Ischemia Sensing Smart Organ Retractor: ____________________________________________________________ - PowerPoint PPT Presentation

1 / 29
About This Presentation
Title:

Ischemia Sensing Smart Organ Retractor: ____________________________________________________________

Description:

Placement of force and ischemia sensors on the Smart Organ Retractor ... Potentiometer. Amplifier. Amplifier/ I-to-V. Converter. LEDs. Photodiode. In-Circuit ... – PowerPoint PPT presentation

Number of Views:174
Avg rating:3.0/5.0
Slides: 30
Provided by: jennife272
Category:

less

Transcript and Presenter's Notes

Title: Ischemia Sensing Smart Organ Retractor: ____________________________________________________________


1
Ischemia Sensing Smart Organ Retractor__________
__________________________________________________
__________________________________________________
_________The Effects of Force on Ischemia over
Time
  • Jen Horwat
  • John Yu
  • Sunipa Saha
  • Mentors
  • Greg Fischer,
  • Mark A.Talamini, MD

MISTC Lab, Johns Hopkins Hospital
2
Overview
  • Problem Detection of ischemia intra-operatively
  • Solution The Smart Organ Retractor
  • Placement of force and ischemia sensors on the
    Smart Organ Retractor will allow measurement of
    the effects of retractor force on ischemia in the
    tissue over time

Photo from 3/5/03 experiments in MISTIC
Photo from 3/5/03 experiments in MISTIC
3
Overview
  • Problem Detection of ischemia intra-operatively
  • Solution The Smart Organ Retractor
  • Placement of force and ischemia sensors on the
    Smart Organ Retractor will allow measurement of
    the effects of retractor force on ischemia in the
    tissue over time

4
Motivation
  • Organs are retracted to
  • provide access to the target
  • area and a clear field of view
  • Clinical Studies
  • intraoperative retraction can
  • produce ischemia in the affected
  • tissue
  • CT Liver Study
  • Spinal Cord Fusion Study

www.adhb.govt.nz/trauma/ presentations/Forums/live
rats/ sld001.htm
5
Motivation
  • Problems with current pulse oximeters
  • Alternate Therapies
  • Modified Liver Retractor
  • Goal
  • Prevention rather than treatment
  • Reduce ischemic damage to organ tissue during
    surgery

www.karlstorz.com/hm/ getframe.html?2_products/
6
Motivation
www.innomed.net/InnoHip1.htm
www.karlstorz.com/hm/getframe.html?2_products/
7
Previous Work
  • Smart Organ Retractor, Greg Fischer
  • Optical technology
  • similar to pulse
  • oximetry
  • Ischemia Sensors
  • Limitations
  • Inconsistent readings for different tissues
  • Calibration
  • No method for measuring force of retraction

8
Previous Work
Deaver Retractor
6 DOF F/T
Emitter Receiver
9
Previous Work
Smart Organ Retractor
10
Previous Work
Strain Gauges
Full Bridge for each axis
Babcock Retractor
Emitter
Receiver
11
Pulse Oximetry
www.wardray-premise.com/engineering_services.htm
12
Pulse Oximetry
  • Measures amount of oxygenated hemoglobin in the
    blood
  • Greatest difference in absorption
  • by oxygenated and deoxygenated
  • hemoglobin at 660nm and 880-940nm
  • SOR - Measures reflected light
  • through tissue

www.wardray-premise.com/ engineering_services.htm
Casciani, J., Pulse Oximeter and Sensor
Optimized for Low Saturation, Aug. 2001.
13
Pulse Oximetry
Casciani, J., Pulse Oximeter and Sensor
Optimized for Low Saturation, United States
Patent No. 6,272,363, Aug. 2001.
14
Pulse Oximetry
15
System Diagram
Controller
In-Circuit Serial Programmer
Serial Interface
PIC Microcontroller
Programming Port
Digital I/O
Retractor
Serial Port Interface
Digital/Analog Converter
16x
8x
Amplifier
LEDs
Analog/Digital Converter
Amplifier/ I-to-V Converter
2x
2x
Photodiode
Programmable Potentiometer
16
System Diagram
Controller
In-Circuit Serial Programmer
Serial Interface
PIC Microcontroller
Programming Port
Digital I/O
Retractor
Serial Port Interface
Digital/Analog Converter
16x
8x
Amplifier
LEDs
Analog/Digital Converter
Amplifier/ I-to-V Converter
2x
2x
Photodiode
Programmable Potentiometer
Force Sensors
17
System Diagram
Hamamatsu Photodiode
www.hamatsu.com
SunLED
www.sun-led.com
18
System Diagram
Photodiode
LED
19
Overview of Deliverables
20
Minimum Deliverables
21
Overview of Deliverables
22
Medium Deliverables
23
Overview of Deliverables
24
Maximum Deliverables
25
Calibration
SPO2 Simulators Index 2XL, Fluke Biomedical,
www.flukebiomedical.com/.
26
Experiments at MISTC
MISTC Lab, Johns Hopkins Hospital
27
Schedule
March
April
May
mid-March
mid-April
Reprogram PIC Preliminary Calibration
Consistent Readings Force sensors on manual
and da Vinci retractors
Preliminary experiments Ischemia sensors on da
Vinci retractor
More experiments with manual retractor
Modifications to manual retractor Experiments
with da Vinci retractor
Sensor pattern on manual retractor
Complete Minimum and Medium Deliverables
Preliminary experiments for pattern
28
Project Management
  • Team Management
  • Funding
  • Funded by Engineering Research Center
  • Materials
  • Free samples from manufacturers, if possible
  • Retractors and da Vinci tools provided by Johns
    Hopkins Hospital and Intuitive Surgical, Inc.
  • Progress Checkpoints
  • Weekly Meeting with Mentor
  • Greg Fischer Monday 11am
  • Dr. Talamini Friday 3pm Surgeons Meeting

29
Main References
  • Adult Cardiac Surgical Instruments Sternal
    Retractor for Valve CABG Surgery, Genesee
    BioMedical, Inc., http//www.geneseebiomedical.com
    /.
  • Casciani, J., Pulse Oximeter and Sensor
    Optimized for Low Saturation, United States
    Patent No. 6,272,363, Aug. 2001.
  • Diab, M., E. Kiani-Azarbayjany, Low Noise
    Optical Probe, United States Patent No.
    5,638,818, June 1997.
  • Diab, M., Manual and Automatic Probe
    Calibration, United States Patent No. 6,397,091,
    May 2002.
  • Germer, C., Optical Properties of Native and
    Coagulated Human Liver Tissue and Liver
    Metastases in the Near Infrared Range, Lasers in
    Surgery and Medicine, 23194-203, 1998.
  • Liver Retractor, Karl Storz Endoskope
    Laparoscopy Products. http//www.karlstorz.com/hm/
    .
  • Lownie, S., X. Wu, S. Karlik, A. Gelb, Brain
    Retractor Edema During Induced Hypotension The
    Effect of the Rate of Return of Blood Pressure,
    Neurosurgery, 27(6)901-5, 1990
  • Ritch, R., W. Shihadeh, J. Liebmann, A
    Conjunctival Retractor for Glaucoma Filtering
    Surgery, Ophthalmic Surgical Lasers Imaging,
    35(3)261-2, May-June 2004.
  • SPO2 Simulators Index 2XL, Fluke Biomedical,
    http//www.flukebiomedical.com/.
  • Yassa, N., J. Peters, CT of Focal Hepatic Injury
    Due to Surgical Retractor, American Journal of
    Roentgenology, 166599-602, 1996.