • BESTemc standard menu offers only 12 of 88 surge
    settings. Each setting must be loaded manually.
  • The WINBEST can not create the complete test by
    starting at 500V up to 4000V.

Line to PE Neutral
to PE Line to Neutral phase deg
0 90 180 270 0 90 180 270 0 90 180
270 500 V no no no no no no no no
no no no no -500 V no no no no no
no no no no no no no 1000 V no no
no no no no no no yes no no no -1000
V no no no no no no no no yes no
no no 2000 V yes no no no yes no no
no yes no no no -2000 V yes no no no
yes no no no yes no no no 4000 V yes
no no no yes no no no -4000 V yes no
no no yes no no no
8
Surge Impulse Switch
  • Surge switch is a mercury Relay - 35A nominal
    current, -Isolation to GND 2,650V acc. to the
    data sheet - Electrical life at 35A limited to
    250000 operations- Operating vertically - 10

Up to 4400 V
Isolation to GND 2650 V
9
Surge BESTemc
  • Voltage impulses over 3kV with severe wave shape
    distortions (steps)

Records of Surge impulse 4.4kV
10
Surge Impulse
  • Jitter of Impulse -15 Deg- Synchronization
    setting 1Deg
  • Inductance is 2.5mHacc. to IEC 61000-4-5
    inductance is 1.5mH- dU 2 x Pi x f x L x
    Inominal x 2LN - dU 50Hz 314 x 2 x 2.5mH
    x 16A 25.13V- dU 60Hz 377 x 2 x 2.5mH x
    16A 30.16V

Jitter
setting at 90 deg
Decoupling network
dU
EUT
2.5mH
11
Power Fail
  • The Power Fail part is designed with relays and
    electronic devices for the DIPS and interrupts.
  • The built-in relays limit the rated current of
    the BESTemc.

100
EUT
40/70
12
Power Fail
  • Only 100, 70, and 40 setting is possible
    Steps of the internal autotransformer 100, 72,
    42
  • Duration of a voltage interruption is 10ms...
    2.5s, minimum step 10ms- 60Hz 0.5cycle
    8.33ms is not possible
  • Repetition Time 10s... 600s, Step 1s- Not
    useful for development

min 10ms
V
t
min 10s
13
Power Fail
  • BESTemc does not control an external motor
    variac.
  • An external variac must be set manually.
  • For connection, a banana plug and the DUT mains
    connector would be required.

14
Power Fail (note from the manual)
  • BESTemc can overheat after a certain time-
    under max load and- max repetition time ( every
    10s) - The test will stop and disconnect the
    DUT mains power supply- This is not
    indicated in the display
  • As a consequencePower down of BESTemc - for
    approx. 1 hour is necessary

15
ESD
  • Built-in is a NSG 435 but without display and
    switches
  • Standard Test only at max. voltage levels 4kV CD
    and 8kV AD- Contact discharge 2kV levels are
    missing- Air discharge 2kV and 4kV levels are
    missing

Contact discharge 2 kV ? 4 kV Air
discharge 2 kV ? 4 kV ? 8 kV
? ?
Built-in standard ESD settings

16
Magnetic Field
  • Calibration is only correct, when the calibration
    meter is in the circuit during the test
  • Option INA 701 single turn coilOption INA 702
    11 turn coil

17
Pulsed Magnetic Field
  • Calibration procedure is very suspect- The
    proposed shunts and clamps have normally a
    frequency range lt100kHz- A current measuring
    coil with bandwidth gt10MHz must be a Pearson
    probe or similar
  • Wave shape with INA 702 is out of tolerance-
    inductance of the antenna is too high

18
Software
  • Installation does not work when MS Office is open
  • Coupling mode- For each coupling mode, a new
    test must be created- no automatic change of 2 /
    12 Ohm
  • Polarity change within a test is not possible-
    Two different test procedures are necessary
  • Report not flexible (no rtf-format capability)
  • The default settings are not usefull- Burst
    1 kHz, 1 pulse/burst , Trep 2s is
    a poor default setting 5 kHz,
    75 pulses/burst , Trep 300ms is better as
    default

19
Accessories
  • Coupling clamp is not according to the standard
  • The delivered ground plane is not according to
    the standard

20
Various
  • To save the new settings of a menu, 20 seconds
    are needed.

Write a Comment
User Comments (0)